Fluke 114,115,117 Manual
Page 1
® 114, 115, and 117 True-rms Multimeters Users Manual PN 2572573 July 2006, Rev. 1, 2/07 © 2006, 2007 Fluke Corporation, All rights reserved. Printed in China All product names are trademarks of their respective companies.
® 114, 115, and 117 True-rms Multimeters Users Manual PN 2572573 July 2006, Rev. 1, 2/07 © 2006, 2007 Fluke Corporation, All rights reserved. Printed in China All product names are trademarks of their respective companies.
Fluke 114,115,117 Manual
Page 3
... Japan: +81-3-3434-0181 Singapore +65-738-5655 Anywhere in the world: +1-425-446-5500 Visit Fluke's web site at www.fluke.com. True-rms Multimeters Introduction The Fluke Model 114, Model 115, and Model 117 are battery-powered, true-rms multimeters (hereafter "the Meter") with a lead in an incorrect terminal. All figures show the Model...
... Japan: +81-3-3434-0181 Singapore +65-738-5655 Anywhere in the world: +1-425-446-5500 Visit Fluke's web site at www.fluke.com. True-rms Multimeters Introduction The Fluke Model 114, Model 115, and Model 117 are battery-powered, true-rms multimeters (hereafter "the Meter") with a lead in an incorrect terminal. All figures show the Model...
Fluke 114,115,117 Manual
Page 5
Contact Fluke or a D AC and DC qualified recycler for continuity before use. Symbols B AC (Alternating Current) F DC (Direct Current) X Hazardous voltage N Battery (Low battery when shown on the display.) I Fuse T Double Insulated W Important Information; True-rms Multimeters Safety Information • Comply with local and national...Use only the replacement fuse specified or the protection may be damaged by this function's low input impedance (≈3 kΩ)(114 and 117 only). Do not use if the readings are high or noisy. • Do not use the Auto Volts function to manual J ...
Contact Fluke or a D AC and DC qualified recycler for continuity before use. Symbols B AC (Alternating Current) F DC (Direct Current) X Hazardous voltage N Battery (Low battery when shown on the display.) I Fuse T Double Insulated W Important Information; True-rms Multimeters Safety Information • Comply with local and national...Use only the replacement fuse specified or the protection may be damaged by this function's low input impedance (≈3 kΩ)(114 and 117 only). Do not use if the readings are high or noisy. • Do not use the Auto Volts function to manual J ...
Fluke 114,115,117 Manual
Page 7
...selected range. 114, 115, & 117 R LEAd W Test lead alert. User sets the Meter's range. 114, 115, & 117 P + Bar graph polarity 114, 115, & 117 Q 0L W The input is too large for best resolution. 114, 115, & 117 Manual Manual ranging. VWX Maximum, ...117 K DC AC Direct current or alternating current 114, 115 & 117 L N Battery low warning. 114, 115, & 117 M 610000 mV Indicates the Meter's range selection. 114, 115, & 117 N (Bar graph) Analog display. 114, 115, & 117 O Auto Volts The Meter is rotated to or from any A position. 115 & 117 5 True-rms Multimeters...
...selected range. 114, 115, & 117 R LEAd W Test lead alert. User sets the Meter's range. 114, 115, & 117 P + Bar graph polarity 114, 115, & 117 Q 0L W The input is too large for best resolution. 114, 115, & 117 Manual Manual ranging. VWX Maximum, ...117 K DC AC Direct current or alternating current 114, 115 & 117 L N Battery low warning. 114, 115, & 117 M 610000 mV Indicates the Meter's range selection. 114, 115, & 117 N (Bar graph) Analog display. 114, 115, & 117 O Auto Volts The Meter is rotated to or from any A position. 115 & 117 5 True-rms Multimeters...
Fluke 114,115,117 Manual
Page 9
...from 6.0 to 20 A, 30 seconds on , 10 minutes off at >250 Ω. 114, 115 & 117 114, 115 & 117 R S j Hz (button) I w Diode Test. Displays OL above 2.0 V. True-rms Multimeters Rotary Switch Positions Rotary Switch Positions Switch Position Measurement Function Model x e Hz (button) D l Automatically selects ...V. Non-contact sensing of ac voltage. 115 & 117 115 & 117 115 & 117 115 & 117 117 Note: All ac functions and Auto-V LoZ are dc-coupled. 7 DC voltage from 0.1 to 600 mV. 114 & 117 114, 115 & 117 115 & 117 114, 115 & 117 114, 115 & 117 Ω Ohms from .001 V to 600 ...
...from 6.0 to 20 A, 30 seconds on , 10 minutes off at >250 Ω. 114, 115 & 117 114, 115 & 117 R S j Hz (button) I w Diode Test. Displays OL above 2.0 V. True-rms Multimeters Rotary Switch Positions Rotary Switch Positions Switch Position Measurement Function Model x e Hz (button) D l Automatically selects ...V. Non-contact sensing of ac voltage. 115 & 117 115 & 117 115 & 117 115 & 117 117 Note: All ac functions and Auto-V LoZ are dc-coupled. 7 DC voltage from 0.1 to 600 mV. 114 & 117 114, 115 & 117 115 & 117 114, 115 & 117 114, 115 & 117 Ω Ohms from .001 V to 600 ...
Fluke 114,115,117 Manual
Page 11
.... 1. When connecting the test leads to make basic measurements. When you turn the Meter on all display segments. Note You cannot manually change . 3. True-rms Multimeters Power-Up Options Button Power-Up Options f p q g Q Turns on , it defaults to Autorange and Auto is displayed when enabled. See page 14. Disables automatic power...
.... 1. When connecting the test leads to make basic measurements. When you turn the Meter on all display segments. Note You cannot manually change . 3. True-rms Multimeters Power-Up Options Button Power-Up Options f p q g Q Turns on , it defaults to Autorange and Auto is displayed when enabled. See page 14. Disables automatic power...
Fluke 114,115,117 Manual
Page 13
...to switch the Meter to ghost voltages. Measuring AC and DC Millivolts Millivolts AC Millivolts DC DC edy03f.eps Using Auto Volts Selection (114 & 117 only) With the function switch in the l position, the Meter measures ac plus dc millivolts. Measuring AC and DC Voltage Volts AC Volts ...DC V True-rms Multimeters Making Basic Measurements This function also sets the Meter's input impedance to approximately 3 kΩ to reduce the possibility of false readings due to...
...to switch the Meter to ghost voltages. Measuring AC and DC Millivolts Millivolts AC Millivolts DC DC edy03f.eps Using Auto Volts Selection (114 & 117 only) With the function switch in the l position, the Meter measures ac plus dc millivolts. Measuring AC and DC Voltage Volts AC Volts ...DC V True-rms Multimeters Making Basic Measurements This function also sets the Meter's input impedance to approximately 3 kΩ to reduce the possibility of false readings due to...
Fluke 114,115,117 Manual
Page 15
Refer to measure currents that exceed the rating of the Meter. True-rms Multimeters Making Basic Measurements Measuring Capacitance (115 & 117 only) edy14f.eps edy05f.eps 13 Measuring Current above 10 Amps The millivolt and voltage function of the Meter can be used with an optional mV/A output Current Probe to a Fluke catalog or contact your local Fluke representative for your current probe. Make sure the Meter has the correct function selected, AC or DC, for compatible current clamps.
Refer to measure currents that exceed the rating of the Meter. True-rms Multimeters Making Basic Measurements Measuring Capacitance (115 & 117 only) edy14f.eps edy05f.eps 13 Measuring Current above 10 Amps The millivolt and voltage function of the Meter can be used with an optional mV/A output Current Probe to a Fluke catalog or contact your local Fluke representative for your current probe. Make sure the Meter has the correct function selected, AC or DC, for compatible current clamps.
Fluke 114,115,117 Manual
Page 17
... used on other styles of the Meter close to switch the Meter into sleep mode. Making Low Impedance Capacitance Measurements (115 & 117 only) For making capacitance measurements on cables with ghost voltage, hold qwhile turning on the VoltAlert detector with voltages as low as ... insulation thickness and type. The "Lo" setting can be effected by differences in the "Hi" setting. True-rms Multimeters Making Basic Measurements Testing Diodes (115 & 117) Good Diode Good Diode Forward Bias Single Beep Bad Diode Reverse Bias Bad Diode Open and Shorted edy07f.eps 15 XW...
... used on other styles of the Meter close to switch the Meter into sleep mode. Making Low Impedance Capacitance Measurements (115 & 117 only) For making capacitance measurements on cables with ghost voltage, hold qwhile turning on the VoltAlert detector with voltages as low as ... insulation thickness and type. The "Lo" setting can be effected by differences in the "Hi" setting. True-rms Multimeters Making Basic Measurements Testing Diodes (115 & 117) Good Diode Good Diode Forward Bias Single Beep Bad Diode Reverse Bias Bad Diode Open and Shorted edy07f.eps 15 XW...
Fluke 114,115,117 Manual
Page 19
... into the case, bottom edge first, until it from the Meter 2. Do not attempt to separate it is then inserted into the case. 5. True-rms Multimeters Maintenance Fuse edy11f.eps 17 Remove the battery door screw. 3. Install and tighten battery door screw. Remove the test leads from the case. The battery...
... into the case, bottom edge first, until it from the Meter 2. Do not attempt to separate it is then inserted into the case. 5. True-rms Multimeters Maintenance Fuse edy11f.eps 17 Remove the battery door screw. 3. Install and tighten battery door screw. Remove the test leads from the case. The battery...
Fluke 114,115,117 Manual
Page 21
...per IEC 61010-1 600V CAT III, Pollution Degree 2 W Fuse for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1 19 True-rms Multimeters General Specifications General Specifications Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with ANSI/ISA...2004, CAN/CSA-C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN 61010-1 2nd Edition for A input (115 & 117 only 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 °...
...per IEC 61010-1 600V CAT III, Pollution Degree 2 W Fuse for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1 19 True-rms Multimeters General Specifications General Specifications Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with ANSI/ISA...2004, CAN/CSA-C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN 61010-1 2nd Edition for A input (115 & 117 only 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 °...
Fluke 114,115,117 Manual
Page 23
... or shorts of Reading] + [Counts]) Beeper on < 20 Ω, off > 250 Ω; Function Continuity Ohms Diode test Capacitance Lo-Z Capacitance (Power-up option) True-rms Multimeters General Specifications Table 1 Accuracy Specifications (cont.) Range 600 Ω 600.0 Ω 6.000 kΩ 60.00 kΩ 600.0 kΩ 6.000 MΩ 40.00 MΩ... 9999 μF 1 nF 0.01 μF 0.1 μF 1 μF 1.9 % + 2 1.9 % + 2 1.9 % + 2 100 μF - 1000 μF: 1.9 % + 2 > 1000 μF: 5 % + 20 1 nF to 500 μF 10% + 2 typical Model 114, 115, 117 114, 115, 117 115, 117 115, 117 115, 117 21
... or shorts of Reading] + [Counts]) Beeper on < 20 Ω, off > 250 Ω; Function Continuity Ohms Diode test Capacitance Lo-Z Capacitance (Power-up option) True-rms Multimeters General Specifications Table 1 Accuracy Specifications (cont.) Range 600 Ω 600.0 Ω 6.000 kΩ 60.00 kΩ 600.0 kΩ 6.000 MΩ 40.00 MΩ... 9999 μF 1 nF 0.01 μF 0.1 μF 1 μF 1.9 % + 2 1.9 % + 2 1.9 % + 2 100 μF - 1000 μF: 1.9 % + 2 > 1000 μF: 5 % + 20 1 nF to 500 μF 10% + 2 typical Model 114, 115, 117 114, 115, 117 115, 117 115, 117 115, 117 21
Fluke 117 Electrician Multimeter Datasheet
Page 1
...+ [ counts ] ) Maximum voltage between any terminal and earth ground Surge protection Fuse for demanding settings like commercial buildings, hospitals and schools. Fluke 117 Electrician's Multimeter with optional magnetic hanger (ToolPak™) for hands free operation • CAT III 600 V safety rated General specifications Accuracy is the ideal meter... signal fluctuations • Compatible with Non-Contact Voltage Technical Data Compact true-rms meter for commercial applications The Fluke 117 is specified for 1 year after calibration, at 0 % to help get the job done faster.
...+ [ counts ] ) Maximum voltage between any terminal and earth ground Surge protection Fuse for demanding settings like commercial buildings, hospitals and schools. Fluke 117 Electrician's Multimeter with optional magnetic hanger (ToolPak™) for hands free operation • CAT III 600 V safety rated General specifications Accuracy is the ideal meter... signal fluctuations • Compatible with Non-Contact Voltage Technical Data Compact true-rms meter for commercial applications The Fluke 117 is specified for 1 year after calibration, at 0 % to help get the job done faster.
Fluke 117 Electrician Multimeter Datasheet
Page 2
...-V LoZ, and ac mV are dc-coupled. 2 Frequency is normal for ac current. Fluke Corporation 117 Electrician's Multimeter with Non-Contact Voltage Ordering information Fluke-117 Electrician's Multimeter with Non-Contact Voltage Included TL75 Test leads, holster, User's manual and 9V battery (installed)....177; ([% of 500 µs or longer. DC amps 6.000 A 10.00 A 0.001 A 0.01 A 1.0 % + 3 20 A continuous overload for 30 seconds max. Fluke. detects opens or shorts of reading] + [counts]) DC millivolts 600.0 mV 0.1 mV 2.0 % + 3 DC volts 6.000 V 0.001 V 60.00 V 0.01 V 600...
...-V LoZ, and ac mV are dc-coupled. 2 Frequency is normal for ac current. Fluke Corporation 117 Electrician's Multimeter with Non-Contact Voltage Ordering information Fluke-117 Electrician's Multimeter with Non-Contact Voltage Included TL75 Test leads, holster, User's manual and 9V battery (installed)....177; ([% of 500 µs or longer. DC amps 6.000 A 10.00 A 0.001 A 0.01 A 1.0 % + 3 20 A continuous overload for 30 seconds max. Fluke. detects opens or shorts of reading] + [counts]) DC millivolts 600.0 mV 0.1 mV 2.0 % + 3 DC volts 6.000 V 0.001 V 60.00 V 0.01 V 600...