Fluke 726 Process Calibrator Datasheet
Page 2
...; All rights reserved. Pt100, 200, 500, 1000 (385), Cu 10 Pressure (requires Fluke 700PXX Modules) Frequency; 726 Specifications Function Measure or Source Voltage mA mV (TC terminals) Resistance Frequency Loop Supply Thermocouples ...manual (print) and user's manual (CD-ROM) in14 languages. 2 Fluke Corporation 726 Precision Multifunction Process Calibrator Fluke. Pt100 (JIS); Printed in ) Weight: 650 g (23 oz) Battery: Four AA alkaline batteries Battery life: 25 hours typical Warranty: Three-years Ordering Information Fluke-726 Precision Multifunction Process Calibrator...
...; All rights reserved. Pt100, 200, 500, 1000 (385), Cu 10 Pressure (requires Fluke 700PXX Modules) Frequency; 726 Specifications Function Measure or Source Voltage mA mV (TC terminals) Resistance Frequency Loop Supply Thermocouples ...manual (print) and user's manual (CD-ROM) in14 languages. 2 Fluke Corporation 726 Precision Multifunction Process Calibrator Fluke. Pt100 (JIS); Printed in ) Weight: 650 g (23 oz) Battery: Four AA alkaline batteries Battery life: 25 hours typical Warranty: Three-years Ordering Information Fluke-726 Precision Multifunction Process Calibrator...
FE 726 Users Manual
Page 1
September 2005 © 2005 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies. ® 726 Multifunction Process Calibrator Users Manual
September 2005 © 2005 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies. ® 726 Multifunction Process Calibrator Users Manual
FE 726 Users Manual
Page 10
Connections for Sourcing Pressure 40 19. Calibrating a Current-to -Current (P/I /P) Transmitter 50 23. SAVE DATA Menu Showing Measurement Memory Location 3, 1 44 20. Calibrating a Chart Recorder 52 24. Calibrating a Thermocouple Transmitter 46 21. Calibrating a Pressure-to -Pressure (I ) Transmitter 48 22. Replacement Parts...55 viii Replacing the Batteries ...53 25. 726 Users Manual 18.
Connections for Sourcing Pressure 40 19. Calibrating a Current-to -Current (P/I /P) Transmitter 50 23. SAVE DATA Menu Showing Measurement Memory Location 3, 1 44 20. Calibrating a Chart Recorder 52 24. Calibrating a Thermocouple Transmitter 46 21. Calibrating a Pressure-to -Pressure (I ) Transmitter 48 22. Replacement Parts...55 viii Replacing the Batteries ...53 25. 726 Users Manual 18.
FE 726 Users Manual
Page 11
...Stores and recalls setups. • Manual and automatic stepping and ramping. • Stores and recalls calibration screens. • Control the Calibrator remotely from a PC running a terminal emulator program. To register your product, visit register.fluke.com 1 The lower display allows the...display. Multifunction Process Calibrator Introduction The Fluke 726 Multifunction Process Calibrator (referred to the functions in the world: +1-425-446-5500 USA Service: 1-888-99-FLUKE (1-888-993-5853) Or, visit Fluke's Web site at www.fluke.com. In addition to as "the Calibrator") is a ...
...Stores and recalls setups. • Manual and automatic stepping and ramping. • Stores and recalls calibration screens. • Control the Calibrator remotely from a PC running a terminal emulator program. To register your product, visit register.fluke.com 1 The lower display allows the...display. Multifunction Process Calibrator Introduction The Fluke 726 Multifunction Process Calibrator (referred to the functions in the world: +1-425-446-5500 USA Service: 1-888-99-FLUKE (1-888-993-5853) Or, visit Fluke's Web site at www.fluke.com. In addition to as "the Calibrator") is a ...
FE 726 Users Manual
Page 13
... test leads • AC72 alligator clips • Stackable alligator clip test leads • 726 Product Overview (not shown in this manual, otherwise the protection provided by the Calibrator may damage the Calibrator or the equipment under test. 3 The items listed below and shown in Figure 1 are...-1, and ISA 82.02.01 XW Warning To avoid possible electric shock or personal injury, use the Calibrator only as specified in Figure 1) • 725/726 CD-ROM (contains Users Manual; A Warning identifies conditions and actions that may be impaired. To order replacement parts, see Table ...
... test leads • AC72 alligator clips • Stackable alligator clip test leads • 726 Product Overview (not shown in this manual, otherwise the protection provided by the Calibrator may damage the Calibrator or the equipment under test. 3 The items listed below and shown in Figure 1 are...-1, and ISA 82.02.01 XW Warning To avoid possible electric shock or personal injury, use the Calibrator only as specified in Figure 1) • 725/726 CD-ROM (contains Users Manual; A Warning identifies conditions and actions that may be impaired. To order replacement parts, see Table ...
FE 726 Users Manual
Page 14
... the test leads are plugged into the current terminals. • Do not use the Calibrator if it operates abnormally. Protection may be impaired. 726 Users Manual XW Warning To avoid possible electric shock or personal injury: • Use the Calibrator only as marked on the probes. • Connect the common test lead before connecting...
... the test leads are plugged into the current terminals. • Do not use the Calibrator if it operates abnormally. Protection may be impaired. 726 Users Manual XW Warning To avoid possible electric shock or personal injury: • Use the Calibrator only as marked on the probes. • Connect the common test lead before connecting...
FE 726 Users Manual
Page 17
International Symbols AC - B F J f P ) Table 2. Important information. See Manual. Precedes Warning. Precedes Warning. Alternating current DC - Conforms to European Union directives W O X Risk of danger. Direct current T M Double insulated Battery Earth ground Pressure Conforms to Canadian Standards Association directives. 7 Power ON/OFF Hazardous Voltage. Multifunction Process Calibrator Safety Information Symbols Symbols used on the Calibrator and in this manual are explained in Table 2.
International Symbols AC - B F J f P ) Table 2. Important information. See Manual. Precedes Warning. Precedes Warning. Alternating current DC - Conforms to European Union directives W O X Risk of danger. Direct current T M Double insulated Battery Earth ground Pressure Conforms to Canadian Standards Association directives. 7 Power ON/OFF Hazardous Voltage. Multifunction Process Calibrator Safety Information Symbols Symbols used on the Calibrator and in this manual are explained in Table 2.
FE 726 Users Manual
Page 18
Table 3 explains their use. 8 1 726 PRECISION CALIBRATOR %Error V mA LOOP ZERO 3 Seconds OPEN/CLOSE SWITCH TEST MEAS SOURCE HART V mA TC RTD FREQ PULSE CONFIG SELECTION EXIT CONFIG 100% SAVE RECALL 25% ENTER 25% TRIGGER/STOP RReReRtteueucrcrnanaltlltoo 0% 8 2 7 6 3 54 bec05f.eps Figure 2. Input/Output Terminals and Connectors 726 Users Manual Getting Acquainted with the Calibrator Input and Output Terminals Figure 2 shows the Calibrator input and output terminals.
Table 3 explains their use. 8 1 726 PRECISION CALIBRATOR %Error V mA LOOP ZERO 3 Seconds OPEN/CLOSE SWITCH TEST MEAS SOURCE HART V mA TC RTD FREQ PULSE CONFIG SELECTION EXIT CONFIG 100% SAVE RECALL 25% ENTER 25% TRIGGER/STOP RReReRtteueucrcrnanaltlltoo 0% 8 2 7 6 3 54 bec05f.eps Figure 2. Input/Output Terminals and Connectors 726 Users Manual Getting Acquainted with the Calibrator Input and Output Terminals Figure 2 shows the Calibrator input and output terminals.
FE 726 Users Manual
Page 20
726 Users Manual Keys Figure 3 shows the Calibrator keys and Table 4 explains their use. 2 1 726 PRECISION CALIBRATOR %Error V mA LOOP ZERO 3 Seconds OPEN/CLOSE SWITCH TEST MEAS SOURCE HART V mA TC RTD FREQ PULSE CONFIG SELECTION SAVE RECALL ENTER TRIGGER/STOP Return ...to Recal EXIT CONFIG 100% 25% 25% 0% 3 4 5 20 19 18 17 16 15 14 13 12 726 PRECISION CALIBRATOR %Error V mA LOOP ZERO 3 Seconds OPEN/CLOSE SWITCH TEST MEAS SOURCE HART V mA TC RTD FREQ PULSE CONFIG SELECTION SAVE RECALL ENTER TRIGGER/STOP Return...
726 Users Manual Keys Figure 3 shows the Calibrator keys and Table 4 explains their use. 2 1 726 PRECISION CALIBRATOR %Error V mA LOOP ZERO 3 Seconds OPEN/CLOSE SWITCH TEST MEAS SOURCE HART V mA TC RTD FREQ PULSE CONFIG SELECTION SAVE RECALL ENTER TRIGGER/STOP Return ...to Recal EXIT CONFIG 100% 25% 25% 0% 3 4 5 20 19 18 17 16 15 14 13 12 726 PRECISION CALIBRATOR %Error V mA LOOP ZERO 3 Seconds OPEN/CLOSE SWITCH TEST MEAS SOURCE HART V mA TC RTD FREQ PULSE CONFIG SELECTION SAVE RECALL ENTER TRIGGER/STOP Return...
FE 726 Users Manual
Page 22
.... T U Selects the pressure measurement and sourcing function. N S ENTER Saves and recalls setups & data. Selects resistance mode. 726 Users Manual Table 4. ENTER is used in the lower display. Repeated pushes cycle through the memory locations of Calibrator setups. R HART V Toggles between voltage, mA sourcing, or mA simulate functions in the configuration menus. Moves through...
.... T U Selects the pressure measurement and sourcing function. N S ENTER Saves and recalls setups & data. Selects resistance mode. 726 Users Manual Table 4. ENTER is used in the lower display. Repeated pushes cycle through the memory locations of Calibrator setups. R HART V Toggles between voltage, mA sourcing, or mA simulate functions in the configuration menus. Moves through...
FE 726 Users Manual
Page 24
... Press S to adjust the contrast up and down. 3. Press S, to exit configuration. Press G/EXIT CONFIG to save the setting. 14 726 PRECISION CALIBRATOR 1 3 %Error V mA LOOP ZERO 3 Seconds OPEN/CLOSE SWITCH TEST MEAS SOURCE HART V mA TC RTD FREQ PULSE CONFIG SELECTION EXIT ...CONFIG 2 100% SAVE RECALL ENTER TRIGGER/STOP RRReeRtteeuuccrrnnaalllttoo 25% 25% 0% Figure 5. 726 Users Manual Configuration Menus Use the configuration menus to set or change these parameters of the Calibrator: • Contrast Adjustment • Shut Down Mode • CJC on/off • ...
... Press S to adjust the contrast up and down. 3. Press S, to exit configuration. Press G/EXIT CONFIG to save the setting. 14 726 PRECISION CALIBRATOR 1 3 %Error V mA LOOP ZERO 3 Seconds OPEN/CLOSE SWITCH TEST MEAS SOURCE HART V mA TC RTD FREQ PULSE CONFIG SELECTION EXIT ...CONFIG 2 100% SAVE RECALL ENTER TRIGGER/STOP RRReeRtteeuuccrrnnaalllttoo 25% 25% 0% Figure 5. 726 Users Manual Configuration Menus Use the configuration menus to set or change these parameters of the Calibrator: • Contrast Adjustment • Shut Down Mode • CJC on/off • ...
FE 726 Users Manual
Page 26
... O to select 1 V for SOURCE mode (lower display). The Calibrator is turned on both mA channels. 16 Getting Started This section details some basic operations of the Calibrator. Press X to turn on the display. 2. 726 Users Manual HART Resistor ON/OFF 1. If necessary, press for the... output value. Connect the Calibrator's voltage output to enter 5 V as the 0 % value. 6....
... O to select 1 V for SOURCE mode (lower display). The Calibrator is turned on both mA channels. 16 Getting Started This section details some basic operations of the Calibrator. Press X to turn on the display. 2. 726 Users Manual HART Resistor ON/OFF 1. If necessary, press for the... output value. Connect the Calibrator's voltage output to enter 5 V as the 0 % value. 6....
FE 726 Users Manual
Page 28
LOOP appears and an internal 24 V loop supply turns on . 2. Press l while the Calibrator is turned on both mA channels. 18 bec42f.eps Figure 7. Measuring Voltage and Current Output To measure current with the current measuring circuit, allowing ...Power The loop power function activates a 24 V supply in series with loop power: 1. Press l to the transmitter current loop terminals as shown in Figure 8. 2. 726 Users Manual Using Measure Mode Measuring Electrical Parameters (Upper Display) To measure the current or voltage output of a transmitter, or to measure the output of a 700 Series...
LOOP appears and an internal 24 V loop supply turns on . 2. Press l while the Calibrator is turned on both mA channels. 18 bec42f.eps Figure 7. Measuring Voltage and Current Output To measure current with the current measuring circuit, allowing ...Power The loop power function activates a 24 V supply in series with loop power: 1. Press l to the transmitter current loop terminals as shown in Figure 8. 2. 726 Users Manual Using Measure Mode Measuring Electrical Parameters (Upper Display) To measure the current or voltage output of a transmitter, or to measure the output of a 700 Series...
FE 726 Users Manual
Page 30
..., and Rfor resistance. 20 726 PRECISION CALIBRATOR Red Black %Error V mA LOOP ZERO 3 Seconds OPEN/CLOSE SWITCH TEST MEAS SOURCE HART V mA TC RTD FREQ PULSE CONFIG SELECTION SAVE RECALL ENTER TRIGGER/STOP Return to Recal EXIT CONFIG 100% 25% 25% 0% bec43f.eps Figure 9. Measuring Electrical Parameters M 2. 726 Users Manual Measuring Electrical Parameters (Lower...
..., and Rfor resistance. 20 726 PRECISION CALIBRATOR Red Black %Error V mA LOOP ZERO 3 Seconds OPEN/CLOSE SWITCH TEST MEAS SOURCE HART V mA TC RTD FREQ PULSE CONFIG SELECTION SAVE RECALL ENTER TRIGGER/STOP Return to Recal EXIT CONFIG 100% 25% 25% 0% bec43f.eps Figure 9. Measuring Electrical Parameters M 2. 726 Users Manual Measuring Electrical Parameters (Lower...
FE 726 Users Manual
Page 34
... RTD measurement inputs in Figure 11. For more information, see the Application Note on the 725/726 CD. 24 726 Users Manual Using Resistance-Temperature Detectors (RTDs) The Calibrator accepts RTD types shown in Table 6. Press R for MEASURE mode. 2. Continue pressing this key to three custom curves can be named and CVD coefficients can...
... RTD measurement inputs in Figure 11. For more information, see the Application Note on the 725/726 CD. 24 726 Users Manual Using Resistance-Temperature Detectors (RTDs) The Calibrator accepts RTD types shown in Table 6. Press R for MEASURE mode. 2. Continue pressing this key to three custom curves can be named and CVD coefficients can...
FE 726 Users Manual
Page 38
...change pressure display units to ¼ ISO adapter if necessary. 2. Press A again to equal the reference pressure. 3. Refer to the Calibrator as shown in the module's Instruction Sheet. The threads on the pressure module or the pressure module instruction sheet for the acceptable material compatibility...Zero the pressure module as follows: 1. The maximum range of the pressure reading. 2. To adjust the Calibrator reading, proceed as described in Figure 13. 726 Users Manual • Never apply pressure above the rated maximum printed on module type, but the 700PA3 module. Modules...
...change pressure display units to ¼ ISO adapter if necessary. 2. Press A again to equal the reference pressure. 3. Refer to the Calibrator as shown in the module's Instruction Sheet. The threads on the pressure module or the pressure module instruction sheet for the acceptable material compatibility...Zero the pressure module as follows: 1. The maximum range of the pressure reading. 2. To adjust the Calibrator reading, proceed as described in Figure 13. 726 Users Manual • Never apply pressure above the rated maximum printed on module type, but the 700PA3 module. Modules...
FE 726 Users Manual
Page 40
... mA and SIM display. 4. Sourcing 4 to 20-mA Transmitter Simulate is a special mode of operation in which the Calibrator is connected into a loop in place of RTD and thermocouple temperature sensors • measures gas pressure from an external source, creating... in the mA terminals (left column). M 2. Connect the test leads in Figure 14. 726 Users Manual Using Source Mode In SOURCE mode, the Calibrator: • generates calibrated signals for testing and calibrating process instruments • supplies voltages, currents, frequencies, and resistances • simulates the electrical ...
... mA and SIM display. 4. Sourcing 4 to 20-mA Transmitter Simulate is a special mode of operation in which the Calibrator is connected into a loop in place of RTD and thermocouple temperature sensors • measures gas pressure from an external source, creating... in the mA terminals (left column). M 2. Connect the test leads in Figure 14. 726 Users Manual Using Source Mode In SOURCE mode, the Calibrator: • generates calibrated signals for testing and calibrating process instruments • supplies voltages, currents, frequencies, and resistances • simulates the electrical ...
FE 726 Users Manual
Page 44
... key to center). M 2. Press Y and Z to select a different digit to force a miniplug into the wrong polarization. If desired, continue pressing this connection. 726 Users Manual Simulating Thermocouples Connect the Calibrator TC input/output to the instrument under test with the thermocouple wire and the appropriate thermocouple mini-connector (polarized thermocouple plug with flat...
... key to center). M 2. Press Y and Z to select a different digit to force a miniplug into the wrong polarization. If desired, continue pressing this connection. 726 Users Manual Simulating Thermocouples Connect the Calibrator TC input/output to the instrument under test with the thermocouple wire and the appropriate thermocouple mini-connector (polarized thermocouple plug with flat...
FE 726 Users Manual
Page 46
... as follows to the instrument under test exceeds the limits of the 726. 36 Note Use the 3W and 4W terminals for measurement only, not for the RTD display. See Figure 17. 3. 726 Users Manual Simulating RTDs Connect the Calibrator to simulate an RTD: M 1. To connect to a 3-wire or... 4-wire transmitter, use the stacking cables to edit. 4. If the 726 display indicates ExI HI, the excitation current from your device under...
... as follows to the instrument under test exceeds the limits of the 726. 36 Note Use the 3W and 4W terminals for measurement only, not for the RTD display. See Figure 17. 3. 726 Users Manual Simulating RTDs Connect the Calibrator to simulate an RTD: M 1. To connect to a 3-wire or... 4-wire transmitter, use the stacking cables to edit. 4. If the 726 display indicates ExI HI, the excitation current from your device under...
FE 726 Users Manual
Page 48
... pressurized system, shut off the valve and slowly bleed off the pressure before attaching the pressure module to a Fluke pressure module which makes it a calibrated source. Refer to be tested. Many ranges and types of torque between the pressure module fittings, or between...apply more than 10 ft.-lb. (13.5 Nm) of pressure modules are available from Fluke, see "Accessories". Before using a pressure module, read its instruction sheet. 726 Users Manual Sourcing Pressure The Calibrator sources pressure by measuring pressure supplied by a pump or other sources, and displaying the pressure...
... pressurized system, shut off the valve and slowly bleed off the pressure before attaching the pressure module to a Fluke pressure module which makes it a calibrated source. Refer to be tested. Many ranges and types of torque between the pressure module fittings, or between...apply more than 10 ft.-lb. (13.5 Nm) of pressure modules are available from Fluke, see "Accessories". Before using a pressure module, read its instruction sheet. 726 Users Manual Sourcing Pressure The Calibrator sources pressure by measuring pressure supplied by a pump or other sources, and displaying the pressure...