Fluke 787 and 789 Process Meter Datasheet
Page 1
..., (selectable at power-up to wwwww.MwyF.MlukyeFSltuokree.Scotmore.com Ordering information Fluke-789 ProcessMeter Fluke-787 ProcessMeter For more while carrying a lot less. Key Fluke 789 features: • 24 V loop power supply • HART mode setting...Fluke-789 ProcessMeter: TL71 Premium Safety-Designed Test Lead Set plus alligator clips, 4 AA alkaline batteries (installed), product overview and users manual (CD-ROM) in 14 languages. Fluke-787 ProcessMeter: C81Y Protective Yellow Holster with mA source. Fluke 787 and 789 ProcessMeter™ Test Tools Safety rated multimeters...
..., (selectable at power-up to wwwww.MwyF.MlukyeFSltuokree.Scotmore.com Ordering information Fluke-789 ProcessMeter Fluke-787 ProcessMeter For more while carrying a lot less. Key Fluke 789 features: • 24 V loop power supply • HART mode setting...Fluke-789 ProcessMeter: TL71 Premium Safety-Designed Test Lead Set plus alligator clips, 4 AA alkaline batteries (installed), product overview and users manual (CD-ROM) in 14 languages. Fluke-787 ProcessMeter: C81Y Protective Yellow Holster with mA source. Fluke 787 and 789 ProcessMeter™ Test Tools Safety rated multimeters...
FE 789 Users Manual
Page 9
... is a handheld, battery-operated tool for information about DMM (digital multimeter) accessories. Contacting Fluke To order accessories, receive operating assistance, or get the location of this manual. The Fluke 789 ProcessMeter™ (referred to test process instruments, and providing a > 24 V loop power supply. Contact a Fluke distributor for measuring electrical parameters, supplying steady or ramping current...
... is a handheld, battery-operated tool for information about DMM (digital multimeter) accessories. Contacting Fluke To order accessories, receive operating assistance, or get the location of this manual. The Fluke 789 ProcessMeter™ (referred to test process instruments, and providing a > 24 V loop power supply. Contact a Fluke distributor for measuring electrical parameters, supplying steady or ramping current...
Calibration Manual
Page 15
... perform the procedures in this manual are using a different calibrator or DMM, make the connections appropriate for use in this manual showing a calibrator or digital multimeter use a Fluke 5500A calibrator or Agilent 3458A.
... perform the procedures in this manual are using a different calibrator or DMM, make the connections appropriate for use in this manual showing a calibrator or digital multimeter use a Fluke 5500A calibrator or Agilent 3458A.
Calibration Manual
Page 16
...new AA alkaline batteries. 6. With a standard blade hand screwdriver, turn the ProcessMeter OFF. 2. 789 Calibration Manual Equipment Calibration Source Digital Process Meter or Digital Process Calibrator Digital Multimeter Test Leads, low leakage, RG-58/U type 1-kΩ shunt Table 2. Remove the test leads... and turn each battery compartment door screw counterclockwise so that is preferable Fluke Model 5500A Fluke 787 ProcessMeter 741,743, or 744...
...new AA alkaline batteries. 6. With a standard blade hand screwdriver, turn the ProcessMeter OFF. 2. 789 Calibration Manual Equipment Calibration Source Digital Process Meter or Digital Process Calibrator Digital Multimeter Test Leads, low leakage, RG-58/U type 1-kΩ shunt Table 2. Remove the test leads... and turn each battery compartment door screw counterclockwise so that is preferable Fluke Model 5500A Fluke 787 ProcessMeter 741,743, or 744...
Calibration Manual
Page 19
...;F and 82.4 °F) and that the person performing the tests has read the 789 Users Manual, knows how to select functions and ranges on the ProcessMeter or the multimeter falls within the acceptable range indicated. ProcessMeter™ Calibration Verification values, and determine if... and knows how to operate the required equipment. Warm up for the digital multimeter identified in the required equipment listed in this manual showing a calibrator or digital multimeter use a Fluke 5500A calibrator or HP 3458A. Preparing to Perform Calibration Verification Note Throughout the ...
...;F and 82.4 °F) and that the person performing the tests has read the 789 Users Manual, knows how to select functions and ranges on the ProcessMeter or the multimeter falls within the acceptable range indicated. ProcessMeter™ Calibration Verification values, and determine if... and knows how to operate the required equipment. Warm up for the digital multimeter identified in the required equipment listed in this manual showing a calibrator or digital multimeter use a Fluke 5500A calibrator or HP 3458A. Preparing to Perform Calibration Verification Note Throughout the ...
Calibration Manual
Page 20
Turn the rotary knob of the UUT and verify it is >29.8 V and 23.8 V and 25 mA and Measure the open circuit voltage of the UUT to LOOP POWER. 3. 789 Calibration Manual Loop Power 1. Enable the dc volts autorange function of the HP3458A multimeter. 2.
Turn the rotary knob of the UUT and verify it is >29.8 V and 23.8 V and 25 mA and Measure the open circuit voltage of the UUT to LOOP POWER. 3. 789 Calibration Manual Loop Power 1. Enable the dc volts autorange function of the HP3458A multimeter. 2.
Calibration Manual
Page 21
...select the 100 mA range. (Do not allow the multimeter to the acceptable readings shown. Current Sourcing Test 789 Range No Range Switching No Range Switching No Range Switching 789 Output Current 4.000 mA 12.000 mA 20.000 mA Minimum Acceptable Multimeter Reading 3.990 mA 11.990 mA 19.990 mA ...Maximum Acceptable Multimeter Reading 4.010 mA 12.010 mA 20....
...select the 100 mA range. (Do not allow the multimeter to the acceptable readings shown. Current Sourcing Test 789 Range No Range Switching No Range Switching No Range Switching 789 Output Current 4.000 mA 12.000 mA 20.000 mA Minimum Acceptable Multimeter Reading 3.990 mA 11.990 mA 19.990 mA ...Maximum Acceptable Multimeter Reading 4.010 mA 12.010 mA 20....
Calibration Manual
Page 23
...Reading 0.103 A 0.406 A Checking the Diode Test Function 1. Turn the UUT rotary switch in the dc mA (autorange) function. 9. Put the multimeter in the Vposition. 3. Apply 2.0 V dc from the UUT. 8. This test just makes sure that the diode test current source is no tolerance ... 0.2 mA. (There is operating.) 15 The multimeter should read close to the COM and terminals on the UUT as shown in Figure 5. 5. ProcessMeter™ Calibration Verification 789 Range No Range Switching No Range Switching No Range Switching Table 5. Put the calibrator in Standby (STBY) ...
...Reading 0.103 A 0.406 A Checking the Diode Test Function 1. Turn the UUT rotary switch in the dc mA (autorange) function. 9. Put the multimeter in the Vposition. 3. Apply 2.0 V dc from the UUT. 8. This test just makes sure that the diode test current source is no tolerance ... 0.2 mA. (There is operating.) 15 The multimeter should read close to the COM and terminals on the UUT as shown in Figure 5. 5. ProcessMeter™ Calibration Verification 789 Range No Range Switching No Range Switching No Range Switching Table 5. Put the calibrator in Standby (STBY) ...
Calibration Manual
Page 33
ProcessMeter™ Calibration Adjustment Note The calibration adjustment procedures assume that the person performing them knows how to do so. the word "multimeter" is reserved for the digital multimeter identified in the required equipment listed in an RF field Calibration adjustment should be performed in Table 2. Do not attempt to calibrate the ProcessMeter unless you are qualified to use the ProcessMeter and the required equipment. Throughout the following, "UUT" (unit under test) refers to the ProcessMeter;
ProcessMeter™ Calibration Adjustment Note The calibration adjustment procedures assume that the person performing them knows how to do so. the word "multimeter" is reserved for the digital multimeter identified in the required equipment listed in an RF field Calibration adjustment should be performed in Table 2. Do not attempt to calibrate the ProcessMeter unless you are qualified to use the ProcessMeter and the required equipment. Throughout the following, "UUT" (unit under test) refers to the ProcessMeter;