Fluke 114,115,117 Manual
Page 1
® 114, 115, and 117 True-rms Multimeters Users Manual PN 2572573 July 2006, Rev. 1, 2/07 © 2006, 2007 Fluke Corporation, All rights reserved. Printed in China All product names are trademarks of their respective companies.
® 114, 115, and 117 True-rms Multimeters Users Manual PN 2572573 July 2006, Rev. 1, 2/07 © 2006, 2007 Fluke Corporation, All rights reserved. Printed in China All product names are trademarks of their respective companies.
Fluke 114,115,117 Manual
Page 4
A "WCaution" statement identifies conditions and actions that could cause bodily harm or death. 114, 115, and 117 Users Manual Safety Information A "XWWarning" statement identifies hazardous conditions and actions that could damage the Meter or the equipment under test. These voltages pose a shock hazard. • ... or between any terminal and earth ground. • Use caution with voltages above 30 V ac rms, 42 V ac peak, or 60 V dc. If in this manual or the protection provided by the Meter might be impaired. • Do not use the Meter or test leads if they appear damaged, or if...
A "WCaution" statement identifies conditions and actions that could cause bodily harm or death. 114, 115, and 117 Users Manual Safety Information A "XWWarning" statement identifies hazardous conditions and actions that could damage the Meter or the equipment under test. These voltages pose a shock hazard. • ... or between any terminal and earth ground. • Use caution with voltages above 30 V ac rms, 42 V ac peak, or 60 V dc. If in this manual or the protection provided by the Meter might be impaired. • Do not use the Meter or test leads if they appear damaged, or if...
Fluke 114,115,117 Manual
Page 6
The Meter function is set to Diode Test Input is in the VoltAlert™ non-contact voltage detect mode. 114, 115, and 117 Users Manual Display No. The Meter function is set to Continuity. X Unsafe voltage. edy02f.eps Model 117 114, 115, & 117 115 & 117 114, 115, & 117 114, 115, & 117 Measured input voltage ≥30 V, or voltage overload condition (OL). A B C D E 4 Symbol w s R O Y 6 5 4 3 2 1 16 15 VoltAlert 7 14 17 8 9 10 11 12 13 18 Meaning The Meter is a negative value.
The Meter function is set to Diode Test Input is in the VoltAlert™ non-contact voltage detect mode. 114, 115, and 117 Users Manual Display No. The Meter function is set to Continuity. X Unsafe voltage. edy02f.eps Model 117 114, 115, & 117 115 & 117 114, 115, & 117 114, 115, & 117 Measured input voltage ≥30 V, or voltage overload condition (OL). A B C D E 4 Symbol w s R O Y 6 5 4 3 2 1 16 15 VoltAlert 7 14 17 8 9 10 11 12 13 18 Meaning The Meter is a negative value.
Fluke 114,115,117 Manual
Page 7
... switch is in the Auto Volts function. 114 & 117 Auto Autoranging. Display freezes present reading. 114, 115, & 117 G M MIN MAX AVG mode enabled. User sets the Meter's range. 114, 115, & 117 P + Bar graph polarity 114, 115, & 117 Q 0L W The input is too large for best resolution. 114, 115, & 117 Manual Manual ranging. VWX Maximum, minimum, average or present...
... switch is in the Auto Volts function. 114 & 117 Auto Autoranging. Display freezes present reading. 114, 115, & 117 G M MIN MAX AVG mode enabled. User sets the Meter's range. 114, 115, & 117 P + Bar graph polarity 114, 115, & 117 Q 0L W The input is too large for best resolution. 114, 115, & 117 Manual Manual ranging. VWX Maximum, minimum, average or present...
Fluke 114,115,117 Manual
Page 8
114, 115, and 117 Users Manual Terminals 1 V A COM 3 10 A FUSED 2 No. edy01f.eps Model 115 & 117 114, 115, & 117 114, 115, & 117 bAtt CAL Err EEPr Err F11 Err Error Messages Battery must be replaced before it will operate. C Input terminal for measuring ac and dc ...
114, 115, and 117 Users Manual Terminals 1 V A COM 3 10 A FUSED 2 No. edy01f.eps Model 115 & 117 114, 115, & 117 114, 115, & 117 bAtt CAL Err EEPr Err F11 Err Error Messages Battery must be replaced before it will operate. C Input terminal for measuring ac and dc ...
Fluke 114,115,117 Manual
Page 10
114, 115, and 117 Users Manual Battery Saver ("Sleep Mode") The Meter automatically enters "Sleep mode" and blanks the display ...f. To disable backlight auto-off, hold down Q while turning the Meter on and off after 40 seconds. Manual and Autoranging The Meter has both Manual and Autorange modes. • In the Autorange mode, the Meter selects the range with the best resolution. ...• In the Manual Range mode, you apply a different voltage. When a new high or low is displayed. • Press p to...
114, 115, and 117 Users Manual Battery Saver ("Sleep Mode") The Meter automatically enters "Sleep mode" and blanks the display ...f. To disable backlight auto-off, hold down Q while turning the Meter on and off after 40 seconds. Manual and Autoranging The Meter has both Manual and Autorange modes. • In the Autorange mode, the Meter selects the range with the best resolution. ...• In the Manual Range mode, you apply a different voltage. When a new high or low is displayed. • Press p to...
Fluke 114,115,117 Manual
Page 12
114, 115, and 117 Users Manual Measuring Resistance Testing for Continuity XWWarning edy04f.eps To avoid electric shock, injury, or damage to the Meter, disconnect circuit power and discharge all highvoltage capacitors before testing resistance, continuity, diodes, or capacitance. 10 edy06f.eps Note The continuity function works best as a fast, convenient method to check for opens and shorts. For maximum accuracy in making resistance measurements, use the Meter's resistance (e) function.
114, 115, and 117 Users Manual Measuring Resistance Testing for Continuity XWWarning edy04f.eps To avoid electric shock, injury, or damage to the Meter, disconnect circuit power and discharge all highvoltage capacitors before testing resistance, continuity, diodes, or capacitance. 10 edy06f.eps Note The continuity function works best as a fast, convenient method to check for opens and shorts. For maximum accuracy in making resistance measurements, use the Meter's resistance (e) function.
Fluke 114,115,117 Manual
Page 14
114, 115, and 117 Users Manual Measuring AC or DC Current (115 & 117) XWWarning To avoid personal injury or damage to the Meter: • Never attempt to make an in-circuit current measurement when the opencircuit potential to ...
114, 115, and 117 Users Manual Measuring AC or DC Current (115 & 117) XWWarning To avoid personal injury or damage to the Meter: • Never attempt to make an in-circuit current measurement when the opencircuit potential to ...
Fluke 114,115,117 Manual
Page 16
Select progressively lower ranges using manual ranging for frequencies >1 kHz. The trigger level is >1 kHz, the bar graph and Z are unspecified. Detecting AC Voltage Presence (117 only) VoltAlert Hot Neutral VoltAlert RANGE Lo edy09f.eps The Meter measures the frequency of a signal by ... the number of the measured signal is 0 V, 0 A for all ranges. 14 VoltAlert edy13f.eps 114, 115, and 117 Users Manual Measuring Frequency (115 & 117 only) XWWarning To avoid electrical shock, disregard the bar graph for a stable reading. AC Voltage Frequency AC Current Frequency VoltAlert ...
Select progressively lower ranges using manual ranging for frequencies >1 kHz. The trigger level is >1 kHz, the bar graph and Z are unspecified. Detecting AC Voltage Presence (117 only) VoltAlert Hot Neutral VoltAlert RANGE Lo edy09f.eps The Meter measures the frequency of a signal by ... the number of the measured signal is 0 V, 0 A for all ranges. 14 VoltAlert edy13f.eps 114, 115, and 117 Users Manual Measuring Frequency (115 & 117 only) XWWarning To avoid electrical shock, disregard the bar graph for a stable reading. AC Voltage Frequency AC Current Frequency VoltAlert ...
Fluke 114,115,117 Manual
Page 18
An input of the selected range. Testing the Fuse (115 & 117 only) Test fuse as shown below ), the major divisions on the scale represent 0, 15, 30, 45, and 60 V. In the 60 V range, for making peak .... Because the bar graph is much faster than the digital display, the bar graph is like the needle on an analog meter. 114, 115, and 117 Users Manual Using the Bargraph The bar graph is useful for example (see below . The bar graph is relative to the full-scale value of −30...
An input of the selected range. Testing the Fuse (115 & 117 only) Test fuse as shown below ), the major divisions on the scale represent 0, 15, 30, 45, and 60 V. In the 60 V range, for making peak .... Because the bar graph is much faster than the digital display, the bar graph is like the needle on an analog meter. 114, 115, and 117 Users Manual Using the Bargraph The bar graph is useful for example (see below . The bar graph is relative to the full-scale value of −30...
Fluke 114,115,117 Manual
Page 20
... holster. Finally, insert the Meter into its holder and replace it with a damp cloth and mild detergent. Remove the Meter from the Meter 2. Use only Fluke PN 803293. 6. Cleaning Wipe the case with an 11 A, 1000 V, FAST fuse having a minimum interrupt rating of 17,000 A. Do not use abrasives, isopropyl alcohol..., first attach the case bottom to clean the case or lens/window. Dirt or moisture in the terminals can affect readings. 18 114, 115, and 117 Users Manual To open the case for fuse replacement: 1.
... holster. Finally, insert the Meter into its holder and replace it with a damp cloth and mild detergent. Remove the Meter from the Meter 2. Use only Fluke PN 803293. 6. Cleaning Wipe the case with an 11 A, 1000 V, FAST fuse having a minimum interrupt rating of 17,000 A. Do not use abrasives, isopropyl alcohol..., first attach the case bottom to clean the case or lens/window. Dirt or moisture in the terminals can affect readings. 18 114, 115, and 117 Users Manual To open the case for fuse replacement: 1.
Fluke 114,115,117 Manual
Page 22
114, 115, and 117 Users Manual Certifications UL, P, CSA, TÜV, ; (N10140), VDE IP Rating (dust and water protection IP42 Table 1. Accuracy Specifications Function DC millivolts DC Volts Range 600.0 mV 6.000 V ... 0.001 V 0.01 V 0.1 V 0.1 V 0.1 mV 0.5 % + 2 0.5 % + 2 DC, 45 to 500 Hz 2.0 % + 3 45 to 500 Hz 1.0 % + 3 500 Hz to 1 kHz 4.0 % + 3 500 Hz to 1 kHz 2.0 % + 3 0.001 V 0.01 V 0.1 V 1.0 % + 3 2.0 % + 3 Model 114, 115, 117 114, 115, 117 114, 117 114, 115, 117 114, 115, 117 20
114, 115, and 117 Users Manual Certifications UL, P, CSA, TÜV, ; (N10140), VDE IP Rating (dust and water protection IP42 Table 1. Accuracy Specifications Function DC millivolts DC Volts Range 600.0 mV 6.000 V ... 0.001 V 0.01 V 0.1 V 0.1 V 0.1 mV 0.5 % + 2 0.5 % + 2 DC, 45 to 500 Hz 2.0 % + 3 45 to 500 Hz 1.0 % + 3 500 Hz to 1 kHz 4.0 % + 3 500 Hz to 1 kHz 2.0 % + 3 0.001 V 0.01 V 0.1 V 1.0 % + 3 2.0 % + 3 Model 114, 115, 117 114, 115, 117 114, 117 114, 115, 117 114, 115, 117 20
Fluke 114,115,117 Manual
Page 24
...decreasing linearly to 1.5 at full scale. AutoV LoZ, AC mV, and AC amps are dc-coupled. [2] AC Volts Hz is ac-coupled. 114, 115, and 117 Users Manual Table 1 Accuracy Specifications (cont.) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model AC Amps Truerms[1] (45 Hz to 500 Hz) 6.000 ...A 10.00 A[3] 20 A for 30 seconds max., 10 minutes rest min. 0.001 A 0.01 A 1.5 % + 3 115, 117 DC Amps 6.000 A 10.00 A[3] 20 A for this and other true-rms meters to display nonzero readings when the test leads are disconnected from 45...
...decreasing linearly to 1.5 at full scale. AutoV LoZ, AC mV, and AC amps are dc-coupled. [2] AC Volts Hz is ac-coupled. 114, 115, and 117 Users Manual Table 1 Accuracy Specifications (cont.) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model AC Amps Truerms[1] (45 Hz to 500 Hz) 6.000 ...A 10.00 A[3] 20 A for 30 seconds max., 10 minutes rest min. 0.001 A 0.01 A 1.5 % + 3 115, 117 DC Amps 6.000 A 10.00 A[3] 20 A for this and other true-rms meters to display nonzero readings when the test leads are disconnected from 45...
Fluke 117 Electrician Multimeter Datasheet
Page 2
... for ac current. Fluke Corporation 117 Electrician's Multimeter with Non-Contact Voltage Ordering information Fluke-117 Electrician's Multimeter with Non-Contact Voltage Included TL75 Test leads, holster, User's manual and 9V battery (installed). Fluke Corporation PO Box 9090, Everett, WA USA 98206 Fluke Europe B.V. Fluke. All rights reserved. Ohms...this and other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com ©2006 Fluke Corporation. detects opens or shorts of range are not specified, it is ac-coupled. For volts, crest factor ...
... for ac current. Fluke Corporation 117 Electrician's Multimeter with Non-Contact Voltage Ordering information Fluke-117 Electrician's Multimeter with Non-Contact Voltage Included TL75 Test leads, holster, User's manual and 9V battery (installed). Fluke Corporation PO Box 9090, Everett, WA USA 98206 Fluke Europe B.V. Fluke. All rights reserved. Ohms...this and other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com ©2006 Fluke Corporation. detects opens or shorts of range are not specified, it is ac-coupled. For volts, crest factor ...