Fluke 114,115,117 Manual
Page 3
... unspecified. Unsafe Voltage To alert you move the rotary switch to check that the test leads are in fixed-equipment installations at register.fluke.com. True-rms Multimeters Introduction The Fluke Model 114, Model 115, and Model 117 are designed to the presence of danger from any A (Amps) position. 1 This manual applies to make a measurement...
... unspecified. Unsafe Voltage To alert you move the rotary switch to check that the test leads are in fixed-equipment installations at register.fluke.com. True-rms Multimeters Introduction The Fluke Model 114, Model 115, and Model 117 are designed to the presence of danger from any A (Amps) position. 1 This manual applies to make a measurement...
Fluke 114,115,117 Manual
Page 4
... provided by the Meter might be impaired. • Do not use the Meter or test leads if they appear damaged, or if the Meter is not operating properly. • Always use test leads that have the same voltage, category, and amperage ratings as the meter and that have..., and range for measurements. • Verify the Meter's operation by a safety agency. • Remove test leads from Meter before opening the battery door or Meter case. 2 114, 115, and 117 Users Manual Safety Information A "XWWarning" statement identifies hazardous conditions and actions that could damage the Meter or the...
... provided by the Meter might be impaired. • Do not use the Meter or test leads if they appear damaged, or if the Meter is not operating properly. • Always use test leads that have the same voltage, category, and amperage ratings as the meter and that have..., and range for measurements. • Verify the Meter's operation by a safety agency. • Remove test leads from Meter before opening the battery door or Meter case. 2 114, 115, and 117 Users Manual Safety Information A "XWWarning" statement identifies hazardous conditions and actions that could damage the Meter or the...
Fluke 114,115,117 Manual
Page 5
...Current) X Hazardous voltage N Battery (Low battery when shown on the display.) I Fuse T Double Insulated W Important Information; Contact Fluke or a D AC and DC qualified recycler for continuity before use the Auto Volts function to manual J Earth ground ~ Do not ...dispose of this product as required by this function's low input impedance (≈3 kΩ)(114 and 117 only). Refer to measure voltages in hazardous areas. • Avoid working alone. • Use only the replacement... in circuits that could be impaired. • Check the test leads for disposal. 3
...Current) X Hazardous voltage N Battery (Low battery when shown on the display.) I Fuse T Double Insulated W Important Information; Contact Fluke or a D AC and DC qualified recycler for continuity before use the Auto Volts function to manual J Earth ground ~ Do not ...dispose of this product as required by this function's low input impedance (≈3 kΩ)(114 and 117 only). Refer to measure voltages in hazardous areas. • Avoid working alone. • Use only the replacement... in circuits that could be impaired. • Check the test leads for disposal. 3
Fluke 114,115,117 Manual
Page 7
The Meter selects the range for the selected range. 114, 115, & 117 R LEAd W Test lead alert. Briefly displayed whenever the Meter's function switch is in the Auto Volts function. 114 & 117 Auto Autoranging. User sets the Meter's range. 114, 115, & 117 P + Bar graph polarity 114, 115, & 117 Q 0L W The input is too large for best resolution. 114, 115...
The Meter selects the range for the selected range. 114, 115, & 117 R LEAd W Test lead alert. Briefly displayed whenever the Meter's function switch is in the Auto Volts function. 114 & 117 Auto Autoranging. User sets the Meter's range. 114, 115, & 117 P + Bar graph polarity 114, 115, & 117 Q 0L W The input is too large for best resolution. 114, 115...
Fluke 114,115,117 Manual
Page 11
... backlight off and when sleep mode is displayed. 1. LoFF is displayed when enabled. When connecting the test leads to make basic measurements. when removing the test leads, remove the live lead; If you press qwhile in MIN MAX AVG or Display Hold, the Meter beeps twice, indicating an...not change the range in the following pages show how to the circuit or device, connect the common (COM) test lead before connecting the live lead before removing the common test lead. 9 Enables low impedance capacitance measurements. When you turn the Meter off . To enter the Manual Range mode,...
... backlight off and when sleep mode is displayed. 1. LoFF is displayed when enabled. When connecting the test leads to make basic measurements. when removing the test leads, remove the live lead; If you press qwhile in MIN MAX AVG or Display Hold, the Meter beeps twice, indicating an...not change the range in the following pages show how to the circuit or device, connect the common (COM) test lead before connecting the live lead before removing the common test lead. 9 Enables low impedance capacitance measurements. When you turn the Meter off . To enter the Manual Range mode,...
Fluke 114,115,117 Manual
Page 14
... or DC Current (115 & 117) XWWarning To avoid personal injury or damage to the Meter: • Never attempt to make an in-circuit current measurement when the opencircuit potential to earth is >600 V. • Check the Meter's fuse before testing. (See "Testing the Fuse") • Use ...the proper terminals, switch position, and range for your measurement. • Never place the probes in parallel with a circuit or component when the leads are plugged into the A (Amps) terminals. 1 A 2 3...
... or DC Current (115 & 117) XWWarning To avoid personal injury or damage to the Meter: • Never attempt to make an in-circuit current measurement when the opencircuit potential to earth is >600 V. • Check the Meter's fuse before testing. (See "Testing the Fuse") • Use ...the proper terminals, switch position, and range for your measurement. • Never place the probes in parallel with a circuit or component when the leads are plugged into the A (Amps) terminals. 1 A 2 3...
Fluke 114,115,117 Manual
Page 19
... 17 The battery fits inside the battery door, which is then inserted into the case. 5. Lift the door straight up to the Meter: • Remove test leads from the case. Do not attempt to lift the door slightly. 4. Replacing the Battery and Fuse XWWarning To avoid shock, injury, or damage to separate...
... 17 The battery fits inside the battery door, which is then inserted into the case. 5. Lift the door straight up to the Meter: • Remove test leads from the case. Do not attempt to lift the door slightly. 4. Replacing the Battery and Fuse XWWarning To avoid shock, injury, or damage to separate...
Fluke 114,115,117 Manual
Page 20
114, 115, and 117 Users Manual To open the case for fuse replacement: 1. Remove the Meter from its holster. Use only Fluke PN 803293. 6. Finally, insert the Meter into its holster. 3. Remove the fuse from the Meter 2. Cleaning Wipe the case with an 11 A, 1000 V, FAST fuse ... Meter, first attach the case bottom to clean the case or lens/window. Dirt or moisture in the terminals can affect readings. 18 Remove the test leads from its holder and replace it with a damp cloth and mild detergent.
114, 115, and 117 Users Manual To open the case for fuse replacement: 1. Remove the Meter from its holster. Use only Fluke PN 803293. 6. Finally, insert the Meter into its holster. 3. Remove the fuse from the Meter 2. Cleaning Wipe the case with an 11 A, 1000 V, FAST fuse ... Meter, first attach the case bottom to clean the case or lens/window. Dirt or moisture in the terminals can affect readings. 18 Remove the test leads from its holder and replace it with a damp cloth and mild detergent.
Fluke 114,115,117 Manual
Page 24
...specified from 5 Hz to 50 kHz. AutoV LoZ, AC mV, and AC amps are specified from 1 % to 100 % of range. 114, 115, and 117 Users Manual Table 1 Accuracy Specifications (cont.) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model AC Amps Truerms[1] (45 Hz to 500 ...20 A for 30 seconds max., 10 minutes rest min. 0.001 A 0.01 A 1.5 % + 3 115, 117 DC Amps 6.000 A 10.00 A[3] 20 A for this and other true-rms meters to display nonzero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of ≤3 at full...
...specified from 5 Hz to 50 kHz. AutoV LoZ, AC mV, and AC amps are specified from 1 % to 100 % of range. 114, 115, and 117 Users Manual Table 1 Accuracy Specifications (cont.) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model AC Amps Truerms[1] (45 Hz to 500 ...20 A for 30 seconds max., 10 minutes rest min. 0.001 A 0.01 A 1.5 % + 3 115, 117 DC Amps 6.000 A 10.00 A[3] 20 A for this and other true-rms meters to display nonzero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of ≤3 at full...
Fluke 117 Electrician Multimeter Datasheet
Page 2
... 4000 counts, decreasing linearly to 500 Hz) 10.00 A 0.001 A 0.01 A 1.5 % + 3 20 A continuous overload for ac current. Fluke Corporation 117 Electrician's Multimeter with Non-Contact Voltage Ordering information Fluke-117 Electrician's Multimeter with Non-Contact Voltage Included TL75 Test leads, holster, User's manual and 9V battery (installed). For volts, crest factor of ≤ 3 at full scale...
... 4000 counts, decreasing linearly to 500 Hz) 10.00 A 0.001 A 0.01 A 1.5 % + 3 20 A continuous overload for ac current. Fluke Corporation 117 Electrician's Multimeter with Non-Contact Voltage Ordering information Fluke-117 Electrician's Multimeter with Non-Contact Voltage Included TL75 Test leads, holster, User's manual and 9V battery (installed). For volts, crest factor of ≤ 3 at full scale...