FE 192,196,199 C Users Manual
Page 1
Fluke 192B - 196B/C - 199B/C ScopeMeter 4822 872 30601 October 2002 Rev. 2, 12/04 © 2002 Fluke Corporation. All product names are trademarks of their respective companies. Printed in the Netherlands. Users Manual All rights reserved.
Fluke 192B - 196B/C - 199B/C ScopeMeter 4822 872 30601 October 2002 Rev. 2, 12/04 © 2002 Fluke Corporation. All product names are trademarks of their respective companies. Printed in the Netherlands. Users Manual All rights reserved.
FE 192,196,199 C Users Manual
Page 6
Fluke 192B - 196B/C - 199B/C Users Manual Analyzing Waveforms ...25 2 Using The Multimeter 27 Making Meter Connections 27 Making Multimeter Measurements 28 Freezing the Readings 31 Selecting Auto/Manual Ranges 31 Making Relative Measurements 32 3 Using The Recorder Functions 33 Opening the Recorder Main Menu 33 Plotting Measurements Over Time (TrendPlot 34 Recording Scope ...
Fluke 192B - 196B/C - 199B/C Users Manual Analyzing Waveforms ...25 2 Using The Multimeter 27 Making Meter Connections 27 Making Multimeter Measurements 28 Freezing the Readings 31 Selecting Auto/Manual Ranges 31 Making Relative Measurements 32 3 Using The Recorder Functions 33 Opening the Recorder Main Menu 33 Plotting Measurements Over Time (TrendPlot 34 Recording Scope ...
FE 192,196,199 C Users Manual
Page 8
Fluke 192B - 196B/C - 199B/C Users Manual Parts and Accessories 88 Troubleshooting ...93 9 Specifications ...95 Introduction ...95 Dual Input Oscilloscope 96 Automatic Scope Measurements 98 Meter ...102 DMM Measurements on Meter Inputs 102 Recorder ...104 Zoom, Replay and Cursors 105 Miscellaneous ...105 Environmental...107 Safety ...107 10:1 Probe ...109 Electromagnetic Immunity 110 iv
Fluke 192B - 196B/C - 199B/C Users Manual Parts and Accessories 88 Troubleshooting ...93 9 Specifications ...95 Introduction ...95 Dual Input Oscilloscope 96 Automatic Scope Measurements 98 Meter ...102 DMM Measurements on Meter Inputs 102 Recorder ...104 Zoom, Replay and Cursors 105 Miscellaneous ...105 Environmental...107 Safety ...107 10:1 Probe ...109 Electromagnetic Immunity 110 iv
FE 192,196,199 C Users Manual
Page 10
ScopeMeter Test Tool Kit 2 Fluke 192B - 196B/C - 199B/C Users Manual Unpacking the Test Tool Kit The following items are included in your test tool kit: 1 2 e (2x) (1x) f (2x) b a (2x) c (2x) (2x) d Note When new, the rechargeable NiMH battery is not fully charged. See Chapter 8. 6 7 3 4 9 10 11 b (2x) a 5 8 Figure 1.
ScopeMeter Test Tool Kit 2 Fluke 192B - 196B/C - 199B/C Users Manual Unpacking the Test Tool Kit The following items are included in your test tool kit: 1 2 e (2x) (1x) f (2x) b a (2x) c (2x) (2x) d Note When new, the rechargeable NiMH battery is not fully charged. See Chapter 8. 6 7 3 4 9 10 11 b (2x) a 5 8 Figure 1.
FE 192,196,199 C Users Manual
Page 11
... Probe for Probe Tip (gray) 5 a) Test Lead Set b) Probe ground lead with 4 mm banana jack 6 Getting Started Manual 7 CD ROM with Users Manual (multi-language) 8 Shipment box (basic version only) Unpacking the Test Tool Kit Fluke 192B, 196B/C and 199B/C S versions include also the following items: # Description 9 Optically Isolated USB Adapter/Cable 10 FlukeView...
... Probe for Probe Tip (gray) 5 a) Test Lead Set b) Probe ground lead with 4 mm banana jack 6 Getting Started Manual 7 CD ROM with Users Manual (multi-language) 8 Shipment box (basic version only) Unpacking the Test Tool Kit Fluke 192B, 196B/C and 199B/C S versions include also the following items: # Description 9 Optically Isolated USB Adapter/Cable 10 FlukeView...
FE 192,196,199 C Users Manual
Page 12
Fluke 192B - 196B/C - 199B/C Users Manual Safety Information: Read First Carefully read the following international symbols are used on the test tool and in this manual: See explanation in manual Double Insulation (Protection Class) Disposal information Earth ground Recycling information Safety Approval Direct Current Conformité Européenne Safety Approval Alternating Current 4 Warning To ...
Fluke 192B - 196B/C - 199B/C Users Manual Safety Information: Read First Carefully read the following international symbols are used on the test tool and in this manual: See explanation in manual Double Insulation (Protection Class) Disposal information Earth ground Recycling information Safety Approval Direct Current Conformité Européenne Safety Approval Alternating Current 4 Warning To ...
FE 192,196,199 C Users Manual
Page 14
... matter should then be impaired if, for mechanical damage and replace damaged test leads! Safety is connected to a voltage different from the line power. Fluke 192B - 196B/C - 199B/C Users Manual The terms 'Isolated' or 'Electrically floating' are fully insulated to protect against electrical shock. The isolated input connectors have no exposed metal and...
... matter should then be impaired if, for mechanical damage and replace damaged test leads! Safety is connected to a voltage different from the line power. Fluke 192B - 196B/C - 199B/C Users Manual The terms 'Isolated' or 'Electrically floating' are fully insulated to protect against electrical shock. The isolated input connectors have no exposed metal and...
FE 192,196,199 C Users Manual
Page 16
The Screen After Reset 8 Now look at the display; Figure 3. The test tool turns on, and you will see a screen that looks like Figure 3. you should hear a double beep, indicating the reset was successful. 4 Release the USER key. Fluke 192B - 196B/C - 199B/C Users Manual Resetting the Test Tool If you want to reset the test tool to the factory settings, do the following: 1 Turn the test tool off. 2 Press and hold the USER key. 3 Press and release.
The Screen After Reset 8 Now look at the display; Figure 3. The test tool turns on, and you will see a screen that looks like Figure 3. you should hear a double beep, indicating the reset was successful. 4 Release the USER key. Fluke 192B - 196B/C - 199B/C Users Manual Resetting the Test Tool If you want to reset the test tool to the factory settings, do the following: 1 Turn the test tool off. 2 Press and hold the USER key. 3 Press and release.
FE 192,196,199 C Users Manual
Page 18
... four signal inputs: two safety BNC jack inputs (red input A and gray input B) and two safety 4-mm banana jack inputs (red and black). Fluke 192B - 196B/C - 199B/C Users Manual Hiding Key Labels and Menus You can hide a menu or key label at the top of the yellow menu keys, e.g. Measurement Connections 10 To...
... four signal inputs: two safety BNC jack inputs (red input A and gray input B) and two safety 4-mm banana jack inputs (red and black). Fluke 192B - 196B/C - 199B/C Users Manual Hiding Key Labels and Menus You can hide a menu or key label at the top of the yellow menu keys, e.g. Measurement Connections 10 To...
FE 192,196,199 C Users Manual
Page 20
...light-gray RANGE, TIME and MOVE keys at the left side of the screen identifies the ground level of virtually any waveform. Fluke 192B - 196B/C - 199B/C Users Manual Displaying an Unknown Signal with Connect-and-View™ The Connect-and-View feature lets the test tool display complex, unknown ...the setup is especially useful for quickly checking several signals. The input A zero - icon ( ) at the bottom of the keypad to select the manual range again. Figure 7. To enable the Connect-and-View feature, do the following: 1 Perform an Auto Set. AUTO appears at the top right ...
...light-gray RANGE, TIME and MOVE keys at the left side of the screen identifies the ground level of virtually any waveform. Fluke 192B - 196B/C - 199B/C Users Manual Displaying an Unknown Signal with Connect-and-View™ The Connect-and-View feature lets the test tool display complex, unknown ...the setup is especially useful for quickly checking several signals. The input A zero - icon ( ) at the bottom of the keypad to select the manual range again. Figure 7. To enable the Connect-and-View feature, do the following: 1 Perform an Auto Set. AUTO appears at the top right ...
FE 192,196,199 C Users Manual
Page 22
Note that the Peak-Peak reading for input B appears next to the input A frequency reading at the top of the reading area. 2 Resume your measurement. HOLD appears at any time. 1 Freeze the screen. Figure 8. Fluke 192B - 196B/C - 199B/C Users Manual Figure 8 shows an example of the screen. Hz and V peak-peak as Scope Readings Note The B versions do not provide Vpwm measurements. 14 Freezing the Screen You can freeze the screen (all readings and waveforms) at the right of the screen.
Note that the Peak-Peak reading for input B appears next to the input A frequency reading at the top of the reading area. 2 Resume your measurement. HOLD appears at any time. 1 Freeze the screen. Figure 8. Fluke 192B - 196B/C - 199B/C Users Manual Figure 8 shows an example of the screen. Hz and V peak-peak as Scope Readings Note The B versions do not provide Vpwm measurements. 14 Freezing the Screen You can freeze the screen (all readings and waveforms) at the right of the screen.
FE 192,196,199 C Users Manual
Page 24
... Persistence: Off , Envelope: On to Signals observe dynamic waveforms (C-versions only). Select Dot-join: On or Off to Waveform: and open the Persistence... Fluke 192B - 196B/C - 199B/C Users Manual Using Persistence to Display Waveforms You can use Persistence to observe dynamic signals. 1 Display the SCOPE key labels. 2 Open the Waveform Options menu. 3 Jump...
... Persistence: Off , Envelope: On to Signals observe dynamic waveforms (C-versions only). Select Dot-join: On or Off to Waveform: and open the Persistence... Fluke 192B - 196B/C - 199B/C Users Manual Using Persistence to Display Waveforms You can use Persistence to observe dynamic signals. 1 Display the SCOPE key labels. 2 Open the Waveform Options menu. 3 Jump...
FE 192,196,199 C Users Manual
Page 26
... display is displayed as positive-going waveform which may provide a more meaningful view. Observe that ac and dc voltages appear on a dc signal. Fluke 192B - 196B/C - 199B/C Users Manual Acquiring Waveforms Selecting AC-Coupling After a reset, the test tool is dc-coupled so that the bottom left of the screen displays the ac...
... display is displayed as positive-going waveform which may provide a more meaningful view. Observe that ac and dc voltages appear on a dc signal. Fluke 192B - 196B/C - 199B/C Users Manual Acquiring Waveforms Selecting AC-Coupling After a reset, the test tool is dc-coupled so that the bottom left of the screen displays the ac...
FE 192,196,199 C Users Manual
Page 28
... To suppress noise without loss of bandwidth, use a Mathematics function, do the following : 1 Display the SCOPE key labels. 2 Open the Waveform Options menu. Fluke 192B - 196B/C - 199B/C Users Manual Working with input A on the vertical axis and input B on the horizontal axis. For the same reason, it improves triggering on the waveform. The...
... To suppress noise without loss of bandwidth, use a Mathematics function, do the following : 1 Display the SCOPE key labels. 2 Open the Waveform Options menu. Fluke 192B - 196B/C - 199B/C Users Manual Working with input A on the vertical axis and input B on the horizontal axis. For the same reason, it improves triggering on the waveform. The...
FE 192,196,199 C Users Manual
Page 30
..., which can result in aliasing, or too fast, which results in less than one signal period on the screen. 6 Perform a spectrum analysis on (toggle function). Fluke 192B - 196B/C - 199B/C Users Manual 3 Jump to open the Mathematics menu. 4 Select Function: Spectrum. 5 Select Window: Auto (automatic windowing), Hanning, Hamming, or None (no windowing).
..., which can result in aliasing, or too fast, which results in less than one signal period on the screen. 6 Perform a spectrum analysis on (toggle function). Fluke 192B - 196B/C - 199B/C Users Manual 3 Jump to open the Mathematics menu. 4 Select Function: Spectrum. 5 Select Window: Auto (automatic windowing), Hanning, Hamming, or None (no windowing).
FE 192,196,199 C Users Manual
Page 32
... only) You can be stored. To use a reference waveform as a reference waveform refer also to be stored Store Pass: each scope screen with Associated Setups. Fluke 192B - 196B/C - 199B/C Users Manual 7 Store the momentary waveform and display it as a test template for the actual waveform.
... only) You can be stored. To use a reference waveform as a reference waveform refer also to be stored Store Pass: each scope screen with Associated Setups. Fluke 192B - 196B/C - 199B/C Users Manual 7 Store the momentary waveform and display it as a test template for the actual waveform.
FE 192,196,199 C Users Manual
Page 36
Measuring Resistance Values To measure a resistance, do the following: 1 Connect the red and black test leads from the 4-mm banana jack inputs to the resistor. 2 Display the METER key labels. 3 Open the Measurement menu. 4 Highlight Ohms. 5 Select Ohms measurement. 28 The resistor value is displayed in ohms. Observe also that the bargraph is displayed. (See Figure 13.) Figure 13. Resistor Value Readings Fluke 192B - 196B/C - 199B/C Users Manual Making Multimeter Measurements The screen displays the numeric readings of the measurements on the meter input.
Measuring Resistance Values To measure a resistance, do the following: 1 Connect the red and black test leads from the 4-mm banana jack inputs to the resistor. 2 Display the METER key labels. 3 Open the Measurement menu. 4 Highlight Ohms. 5 Select Ohms measurement. 28 The resistor value is displayed in ohms. Observe also that the bargraph is displayed. (See Figure 13.) Figure 13. Resistor Value Readings Fluke 192B - 196B/C - 199B/C Users Manual Making Multimeter Measurements The screen displays the numeric readings of the measurements on the meter input.
FE 192,196,199 C Users Manual
Page 38
Now, you will see a screen like in the menu, e.g. 10 mV/A. 7 Accept the current measurement. Fluke 192B - 196B/C - 199B/C Users Manual 6 Observe the sensitivity of the current probe. Ampere Measurement Readings 30 Highlight the corresponding sensitivity in Figure 15 Figure 15.
Now, you will see a screen like in the menu, e.g. 10 mV/A. 7 Accept the current measurement. Fluke 192B - 196B/C - 199B/C Users Manual 6 Observe the sensitivity of the current probe. Ampere Measurement Readings 30 Highlight the corresponding sensitivity in Figure 15 Figure 15.
FE 192,196,199 C Users Manual
Page 40
... of the screen after the word REFERENCE. 4 Measure the voltage to be used as reference value. 3 Set RELATIVE to a known good value. 32 Fluke 192B - 196B/C - 199B/C Users Manual Making Relative Measurements A relative measurement displays the present measurement result relative to perform a relative voltage measurement. Figure 16. Observe that the main reading is...
... of the screen after the word REFERENCE. 4 Measure the voltage to be used as reference value. 3 Set RELATIVE to a known good value. 32 Fluke 192B - 196B/C - 199B/C Users Manual Making Relative Measurements A relative measurement displays the present measurement result relative to perform a relative voltage measurement. Figure 16. Observe that the main reading is...