Operation Manual
Page 2
... test set -up . SAFETY INFORMATION The Operator's Manual contains cautions and warnings alerting the user to the test leads. 1 The kit has been specifically designed for Static Control Surfaces includes a hand-held megohmmeter (here-in after referred to as instructed in this publication, where applicable, and is suspected, open circuit output test voltage. 6. Available replacement parts are...
... test set -up . SAFETY INFORMATION The Operator's Manual contains cautions and warnings alerting the user to the test leads. 1 The kit has been specifically designed for Static Control Surfaces includes a hand-held megohmmeter (here-in after referred to as instructed in this publication, where applicable, and is suspected, open circuit output test voltage. 6. Available replacement parts are...
Operation Manual
Page 3
...replace the battery and retest. If the pointer is connected to. Press TEST. The pointer should come to be used . PERIODIC PERFORMANCE TESTING OF INSTALLED...scale. If the pointer is used with any accessories not specifically designed to be an electrical ... table top or other stable surface. Set the Main Selector Switch to BATTERY TEST. CONTINUITY TEST Place the Instrument ...BATTERY CONTINUITY TAUT BAND SUSPENSION BATTERY TEST CONTINUITY TEST OFF SURFACE TEST 10V 100V TEST CALIBRATION PLATE READ MANUAL BEFORE USE Figure 4-1: Continuity Test OPERATING INSTRUCTIONS...
...replace the battery and retest. If the pointer is connected to. Press TEST. The pointer should come to be used . PERIODIC PERFORMANCE TESTING OF INSTALLED...scale. If the pointer is used with any accessories not specifically designed to be an electrical ... table top or other stable surface. Set the Main Selector Switch to BATTERY TEST. CONTINUITY TEST Place the Instrument ...BATTERY CONTINUITY TAUT BAND SUSPENSION BATTERY TEST CONTINUITY TEST OFF SURFACE TEST 10V 100V TEST CALIBRATION PLATE READ MANUAL BEFORE USE Figure 4-1: Continuity Test OPERATING INSTRUCTIONS...
Operation Manual
Page 4
...the test setup is used to verify the correct installation of ... Complete BATTERY TEST and CONTINUITY TEST. Set the ...using the 100 volts SURFACE TEST. Allow the surface to ground; repair and retest. 2. QUALIFICATION OF INSTALLED...using one of six samples should be followed when testing installed static control surfaces. NOTE: Values obtained by the user...BATTERY CONTINUITY TAUT BAND SUSPENSION BATTERY TEST CONTINUITY TEST OFF SURFACE TEST 10V 100V TEST TO TEST WEIGHT AT TEST POINT A, B, C, D & E READ MANUAL BEFORE USE Figure 5-1: Periodic Performance Tests of Installed...
...the test setup is used to verify the correct installation of ... Complete BATTERY TEST and CONTINUITY TEST. Set the ...using the 100 volts SURFACE TEST. Allow the surface to ground; repair and retest. 2. QUALIFICATION OF INSTALLED...using one of six samples should be followed when testing installed static control surfaces. NOTE: Values obtained by the user...BATTERY CONTINUITY TAUT BAND SUSPENSION BATTERY TEST CONTINUITY TEST OFF SURFACE TEST 10V 100V TEST TO TEST WEIGHT AT TEST POINT A, B, C, D & E READ MANUAL BEFORE USE Figure 5-1: Periodic Performance Tests of Installed...
Operation Manual
Page 5
...clean, electrodes as (RTS-GP). Connect the test leads to an appropriate electrical ground. TEST DESCRIPTION PERIODIC PERFORMANCE TESTING OF INSTALLED OR APPLIED FLOORING MATERIALS(Measurement of the test methods and procedures outlined in Figure 5-5; The ground point may be done on...10M 1M FAIL 100V 105 0 1M 100K 100K PASS 10K 0 0 BATTERY CONTINUITY TAUT BAND SUSPENSION TO TEST WEIGHT AT TEST POINT A & B BATTERY TEST CONTINUITY TEST OFF SURFACE TEST 10V 100V TEST READ MANUAL BEFORE USE Figure 5-5: Material Evaluation Surface to 1.0 x 1011 ohms. The following ...
...clean, electrodes as (RTS-GP). Connect the test leads to an appropriate electrical ground. TEST DESCRIPTION PERIODIC PERFORMANCE TESTING OF INSTALLED OR APPLIED FLOORING MATERIALS(Measurement of the test methods and procedures outlined in Figure 5-5; The ground point may be done on...10M 1M FAIL 100V 105 0 1M 100K 100K PASS 10K 0 0 BATTERY CONTINUITY TAUT BAND SUSPENSION TO TEST WEIGHT AT TEST POINT A & B BATTERY TEST CONTINUITY TEST OFF SURFACE TEST 10V 100V TEST READ MANUAL BEFORE USE Figure 5-5: Material Evaluation Surface to 1.0 x 1011 ohms. The following ...
Operation Manual
Page 6
.... D. For testing of floor finishes or monitoring of existing floor materials, test in Figure 6-1. Place the Instrument and test weight at different locations. Complete BATTERY TEST and CONTINUITY TEST. I . Performance by the user of use . Make sure that are visibly dirty. Install new batteries as described in the OFF position before use , the conductive rubber pads on...
.... D. For testing of floor finishes or monitoring of existing floor materials, test in Figure 6-1. Place the Instrument and test weight at different locations. Complete BATTERY TEST and CONTINUITY TEST. I . Performance by the user of use . Make sure that are visibly dirty. Install new batteries as described in the OFF position before use , the conductive rubber pads on...
Operation Manual
Page 7
... (Eveready 505)* *(See replacement parts above for batteries equivalent to the Eveready 505.) Physical Data Instrument 4.6"(H) x 3.3"(W) x 1.8"(D) (11,7 x 8,4 x 4,6 cm) Weight: 12 oz (340 gms) Test Weights(2) Weight: 5 lbs. (2.27 Kg) each ) Test Leads (pair) Carrying Case Insulated "Bulldog" Clip Noninsulated Alligator Clip Operator's Manual Batteries (2) Part Number 3M 701-M, 98-0798-2290-8 3M 701-W, 98-0798-2291-6 3M 701-L, 98-0798...
... (Eveready 505)* *(See replacement parts above for batteries equivalent to the Eveready 505.) Physical Data Instrument 4.6"(H) x 3.3"(W) x 1.8"(D) (11,7 x 8,4 x 4,6 cm) Weight: 12 oz (340 gms) Test Weights(2) Weight: 5 lbs. (2.27 Kg) each ) Test Leads (pair) Carrying Case Insulated "Bulldog" Clip Noninsulated Alligator Clip Operator's Manual Batteries (2) Part Number 3M 701-M, 98-0798-2290-8 3M 701-W, 98-0798-2291-6 3M 701-L, 98-0798...
Operation Manual
Page 8
...instrument. Except where prohibited by law, 3M will be given complete instructions, including a RMA number for the Service and Repair Center. Austin, TX 78726-9000 © 3M 2003 78-8132-1177-4 Limited Liability...8226; Complete return address • Description of the problem You will not be liable for a period of 1 year from this 3M product, whether direct, indirect, special, incidental or ...liability associated with such use. Important Notice Before using this product, you must evaluate it and determine if it prepaid and insured to replace or repair the 3M product or refund ...
...instrument. Except where prohibited by law, 3M will be given complete instructions, including a RMA number for the Service and Repair Center. Austin, TX 78726-9000 © 3M 2003 78-8132-1177-4 Limited Liability...8226; Complete return address • Description of the problem You will not be liable for a period of 1 year from this 3M product, whether direct, indirect, special, incidental or ...liability associated with such use. Important Notice Before using this product, you must evaluate it and determine if it prepaid and insured to replace or repair the 3M product or refund ...