FE 983 Users Manual
Page 1
All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies. ® Fluke 983 Particle Counter Users Manual PN 2414721 January 2005 Rev. 2, 1/07 © 2005-2007 Fluke Corporation.
All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies. ® Fluke 983 Particle Counter Users Manual PN 2414721 January 2005 Rev. 2, 1/07 © 2005-2007 Fluke Corporation.
FE 983 Users Manual
Page 4
Fluke 983 Users Manual Purging the Particle Counter 12 Set Up ...12 Setting the Particle Counting Mode 14 Setting the Air Sampling Volume or Sampling Time 14 Setting the ...
Fluke 983 Users Manual Purging the Particle Counter 12 Set Up ...12 Setting the Particle Counting Mode 14 Setting the Air Sampling Volume or Sampling Time 14 Setting the ...
FE 983 Users Manual
Page 12
Fluke 983 Users Manual Safety Information A Warning identifies a condition or action that may damage the Particle Counter or the equipment under test. Safety symbols used in Table 1. P Conforms to Fluke's website for recycling information. Go to relevant European Union directives. 2 Refer to the user; Hazardous voltage. Symbol W X Table 1. Important information. Symbols Meaning Risk of this...
Fluke 983 Users Manual Safety Information A Warning identifies a condition or action that may damage the Particle Counter or the equipment under test. Safety symbols used in Table 1. P Conforms to Fluke's website for recycling information. Go to relevant European Union directives. 2 Refer to the user; Hazardous voltage. Symbol W X Table 1. Important information. Symbols Meaning Risk of this...
FE 983 Users Manual
Page 13
...the country or location in explosive atmospheres. • The Particle Counter contains no userserviceable parts. The connection should be sent to Fluke. Excessive particle intake can damage the Particle Counter. • Do not use a wrench to connect or disconnect the Isokinetic Probe ... or damage to the Particle Counter, follow these safety guidelines: • Use the Particle Counter only as described in the Users Manual, otherwise the protection provided by qualified service personnel. • Inspect the Particle Counter before use. Particle Counter Safety Information WCaution To...
...the country or location in explosive atmospheres. • The Particle Counter contains no userserviceable parts. The connection should be sent to Fluke. Excessive particle intake can damage the Particle Counter. • Do not use a wrench to connect or disconnect the Isokinetic Probe ... or damage to the Particle Counter, follow these safety guidelines: • Use the Particle Counter only as described in the Users Manual, otherwise the protection provided by qualified service personnel. • Inspect the Particle Counter before use. Particle Counter Safety Information WCaution To...
FE 983 Users Manual
Page 14
Remove the Isokinetic Probe Cover (the black cap) 3. To immediately use the Particle Counter: 1. Read "Safety Information" 2. Fluke 983 Users Manual Compliance The Particle Counter complies with: P, JIS B9921: 1997, 89/336/EEC, 73/23/EEC, EN61326: 1997, EN60825-1:2001, Federal Standard 209E Quick Start The Particle Counter may be used directly out of the box. Press the POWER keypad 4. Press 1 to start sampling Be sure to read the entire Users Manual for more detailed information. 4
Remove the Isokinetic Probe Cover (the black cap) 3. To immediately use the Particle Counter: 1. Read "Safety Information" 2. Fluke 983 Users Manual Compliance The Particle Counter complies with: P, JIS B9921: 1997, 89/336/EEC, 73/23/EEC, EN61326: 1997, EN60825-1:2001, Federal Standard 209E Quick Start The Particle Counter may be used directly out of the box. Press the POWER keypad 4. Press 1 to start sampling Be sure to read the entire Users Manual for more detailed information. 4
FE 983 Users Manual
Page 15
Particle Counter Shipping Contents A Fluke 983 Particle Counter and Holster with Hand Strap I Users Manual (English) bak07f.eps B Isokinetic Probe and Cap J Safety Information (Multi-language) C AC Adapter/Charger K Intake Nozzle and Cap D RS-232 Modular Cable L Power Cord E ¼ ... shown in . High Purity Tubing (2 ft.) M Hardshell Carrying Case F RJ-45 to DB-9 Connector N Zero Count Filter Adapter G Zero Count Filter O Isokinetic Probe with Barb H 983 Users Manual CD P FlukeView® Forms Documenting Software CD Figure 1. Fluke 983 Shipping Contents 5
Particle Counter Shipping Contents A Fluke 983 Particle Counter and Holster with Hand Strap I Users Manual (English) bak07f.eps B Isokinetic Probe and Cap J Safety Information (Multi-language) C AC Adapter/Charger K Intake Nozzle and Cap D RS-232 Modular Cable L Power Cord E ¼ ... shown in . High Purity Tubing (2 ft.) M Hardshell Carrying Case F RJ-45 to DB-9 Connector N Zero Count Filter Adapter G Zero Count Filter O Isokinetic Probe with Barb H 983 Users Manual CD P FlukeView® Forms Documenting Software CD Figure 1. Fluke 983 Shipping Contents 5
FE 983 Users Manual
Page 16
Power Key Arrow Keys Press to navigate up . Figure 2. Fluke 983 Users Manual The Keypad Use the keypad to navigate among different screens on the display, refer to control the display contrast when on the Main Screen. Each ... automatic backlight that stays on and off to a F1 F2 F2 F4 name above the keys. Fluke 983 Keypad If the backlight goes out, press any function or arrow key to turn it back on. 6 983 bak01f.eps Turns the Particle Counter on for two minutes when the unit is powered up and...
Power Key Arrow Keys Press to navigate up . Figure 2. Fluke 983 Users Manual The Keypad Use the keypad to navigate among different screens on the display, refer to control the display contrast when on the Main Screen. Each ... automatic backlight that stays on and off to a F1 F2 F2 F4 name above the keys. Fluke 983 Keypad If the backlight goes out, press any function or arrow key to turn it back on. 6 983 bak01f.eps Turns the Particle Counter on for two minutes when the unit is powered up and...
FE 983 Users Manual
Page 18
F1 F2 F3 F4 Figure 4. Fluke 983 Users Manual The Main Screen The Main Screen (see Figure 4) appears after the opening screen appears. Main Screen bak03f.eps 8 The Main Screen includes: • Six particle-...
F1 F2 F3 F4 Figure 4. Fluke 983 Users Manual The Main Screen The Main Screen (see Figure 4) appears after the opening screen appears. Main Screen bak03f.eps 8 The Main Screen includes: • Six particle-...
FE 983 Users Manual
Page 20
... rooms. Use the Label Menu to move the cursor up and down to store the choice. Select 1 (PROGRAM). 3. The clock will switch to Date. 2. Fluke 983 Users Manual The Clock Setup Menu To use and to enter the edit mode. Once the location number is selected, press 1 (EDIT). From the Main Screen, press...
... rooms. Use the Label Menu to move the cursor up and down to store the choice. Select 1 (PROGRAM). 3. The clock will switch to Date. 2. Fluke 983 Users Manual The Clock Setup Menu To use and to enter the edit mode. Once the location number is selected, press 1 (EDIT). From the Main Screen, press...
FE 983 Users Manual
Page 22
...;m in five minutes or not more than one particle per 0.5 cubic foot should be apparent. • Once the unit has been purged, resume normal use . Fluke 983 Users Manual Using the Particle Counter The following sections give detailed information about using the Zero Count Filter.
...;m in five minutes or not more than one particle per 0.5 cubic foot should be apparent. • Once the unit has been purged, resume normal use . Fluke 983 Users Manual Using the Particle Counter The following sections give detailed information about using the Zero Count Filter.
FE 983 Users Manual
Page 24
...the Particle Counter (liter, cubic foot, etc.). In Audio Mode, the display shows counts as they accumulate until the unit is stopped manually. For example, Concentration Mode might be substituted for full sampling. On Page 1 of the Setup Screen, use and to continuously sample ... the end of the sample. To select a sample volume, sample time, or to manually stop the Particle Counter, press 2 (left arrow) and 3 (right arrow) to select other modes. 3. Fluke 983 Users Manual Setting the Particle Counting Mode The Particle Counting Mode choices are: Concentration Mode In Concentration ...
...the Particle Counter (liter, cubic foot, etc.). In Audio Mode, the display shows counts as they accumulate until the unit is stopped manually. For example, Concentration Mode might be substituted for full sampling. On Page 1 of the Setup Screen, use and to continuously sample ... the end of the sample. To select a sample volume, sample time, or to manually stop the Particle Counter, press 2 (left arrow) and 3 (right arrow) to select other modes. 3. Fluke 983 Users Manual Setting the Particle Counting Mode The Particle Counting Mode choices are: Concentration Mode In Concentration ...
FE 983 Users Manual
Page 26
... Existing Password To change the selected character of samples by pressing 2 (left arrow) or 3 (right arrow). 4. The next time the Particle Counter is pressed. Fluke 983 Users Manual Setting the Number of the display. 5. Continue to the main screen. Select the number of the password. 7. Use and to the Login screen. The screen...
... Existing Password To change the selected character of samples by pressing 2 (left arrow) or 3 (right arrow). 4. The next time the Particle Counter is pressed. Fluke 983 Users Manual Setting the Number of the display. 5. Continue to the main screen. Select the number of the password. 7. Use and to the Login screen. The screen...
FE 983 Users Manual
Page 28
... on the PC. Transferring Data to a personal computer (PC) using the included FlukeView® Forms Documenting Software and RS-232 cable and RJ-45 converter. Fluke 983 Users Manual Setting Temperature Mode To set delay time, and start and stop a sample that is the default I/O Port setting for software version 1.1 and cannot be...
... on the PC. Transferring Data to a personal computer (PC) using the included FlukeView® Forms Documenting Software and RS-232 cable and RJ-45 converter. Fluke 983 Users Manual Setting Temperature Mode To set delay time, and start and stop a sample that is the default I/O Port setting for software version 1.1 and cannot be...
FE 983 Users Manual
Page 30
... the label in the FlukeView® Forms Documenting Software. Printing the Samples to either schedule a datalogging session or begin datalogging remotely. The Fluke 983 Thermal Printer kit includes: • Seiko DPU-414 Thermal Printer • Seiko PW-4007-U1 120 V ac adapter or PW-4007-...E1 220 V ac adapter • Seiko BP-4005 battery pack • 1 roll thermal paper • RJ-45 to "I /O port. Fluke 983 Users Manual Click "Meter" then "Logging" then press the "Schedule" tab to a Serial Printer Note The serial printer capability requires software version 1.2 or higher....
... the label in the FlukeView® Forms Documenting Software. Printing the Samples to either schedule a datalogging session or begin datalogging remotely. The Fluke 983 Thermal Printer kit includes: • Seiko DPU-414 Thermal Printer • Seiko PW-4007-U1 120 V ac adapter or PW-4007-...E1 220 V ac adapter • Seiko BP-4005 battery pack • 1 roll thermal paper • RJ-45 to "I /O port. Fluke 983 Users Manual Click "Meter" then "Logging" then press the "Schedule" tab to a Serial Printer Note The serial printer capability requires software version 1.2 or higher....
FE 983 Users Manual
Page 32
... to "Trickle Charge" mode to clean the Particle Counter case. To connect the ac adapter/charger: 1. Once 95 % of the charge is fully charged. Fluke 983 Users Manual WCaution To avoid damaging the Particle Counter, do not use abrasives or solvents to charge the remaining 5 % of the battery. Charging the Battery and Using...
... to "Trickle Charge" mode to clean the Particle Counter case. To connect the ac adapter/charger: 1. Once 95 % of the charge is fully charged. Fluke 983 Users Manual WCaution To avoid damaging the Particle Counter, do not use abrasives or solvents to charge the remaining 5 % of the battery. Charging the Battery and Using...
FE 983 Users Manual
Page 34
...". Try turning the unit on , press to darken the display or to lighten it. 24 If the problem persists, contact Fluke. The Particle Counter needs to update the battery status symbol. Troubleshooting Problem Particle Counter does not zero count Possible Reason and Recommended... not show a full battery charge. After the Particle Counter is indicated. Fluke 983 Users Manual Troubleshooting Use Table 2 to "Charging the Battery and Using the AC Adapter". Table 2. Internal air leak. Contact Fluke for assistance. Disconnect the Particle Counter from the ac adapter/charger. Refer ...
...". Try turning the unit on , press to darken the display or to lighten it. 24 If the problem persists, contact Fluke. The Particle Counter needs to update the battery status symbol. Troubleshooting Problem Particle Counter does not zero count Possible Reason and Recommended... not show a full battery charge. After the Particle Counter is indicated. Fluke 983 Users Manual Troubleshooting Use Table 2 to "Charging the Battery and Using the AC Adapter". Table 2. Internal air leak. Contact Fluke for assistance. Disconnect the Particle Counter from the ac adapter/charger. Refer ...
FE 983 Users Manual
Page 36
...-condensing Power AC Adapter, 12 VDC at 2.5 A, 90 to 250 VAC, 50 to 60 Hz Rechargeable Battery NiMH, 4.8 V at 2,000,000 particles per cubic ft. Fluke 983 Users Manual Specifications 6 Particle-Size Channels: 0.3, 0.5, 1.0, 2.0, 5.0, 10.0 µm Flow Rate 0.1 cfm (2.83 L/min) Light Source Laser diode;
...-condensing Power AC Adapter, 12 VDC at 2.5 A, 90 to 250 VAC, 50 to 60 Hz Rechargeable Battery NiMH, 4.8 V at 2,000,000 particles per cubic ft. Fluke 983 Users Manual Specifications 6 Particle-Size Channels: 0.3, 0.5, 1.0, 2.0, 5.0, 10.0 µm Flow Rate 0.1 cfm (2.83 L/min) Light Source Laser diode;
FE 983 Users Manual
Page 38
Fluke 983 Users Manual Callout A B C D E F G H I J K L M N O P Q Part Number 2414654 2421437 2421481 2421350 2421377 2421455 2421392 2421428 2414717 2414721 2414700 2421443 2421361 2414693 2414668 2424376 1578384 Table 3. Replaceable Parts and Accessories...Isokinetic Probe Cap AC Adapter/Charger RS-232 Modular Cable ¼" High Purity Tubing (2 ft.) RJ-45 to DB-9 Connector Zero Count Filter 983 Users Manual CD Users Manual (English) Safety Information (Multi-language) Intake Nozzle and Cap US Power Cord Hardshell Carrying Case Zero Count Filter Adapter Barbed Isokinetic Probe FlukeView®...
Fluke 983 Users Manual Callout A B C D E F G H I J K L M N O P Q Part Number 2414654 2421437 2421481 2421350 2421377 2421455 2421392 2421428 2414717 2414721 2414700 2421443 2421361 2414693 2414668 2424376 1578384 Table 3. Replaceable Parts and Accessories...Isokinetic Probe Cap AC Adapter/Charger RS-232 Modular Cable ¼" High Purity Tubing (2 ft.) RJ-45 to DB-9 Connector Zero Count Filter 983 Users Manual CD Users Manual (English) Safety Information (Multi-language) Intake Nozzle and Cap US Power Cord Hardshell Carrying Case Zero Count Filter Adapter Barbed Isokinetic Probe FlukeView®...