Fluke 51II, 52II, 53II, and 54II Thermometer Datasheet
Page 1
...battery life • Battery door allows easy battery replacement without breaking the calibration seal Additional features for the 53 and 54 Series II: • Data Logging up to 500 points of data with user adjustable recording interval • Real time clock ...absorbing holster Two bead probe thermocouples 80PK-1 (54+52) One bead probe thermocouple 80PK-1 (51+53) Ordering Information Fluke 51 II Thermometer Fluke 52 II Thermometer Fluke 53 II Thermometer Fluke 54 II Thermometer FVF-SC1 FlukeView Forms Software including interface cable Specifications Temperature range...
...battery life • Battery door allows easy battery replacement without breaking the calibration seal Additional features for the 53 and 54 Series II: • Data Logging up to 500 points of data with user adjustable recording interval • Real time clock ...absorbing holster Two bead probe thermocouples 80PK-1 (54+52) One bead probe thermocouple 80PK-1 (51+53) Ordering Information Fluke 51 II Thermometer Fluke 52 II Thermometer Fluke 53 II Thermometer Fluke 54 II Thermometer FVF-SC1 FlukeView Forms Software including interface cable Specifications Temperature range...
FE 51 & 52 II Users Manual
Page 1
Printed in USA All product names are trademarks of their respective companies. ® 51 & 52 Series II Thermometer English September 1999 Rev.1, 6/01 © 1999-2001 Fluke Corporation, All rights reserved. Users Manual
Printed in USA All product names are trademarks of their respective companies. ® 51 & 52 Series II Thermometer English September 1999 Rev.1, 6/01 © 1999-2001 Fluke Corporation, All rights reserved. Users Manual
FE 51 & 52 II Users Manual
Page 4
51 & 52 Series II Users Manual Maintenance...13 Replacing the Batteries 13 Cleaning the Case and Holster 13 Calibration ...13 Specifications ...13 Environmental...13 General...14 80 PK-1 Thermocouple (supplied with thermometer 14 Electrical...14 Replacement Parts and Accessories 15 ii
51 & 52 Series II Users Manual Maintenance...13 Replacing the Batteries 13 Cleaning the Case and Holster 13 Calibration ...13 Specifications ...13 Environmental...13 General...14 80 PK-1 Thermocouple (supplied with thermometer 14 Electrical...14 Replacement Parts and Accessories 15 ii
FE 51 & 52 II Users Manual
Page 5
... in Table 1 and meter symbols in this manual. P.O. Contacting Fluke To order accessories, receive assistance, or locate the nearest Fluke distributor or Service Center, call: 1-888-99-FLUKE (1-888-993-5853) in USA 1-800-36-FLUKE (1-800-363-5853) in Canada +31-402-678-200 in Europe...-5500 from other countries Address correspondence to use external J-, K-, T-, and Etype thermocouples (temperature probes) as specified in Table 2. 51 & 52 Series II Introduction The Fluke Model 51 and Model 52 Thermometers ("the thermometer") are microprocessor-based, digital thermometers designed to...
... in Table 1 and meter symbols in this manual. P.O. Contacting Fluke To order accessories, receive assistance, or locate the nearest Fluke distributor or Service Center, call: 1-888-99-FLUKE (1-888-993-5853) in USA 1-800-36-FLUKE (1-800-363-5853) in Canada +31-402-678-200 in Europe...-5500 from other countries Address correspondence to use external J-, K-, T-, and Etype thermocouples (temperature probes) as specified in Table 2. 51 & 52 Series II Introduction The Fluke Model 51 and Model 52 Thermometers ("the thermometer") are microprocessor-based, digital thermometers designed to...
FE 51 & 52 II Users Manual
Page 6
The possibility of false readings can lead to the user. Protection may be impaired. 51 & 52 Series II Users Manual Table 1. Safety Information WWarning A Warning identifies conditions and actions that pose hazards to personal injury. • Do not use the thermometer if ...
The possibility of false readings can lead to the user. Protection may be impaired. 51 & 52 Series II Users Manual Table 1. Safety Information WWarning A Warning identifies conditions and actions that pose hazards to personal injury. • Do not use the thermometer if ...
FE 51 & 52 II Users Manual
Page 7
... laws or regulations when disposing of batteries. • Match the + and − polarities of the battery with any part of the case or cover removed. 51 & 52 Series II Introduction Table 1.
... laws or regulations when disposing of batteries. • Match the + and − polarities of the battery with any part of the case or cover removed. 51 & 52 Series II Introduction Table 1.
FE 51 & 52 II Users Manual
Page 8
... following sections. 4 Complies with relevant Canadian Standards Association directives. International Symbols W M Refer to the manual for information about this Users Manual applies both to Models 51 and 52 unless otherwise indicated. Then read the following : • Figure 1 and Table 3 describe the components. • Figure 2 and Table 4 describe the display. • Table...
... following sections. 4 Complies with relevant Canadian Standards Association directives. International Symbols W M Refer to the manual for information about this Users Manual applies both to Models 51 and 52 unless otherwise indicated. Then read the following : • Figure 1 and Table 3 describe the components. • Figure 2 and Table 4 describe the display. • Table...
FE 51 & 52 II Users Manual
Page 9
Components A Thermocouple T1 input B Model 52: Thermocouple T2 input C Holster D Display E Buttons F Battery door G Batteries Figure 1. Components aas01f.eps 5 Components 3 4 5 6 7 xx 2 1 51 & 52 Series II Getting Started Table 3.
Components A Thermocouple T1 input B Model 52: Thermocouple T2 input C Holster D Display E Buttons F Battery door G Batteries Figure 1. Components aas01f.eps 5 Components 3 4 5 6 7 xx 2 1 51 & 52 Series II Getting Started Table 3.
FE 51 & 52 II Users Manual
Page 10
... includes an offset. D Setup is in progress. C A shift function is in progress. E Low battery. I The thermocouple type. Model 52: T1, T2, or T1-T2 reading. 51 & 52 Series II Users Manual Display Elements 23 4 1 10 9 Figure 2. See "Changing Setup Options." H Secondary Display: MAX, MIN, AVG, or offset. J Time Display: The elapsed time...
... includes an offset. D Setup is in progress. C A shift function is in progress. E Low battery. I The thermocouple type. Model 52: T1, T2, or T1-T2 reading. 51 & 52 Series II Users Manual Display Elements 23 4 1 10 9 Figure 2. See "Changing Setup Options." H Secondary Display: MAX, MIN, AVG, or offset. J Time Display: The elapsed time...
FE 51 & 52 II Users Manual
Page 11
... off . Press h when turning on and off this display. Press h to stop displaying the minimum, maximum, and average readings in the primary or secondary display. 7 51 & 52 Series II Getting Started Buttons A G (Shift function) Q M C h T Table 5. Buttons Press A to turn the backlight on the thermometer to switch between Celsius (oC), Fahrenheit (oF), and...
... off . Press h when turning on and off this display. Press h to stop displaying the minimum, maximum, and average readings in the primary or secondary display. 7 51 & 52 Series II Getting Started Buttons A G (Shift function) Q M C h T Table 5. Buttons Press A to turn the backlight on the thermometer to switch between Celsius (oC), Fahrenheit (oF), and...
FE 51 & 52 II Users Manual
Page 12
51 & 52 Series II Users Manual D K N E Table 5. Press E to decrease the displayed setting. Press N to enter a Setup option. Press E again to increase the displayed setting. Buttons (cont.) Press D to start or exit Setup. (See "Changing Setup Options.") Press K to scroll to the Setup option you want to change . Press N to scroll to the Setup option you want to change . Press K to store the displayed setting in memory. 8
51 & 52 Series II Users Manual D K N E Table 5. Press E to decrease the displayed setting. Press N to enter a Setup option. Press E again to increase the displayed setting. Buttons (cont.) Press D to start or exit Setup. (See "Changing Setup Options.") Press K to scroll to the Setup option you want to change . Press N to scroll to the Setup option you want to change . Press K to store the displayed setting in memory. 8
FE 51 & 52 II Users Manual
Page 13
... T2 (Model 52) Sleep Mode SLP on (sleep mode on the thermometer. Setup settings reset only when the batteries are removed for more than 2 minutes. 51 & 52 Series II Using the Thermometer Entering or Exiting Setup When the thermometer is "open," the display shows "- - - -" Changing Setup Options Use Setup to start or...
... T2 (Model 52) Sleep Mode SLP on (sleep mode on the thermometer. Setup settings reset only when the batteries are removed for more than 2 minutes. 51 & 52 Series II Using the Thermometer Entering or Exiting Setup When the thermometer is "open," the display shows "- - - -" Changing Setup Options Use Setup to start or...
FE 51 & 52 II Users Manual
Page 14
... the setup option you turn on the display. 4. Offset: The primary display shows the temperature plus the offset and the secondary display shows the offset. 51 & 52 Series II Users Manual Changing a Setup Option 1. Line frequency: For optimum rejection of line noise, the thermometer must be set for the local line frequency...
... the setup option you turn on the display. 4. Offset: The primary display shows the temperature plus the offset and the secondary display shows the offset. 51 & 52 Series II Users Manual Changing a Setup Option 1. Line frequency: For optimum rejection of line noise, the thermometer must be set for the local line frequency...
FE 51 & 52 II Users Manual
Page 15
... Blue Orange 1. Press C to the measurement location. T 3. Model 52: If only thermocouple T2 is connected, the T2 reading appears in the primary or secondary display. 51 & 52 Series II Measuring Temperatures Notes The display shows "- - - -" when a thermocouple is outside the thermocouple's valid range. Measuring Temperatures Connecting a Thermocouple To change the thermocouple type...
... Blue Orange 1. Press C to the measurement location. T 3. Model 52: If only thermocouple T2 is connected, the T2 reading appears in the primary or secondary display. 51 & 52 Series II Measuring Temperatures Notes The display shows "- - - -" when a thermocouple is outside the thermocouple's valid range. Measuring Temperatures Connecting a Thermocouple To change the thermocouple type...
FE 51 & 52 II Users Manual
Page 16
... offset option in the primary or secondary display. 3. Place the thermocouple in a known, stable temperature environment (such as an ice bath or a dry well calibrator). 3. 51 & 52 Series II Users Manual Holding the Displayed Readings 1.
... offset option in the primary or secondary display. 3. Place the thermocouple in a known, stable temperature environment (such as an ice bath or a dry well calibrator). 3. 51 & 52 Series II Users Manual Holding the Displayed Readings 1.
FE 51 & 52 II Users Manual
Page 17
To calibrate the thermometer, contact Fluke for the Service Center nearest you calibrate the thermometer annually, starting one year after purchase. Maintenance Replacing the Batteries Refer to +140 oF) Humidity Non ...;40 oF to the safety information in "Replacement Parts and Accessories." Wipe with a damp sponge or soft rag. 51 & 52 Series II Maintenance Calibration To ensure that the thermometer performs to its accuracy specifications, Fluke recommends that you or follow the calibration procedure in the service manual listed in Table 1 before replacing the...
To calibrate the thermometer, contact Fluke for the Service Center nearest you calibrate the thermometer annually, starting one year after purchase. Maintenance Replacing the Batteries Refer to +140 oF) Humidity Non ...;40 oF to the safety information in "Replacement Parts and Accessories." Wipe with a damp sponge or soft rag. 51 & 52 Series II Maintenance Calibration To ensure that the thermometer performs to its accuracy specifications, Fluke recommends that you or follow the calibration procedure in the service manual listed in Table 1 before replacing the...
FE 51 & 52 II Users Manual
Page 18
... to +752 oF) E-type: −150 oC to +1000 oC (−238 oF to +1832 oF) 0.1 oC / oF / K < 1000o 1.0 oC / oF / K ≥ 1000o 14 51 & 52 Series II Users Manual General Weight 280 g (10 oz) Dimensions (without holster) 2.8 cm × 7.8 cm × 16.2 cm (1.1 in × 3 in × 6.4 in which measures are...
... to +752 oF) E-type: −150 oC to +1000 oC (−238 oF to +1832 oF) 0.1 oC / oF / K < 1000o 1.0 oC / oF / K ≥ 1000o 14 51 & 52 Series II Users Manual General Weight 280 g (10 oz) Dimensions (without holster) 2.8 cm × 7.8 cm × 16.2 cm (1.1 in × 3 in × 6.4 in which measures are...
FE 51 & 52 II Users Manual
Page 19
... reading for T-type] Susceptibility: ±2 oC (±3.6 oF) for 80 MHz to 200 MHz in 3 V/m field. The above specifications do not include thermocouple error. 51 & 52 Series II Replacement Parts and Accessories Replacement Parts and Accessories Accessory Part Number Holster and Flex Stand™ Assembly 1272438 AA NEDA 15A IEC LR6 batteries...
... reading for T-type] Susceptibility: ±2 oC (±3.6 oF) for 80 MHz to 200 MHz in 3 V/m field. The above specifications do not include thermocouple error. 51 & 52 Series II Replacement Parts and Accessories Replacement Parts and Accessories Accessory Part Number Holster and Flex Stand™ Assembly 1272438 AA NEDA 15A IEC LR6 batteries...