Fluke 741B, 743B, and 744 Process Calibrator Datasheet
Page 1
... by -step procedure showing how to use PCs or that don't use the Fluke 744 to calibrate, maintain, and troubleshoot HART instrumentation. Additional 744 accessories BP7235 NiMH Battery Pack, HART communications cable, HART user's manual, NIST Traceable calibration certificate and data, three-year warranty. Ordering information Fluke-741B Documenting Process Calibrator Fluke-743B Documenting Process Calibrator Fluke-744 Documenting Process Calibrator For more...
... by -step procedure showing how to use PCs or that don't use the Fluke 744 to calibrate, maintain, and troubleshoot HART instrumentation. Additional 744 accessories BP7235 NiMH Battery Pack, HART communications cable, HART user's manual, NIST Traceable calibration certificate and data, three-year warranty. Ordering information Fluke-741B Documenting Process Calibrator Fluke-743B Documenting Process Calibrator Fluke-744 Documenting Process Calibrator For more...
Fluke 744 Users Manual
Page 12
...calibration procedures for testing limit switches. Display of measurements in the world: +1-425-356-5500 Or visit us on the World Wide Web: www.fluke.com Built-in which the calibrator can be configured to engineering units, percent of a process instrument. For performance testing and calibration instructions order the 74X Series Calibration...serial computer interface for entering alphanumeric characters. • Measuring features: Damping (smoothing of damped status. 744 Users Manual The ability to automatically log up to the next. Transmitter mode in calculator with display ...
...calibration procedures for testing limit switches. Display of measurements in the world: +1-425-356-5500 Or visit us on the World Wide Web: www.fluke.com Built-in which the calibrator can be configured to engineering units, percent of a process instrument. For performance testing and calibration instructions order the 74X Series Calibration...serial computer interface for entering alphanumeric characters. • Measuring features: Damping (smoothing of damped status. 744 Users Manual The ability to automatically log up to the next. Transmitter mode in calculator with display ...
Fluke 744 Users Manual
Page 42
...¼ ISO adapter if necessary. 2. Press eor the Choices softkey with u or d. Press e. Connect a pressure module to the calibrator as described in Figure 12. If desired, you execute a task that sources or measures pressure. 5. Metric units (kPa, mmHg, etc....) are shown in Setup mode in zeroing procedures depending on module type. Press s. c. f. 744 Users Manual 1. Press p. The calibrator automatically senses which pressure module is attached and sets its range accordingly. 4. d. e. Press M for MEASURE ...
...¼ ISO adapter if necessary. 2. Press eor the Choices softkey with u or d. Press e. Connect a pressure module to the calibrator as described in Figure 12. If desired, you execute a task that sources or measures pressure. 5. Metric units (kPa, mmHg, etc....) are shown in Setup mode in zeroing procedures depending on module type. Press s. c. f. 744 Users Manual 1. Press p. The calibrator automatically senses which pressure module is attached and sets its range accordingly. 4. d. e. Press M for MEASURE ...
Fluke 744 Users Manual
Page 44
...as steps a-f above. 34 The display prompts you can change between ITS-90 or IPTS-68 Temperature Scale in Figure 13. b. The procedure is wider than the other. Table 5 summarizes the ranges and characteristics of the supported thermocouples. f. Use the u and d keys to... for MEASURE mode. 3. If necessary, press M for that parameter. Press eto go back to the desired parameter. 744 Users Manual Measuring Temperature Using Thermocouples The calibrator supports eleven standard thermocouples, each identified with an alpha character: E, N, J, K, T, B, R, S, C, L, or U.
...as steps a-f above. 34 The display prompts you can change between ITS-90 or IPTS-68 Temperature Scale in Figure 13. b. The procedure is wider than the other. Table 5 summarizes the ranges and characteristics of the supported thermocouples. f. Use the u and d keys to... for MEASURE mode. 3. If necessary, press M for that parameter. Press eto go back to the desired parameter. 744 Users Manual Measuring Temperature Using Thermocouples The calibrator supports eleven standard thermocouples, each identified with an alpha character: E, N, J, K, T, B, R, S, C, L, or U.
Fluke 744 Users Manual
Page 49
c. e. Press eto go back to the desired setting. The procedure is the same as follows: a. V mA mA V RTD RTD CAT 30V MAX SOURCE 30V MEAS 30V 300V MAX TC MAX MAX WRONG V mA mA V RTD ... mode as steps a through f above. 39 Press u or d to move the cursor to choose a setting for that parameter. f. Using a Jumper Correctly gj14f.eps Documenting Process Calibrator Using Measure Mode 8. Press the Next Page softkey twice. If necessary, you which to exit Setup mode. 9. Press the Done softkey or s to change between...
c. e. Press eto go back to the desired setting. The procedure is the same as follows: a. V mA mA V RTD RTD CAT 30V MAX SOURCE 30V MEAS 30V 300V MAX TC MAX MAX WRONG V mA mA V RTD ... mode as steps a through f above. 39 Press u or d to move the cursor to choose a setting for that parameter. f. Using a Jumper Correctly gj14f.eps Documenting Process Calibrator Using Measure Mode 8. Press the Next Page softkey twice. If necessary, you which to exit Setup mode. 9. Press the Done softkey or s to change between...
Fluke 744 Users Manual
Page 53
... function. 5. When measuring or sourcing the function of your custom units, for the input of the custom unit. Enter the 0% and 100% scale points for calibration procedures in native engineering units. To do this, you select a function, for example mV dc, scale it as follows to the right of your choice, press...
... function. 5. When measuring or sourcing the function of your custom units, for the input of the custom unit. Enter the 0% and 100% scale points for calibration procedures in native engineering units. To do this, you select a function, for example mV dc, scale it as follows to the right of your choice, press...
Fluke 744 Users Manual
Page 63
..., inH2O, inH2O@60°F, ftH2O, bar, g/cm2, or Pa. The threads on the display. 6. Use the supplied ¼ NPT to the calibrator as shown on the pressure modules accept ¼ NPT fittings. Press p. b. e. Connect a pressure module and pressure source to ¼ ISO adapter... if necessary. 2. c. Press the Done softkey. 53 Modules vary in zeroing procedures depending on Pressure Units. Change the pressure display units as described in their Documenting Process Calibrator Using Source Mode base units (Pa, mHg, etc.). d. Press s. 1. If necessary, press...
..., inH2O, inH2O@60°F, ftH2O, bar, g/cm2, or Pa. The threads on the display. 6. Use the supplied ¼ NPT to the calibrator as shown on the pressure modules accept ¼ NPT fittings. Press p. b. e. Connect a pressure module and pressure source to ¼ ISO adapter... if necessary. 2. c. Press the Done softkey. 53 Modules vary in zeroing procedures depending on Pressure Units. Change the pressure display units as described in their Documenting Process Calibrator Using Source Mode base units (Pa, mHg, etc.). d. Press s. 1. If necessary, press...
Fluke 744 Users Manual
Page 78
...the corresponding output current. 2. If necessary, press M for SOURCE mode. 5. Switch tests. Press m. 4. 744 Users Manual Calibrating a Process Instrument Note To calibrate a HART-capable transmitter using a host computer and compatible application software. Press M for MEASURE mode. 3. Generating... The following example shows how to MEASUREMENT or SOURCE mode and change the operating parameters before it is calibrated.) The calibrator runs preloaded tasks (procedures) that follows. In this same method. Enter a source value, for instructions. The way you press...
...the corresponding output current. 2. If necessary, press M for SOURCE mode. 5. Switch tests. Press m. 4. 744 Users Manual Calibrating a Process Instrument Note To calibrate a HART-capable transmitter using a host computer and compatible application software. Press M for MEASURE mode. 3. Generating... The following example shows how to MEASUREMENT or SOURCE mode and change the operating parameters before it is calibrated.) The calibrator runs preloaded tasks (procedures) that follows. In this same method. Enter a source value, for instructions. The way you press...
Fluke 744 Users Manual
Page 81
... as follows: gj45s.eps 71 The Strategy is indicated by the up arrow on the display and in this line. If the instrument calibration procedure requires you to enter 0% and 100% values for "user entered values." Rising is the number of strategies appears from which test points... for Delay. 12. Select one, then press Done. 15. See "Creating Custom Measurement Units," earlier in results. When you finish selecting the calibration parameters, the display appears as "PH." When you use custom units, the w symbol appears next to another test strategy by hand at each step...
... as follows: gj45s.eps 71 The Strategy is indicated by the up arrow on the display and in this line. If the instrument calibration procedure requires you to enter 0% and 100% values for "user entered values." Rising is the number of strategies appears from which test points... for Delay. 12. Select one, then press Done. 15. See "Creating Custom Measurement Units," earlier in results. When you finish selecting the calibration parameters, the display appears as "PH." When you use custom units, the w symbol appears next to another test strategy by hand at each step...
Fluke 744 Users Manual
Page 82
....eps The error of starting an automatic test or stepping through the tests automatically. (Abort gets you are sourcing pressure, the calibrator pauses at the zero point, sourcing the correct temperature (a voltage) and measuring the corresponding current from the transmitter. The... such as a measurement has settled and been captured, the calibrator moves to accept the calibration parameters. If you out of the calibration procedure.) The tests begin at each step for instruments with built-in the top left of points. 744 Users Manual 16. The display changes to the remaining set ...
....eps The error of starting an automatic test or stepping through the tests automatically. (Abort gets you are sourcing pressure, the calibrator pauses at the zero point, sourcing the correct temperature (a voltage) and measuring the corresponding current from the transmitter. The... such as a measurement has settled and been captured, the calibrator moves to accept the calibration parameters. If you out of the calibration procedure.) The tests begin at each step for instruments with built-in the top left of points. 744 Users Manual 16. The display changes to the remaining set ...
Fluke 744 Users Manual
Page 85
... are developed using a host computer and compatible application software. Test Comments The calibrator runs tasks (custom procedures) that are in a menu after the As Found calibration template is displayed, select a comment to compute instrument errors. Calibrating a Delta-Pressure Flow Instrument The procedure to calibrate a √ instrument is the same as just described, with the test results...
... are developed using a host computer and compatible application software. Test Comments The calibrator runs tasks (custom procedures) that are in a menu after the As Found calibration template is displayed, select a comment to compute instrument errors. Calibrating a Delta-Pressure Flow Instrument The procedure to calibrate a √ instrument is the same as just described, with the test results...
Fluke 744 Users Manual
Page 86
744 Users Manual Calibrating a Limit Switch The procedure to set up the limit switch procedure lets you press the As Found softkey. Figure 24 defines the terminology used in a menu after you select the following parameters: • Switch sense (normally ... Switch B Open Deadband Switch B Closed Switch B Setpoint Reset Switch A Closed Reset Deadband Switch A Open Switch A Setpoint Switch A: Low Limit Switch B: High Limit Figure 24. Switch procedure in calibrating limit switches. Select either the 1 Pt. Switch or 2 Pt. Limit Switch Terminology gj24f.eps The template to...
744 Users Manual Calibrating a Limit Switch The procedure to set up the limit switch procedure lets you press the As Found softkey. Figure 24 defines the terminology used in a menu after you select the following parameters: • Switch sense (normally ... Switch B Open Deadband Switch B Closed Switch B Setpoint Reset Switch A Closed Reset Deadband Switch A Open Switch A Setpoint Switch A: Low Limit Switch B: High Limit Figure 24. Switch procedure in calibrating limit switches. Select either the 1 Pt. Switch or 2 Pt. Limit Switch Terminology gj24f.eps The template to...
Fluke 744 Users Manual
Page 87
... and press e. 11. Do not set state is a closed switch contact. For pressure switches, you can use the Manual Test choice. The procedure for the pressure source function. Highlight 1 Pt. Make the following selections: Setpoint 1 = 10.000 psi Setpoint Type = High Set State .... 14. Set the Tolerance to atmosphere. 3. Press e to the limit switch. Press q q for MEASURE mode. 4. Connect the pressure module to the calibrator, and connect a pressure line to modify the parameters for SOURCE mode. 6. Press c to zero the pressure module. 8. Press M for Setpoint 1. 12....
... and press e. 11. Do not set state is a closed switch contact. For pressure switches, you can use the Manual Test choice. The procedure for the pressure source function. Highlight 1 Pt. Make the following selections: Setpoint 1 = 10.000 psi Setpoint Type = High Set State .... 14. Set the Tolerance to atmosphere. 3. Press e to the limit switch. Press q q for MEASURE mode. 4. Connect the pressure module to the calibrator, and connect a pressure line to modify the parameters for SOURCE mode. 6. Press c to zero the pressure module. 8. Press M for Setpoint 1. 12....
Fluke 744 Users Manual
Page 88
...again. 25. When you simply follow the prompts on the display for later viewing or uploading. Press the Done softkey and view the results. 22. 744 Users Manual 16. When the switch sets, slowly bring the pressure up to other parameters work similarly. Press the Done softkey. 24. Results from ...the As Found and As Left tests are saved in the calibrator memory for testing the first switch, changing test leads, and testing the second limit switch. 78 The procedure for limit switches that respond to the trip point. 20. Press the Done softkey. 18. ...
...again. 25. When you simply follow the prompts on the display for later viewing or uploading. Press the Done softkey and view the results. 22. 744 Users Manual 16. When the switch sets, slowly bring the pressure up to other parameters work similarly. Press the Done softkey. 24. Results from ...the As Found and As Left tests are saved in the calibrator memory for testing the first switch, changing test leads, and testing the second limit switch. 78 The procedure for limit switches that respond to the trip point. 20. Press the Done softkey. 18. ...
Fluke 744 Users Manual
Page 96
A task configures the calibrator for transmitter calibration with a procedure name, for example the type and manufacturer of tasks (procedures) downloaded from a host computer. Press the Min Max softkey again to revert to turn on : gj55s.eps 86 Running a Preloaded Task Press the More Choices... softkey twice, then press the Min Max softkey to the normal display. Press the c key to record and show the maximum and minimum readings. 744 Users Manual Recording Min and Max Measurements You can set the display to reset the Min Max registers. The following figure shows the display with...
A task configures the calibrator for transmitter calibration with a procedure name, for example the type and manufacturer of tasks (procedures) downloaded from a host computer. Press the Min Max softkey again to revert to turn on : gj55s.eps 86 Running a Preloaded Task Press the More Choices... softkey twice, then press the Min Max softkey to the normal display. Press the c key to record and show the maximum and minimum readings. 744 Users Manual Recording Min and Max Measurements You can set the display to reset the Min Max registers. The following figure shows the display with...
Fluke 744 Users Manual
Page 108
...procedures and saved results to 1000 charge/discharge cycles. Spent batteries should be disposed of replaceable parts is included with DPC/TRACK. A custom serial interface cable for the calibrator is available in the 74X Series Calibration Manual (PN 602505). Contact your authorized Fluke Service Center for up to a PC. 744...PCTM-compatible personal computer, Microsoft Windows, and Fluke DPC/TRACK™ software, or a qualified Fluke partner's software are required. Maintenance Note Additional maintenance instructions, including a calibration procedure and a list of by a qualified ...
...procedures and saved results to 1000 charge/discharge cycles. Spent batteries should be disposed of replaceable parts is included with DPC/TRACK. A custom serial interface cable for the calibrator is available in the 74X Series Calibration Manual (PN 602505). Contact your authorized Fluke Service Center for up to a PC. 744...PCTM-compatible personal computer, Microsoft Windows, and Fluke DPC/TRACK™ software, or a qualified Fluke partner's software are required. Maintenance Note Additional maintenance instructions, including a calibration procedure and a list of by a qualified ...
Fluke 744 Users Manual
Page 109
... water and mild soap. STATUS number on the sticker should always match the Calibration Status number in doubt, have the calibrator serviced. Calibration of the 744 is to adjust the contrast. 99 When in the calibration s screen. Press the u and d keys to be impaired. You cannot... Lithium Backup Battery A lithium battery maintains the memory contents and Setup settings. Return the calibrator to an authorized Fluke service center listed at the next calibration interval as a preventive maintenance procedure. Caution To avoid damaging the plastic lens and case, do not use it at the...
... water and mild soap. STATUS number on the sticker should always match the Calibration Status number in doubt, have the calibrator serviced. Calibration of the 744 is to adjust the contrast. 99 When in the calibration s screen. Press the u and d keys to be impaired. You cannot... Lithium Backup Battery A lithium battery maintains the memory contents and Setup settings. Return the calibrator to an authorized Fluke service center listed at the next calibration interval as a preventive maintenance procedure. Caution To avoid damaging the plastic lens and case, do not use it at the...
Fluke 744 Users Manual
Page 112
...: 0 to 100 psi (gage, wet) Fluke-700P07: 0 to 500 psi (gage, wet) Fluke-700P08: 0 to 1,000 psi (gage, wet) Fluke-700P09: 0 to 1,500 psi (gage, wet) Fluke-700P29: 0 to 3,000 psi (gage, wet) Fluke-700P30: 0 to 5,000 psi (gage, wet) Fluke-700P31: 0 to interface the calibrator with the 744 calibrator. 744 Users Manual Accessories The Fluke accessories listed below . (Differential models...
...: 0 to 100 psi (gage, wet) Fluke-700P07: 0 to 500 psi (gage, wet) Fluke-700P08: 0 to 1,000 psi (gage, wet) Fluke-700P09: 0 to 1,500 psi (gage, wet) Fluke-700P29: 0 to 3,000 psi (gage, wet) Fluke-700P30: 0 to 5,000 psi (gage, wet) Fluke-700P31: 0 to interface the calibrator with the 744 calibrator. 744 Users Manual Accessories The Fluke accessories listed below . (Differential models...