Fluke 718ex Users Manual
Page 1
Printed in U.S.A. All product names are trademarks of their respective companies. Users Manual ® 718Ex 30G/100G Pressure Calibrator May 2004 © 2004 Fluke Corporation. All rights reserved.
Printed in U.S.A. All product names are trademarks of their respective companies. Users Manual ® 718Ex 30G/100G Pressure Calibrator May 2004 © 2004 Fluke Corporation. All rights reserved.
Fluke 718ex Users Manual
Page 4
718Ex 30G/100G Users Manual Calibration ...21 Replacing the Battery 22 Approved Batteries...23 Parts and Accessories 24 Specifications...25 Pressure Sensor Input, 718Ex 30G 25 Pressure Sensor Input, 718Ex 100G 25 Pressure Sensor Range and Resolution 25 Pressure Module Input, 718Ex 30G and 718Ex 100G 26 DC mA Input, 718Ex 30G and 718Ex 100G 26 General Specifications 27 Product Compliance Markings 27 ii
718Ex 30G/100G Users Manual Calibration ...21 Replacing the Battery 22 Approved Batteries...23 Parts and Accessories 24 Specifications...25 Pressure Sensor Input, 718Ex 30G 25 Pressure Sensor Input, 718Ex 100G 25 Pressure Sensor Range and Resolution 25 Pressure Module Input, 718Ex 30G and 718Ex 100G 26 DC mA Input, 718Ex 30G and 718Ex 100G 26 General Specifications 27 Product Compliance Markings 27 ii
Fluke 718ex Users Manual
Page 10
718Ex 30G/100G Users Manual measured for information about accessories. The Calibrator is missing, contact the place of the nearest Fluke distributor or Service Center, call: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +...65-738-5655 Anywhere in the world: +1-425-446-5500 Or, visit Fluke's Web site at www.fluke.com. See "Contacting Fluke." To order replacement parts or spares, see "Parts and Accessories." To register this manual and on the Calibrator are displayed in bar and kg/cm2 units. Contact...
718Ex 30G/100G Users Manual measured for information about accessories. The Calibrator is missing, contact the place of the nearest Fluke distributor or Service Center, call: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +...65-738-5655 Anywhere in the world: +1-425-446-5500 Or, visit Fluke's Web site at www.fluke.com. See "Contacting Fluke." To order replacement parts or spares, see "Parts and Accessories." To register this manual and on the Calibrator are displayed in bar and kg/cm2 units. Contact...
Fluke 718ex Users Manual
Page 11
Refer to relevant Canadian and US Standards. Double insulated Conforms to manual. Battery Hazardous Voltage Risk of Danger. International Electrical Symbols Meaning Power ON/OFF Earth ground Conforms to relevant European Union directives. Conforms to ATEX requirements. Important information. Pressure Pressure Calibrator Safety Information 3 Symbol + J ( M X W T ) P f Table 1.
Refer to relevant Canadian and US Standards. Double insulated Conforms to manual. Battery Hazardous Voltage Risk of Danger. International Electrical Symbols Meaning Power ON/OFF Earth ground Conforms to relevant European Union directives. Conforms to ATEX requirements. Important information. Pressure Pressure Calibrator Safety Information 3 Symbol + J ( M X W T ) P f Table 1.
Fluke 718ex Users Manual
Page 12
...measuring or sourcing application. • To prevent damage to the Calibrator: • Use the Calibrator only as described in this User Manual and the Fluke 718Ex CCD (Concept Control Drawing) or the protection provided by the Calibrator may be sure the Calibrator is closed before using probes, keep...probe before the COM probe. • Never use the Calibrator with the red holster removed. • Never open the Calibrator case. 718Ex 30G/100G Users Manual XW Warning To avoid electric shock, injury, or damage to the unit under test, be impaired. • Inspect the Calibrator before use...
...measuring or sourcing application. • To prevent damage to the Calibrator: • Use the Calibrator only as described in this User Manual and the Fluke 718Ex CCD (Concept Control Drawing) or the protection provided by the Calibrator may be sure the Calibrator is closed before using probes, keep...probe before the COM probe. • Never use the Calibrator with the red holster removed. • Never open the Calibrator case. 718Ex 30G/100G Users Manual XW Warning To avoid electric shock, injury, or damage to the unit under test, be impaired. • Inspect the Calibrator before use...
Fluke 718ex Users Manual
Page 14
718Ex 30G/100G Users Manual • To avoid overpressure damage, do not apply pressure to the internal pressure...of leakage: • Confirm that all pressure connections are properly sealed. • Confirm that exceeds the following: • Model 718Ex 30G: 30.000 psi, 206.85 kPa, or 2.0685 bar. W Caution To avoid mechanically damaging the Calibrator: • ... components. "OL" appears at 33 psi. • Model 718Ex 100G: 100.00 psi, 689.5 kPa, or 6.895 bar. See Figure 1 for leaks to confirm the integrity of the included Fluke 700-ILF In-Line Filter will help isolate the pump from ...
718Ex 30G/100G Users Manual • To avoid overpressure damage, do not apply pressure to the internal pressure...of leakage: • Confirm that all pressure connections are properly sealed. • Confirm that exceeds the following: • Model 718Ex 30G: 30.000 psi, 206.85 kPa, or 2.0685 bar. W Caution To avoid mechanically damaging the Calibrator: • ... components. "OL" appears at 33 psi. • Model 718Ex 100G: 100.00 psi, 689.5 kPa, or 6.895 bar. See Figure 1 for leaks to confirm the integrity of the included Fluke 700-ILF In-Line Filter will help isolate the pump from ...
Fluke 718ex Users Manual
Page 15
... in an Exhazardous area. Hold in fixed position Pressure Calibrator Safety Information Faults and Damage Applying a voltage greater than 30 V to the original English users manual. Figure 1. If there is any reason to suspect that the safe operation of the Calibrator has been affected, it must be immediately withdrawn from use... precautionary measures must be taken to prevent any further use of the Calibrator invalidates its Ex Approval and may impair its safe operation in this manual. Connection Technique wh001f.eps 7
... in an Exhazardous area. Hold in fixed position Pressure Calibrator Safety Information Faults and Damage Applying a voltage greater than 30 V to the original English users manual. Figure 1. If there is any reason to suspect that the safe operation of the Calibrator has been affected, it must be immediately withdrawn from use... precautionary measures must be taken to prevent any further use of the Calibrator invalidates its Ex Approval and may impair its safe operation in this manual. Connection Technique wh001f.eps 7
Fluke 718ex Users Manual
Page 16
... • Incorrect storage of the unit • Damage sustained in transit • Correct certification is illegible • Using the product with compatible entity parameters. 718Ex 30G/100G Users Manual The safety features and integrity of the unit may be compromised by the Calibrator • Opening the case 8 Safety Regulations The use of the...
... • Incorrect storage of the unit • Damage sustained in transit • Correct certification is illegible • Using the product with compatible entity parameters. 718Ex 30G/100G Users Manual The safety features and integrity of the unit may be compromised by the Calibrator • Opening the case 8 Safety Regulations The use of the...
Fluke 718ex Users Manual
Page 18
... pushbutton. The g symbol appears on and off. Press to read the minimum pressure and current readings since power was pressed. All units are not available. 718Ex 30G/100G Users Manual Table 2.
... pushbutton. The g symbol appears on and off. Press to read the minimum pressure and current readings since power was pressed. All units are not available. 718Ex 30G/100G Users Manual Table 2.
Fluke 718ex Users Manual
Page 20
... (clockwise) for pressure, backward (counter-clockwise) for partial release.) Rotate fully forward (clockwise) to close valve. Internal Pump Increase pressure on the outward stroke. 718Ex 30G/100G Users Manual ZERO MMINAX BACROLMRETHROICLADDJ. Fine Adjustment Knob Rotate either direction for precise adjustment of applied pressure or vacuum. Full rotation is about 30 turns. SWTITECSHT...
... (clockwise) for pressure, backward (counter-clockwise) for partial release.) Rotate fully forward (clockwise) to close valve. Internal Pump Increase pressure on the outward stroke. 718Ex 30G/100G Users Manual ZERO MMINAX BACROLMRETHROICLADDJ. Fine Adjustment Knob Rotate either direction for precise adjustment of applied pressure or vacuum. Full rotation is about 30 turns. SWTITECSHT...
Fluke 718ex Users Manual
Page 22
... the release valve. 8. Forward (clockwise) is in and out to Figure 3 and perform the following steps: 1. Work the pump handle in the desired position. 718Ex 30G/100G Users Manual To use the Calibrator's internal pump, refer to apply incrementally larger pressure/vacuum changes. Depressurize the system before connecting the Calibrator. 2. Turn the pressure...
... the release valve. 8. Forward (clockwise) is in and out to Figure 3 and perform the following steps: 1. Work the pump handle in the desired position. 718Ex 30G/100G Users Manual To use the Calibrator's internal pump, refer to apply incrementally larger pressure/vacuum changes. Depressurize the system before connecting the Calibrator. 2. Turn the pressure...
Fluke 718ex Users Manual
Page 26
Pressure Module with External Pump alz006f.eps 18 TEST Black P 718ExH1O0LD0G RCEASL ISBURRAET O R Pressure Module Reading DAMP 1 UNITS )LRAE11xID0.iSai4v.6IC1I0ClGaTsro4su1p A-D ZERO MMINAX BACROLMRETHROICLADDJ. SWTITECSHT mA 10R0ApNsGi E689kPa 3M0VAX FLUKEO7N0L0YPEx 200psi 1M3A7X9kPa Red External Loop Supply Figure 6. 718Ex 30G/100G Users Manual Pressure Module SIGNAL + -
Pressure Module with External Pump alz006f.eps 18 TEST Black P 718ExH1O0LD0G RCEASL ISBURRAET O R Pressure Module Reading DAMP 1 UNITS )LRAE11xID0.iSai4v.6IC1I0ClGaTsro4su1p A-D ZERO MMINAX BACROLMRETHROICLADDJ. SWTITECSHT mA 10R0ApNsGi E689kPa 3M0VAX FLUKEO7N0L0YPEx 200psi 1M3A7X9kPa Red External Loop Supply Figure 6. 718Ex 30G/100G Users Manual Pressure Module SIGNAL + -
Fluke 718ex Users Manual
Page 28
... will display CLOSE instead of the terminals does not matter. 2. OPEN is displayed once the switch is increased to reset the switch test. 20 718Ex 30G/100G Users Manual Switch Test To perform a switch test, follow these steps: Note This example used a normally closed , and for the deadband. The procedure is open and...
... will display CLOSE instead of the terminals does not matter. 2. OPEN is displayed once the switch is increased to reset the switch test. 20 718Ex 30G/100G Users Manual Switch Test To perform a switch test, follow these steps: Note This example used a normally closed , and for the deadband. The procedure is open and...
Fluke 718ex Users Manual
Page 29
In Case of pressure, review "Safety Information" earlier in this manual, or if the Calibrator needs repair, contact a Fluke Service Center. do not use abrasives or solvents. A calibration manual is under warranty, see the warranty statement for terms. If the warranty has... tubing. and Canada. Follow replacement and connection instructions properly. • Review this manual before proceeding. Pressure Calibrator Maintenance Cleaning Periodically wipe the case with a damp cloth; Calibration Fluke recommends that the Calibrator be repaired and returned for a fixed fee. If the ...
In Case of pressure, review "Safety Information" earlier in this manual, or if the Calibrator needs repair, contact a Fluke Service Center. do not use abrasives or solvents. A calibration manual is under warranty, see the warranty statement for terms. If the warranty has... tubing. and Canada. Follow replacement and connection instructions properly. • Review this manual before proceeding. Pressure Calibrator Maintenance Cleaning Periodically wipe the case with a damp cloth; Calibration Fluke recommends that the Calibrator be repaired and returned for a fixed fee. If the ...
Fluke 718ex Users Manual
Page 30
Battery Replacement alz008f.eps 22 When the B symbol appears on the display, replace the 9 V alkaline battery. Hex Head Wrench Figure 7. Refer to possible electric shock or personal injury, replace the battery as soon as the battery indicator B appears. • Remove the Calibrator from the Exhazardous area before opening the battery door. • Use only the battery types listed in the Approved Battery Table. 718Ex 30G/100G Users Manual Replacing the Battery XW Warning • To avoid false readings, which could lead to Figure 7.
Battery Replacement alz008f.eps 22 When the B symbol appears on the display, replace the 9 V alkaline battery. Hex Head Wrench Figure 7. Refer to possible electric shock or personal injury, replace the battery as soon as the battery indicator B appears. • Remove the Calibrator from the Exhazardous area before opening the battery door. • Use only the battery types listed in the Approved Battery Table. 718Ex 30G/100G Users Manual Replacing the Battery XW Warning • To avoid false readings, which could lead to Figure 7.
Fluke 718ex Users Manual
Page 32
... 6. 718Ex 30G/100G Users Manual Parts and Accessories Refer to Table 6 for a list of replacement parts and accessories. AC72 AC72 BT1 Description Alligator clips(Black) Alligator clips(Red) 9 V battery, ANSI/NEDA 1604A or IEC 6LR61 Holster 700-ILF Holster, Red Battery Door Assembly Test lead set 718Ex CD-ROM (contains Users Manual) 71X Series Calibration Manual 718Ex...
... 6. 718Ex 30G/100G Users Manual Parts and Accessories Refer to Table 6 for a list of replacement parts and accessories. AC72 AC72 BT1 Description Alligator clips(Black) Alligator clips(Red) 9 V battery, ANSI/NEDA 1604A or IEC 6LR61 Holster 700-ILF Holster, Red Battery Door Assembly Test lead set 718Ex CD-ROM (contains Users Manual) 71X Series Calibration Manual 718Ex...
Fluke 718ex Users Manual
Page 34
718Ex 30G/100G Users Manual Pressure Module Input, 718Ex 30G and 718Ex 100G Range Resolution Accuracy (determined by Pressure Module) DC mA Input, 718Ex 30G and 718Ex 100G Range Resolution Accuracy, ±(% of Reading + Counts) 24 mA 0.001 mA .02 + 2 Temperature coefficient: 0.005 % of range per °C for temperature ranges -10 °C to 18 °C and 28 °C to 55 °C 26
718Ex 30G/100G Users Manual Pressure Module Input, 718Ex 30G and 718Ex 100G Range Resolution Accuracy (determined by Pressure Module) DC mA Input, 718Ex 30G and 718Ex 100G Range Resolution Accuracy, ±(% of Reading + Counts) 24 mA 0.001 mA .02 + 2 Temperature coefficient: 0.005 % of range per °C for temperature ranges -10 °C to 18 °C and 28 °C to 55 °C 26