Product Manual
Page 19
2638A Users Manual General Specifications Mains Voltage 100 V Setting 90 V to 110 V 120 V Setting 108 V to 132 V 220 V Setting 198 V to 242 V 240 V Setting 216 V to 264 V Frequency 47 Hz to 440 Hz Power Consumption 36 VA peak (24 W average) Environment Temperature Operating 0 °C to 50 °C Full accuracy 18 °C to 28 °C Storage 20 °C to 70 °C Warm-up 1 hour to full accuracy specifications Relative Humidity (non-condensing) Operating 0 °C to 28 °C
2638A Users Manual General Specifications Mains Voltage 100 V Setting 90 V to 110 V 120 V Setting 108 V to 132 V 220 V Setting 198 V to 242 V 240 V Setting 216 V to 264 V Frequency 47 Hz to 440 Hz Power Consumption 36 VA peak (24 W average) Environment Temperature Operating 0 °C to 50 °C Full accuracy 18 °C to 28 °C Storage 20 °C to 70 °C Warm-up 1 hour to full accuracy specifications Relative Humidity (non-condensing) Operating 0 °C to 28 °C
Product Manual
Page 31
... for instructions on how to set the voltage selector. Fuses Voltage Selector 100 V 120 V 220 V 240 V Fuse 0.25 A, 250 V (slow blow) 0.25 A, 250 V (slow blow) 160 mA, 250 V (slow blow) 160 mA, 250 V (slow blow) Fluke Calibration Part Number 166306 166306 4394437 4394437 WCaution To prevent damage to the Product, ...make sure the regional voltage selector is in the correct position for Fluke products online at: www. .com 1.888.610.7664 The selector can be set for the region of use before the mains power ...
... for instructions on how to set the voltage selector. Fuses Voltage Selector 100 V 120 V 220 V 240 V Fuse 0.25 A, 250 V (slow blow) 0.25 A, 250 V (slow blow) 160 mA, 250 V (slow blow) 160 mA, 250 V (slow blow) Fluke Calibration Part Number 166306 166306 4394437 4394437 WCaution To prevent damage to the Product, ...make sure the regional voltage selector is in the correct position for Fluke products online at: www. .com 1.888.610.7664 The selector can be set for the region of use before the mains power ...
Product Manual
Page 33
...regional voltage selector is in the correct position for the region of use before mains power is connected. WCaution To prevent damage to a 100 V ac, 120 V ac, or 230 V ac nominal outlet as shown in Figure 2-2. Disruption of wear. • Make sure the ground conductor in your country and... rated for Fluke products online at: www. .com 1.888.610.7664 WXWarning To prevent possible electrical shock, fire, or personal injury: • Use only the mains ...
...regional voltage selector is in the correct position for the region of use before mains power is connected. WCaution To prevent damage to a 100 V ac, 120 V ac, or 230 V ac nominal outlet as shown in Figure 2-2. Disruption of wear. • Make sure the ground conductor in your country and... rated for Fluke products online at: www. .com 1.888.610.7664 WXWarning To prevent possible electrical shock, fire, or personal injury: • Use only the mains ...
Product Manual
Page 97
... how to clean the Product, reset the memory, and replace the fuse in Table 6-1. 6-1 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664 Fuses Voltage Selector Fuse Fluke Calibration Part Number 100 V 120 V 0.25 A, 250 V (slow blow) 0.25 A, 250 V (slow blow) 166306 166306 220 V 160 mA, 250 V (slow blow) 4394437...
... how to clean the Product, reset the memory, and replace the fuse in Table 6-1. 6-1 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664 Fuses Voltage Selector Fuse Fluke Calibration Part Number 100 V 120 V 0.25 A, 250 V (slow blow) 0.25 A, 250 V (slow blow) 166306 166306 220 V 160 mA, 250 V (slow blow) 4394437...