FE 725ex Users Manual
Page 13
...725Ex calibrator to equipment that is used within intrinsically safe circuits will generally come from intrinsic safety barriers that may be connected to the Model 725Ex terminals without compromising the intrinsic safety. If the Calibrator is damaged or something is designed as intrinsically safe...8226; AC72 alligator clips (one set) • Stackable alligator clip test leads (one set) • Fluke 725Ex CD-ROM (contains Fluke 725Ex Users Manual) • Fluke 725Ex CCD • Fluke 725Ex Safety Information • 4 AA Batteries (installed) • Hex Key, 5/64 in Ex Hazardous Areas...
...725Ex calibrator to equipment that is used within intrinsically safe circuits will generally come from intrinsic safety barriers that may be connected to the Model 725Ex terminals without compromising the intrinsic safety. If the Calibrator is damaged or something is designed as intrinsically safe...8226; AC72 alligator clips (one set) • Stackable alligator clip test leads (one set) • Fluke 725Ex CD-ROM (contains Fluke 725Ex Users Manual) • Fluke 725Ex CCD • Fluke 725Ex Safety Information • 4 AA Batteries (installed) • Hex Key, 5/64 in Ex Hazardous Areas...
FE 725ex Users Manual
Page 14
...(Ci) plus the capacitance of the cable in the intrinsically safe circuit. As an example, Fluke 725Ex CCD explains that Voc of the 725Ex calibrator Isc and the barrier Isc. when the circuit is powered by an intrinsic safety barrier, then the maximum circuit voltage used . Each... higher of terminals has a Voc and an Isc rating as shown on the 725Ex calibrator ratings for the specific terminals used for Hazardous (Classified) Locations Instrumentation Part 1: Intrinsic Safety. 4 Again, Fluke 725Ex CCD identifies the maximum capacitance (Ca) and maximum inductance (La) that capacitance ...
...(Ci) plus the capacitance of the cable in the intrinsically safe circuit. As an example, Fluke 725Ex CCD explains that Voc of the 725Ex calibrator Isc and the barrier Isc. when the circuit is powered by an intrinsic safety barrier, then the maximum circuit voltage used . Each... higher of terminals has a Voc and an Isc rating as shown on the 725Ex calibrator ratings for the specific terminals used for Hazardous (Classified) Locations Instrumentation Part 1: Intrinsic Safety. 4 Again, Fluke 725Ex CCD identifies the maximum capacitance (Ca) and maximum inductance (La) that capacitance ...