CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 1

CS547A Conductivity and Temperature Probe and A547 Interface Revision: 10/11 Copyright © 1994-2011 Campbell Scientific, Inc.
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 6

... and CS547A Conductivity and Temperature Probe ...... 2 4-1. Long Lead Lengths Temperature 20 13. Polynomial Coefficients 19 ii Temperature , Resistance, and Datalogger Output 19 10-2. Therm107 / P11 Instruction Details 18 11. Thermistor Interchangeability Specification Temperature 17 8.2-2. Error produced by polynomial fit to published values 17 13-1. CS547A Conductivity and Temperature...
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 7

... effects. Authoritative sources to EC not yet compensated for measuring the electrical conductivity, dissolved solids, and temperature of the unit. 1 This electrode configuration eliminates the ground looping problems associated with Campbell Scientific dataloggers. Exceptions include the CR200-series, the BDR301 and BDR320 loggers which are typical calibration compounds, and to consult...
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 9

... parallel to the direction of rapid changes in EC requires that the probe be avoided. Wiring WARNING The voltage excitation channel used for temperature or measurement errors will be flushed continuously. This is recommended for each conductivity excitation must be separate from Campbell Scientific as leaving it in the sun and then submersing it in a cold...
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 10

...specific application's wiring and programming needs. CS547A Conductivity and Temperature Probe...temperature in this manual are as follows: the LO COND lead is computed from the results of the BrFull (P6) or BrHalf (P5) instructions (chosen automatically as part of four parts: 1. CS547A wiring diagram for temperature 5.1 Programming Overview Typical datalogger programs to measure the CS547A...
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 11

... when the temperature coefficient is not possible, use the Sample instruction in CRBasic or Instruction 70 in standard...errors in Edlog. Output processing Over large ranges, EC is not linear and is best to 3% of 2%/°C as a rough estimate. 4. CS547A Conductivity and Temperature Probe and A547 Interface 2. Campbell Scientific has developed a linear correction for conductivity...
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 12

CS547A Conductivity and Temperature Probe and A547 Interface PROGRAM BeginProg 'evaluate and edit each of these 3 user specific values Rcable=25 'edit this value to the actual footage of cable on your sensor CellConstant=1.50 'edit this value with the Cell Constant (Kc) printed 'on the label of each sensor TempCoef=2 'see section 9 of the manual...'Subtract resistance errors (Rp) caused...
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 13

CS547A Conductivity and Temperature Probe and A547 Interface 'correct errors in the EC measurement due to temperature Therm107 (TempDeg_C,1,3,Vx2,0,250,1,0) C25mScm = (Ct * 100)/(((TempDeg_C-25) * TempCoef) + 100) 'end scan loop by calling output table CallTable ECSample NextScan EndProg 5.2.2 Edlog *Table 1 Program ...
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 15

... (P37) 1: 2 X Loc [ one_ovrRs ] 2: nnn F 3: 3 Z Loc [ Ct ] Enter cell constant. ; ;The following program set corrects for errors of ionization in feet. 18: Z=X*F (P37) 1: 5 X Loc [ Rp ] 2: .00032 F 3: 5 Z Loc [ Rp ] 19...length in the EC ;measurement. ; 9 CS547A Conductivity and Temperature Probe and A547 Interface ; ;Subtract resistance errors (Rp) caused by the cell constant....
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 16

CS547A Conductivity and Temperature Probe and A547 Interface 24: IF (XF) (P89) 1: 3 X Loc [ Ct ] 2: 4 < 3: .474 F 4: 30 Then Do 25: Z=X*F ...02846 C2 7: .000000 C3 8: .000000 C4 9: .000000 C5 29: End (P95) ;This next program set will correct errors in the EC measurement resulting ;from temperature differences. ; 30: Temp (107) (P11) 1: 1 Reps 2: 3 SE Channel 3: 2 Excite all ...
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 17

CS547A Conductivity and Temperature Probe and A547 Interface 33: Z=X*F (P37) 1: 6 X Loc [ A ] 2: nnn F 3: 8Z Loc [ TC_Proces ] Enter TC (%/°C) to correct cond. reading. 34: Z=X+F (P34) 1: 8 X Loc [ TC_Proces ] 2: 100 F 3: 8 Z Loc [ TC_Proces ] 35: Z=X/Y (P38) 1: 7 2: 8 3: 9 X Loc [ Ct100 ] Y Loc [ TC_Proces ] Z Loc [ C25mScm_l ] EC corrected for temperature. ;Output ...
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 21

...KCl and Na+ Solutions After Correction: < 2.0%, 0.45 to 7.0 mS cm-1 < 8.0%, 0.005 to compensate for ionization effects on the measurements. Bridge Measurement Error: < 1.0% 2. CS547A Conductivity and Temperature Probe and A547 Interface 11: Polynomial (P55) 1: 1 Rep 2: 3 X Loc [T25_01] 3: 4 F(X) Loc [f_of_T ] 4: .99124 C0 5: -1.8817 C1 6: 3.4789 C2 7: -3.51 C3 8: -1.2 C4...
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 22

FIGURE 8.1-1. Plot of ideal and actual correction between 0 and 0.44 mS cm-1 FIGURE 8.1-2. Plot of 0.0234, 0.07, 0.4471, 07, 1.413, 2.070, 3.920, and 7.0 mS cm-1. CS547A Conductivity and Temperature Probe and A547 Interface derived by measuring the standard solutions described in Section 2.2 with values of ideal and actual correction between 0.44 and 7.0 mS cm-1 16
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 23

... tolerance with an offset entered in Table 8.2-1. TABLE 8.2-1. For the range of 0° to +50 Tolerance (±°C) 0.40 0.40 0.32 0.25 0.20 TABLE 8.2-2. CS547A Conductivity and Temperature Probe and A547 Interface 8.2 Temperature Measurement Error The overall probe accuracy is the interchangeability specification of the thermistor, tabulated in the measurement instruction.
CS547A-L Conductivity/Temperature Probe and A547A Interface - Page 24

... from Section 5.2 in Edlog) outputs a precise 2 VAC excitation (4 V with temperature. CS547A Conductivity and Temperature Probe and A547 Interface 9. Place the CS547A in this section is the resistance of voltage measured to excitation voltage (Vs/Vx) which field measurements will be made. The instruction calculates the ratio of the thermistor. Resistance and datalogger...

Campbell Scientific CS547A Reviews

Do you have an experience with the Campbell Scientific CS547A that you would like to share?
Earn 750 points for your review!
We have not received any reviews for Campbell Scientific yet.

Popular Campbell Scientific CS547A Search Terms

The following terms are frequently used to search for Campbell Scientific CS547A support:

Ask a New Question
Use the box below to post a new question about Campbell Scientific CS547A.
Manuals / Documents
Download any of our Campbell Scientific CS547A manuals for free!

We have the following 1 documents available for the Campbell Scientific CS547A:
  • CS547A-L Conductivity/Temperature Probe and A547A Interface
Related Items
View other Campbell Scientific Electronics Accessories:

Points & Prizes
  • You can earn points for nearly everything you do on HelpOwl.com
  • You can trade in those points for gift cards at leading retailers such as Amazon.com and Walmart
  • It's that simple!
See How it Works
Create a Free Account

Campbell Scientific Manuals

Find free Campbell Scientific CS547A manuals and user guides available at ManualOwl.com. Try out our unique manual viewer allowing you to interact with manuals from directly within your browser!

Campbell Scientific Reviews

View thousands of Campbell Scientific CS547A user reviews and customer ratings available at ReviewOwl.com.

Contact Information

Complete Campbell Scientific customer service contact information including steps to reach representatives, hours of operation, customer support links and more from ContactHelp.com.

Scoreboard Ratings

See detailed Campbell Scientific customer service rankings, employee comments and much more from our sister site.