Fluke 114,115,117 Manual
Page 1
® 114, 115, and 117 True-rms Multimeters Users Manual PN 2572573 July 2006, Rev. 1, 2/07 © 2006, 2007 Fluke Corporation, All rights reserved. Printed in China All product names are trademarks of their respective companies.
® 114, 115, and 117 True-rms Multimeters Users Manual PN 2572573 July 2006, Rev. 1, 2/07 © 2006, 2007 Fluke Corporation, All rights reserved. Printed in China All product names are trademarks of their respective companies.
Fluke 114,115,117 Manual
Page 3
... Meter at the distribution level. True-rms Multimeters Introduction The Fluke Model 114, Model 115, and Model 117 are designed to protect against transients in the world: +1-425-446-5500 Visit Fluke's web site at www.fluke.com. Contacting Fluke To contact Fluke, call: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31...
... Meter at the distribution level. True-rms Multimeters Introduction The Fluke Model 114, Model 115, and Model 117 are designed to protect against transients in the world: +1-425-446-5500 Visit Fluke's web site at www.fluke.com. Contacting Fluke To contact Fluke, call: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31...
Fluke 114,115,117 Manual
Page 5
... the test leads for disposal. 3 Contact Fluke or a D AC and DC qualified recycler for continuity before use the Auto Volts function to manual J Earth ground ~ Do not dispose of this product as required by this function's low input impedance (≈3 kΩ)(114 and 117 only). True-rms Multimeters Safety Information • Comply with local...
... the test leads for disposal. 3 Contact Fluke or a D AC and DC qualified recycler for continuity before use the Auto Volts function to manual J Earth ground ~ Do not dispose of this product as required by this function's low input impedance (≈3 kΩ)(114 and 117 only). True-rms Multimeters Safety Information • Comply with local...
Fluke 114,115,117 Manual
Page 7
.... 114, 115, & 117 N (Bar graph) Analog display. 114, 115, & 117 O Auto Volts The Meter is rotated to or from any A position. 115 & 117 5 Briefly displayed whenever the Meter's function switch is in the Auto Volts function. 114 & 117 Auto Autoranging. Display freezes present reading. 114, 115, & 117 G M MIN MAX AVG mode enabled. True-rms Multimeters Display F K Display...
.... 114, 115, & 117 N (Bar graph) Analog display. 114, 115, & 117 O Auto Volts The Meter is rotated to or from any A position. 115 & 117 5 Briefly displayed whenever the Meter's function switch is in the Auto Volts function. 114 & 117 Auto Autoranging. Display freezes present reading. 114, 115, & 117 G M MIN MAX AVG mode enabled. True-rms Multimeters Display F K Display...
Fluke 114,115,117 Manual
Page 9
DC voltage from 0.1 Ω to 40 MΩ. DC voltage from 0.1 to 600 mV. 114 & 117 114, 115 & 117 115 & 117 114, 115 & 117 114, 115 & 117 Ω Ohms from .001 V to 9999 μF. DC-coupled. True-rms Multimeters Rotary Switch Positions Rotary Switch Positions Switch Position Measurement Function Model x e Hz (button) D l Automatically selects ac or dc volts based...
DC voltage from 0.1 Ω to 40 MΩ. DC voltage from 0.1 to 600 mV. 114 & 117 114, 115 & 117 115 & 117 114, 115 & 117 114, 115 & 117 Ω Ohms from .001 V to 9999 μF. DC-coupled. True-rms Multimeters Rotary Switch Positions Rotary Switch Positions Switch Position Measurement Function Model x e Hz (button) D l Automatically selects ac or dc volts based...
Fluke 114,115,117 Manual
Page 11
... common (COM) test lead before removing the common test lead. 9 If you turn the Meter off . The Meter returns to Autorange and Auto is displayed. 1. True-rms Multimeters Power-Up Options Button Power-Up Options f p q g Q Turns on the following table while turning the Meter on. When you turn the Meter on, it...
... common (COM) test lead before removing the common test lead. 9 If you turn the Meter off . The Meter returns to Autorange and Auto is displayed. 1. True-rms Multimeters Power-Up Options Button Power-Up Options f p q g Q Turns on the following table while turning the Meter on. When you turn the Meter on, it...
Fluke 114,115,117 Manual
Page 13
... AC Volts DC V True-rms Multimeters Making Basic Measurements This function also sets the Meter's input impedance to approximately 3 kΩ to reduce the possibility of false readings due to dc millivolts. 11 Measuring AC and DC Millivolts Millivolts AC Millivolts DC DC edy03f.eps Using Auto Volts Selection (114 & 117 only) With the...
... AC Volts DC V True-rms Multimeters Making Basic Measurements This function also sets the Meter's input impedance to approximately 3 kΩ to reduce the possibility of false readings due to dc millivolts. 11 Measuring AC and DC Millivolts Millivolts AC Millivolts DC DC edy03f.eps Using Auto Volts Selection (114 & 117 only) With the...
Fluke 114,115,117 Manual
Page 15
Refer to measure currents that exceed the rating of the Meter can be used with an optional mV/A output Current Probe to a Fluke catalog or contact your local Fluke representative for your current probe. Make sure the Meter has the correct function selected, AC or DC, for compatible current clamps. True-rms Multimeters Making Basic Measurements Measuring Capacitance (115 & 117 only) edy14f.eps edy05f.eps 13 Measuring Current above 10 Amps The millivolt and voltage function of the Meter.
Refer to measure currents that exceed the rating of the Meter can be used with an optional mV/A output Current Probe to a Fluke catalog or contact your local Fluke representative for your current probe. Make sure the Meter has the correct function selected, AC or DC, for compatible current clamps. True-rms Multimeters Making Basic Measurements Measuring Capacitance (115 & 117 only) edy14f.eps edy05f.eps 13 Measuring Current above 10 Amps The millivolt and voltage function of the Meter.
Fluke 114,115,117 Manual
Page 17
..., insulation thickness and type. XW Warning If there is recessed within the connector itself. Making Low Impedance Capacitance Measurements (115 & 117 only) For making capacitance measurements on cables with ghost voltage, hold qwhile turning on other styles of the Meter close to switch ... as visual indication when voltage is turned off or goes into LoZ, (low input impedance) Capacitance mode. True-rms Multimeters Making Basic Measurements Testing Diodes (115 & 117) Good Diode Good Diode Forward Bias Single Beep Bad Diode Reverse Bias Bad Diode Open and Shorted edy07f.eps...
..., insulation thickness and type. XW Warning If there is recessed within the connector itself. Making Low Impedance Capacitance Measurements (115 & 117 only) For making capacitance measurements on cables with ghost voltage, hold qwhile turning on other styles of the Meter close to switch ... as visual indication when voltage is turned off or goes into LoZ, (low input impedance) Capacitance mode. True-rms Multimeters Making Basic Measurements Testing Diodes (115 & 117) Good Diode Good Diode Forward Bias Single Beep Bad Diode Reverse Bias Bad Diode Open and Shorted edy07f.eps...
Fluke 114,115,117 Manual
Page 19
... from the Meter 2. Install and tighten battery door screw. To remove the battery door for battery replacement: 1. Do not attempt to lift the door slightly. 4. True-rms Multimeters Maintenance Fuse edy11f.eps 17 Replacing the Battery and Fuse XWWarning To avoid shock, injury, or damage to separate it is fully seated. Maintenance...
... from the Meter 2. Install and tighten battery door screw. To remove the battery door for battery replacement: 1. Do not attempt to lift the door slightly. 4. True-rms Multimeters Maintenance Fuse edy11f.eps 17 Replacing the Battery and Fuse XWWarning To avoid shock, injury, or damage to separate it is fully seated. Maintenance...
Fluke 114,115,117 Manual
Page 21
...Alkaline, NEDA 1604A / IEC 6LR61 Battery Life Alkaline: 400 hours typical, without backlight Safety Compliances Complies with relative humidity at www.Fluke.com. True-rms Multimeters General Specifications General Specifications Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 ... CAN/CSA-C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN 61010-1 2nd Edition for A input (115 & 117 only 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 ...
...Alkaline, NEDA 1604A / IEC 6LR61 Battery Life Alkaline: 400 hours typical, without backlight Safety Compliances Complies with relative humidity at www.Fluke.com. True-rms Multimeters General Specifications General Specifications Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 ... CAN/CSA-C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN 61010-1 2nd Edition for A input (115 & 117 only 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 ...
Fluke 114,115,117 Manual
Page 22
...[1] True-rms 600.0 V AC millivolts[1] Truerms AC Volts[1] True-rms 600.0 mV 6.000 V 60.00 V 600.0 V Resolution Accuracy ± ([% of Reading] + [Counts]) 0.1 mV 0.001 V 0.01 V 0.1 V 0.1 V 0.1 mV 0.5 % + 2 0.5 % + 2 DC, 45 to 500 Hz 2.0 % + 3 45 to 500 Hz 1.0 % + 3 500 Hz to 1 kHz 4.0 % + 3 500 Hz to 1 kHz 2.0 % + 3 0.001 V 0.01 V 0.1 V 1.0 % + 3 2.0 % + 3 Model 114, 115, 117 114, 115, 117 114, 117 114, 115, 117 114...
...[1] True-rms 600.0 V AC millivolts[1] Truerms AC Volts[1] True-rms 600.0 mV 6.000 V 60.00 V 600.0 V Resolution Accuracy ± ([% of Reading] + [Counts]) 0.1 mV 0.001 V 0.01 V 0.1 V 0.1 V 0.1 mV 0.5 % + 2 0.5 % + 2 DC, 45 to 500 Hz 2.0 % + 3 45 to 500 Hz 1.0 % + 3 500 Hz to 1 kHz 4.0 % + 3 500 Hz to 1 kHz 2.0 % + 3 0.001 V 0.01 V 0.1 V 1.0 % + 3 2.0 % + 3 Model 114, 115, 117 114, 115, 117 114, 117 114, 115, 117 114...
Fluke 114,115,117 Manual
Page 23
detects opens or shorts of Reading] + [Counts]) Beeper on < 20 Ω, off > 250 Ω; Function Continuity Ohms Diode test Capacitance Lo-Z Capacitance (Power-up option) True-rms Multimeters General Specifications Table 1 Accuracy Specifications (cont.) Range 600 Ω 600.0 Ω 6.000 kΩ 60.00 kΩ 600.0 kΩ 6.000 MΩ 40.00...μF 1 nF 0.01 μF 0.1 μF 1 μF 1.9 % + 2 1.9 % + 2 1.9 % + 2 100 μF - 1000 μF: 1.9 % + 2 > 1000 μF: 5 % + 20 1 nF to 500 μF 10% + 2 typical Model 114, 115, 117 114, 115, 117 115, 117 115, 117 115, 117 21
detects opens or shorts of Reading] + [Counts]) Beeper on < 20 Ω, off > 250 Ω; Function Continuity Ohms Diode test Capacitance Lo-Z Capacitance (Power-up option) True-rms Multimeters General Specifications Table 1 Accuracy Specifications (cont.) Range 600 Ω 600.0 Ω 6.000 kΩ 60.00 kΩ 600.0 kΩ 6.000 MΩ 40.00...μF 1 nF 0.01 μF 0.1 μF 1 μF 1.9 % + 2 1.9 % + 2 1.9 % + 2 100 μF - 1000 μF: 1.9 % + 2 > 1000 μF: 5 % + 20 1 nF to 500 μF 10% + 2 typical Model 114, 115, 117 114, 115, 117 115, 117 115, 117 115, 117 21
Fluke 114,115,117 Manual
Page 24
...volts, crest factor of ≤3. For amps, crest factor of ≤3 at 4000 counts, decreasing linearly to 1.5 at full scale. 114, 115, and 117 Users Manual Table 1 Accuracy Specifications (cont.) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model AC Amps Truerms[1] (45 Hz to 500... A 10.00 A[3] 20 A for 30 seconds max., 10 minutes rest min. 0.001 A 0.01 A 1.5 % + 3 115, 117 DC Amps 6.000 A 10.00 A[3] 20 A for this and other true-rms meters to display nonzero readings when the test leads are disconnected from a circuit or are specified from 1 % to 100 % of...
...volts, crest factor of ≤3. For amps, crest factor of ≤3 at 4000 counts, decreasing linearly to 1.5 at full scale. 114, 115, and 117 Users Manual Table 1 Accuracy Specifications (cont.) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model AC Amps Truerms[1] (45 Hz to 500... A 10.00 A[3] 20 A for 30 seconds max., 10 minutes rest min. 0.001 A 0.01 A 1.5 % + 3 115, 117 DC Amps 6.000 A 10.00 A[3] 20 A for this and other true-rms meters to display nonzero readings when the test leads are disconnected from a circuit or are specified from 1 % to 100 % of...
Fluke 117 Electrician Multimeter Datasheet
Page 1
... [ % of 18 °C to 28 °C, with Non-Contact Voltage Technical Data Compact true-rms meter for commercial applications The Fluke 117 is specified for 1 year after calibration, at 0 % to 95 %. Fluke 117 Electrician's Multimeter with relative humidity at operating temperatures of reading ] + [ counts ] ) Maximum... LoZ: helps prevent false readings due to ghost voltage • Large white LED backlight to work in poorly lit areas • True-rms for accurate measurements on non-linear loads • Measures 20 A (30 seconds momentary; 10 A continuous) • Resistance, continuity...
... [ % of 18 °C to 28 °C, with Non-Contact Voltage Technical Data Compact true-rms meter for commercial applications The Fluke 117 is specified for 1 year after calibration, at 0 % to 95 %. Fluke 117 Electrician's Multimeter with relative humidity at operating temperatures of reading ] + [ counts ] ) Maximum... LoZ: helps prevent false readings due to ghost voltage • Large white LED backlight to work in poorly lit areas • True-rms for accurate measurements on non-linear loads • Measures 20 A (30 seconds momentary; 10 A continuous) • Resistance, continuity...
Fluke 117 Electrician Multimeter Datasheet
Page 2
..., it is dc coupled, 45 Hz to 1.5 at 4000 counts, decreasing linearly to 5 kHz for ac current. Fluke Corporation 117 Electrician's Multimeter with Non-Contact Voltage Ordering information Fluke-117 Electrician's Multimeter with Non-Contact Voltage Included TL75 Test leads, holster, User's manual and 9V battery (installed). All rights reserved...Fax (425) 446-5116 In Europe/M-East/Africa (31 40) 2 675 200 or Fax (31 40) 2 675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other true-rms meters to display non-zero readings when the test leads are shorted together.
..., it is dc coupled, 45 Hz to 1.5 at 4000 counts, decreasing linearly to 5 kHz for ac current. Fluke Corporation 117 Electrician's Multimeter with Non-Contact Voltage Ordering information Fluke-117 Electrician's Multimeter with Non-Contact Voltage Included TL75 Test leads, holster, User's manual and 9V battery (installed). All rights reserved...Fax (425) 446-5116 In Europe/M-East/Africa (31 40) 2 675 200 or Fax (31 40) 2 675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other true-rms meters to display non-zero readings when the test leads are shorted together.