Fluke 725 Process Calibrator Datasheet
Page 1
... Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters • Measure/source1 pressure using any of 29 Fluke 700Pxx Pressure Modules • Source mA with simultaneous pressure measurement to conduct valve and I/P tests • Support flow meter testing with frequency... M 100.00 mV DC M or S 30.000 V DC measure M 20.000 V DC measure 10.000 VDC Source M or S 15 to 3200 Ohms M or S Specifications - 724 & 725 Calibrators Function Measure or Source Range or Type Voltage mA1 0 to 100 mV 0 to 10 V (source) 0 to 30 V (measure) 0 to 24 ...
... Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters • Measure/source1 pressure using any of 29 Fluke 700Pxx Pressure Modules • Source mA with simultaneous pressure measurement to conduct valve and I/P tests • Support flow meter testing with frequency... M 100.00 mV DC M or S 30.000 V DC measure M 20.000 V DC measure 10.000 VDC Source M or S 15 to 3200 Ohms M or S Specifications - 724 & 725 Calibrators Function Measure or Source Range or Type Voltage mA1 0 to 100 mV 0 to 10 V (source) 0 to 30 V (measure) 0 to 24 ...
FE 725ex Users Manual
Page 2
... will provide an estimate of repair costs and obtain authorization before commencing the work. Fluke Corporation P.O. Fluke does not warrant that service center, with its functional specifications for 90 days. To obtain warranty service, contact your product online, visit register.fluke.com Following warranty repair, the product will not affect the validity or enforceability...
... will provide an estimate of repair costs and obtain authorization before commencing the work. Fluke Corporation P.O. Fluke does not warrant that service center, with its functional specifications for 90 days. To obtain warranty service, contact your product online, visit register.fluke.com Following warranty repair, the product will not affect the validity or enforceability...
FE 725ex Users Manual
Page 5
Contents (continued) Service Center Calibration or Repair 50 Replacement Parts ...51 Accessories ...53 Specifications ...55 DC Voltage Measurement 55 DC Voltage Source...55 Millivolt Measurement and Source 55 DC mA Measurement and Source 56 Ohms Measurement 56 Ohms Source...56 Frequency Measurement 56 Frequency Source ...57 Temperature, Thermocouples 57 Loop Power Supply 57 RTD Excitation (simulation 58 Temperature, RTD Ranges, and Accuracies 58 Pressure Measurement 59 General Specifications 59 iii
Contents (continued) Service Center Calibration or Repair 50 Replacement Parts ...51 Accessories ...53 Specifications ...55 DC Voltage Measurement 55 DC Voltage Source...55 Millivolt Measurement and Source 55 DC mA Measurement and Source 56 Ohms Measurement 56 Ohms Source...56 Frequency Measurement 56 Frequency Source ...57 Temperature, Thermocouples 57 Loop Power Supply 57 RTD Excitation (simulation 58 Temperature, RTD Ranges, and Accuracies 58 Pressure Measurement 59 General Specifications 59 iii
FE 725ex Users Manual
Page 14
...on Fluke 725Ex CCD. Each set of the cable in Hazardous (Classified) Locations and to ANSI/ISARP12.06.01-2003 Recommended Practice for Wiring Methods for the terminals connected to other equipment, the Vmax and Imax ratings on the 725Ex calibrator. Similarly for the specific ...parameter (Voc) and a maximum short circuit current parameter (Isc). As an example, Fluke 725Ex CCD explains that is powered by an intrinsic safety barrier, then the maximum circuit voltage used . Again, Fluke 725Ex CCD identifies the maximum capacitance (Ca) and maximum inductance (La) that the capacitance ...
...on Fluke 725Ex CCD. Each set of the cable in Hazardous (Classified) Locations and to ANSI/ISARP12.06.01-2003 Recommended Practice for Wiring Methods for the terminals connected to other equipment, the Vmax and Imax ratings on the 725Ex calibrator. Similarly for the specific ...parameter (Voc) and a maximum short circuit current parameter (Isc). As an example, Fluke 725Ex CCD explains that is powered by an intrinsic safety barrier, then the maximum circuit voltage used . Again, Fluke 725Ex CCD identifies the maximum capacitance (Ca) and maximum inductance (La) that the capacitance ...
FE 725ex Users Manual
Page 65
...% + 2 Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ±0.005 % of range per °C Multifunction Process Calibrator Specifications DC Voltage Source Range Resolution Accuracy, (% of Reading + Counts) 100 mV 0.01 mV 0.02 % + 2 10 V 0.001 V 0.02 % + 2 Temperature ...;0.005 % of range per °C *Select this function by pressing T. The signal is available at the thermocouple miniplug connector. 55 Specifications All specifications apply from +18 °C to 55 °C: ±0.005 % of range per °C Maximum load: 1 mA Millivolt Measurement...
...% + 2 Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ±0.005 % of range per °C Multifunction Process Calibrator Specifications DC Voltage Source Range Resolution Accuracy, (% of Reading + Counts) 100 mV 0.01 mV 0.02 % + 2 10 V 0.001 V 0.02 % + 2 Temperature ...;0.005 % of range per °C *Select this function by pressing T. The signal is available at the thermocouple miniplug connector. 55 Specifications All specifications apply from +18 °C to 55 °C: ±0.005 % of range per °C Maximum load: 1 mA Millivolt Measurement...
FE 725ex Users Manual
Page 67
...;C 500 to 1750 °C Measure and Source Accuracies 1.0 °C 0.7 °C 1.2 °C 0.8 °C 1.2 °C 0.8 °C 0.9 °C 0.7 °C 2.5 °C 1.8 °C 1.4 °C 2.5 °C 1.8 °C 1.5 °C Multifunction Process Calibrator Specifications Type Range B 600 to 800 °C 800 to 1000 °C 1000 to 1800 °C L -200 to 0 °C 0 to 900 °C U -200 to 0 °C 0 to 400...
...;C 500 to 1750 °C Measure and Source Accuracies 1.0 °C 0.7 °C 1.2 °C 0.8 °C 1.2 °C 0.8 °C 0.9 °C 0.7 °C 2.5 °C 1.8 °C 1.4 °C 2.5 °C 1.8 °C 1.5 °C Multifunction Process Calibrator Specifications Type Range B 600 to 800 °C 800 to 1000 °C 1000 to 1800 °C L -200 to 0 °C 0 to 900 °C U -200 to 0 °C 0 to 400...
FE 725ex Users Manual
Page 69
... temperature Storage temperature Operating altitude Relative Humidity (% RH operating without condensation) Vibration Product Compliance Markings EMC Power requirements Size Weight Multifunction Process Calibrator Specifications Units psi, inH2O@4 °C, inH2O@20 °C, inH2O@60 °F, kPa, cmH2O@4 °C, cmH2O@20 °C, bar, mbar, kg/cm2, mmHg, inHg -10 °C to ...
... temperature Storage temperature Operating altitude Relative Humidity (% RH operating without condensation) Vibration Product Compliance Markings EMC Power requirements Size Weight Multifunction Process Calibrator Specifications Units psi, inH2O@4 °C, inH2O@20 °C, inH2O@60 °F, kPa, cmH2O@4 °C, cmH2O@20 °C, bar, mbar, kg/cm2, mmHg, inHg -10 °C to ...