Fluke 1623 and 1625 Geo Earth Ground Testers Datasheet - Page 2
... coefficient Intrinsic error Operating error Climatic class Protective type Safety EMC (Emission Immunity) Quality system External voltage V ext rejection Measuring time Max. and 4-Pole Fall of earth ground measurement:
• 3- The complete model kit comes with the 1623 or 1625 tester, test leads, 4 earth ground stakes, 3 cable reels with wire, 2 clamps, batteries, and manual-all four...
Fluke 1623 and 1625 Geo Earth Ground Testers Datasheet - Page 5
... and between 99 % and 101 % of range/Kelvin
Operating errors
Refer to 140° F)
Note: The chart of four temperature ranges for measuring equipment
with a mains supply and/or measuring equipment deriving its output voltage directly from the distribution system.
5 Fluke Corporation 1623 and 1625 GEO Earth Ground Testers Temperature coefficient
± 0.1 % of the nominal system frequency...
Fluke 1623 and 1625 Geo Earth Ground Testers Datasheet - Page 8
...48 V ac
250 mA ac
94, 105, 111, 128 Hz selected manually or automatically (AFC), 55 Hz (R*)
120 dB (162/3, 50, 60...error because of RETOTAL + RS + RH
Typ. 8 sec. with AFC and complete cycle of all measuring frequencies
0.5 mA
With transformer (1000:1)
0.1 mA
With transformer (200:1)
3A
With a transformer (1000:1)
8 Fluke Corporation 1623 and 1625 GEO Earth Ground Testers and auxiliary earth...
Fluke 1623 and 1625 Geo Earth Ground Testers Datasheet - Page 9
...measurement 20 V ac, square pulse > 250 mA ac 94, 105, 111, 128 Hz selected manually or automatically (AFC)
Measuring range 0.020 7 to 300 k7
Display range 0.001 7 to ...errors ± (5 % of rdg + 5 digit)
Resistance measurement (R )
Measuring method Open circuit voltage Short circuit current Formation of measuring adjustment.
2-pole
9 Fluke Corporation 1623 and 1625 GEO Earth Ground Testers...
FE 1625 Users Manual - Page 1
®
1625
Earth/Ground Tester
Users Manual
January 2006
© 2006 Fluke Corporation, All rights reserved. Printed in USA All product names are trademarks of their respective companies.
FE 1625 Users Manual - Page 10
...
Users Manual
Notes • The terms earth and earthing also refer to Appendix A for a complete set of operating information including specifications.
• Selective measurements are described in the main section of this manual.
• For stakeless earth resistance measurements, the EI-1625 must be purchased. (The EI-1625 comes standard with the 1625 Kit). Fluke 1625 Earth/Ground Tester...
FE 1625 Users Manual - Page 11
... w/50m Wire
1625 Users Manual
Item/Part Number Fluke-1625 Fluke-1625 Kit
Fluke-162x-7001 ES-162P3
ES-162P4
EI-1625
EI-162X 2539195 EI-162AC EI-162BN 2577171 2539121 2539100 2539117 2560348
3 Earth/Ground Tester Models and Accessories
Table 1. Basic (Includes manual, 2 leads and 2 clips)
Earth Ground Tester - Models and Accessories
Description
Earth Ground Tester - Models and Accessories...
FE 1625 Users Manual - Page 40
... be carried out for a specific measurement can be called and displayed with automatic switch over if necessary (AFC - At the activation of Interference - The corresponding measuring frequency used for up to the mains frequency (55 Hz) is activated automatically.
32 1625
Users Manual
START TEST
1625 EARTH / GROUND TESTER
H /C2
DISPLAY MENU
Earth/Ground Resistance 300 k RA
CHANGE...
FE 1625 Users Manual - Page 42
... like auxiliary earth electrode, probe- Remarks for the setting of earth spikes:
Before setting the earth spikes for probe and auxiliary earth electrode make ...distance of > 20 m between the earth electrode and the earth spikes as well as of Earthing Resistance - 1625
Users Manual
START TEST
1625 EARTH / GROUND TESTER
H /C2
DISPLAY MENU
Earth/Ground Resistance 300 k RA
CHANGE ITEM
4...
FE 1625 Users Manual - Page 48
... given on the instrument corresponds to an incorrect or incomplete connection of the sockets symbols EFGH or B, points to the current transformer used. Measuring the Earthing Resistance without Disengaging the Overhead Earth Wire
1. 1625
Users Manual
Accordingly, these equivalent resistances are displayed as a parallel circuit of All Data Settings with Personalised CODE")
40
FE 1625 Users Manual - Page 52
... ratio to be set calculates:
trnew = 1000 × 1.175 = 1195 0.983
Compensation of Earth Electrode Connecting Lead
If the line resistance to the earth electrode can not be ignored, a compensation of the connecting lead resistance to RE = 0.983 Ω an error evolves as described below:
Measuring process:
START TEST
1625 EARTH / GROUND TESTER
H /C2
DISPLAY MENU
Earth/Ground Resistance 300...
FE 1625 Users Manual - Page 56
1625
Users Manual
90O
E
ES
S
H
I
Measurement of Measuring Lead Resistance").
Resistance Measurement (R~)
1. For measurements of very low resistances a compensation of the connecting leads is done with AC voltage. START TEST
1625 EARTH / GROUND TESTER
H /C2
DISPLAY MENU
Earth/Ground Resistance 300 k RA
CHANGE ITEM
4 POLE 3 POLE 4 POLE
3 POLE
AC Resistance
300 k R
DC Low ...
FE 1625 Users Manual - Page 72
... range 0 °C ... +30 °C.
Quality standard:
Developed, designed and manufactured to this section exclusively refers to comply with ground testers in the mode: RE D
3 pole (Order No. Principle:
Ground testing without using earth stakes for use with DIN ISO 9001.
64
1625
Users Manual
edw060.eps
Specifications
General:
Adapter unit for the probe and the auxiliary...
FE 1625 Users Manual - Page 76
START TEST
1625 EARTH / GROUND TESTER
H /C2
DISPLAY MENU
Earth/Ground Resistance 300 k RA
CHANGE ITEM
4 POLE 3 POLE 4 POLE
3 POLE
AC Resistance
300 k R
DC Low Resistance 3k R
2 POLE
2 POLE
4 POLE
SELECT
OFF
S/P2
ES/P1 E /C1
H S
E
Black EI-162AC
Red EI-162X
edw063.eps
Use the test cable contained in this manual.
Clamp both transformers around the ground conductor...
FE 1625 Users Manual - Page 80
... uring range or adapter cable is on/off for below meas-
Display
Significance Note
Polarity of Displays
With stakeless ground testing some display combinations have certain significance. 1625
Users Manual
Description of CT's is reversed
Turn one CT
E
edw067.eps
edw068.eps
Legend: W= displayed flashing
Resistance Switch under test is plugged into ground tester incorrectly