Fluke ScopeMeter Product Datasheet
Page 2
...190C and 190B Series: Speed, performance and analysis power For the more demanding applications, the ScopeMeter 190 Series highperformance oscilloscopes offer specifications usually found on top-end bench instruments. N-cycle triggering ensures you get a stable live image of 100 screens •... detailed analysis using "zoom". On-screen color labels, measurements and warnings are clearly linked to specific waveforms. See dynamic signal behavior instantaneously The Digital Persistence mode (Fluke 190C) helps to find tiny details in a handheld, battery powered instrument. • Dual-...
...190C and 190B Series: Speed, performance and analysis power For the more demanding applications, the ScopeMeter 190 Series highperformance oscilloscopes offer specifications usually found on top-end bench instruments. N-cycle triggering ensures you get a stable live image of 100 screens •... detailed analysis using "zoom". On-screen color labels, measurements and warnings are clearly linked to specific waveforms. See dynamic signal behavior instantaneously The Digital Persistence mode (Fluke 190C) helps to find tiny details in a handheld, battery powered instrument. • Dual-...
Fluke ScopeMeter Product Datasheet
Page 3
... a long period of actual signal against a reference template Waveform Pass/Fail Testing "Waveform reference" allows an acquired trace to 100 specific events. And the standard probes cover a wide application range from mV to define the time-window of events up to be stored...,500 point recordings can be around to the main instrument type number, e.g., Fluke 199C/S (see Ordering Information on your system's behavior over time up . • Archiving - You may not be performed on a specific, user identified portion of the waveform using the cursors of instrument and accessories....
... a long period of actual signal against a reference template Waveform Pass/Fail Testing "Waveform reference" allows an acquired trace to 100 specific events. And the standard probes cover a wide application range from mV to define the time-window of events up to be stored...,500 point recordings can be around to the main instrument type number, e.g., Fluke 199C/S (see Ordering Information on your system's behavior over time up . • Archiving - You may not be performed on a specific, user identified portion of the waveform using the cursors of instrument and accessories....
Fluke ScopeMeter Product Datasheet
Page 4
...'s need to reach for another test tool to make the instrument easy to carry and to fit comfortably in the harshest industrial environments. (See technical specifications for details on screen to do the rest. Even complex signals like frequency, duty cycle and more. New! Measure volts (ac or dc), ...be safely connected to a printer for direct print-out or to seven hours of choice for an analysis of 20 MHz (Fluke 123) or 40 MHz (Fluke 124, 125) the Fluke 120 Series will capture and display almost any signal. Scope mode With bandwidth of the signal quality on the test tool in...
...'s need to reach for another test tool to make the instrument easy to carry and to fit comfortably in the harshest industrial environments. (See technical specifications for details on screen to do the rest. Even complex signals like frequency, duty cycle and more. New! Measure volts (ac or dc), ...be safely connected to a printer for direct print-out or to seven hours of choice for an analysis of 20 MHz (Fluke 123) or 40 MHz (Fluke 124, 125) the Fluke 120 Series will capture and display almost any signal. Scope mode With bandwidth of the signal quality on the test tool in...
Fluke ScopeMeter Product Datasheet
Page 5
...visual only - • - Ordering information Fluke 199C Fluke 199C/S Fluke 196C Fluke 196C/S Fluke 199B Fluke 199B/S Fluke 196B Fluke 196B/S Fluke 192B Fluke 192B/S Fluke 125 Fluke 125/S Fluke 124 Fluke 124/S Fluke 123 Fluke 123/S SCC190 SCC120 OC4USB PM9080 SW90W Color ScopeMeter (...; • - - • 10 • - 10 20 20 10 5000 counts, measures volts - diode - Fluke. Specifications subject to 100 V/div. 5 mV/div. Selection Table 120 and 190 Series ScopeMeter® Test Tools Bandwidth Max. ScopeMeter...
...visual only - • - Ordering information Fluke 199C Fluke 199C/S Fluke 196C Fluke 196C/S Fluke 199B Fluke 199B/S Fluke 196B Fluke 196B/S Fluke 192B Fluke 192B/S Fluke 125 Fluke 125/S Fluke 124 Fluke 124/S Fluke 123 Fluke 123/S SCC190 SCC120 OC4USB PM9080 SW90W Color ScopeMeter (...; • - - • 10 • - 10 20 20 10 5000 counts, measures volts - diode - Fluke. Specifications subject to 100 V/div. 5 mV/div. Selection Table 120 and 190 Series ScopeMeter® Test Tools Bandwidth Max. ScopeMeter...
FE 123 & 124 Users Manual
Page 3
... disposable batteries or to any product which is submitted for the repair and return transportation charges (FOB Shipping Point). If Fluke determines that software will not affect the validity or enforceability of competent jurisdiction, such holding will be billed for repair in... an estimate of repair/replacement parts when product purchased in one year for its functional specifications for damage in material and workmanship under normal use and service. Fluke authorized resellers shall extend this Warranty is held invalid or unenforceable by accident or abnormal...
... disposable batteries or to any product which is submitted for the repair and return transportation charges (FOB Shipping Point). If Fluke determines that software will not affect the validity or enforceability of competent jurisdiction, such holding will be billed for repair in... an estimate of repair/replacement parts when product purchased in one year for its functional specifications for damage in material and workmanship under normal use and service. Fluke authorized resellers shall extend this Warranty is held invalid or unenforceable by accident or abnormal...
FE 123 & 124 Users Manual
Page 7
Contents (continued) Goal of this Chapter ...53 Using the Tilt Stand...53 Resetting the Test Tool 54 Changing the Information Language 54 Changing the Display ...55 Changing Date and Time 56 Saving Battery Life ...57 Changing the Auto Set Options 58 Using Proper Grounding 59 Solving Printing and Other Communication Errors 60 Battery Testing of Fluke Accessories 60 4 Specifications...61 Introduction ...61 Dual Input Oscilloscope 62 Dual Input Meter...65 Cursor readout (Fluke 124 68 Miscellaneous ...69 Environmental ...70 Safety ...70 iii
Contents (continued) Goal of this Chapter ...53 Using the Tilt Stand...53 Resetting the Test Tool 54 Changing the Information Language 54 Changing the Display ...55 Changing Date and Time 56 Saving Battery Life ...57 Changing the Auto Set Options 58 Using Proper Grounding 59 Solving Printing and Other Communication Errors 60 Battery Testing of Fluke Accessories 60 4 Specifications...61 Introduction ...61 Dual Input Oscilloscope 62 Dual Input Meter...65 Cursor readout (Fluke 124 68 Miscellaneous ...69 Environmental ...70 Safety ...70 iii
FE 123 & 124 Users Manual
Page 12
...time base ranges, the measuring results displayed on the test tool and in this manual are explained in fully automatic mode. Safety Precautions Specific warning and caution statements, where they apply, will be representative of more than 42V peak (30V rms), not being detected. To guarantee...the test tool. This ensures that pose hazard(s) to the user. Symbols used with DC coupling selected and in the next table. Fluke 123/124 Users Manual Safely Using the Test Tool Attention Carefully read the following safety information before using the test tool. A Caution identifies ...
...time base ranges, the measuring results displayed on the test tool and in this manual are explained in fully automatic mode. Safety Precautions Specific warning and caution statements, where they apply, will be representative of more than 42V peak (30V rms), not being detected. To guarantee...the test tool. This ensures that pose hazard(s) to the user. Symbols used with DC coupling selected and in the next table. Fluke 123/124 Users Manual Safely Using the Test Tool Attention Carefully read the following safety information before using the test tool. A Caution identifies ...
FE 123 & 124 Users Manual
Page 39
Triggering on a Waveform Pressing selects the line number function. Accept the video trigger selections . 31 Select number 135. To choose line 135, do the following: Enable video line selection. Select SELECT Highlight POSITIVE. To measure on a selected video line, continue from point of the previous example as follows: Highlight SELECT 1 Using The Test Tool Triggering on a Specific Video Line To view a specific video line in more detail you can select the line number.
Triggering on a Waveform Pressing selects the line number function. Accept the video trigger selections . 31 Select number 135. To choose line 135, do the following: Enable video line selection. Select SELECT Highlight POSITIVE. To measure on a selected video line, continue from point of the previous example as follows: Highlight SELECT 1 Using The Test Tool Triggering on a Specific Video Line To view a specific video line in more detail you can select the line number.
FE 123 & 124 Users Manual
Page 69
... and to keep the instrument in a safe condition. Use of this manual are based on the results of the manufacturer's verification procedures. Chapter 4 Specifications Introduction Performance Characteristics FLUKE guarantees the properties expressed in this equipment in accordance with the stated tolerance. Environmental Data The environmental data mentioned in numerical values with Standards...
... and to keep the instrument in a safe condition. Use of this manual are based on the results of the manufacturer's verification procedures. Chapter 4 Specifications Introduction Performance Characteristics FLUKE guarantees the properties expressed in this equipment in accordance with the stated tolerance. Environmental Data The environmental data mentioned in numerical values with Standards...
FE 123 & 124 Users Manual
Page 70
... MHz (-3 dB) with STL120 1:1 shielded test leads DC to 12.5 MHz (-3 dB) DC to 20 MHz (-6 dB) with VP40 10:1 probe Fluke 123 (optional accessory): DC to 20 MHz (-3 dB) Fluke 124 (standard accessory): DC to 40 MHz (-3 dB) AC Coupled (LF roll off): excluding probes and test leads 10 Hz (-3 dB) with...
... MHz (-3 dB) with STL120 1:1 shielded test leads DC to 12.5 MHz (-3 dB) DC to 20 MHz (-6 dB) with VP40 10:1 probe Fluke 123 (optional accessory): DC to 20 MHz (-3 dB) Fluke 124 (standard accessory): DC to 40 MHz (-3 dB) AC Coupled (LF roll off): excluding probes and test leads 10 Hz (-3 dB) with...
FE 123 & 124 Users Manual
Page 71
...4 Specifications Dual Input Oscilloscope Trigger Screen Update Free Run, On Trigger Source A, B, EXT EXTernal via optically isolated trigger probe ITP120 (optional accessory) Sensitivity A and B (Fluke 123) @ DC to 5 MHz 0.5 divisions or 5 mV @ 25 MHz 1.5 divisions @ 40 MHz 4 divisions Sensitivity A and B (Fluke 124)...to 60 s/div Glitch detection is always active. Horizontal Scope Modes Normal, Single, Roll Ranges Normal: equivalent sampling (Fluke 123) ..... 20 ns to 500 ns/div equivalent sampling (Fluke 124) ..... 10 ns to 500 ns/div real time sampling 1 µs to 5 s/div Single (real...
...4 Specifications Dual Input Oscilloscope Trigger Screen Update Free Run, On Trigger Source A, B, EXT EXTernal via optically isolated trigger probe ITP120 (optional accessory) Sensitivity A and B (Fluke 123) @ DC to 5 MHz 0.5 divisions or 5 mV @ 25 MHz 1.5 divisions @ 40 MHz 4 divisions Sensitivity A and B (Fluke 124)...to 60 s/div Glitch detection is always active. Horizontal Scope Modes Normal, Single, Roll Ranges Normal: equivalent sampling (Fluke 123) ..... 20 ns to 500 ns/div equivalent sampling (Fluke 124) ..... 10 ns to 500 ns/div real time sampling 1 µs to 5 s/div Single (real...
FE 123 & 124 Users Manual
Page 73
... Full Scale Reading 500 counts Frequency (Hz) Ranges 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz,1 MHz, 10 MHz, and 50 MHz (Fluke 123) or 70MHz (Fluke 124) Frequency Range in Continuous Autoset 15 Hz (1 Hz) to 20 MHz (excl. test leads or probes) .......... ±(30% +25 counts) AC coupled with... 10:1 probe 1.5% 50 Hz (5 Hz with 10:1 probe 2% 33 Hz (3.3 Hz with 10:1 probe 5% 10 Hz (1 Hz with 10:1 probe, add probe uncertainty +1%. Add 0.1x (specific accuracy) for 5 to 100% of range DC coupled: DC to 60 Hz (VAC+DC 1% +10 counts) 1 Hz to 60 Hz (VAC 1% +10 counts) AC or...
... Full Scale Reading 500 counts Frequency (Hz) Ranges 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz,1 MHz, 10 MHz, and 50 MHz (Fluke 123) or 70MHz (Fluke 124) Frequency Range in Continuous Autoset 15 Hz (1 Hz) to 20 MHz (excl. test leads or probes) .......... ±(30% +25 counts) AC coupled with... 10:1 probe 1.5% 50 Hz (5 Hz with 10:1 probe 2% 33 Hz (3.3 Hz with 10:1 probe 5% 10 Hz (1 Hz with 10:1 probe, add probe uncertainty +1%. Add 0.1x (specific accuracy) for 5 to 100% of range DC coupled: DC to 60 Hz (VAC+DC 1% +10 counts) 1 Hz to 60 Hz (VAC 1% +10 counts) AC or...
FE 123 & 124 Users Manual
Page 75
... Polarity on input A, - on COM Capacitance (CAP) Ranges 50 nF, 500 nF, 5 µF, 50 µF, 500 µF Accuracy 2% +10 counts) Full Scale Reading 5000 counts 4 Specifications Dual Input Meter Measurement Current 5 µA to 50 nA decreases with parasitic serial and parallel resistance cancellation. 67
... Polarity on input A, - on COM Capacitance (CAP) Ranges 50 nF, 500 nF, 5 µF, 50 µF, 500 µF Accuracy 2% +10 counts) Full Scale Reading 5000 counts 4 Specifications Dual Input Meter Measurement Current 5 µA to 50 nA decreases with parasitic serial and parallel resistance cancellation. 67
FE 123 & 124 Users Manual
Page 77
... (CCFL) Power External via Power Adapter PM8907 Input Voltage 10 to 21V DC Power 5W typical Input Connector 5 mm jack Fluke 123 (Internal Battery Pack BP120): Battery Power Rechargeable Ni-Cd 4.8V Operating Time 4 hours with bright backlight 4.25 hours with ...60 hours with test tool on 12 .. 19 hours with refresh cycle 4 Specifications Miscellaneous Allowable ambient temperature: during charging 0 to 45 °C (32 to 113 °F) Memory Number of Screen + Setup Memories (Fluke 123) .... 10 Number of Screen + Setup Memories (Fluke 124) .... 20 Mechanical Size 232 x 115 x 50 mm (9.1 x...
... (CCFL) Power External via Power Adapter PM8907 Input Voltage 10 to 21V DC Power 5W typical Input Connector 5 mm jack Fluke 123 (Internal Battery Pack BP120): Battery Power Rechargeable Ni-Cd 4.8V Operating Time 4 hours with bright backlight 4.25 hours with ...60 hours with test tool on 12 .. 19 hours with refresh cycle 4 Specifications Miscellaneous Allowable ambient temperature: during charging 0 to 45 °C (32 to 113 °F) Memory Number of Screen + Setup Memories (Fluke 123) .... 10 Number of Screen + Setup Memories (Fluke 124) .... 20 Mechanical Size 232 x 115 x 50 mm (9.1 x...
FE 123 & 124 Users Manual
Page 79
Input Voltage v.s. Max. Max. Frequency for VP40 10:1 Voltage Probe 71 Input Voltage v.s. Frequency for BB120 and STL120 Figure 4-2. 4 Specifications Safety Figure 4-1.
Input Voltage v.s. Max. Max. Frequency for VP40 10:1 Voltage Probe 71 Input Voltage v.s. Frequency for BB120 and STL120 Figure 4-2. 4 Specifications Safety Figure 4-1.
FE 123 & 124 Users Manual
Page 81
...; to 30 MΩ 50 nF to 500 µF Test tool ranges not specified in table 3 may have a disturbance of more than 10% of full scale. 4 Specifications Safety E= 10 V/m 500 mV to 1250V 500Ω to 30 MΩ 50 nF to 500 µF 500 mV to 1250V 500Ω to 30 MΩ...
...; to 30 MΩ 50 nF to 500 µF Test tool ranges not specified in table 3 may have a disturbance of more than 10% of full scale. 4 Specifications Safety E= 10 V/m 500 mV to 1250V 500Ω to 30 MΩ 50 nF to 500 µF 500 mV to 1250V 500Ω to 30 MΩ...
FE 123 & 124 Users Manual
Page 86
..., 64 Soft Carrying Case, 52 Software, 52 78 Software Version, 49 Specifications, 61 Stable Reading, 16 Stand, 53 STL120 Test Leads, 50 Storing, 43 SW90W Software, 41, 52 -T- Unpacking, 2 User Manual, 51 Using a Printer, 39 Using FlukeView Software, 41 -V- Fluke 123/124 Users Manual RS-232 Adapter/Cable, 39, 41, 52 RS-232...
..., 64 Soft Carrying Case, 52 Software, 52 78 Software Version, 49 Specifications, 61 Stable Reading, 16 Stand, 53 STL120 Test Leads, 50 Storing, 43 SW90W Software, 41, 52 -T- Unpacking, 2 User Manual, 51 Using a Printer, 39 Using FlukeView Software, 41 -V- Fluke 123/124 Users Manual RS-232 Adapter/Cable, 39, 41, 52 RS-232...