Fluke ScopeMeter Product Datasheet
Page 2
... color labels, measurements and warnings are clearly linked to specific waveforms. See dynamic signal behavior instantaneously The Digital Persistence mode (Fluke 190C) helps to find tiny details in a long waveform, for detailed signal analysis. Advanced trigger modes The ScopeMeter 190 Series...multiple intensity levels and user selectable decay time-it again. This means that 's standard on a digital datastream. like on an analog scope • F ast display update rate for seeing dynamic behavior instantaneously • A utomatic capture and replay of what 's really happening ...
... color labels, measurements and warnings are clearly linked to specific waveforms. See dynamic signal behavior instantaneously The Digital Persistence mode (Fluke 190C) helps to find tiny details in a long waveform, for detailed signal analysis. Advanced trigger modes The ScopeMeter 190 Series...multiple intensity levels and user selectable decay time-it again. This means that 's standard on a digital datastream. like on an analog scope • F ast display update rate for seeing dynamic behavior instantaneously • A utomatic capture and replay of what 's really happening ...
Fluke ScopeMeter Product Datasheet
Page 3
...circuits having to a different "low" level. This way, the instrument will . Store complete Replay cycles for an instant, stable display Scope users know how frustrating it can look back in a while -intermittents. Cursors can also cause a machine to 100 specific events. fully... Kit SCC120 Kit FlukeView® Software for documenting, archiving and analysis FlukeView for visual comparisons, or it is measured on your Fluke ScopeMeter will automatically recognize a power failure and store the waveform data preceding it your PC for automatic "Pass/Fail" testing (...
...circuits having to a different "low" level. This way, the instrument will . Store complete Replay cycles for an instant, stable display Scope users know how frustrating it can look back in a while -intermittents. Cursors can also cause a machine to 100 specific events. fully... Kit SCC120 Kit FlukeView® Software for documenting, archiving and analysis FlukeView for visual comparisons, or it is measured on your Fluke ScopeMeter will automatically recognize a power failure and store the waveform data preceding it your PC for automatic "Pass/Fail" testing (...
Fluke ScopeMeter Product Datasheet
Page 4
... the ScopeMeter 120 can only make the instrument easy to carry and to fit comfortably in the harshest of 20 MHz (Fluke 123) or 40 MHz (Fluke 124, 125) the Fluke 120 Series will capture and display almost any signal. Bus Health mode allows for true on 600 V CAT III industrial ...up to the 33rd harmonic to IEC 529. ScopeMeter® 120 Series: As simple as connecting to the test point and letting the scope do the job right. scope, resistance and continuity measurements • 10:1 Voltage Probe included with an oscilloscope. Measure volts (ac or dc), resistance, capacitance, current...
... the ScopeMeter 120 can only make the instrument easy to carry and to fit comfortably in the harshest of 20 MHz (Fluke 123) or 40 MHz (Fluke 124, 125) the Fluke 120 Series will capture and display almost any signal. Bus Health mode allows for true on 600 V CAT III industrial ...up to the 33rd harmonic to IEC 529. ScopeMeter® 120 Series: As simple as connecting to the test point and letting the scope do the job right. scope, resistance and continuity measurements • 10:1 Voltage Probe included with an oscilloscope. Measure volts (ac or dc), resistance, capacitance, current...
Fluke ScopeMeter Product Datasheet
Page 5
...Fluke 199B Fluke 196B Fluke 192B Fluke 125 Fluke 124 Fluke 123 200 MHz 100 MHz 200 MHz 100 MHz 60 MHz 40 MHz 40 MHz 20 MHz 2.5 GS/s 1.0 GS/s 2.5 GS/s 1.0 GS/s 500 MS/s 25 MS/s 25 MS/s 25 MS/s (covered by real time sample rate) 2.5 GS/s 2.5 GS/s 1.25 GS/s 3000 points 512 points (min/max pairs) 2 scope...- amps - Ordering information Fluke 199C Fluke 199C/S Fluke 196C Fluke 196C/S Fluke 199B Fluke 199B/S Fluke 196B Fluke 196B/S Fluke 192B Fluke 192B/S Fluke 125 Fluke 125/S Fluke 124 Fluke 124/S Fluke 123 Fluke 123/S SCC190 SCC120 OC4USB PM9080 ...
...Fluke 199B Fluke 196B Fluke 192B Fluke 125 Fluke 124 Fluke 123 200 MHz 100 MHz 200 MHz 100 MHz 60 MHz 40 MHz 40 MHz 20 MHz 2.5 GS/s 1.0 GS/s 2.5 GS/s 1.0 GS/s 500 MS/s 25 MS/s 25 MS/s 25 MS/s (covered by real time sample rate) 2.5 GS/s 2.5 GS/s 1.25 GS/s 3000 points 512 points (min/max pairs) 2 scope...- amps - Ordering information Fluke 199C Fluke 199C/S Fluke 196C Fluke 196C/S Fluke 199B Fluke 199B/S Fluke 196B Fluke 196B/S Fluke 192B Fluke 192B/S Fluke 125 Fluke 125/S Fluke 124 Fluke 124/S Fluke 123 Fluke 123/S SCC190 SCC120 OC4USB PM9080 ...
Fluke 123 and 124 Scopemeter Datasheet
Page 1
... included accessories allow hook-up correct triggering. without touching a button. Ordering information FLUKE-123/003 Industrial ScopeMeter 20 MHz FLUKE-123/003S Industrial ScopeMeter 20 MHz with a fast answers wherever you work. operation • Shielded test leads for an instant, stable display Scope users know how difficult triggering can be safely connected to a printer Connect...
... included accessories allow hook-up correct triggering. without touching a button. Ordering information FLUKE-123/003 Industrial ScopeMeter 20 MHz FLUKE-123/003S Industrial ScopeMeter 20 MHz with a fast answers wherever you work. operation • Shielded test leads for an instant, stable display Scope users know how difficult triggering can be safely connected to a printer Connect...
FE 123 & 124 Users Manual
Page 6
Fluke 123/124 Users Manual Freezing the Screen...16 Holding a Stable Reading 16 Making Relative Measurements 17 Selecting Auto/Manual Ranges 18 Changing the Graphic Representation on ... Charging the Rechargeable Battery Pack 44 Keeping Batteries in Optimal Condition 45 Replacing and Disposing of the Rechargeable Battery Pack 46 Using and Adjusting 10:1 Scope Probes 47 Calibrating the Test Tool 49 Parts and Accessories 49 3 Tips and Troubleshooting 53 ii
Fluke 123/124 Users Manual Freezing the Screen...16 Holding a Stable Reading 16 Making Relative Measurements 17 Selecting Auto/Manual Ranges 18 Changing the Graphic Representation on ... Charging the Rechargeable Battery Pack 44 Keeping Batteries in Optimal Condition 45 Replacing and Disposing of the Rechargeable Battery Pack 46 Using and Adjusting 10:1 Scope Probes 47 Calibrating the Test Tool 49 Parts and Accessories 49 3 Tips and Troubleshooting 53 ii
FE 123 & 124 Users Manual
Page 19
... and to choose an item. Press the blue 'ENTER' function key to confirm selection Press 'ENTER' until you to normal mode. Figure 1-4. Note Pressing the SCOPE MENU key a second time closes this menu and resumes normal measurement. Making Selections in a Menu Figure 1-4 shows the basic navigation of the test tool. 11...
... and to choose an item. Press the blue 'ENTER' function key to confirm selection Press 'ENTER' until you to normal mode. Figure 1-4. Note Pressing the SCOPE MENU key a second time closes this menu and resumes normal measurement. Making Selections in a Menu Figure 1-4 shows the basic navigation of the test tool. 11...
FE 123 & 124 Users Manual
Page 21
..., the setup will track these changes. This function optimizes the position, range, time base, and triggering and assures a stable display on left of the waveform. - A scope trace gives a graphical representation of the waveform area. The zero icon ( ) identifies the ground level of the waveform. To enable the Connect-and-View™...
..., the setup will track these changes. This function optimizes the position, range, time base, and triggering and assures a stable display on left of the waveform. - A scope trace gives a graphical representation of the waveform area. The zero icon ( ) identifies the ground level of the waveform. To enable the Connect-and-View™...
FE 123 & 124 Users Manual
Page 28
You can use waveform smooth to WAVEFORM MODE. Jump to suppress noise without smoothing are shown in Figure 1-11. 20 Figure 1-11. Highlight SMOOTH. Waveform samples with and without loss of bandwidth. Accept waveform smooth. Open the SCOPE OPTIONS submenu. Smoothing the Waveform Fluke 123/124 Users Manual Smoothing the Waveform To smooth the waveform, do the following: Open the SCOPE INPUTS menu.
You can use waveform smooth to WAVEFORM MODE. Jump to suppress noise without smoothing are shown in Figure 1-11. 20 Figure 1-11. Highlight SMOOTH. Waveform samples with and without loss of bandwidth. Accept waveform smooth. Open the SCOPE OPTIONS submenu. Smoothing the Waveform Fluke 123/124 Users Manual Smoothing the Waveform To smooth the waveform, do the following: Open the SCOPE INPUTS menu.
FE 123 & 124 Users Manual
Page 31
... To catch single events, you can perform a single shot. (One time screen update.) To set up the test tool for a Single Shot. 23 Open the SCOPE OPTIONS submenu. Changing the TrendPlot Reading To toggle the TrendPlot reading between MIN (minimum), MAX (maximum), and AVERAGE, do the following : Change MIN into AVG...
... To catch single events, you can perform a single shot. (One time screen update.) To set up the test tool for a Single Shot. 23 Open the SCOPE OPTIONS submenu. Changing the TrendPlot Reading To toggle the TrendPlot reading between MIN (minimum), MAX (maximum), and AVERAGE, do the following : Change MIN into AVG...
FE 123 & 124 Users Manual
Page 33
... MODE. (2x) Start Recording. Observe that during recording no measurements are only displayed after is especially useful when you measure lower frequency waveforms. Open the SCOPE INPUTS menu. The waveform moves accross the screen from right to left like a normal chart recorder. Recording Waveforms over a Long Period of Time The roll...
... MODE. (2x) Start Recording. Observe that during recording no measurements are only displayed after is especially useful when you measure lower frequency waveforms. Open the SCOPE INPUTS menu. The waveform moves accross the screen from right to left like a normal chart recorder. Recording Waveforms over a Long Period of Time The roll...
FE 123 & 124 Users Manual
Page 34
...the following: Open the SCOPE INPUTS menu. Highlight INVERT. (3x) Accept inverted waveform display. Reversing the Polarity of INPUT A). For example, a negative-going waveform is identified by trace identifier A on a DC signal. Highlight AC. (4x) Accept AC-coupling for INPUT A. Fluke 123/124 Users Manual Selecting... AC-Coupling Use AC-coupling when you wish to observe a small AC signal that rides on left of the waveform area. 26 Open the SCOPE INPUTS menu.
...the following: Open the SCOPE INPUTS menu. Highlight INVERT. (3x) Accept inverted waveform display. Reversing the Polarity of INPUT A). For example, a negative-going waveform is identified by trace identifier A on a DC signal. Highlight AC. (4x) Accept AC-coupling for INPUT A. Fluke 123/124 Users Manual Selecting... AC-Coupling Use AC-coupling when you wish to observe a small AC signal that rides on left of the waveform area. 26 Open the SCOPE INPUTS menu.
FE 123 & 124 Users Manual
Page 36
... auto range triggering for waveforms from 1 Hz, do the following: Open the SCOPE INPUTS menu. Select FREE RUN. Note Setting the automatic triggering to normal measurement. Highlight FREE RUN. Highlight Input 'A'. Highlight >1 Hz 28 Open the TRIGGER submenu. Fluke 123/124 Users Manual Selecting the Trigger Parameters To trigger on bottom of the...
... auto range triggering for waveforms from 1 Hz, do the following: Open the SCOPE INPUTS menu. Select FREE RUN. Note Setting the automatic triggering to normal measurement. Highlight FREE RUN. Highlight Input 'A'. Highlight >1 Hz 28 Open the TRIGGER submenu. Fluke 123/124 Users Manual Selecting the Trigger Parameters To trigger on bottom of the...
FE 123 & 124 Users Manual
Page 43
... cursor. } Move the lower cursor to make precise digital measurements on waveforms. This can still use the cursors for a voltage measurement, do the following: From Scope mode, display the Cursor Key functions. Voltage Measurement with Cursors 35 Cursors allow you can be done on live waveforms and on saved waveforms. Using... Horizontal Cursors on the screen. 1 Using The Test Tool Making Cursor Measurements Note Even when the key labels are displayed. Making Cursor Measurements Fluke 124 has cursors.
... cursor. } Move the lower cursor to make precise digital measurements on waveforms. This can still use the cursors for a voltage measurement, do the following: From Scope mode, display the Cursor Key functions. Voltage Measurement with Cursors 35 Cursors allow you can be done on live waveforms and on saved waveforms. Using... Horizontal Cursors on the screen. 1 Using The Test Tool Making Cursor Measurements Note Even when the key labels are displayed. Making Cursor Measurements Fluke 124 has cursors.
FE 123 & 124 Users Manual
Page 44
... Cursors on the waveform. 36 } Highlight the right cursor. ~ Move the right cursor to measure the amplitude, high and low value, or overshoot of a waveform. Fluke 123/124 Users Manual The readout shows the voltage difference between the two cursors and the voltages at the cursors in relation to highlight . Use horizontal... are displayed. Figure 1-20. Highlight the left cursor to the desired position on a Waveform To use the cursors for a time measurement, do the following: From Scope mode, display the Cursor Key functions. See Figure 1-19). Move the left cursor.
... Cursors on the waveform. 36 } Highlight the right cursor. ~ Move the right cursor to measure the amplitude, high and low value, or overshoot of a waveform. Fluke 123/124 Users Manual The readout shows the voltage difference between the two cursors and the voltages at the cursors in relation to highlight . Use horizontal... are displayed. Figure 1-20. Highlight the left cursor to the desired position on a Waveform To use the cursors for a time measurement, do the following: From Scope mode, display the Cursor Key functions. See Figure 1-19). Move the left cursor.
FE 123 & 124 Users Manual
Page 45
... two markers (See Figure 1-20). AUTO automatically does steps 4 to highlight (rise time). Making Rise Time Measurements To measure rise time, do the following: From Scope mode, display the Cursor Key functions. Move the upper cursor to 0% of the trace height. Observe that two horizontal cursors are displayed. The signal frequency...
... two markers (See Figure 1-20). AUTO automatically does steps 4 to highlight (rise time). Making Rise Time Measurements To measure rise time, do the following: From Scope mode, display the Cursor Key functions. Move the upper cursor to 0% of the trace height. Observe that two horizontal cursors are displayed. The signal frequency...
FE 123 & 124 Users Manual
Page 46
... a 10:1 Probe: Probe Attenuation. Proceed as follows to adapt the Test Tool's voltage readout to input B: Open the Scope menu. Use of this Probe is for High Frequency Measurements. Open the Probes menu. Fluke 123/124 Users Manual The reading now shows the risetime from 10%-90% of the trace amplitude and the...
... a 10:1 Probe: Probe Attenuation. Proceed as follows to adapt the Test Tool's voltage readout to input B: Open the Scope menu. Use of this Probe is for High Frequency Measurements. Open the Probes menu. Fluke 123/124 Users Manual The reading now shows the risetime from 10%-90% of the trace amplitude and the...
FE 123 & 124 Users Manual
Page 47
... the OPTICAL PORT of the test tool. Observe that the 10 times attenuation of the Probe is compensated in Chapter 2 under 'Using and Adjusting 10:1 Scope Probes'. 1 Using The Test Tool Using a Printer Using a Printer To print a (graphic) hard copy of the present screen, you need to use one of the...
... the OPTICAL PORT of the test tool. Observe that the 10 times attenuation of the Probe is compensated in Chapter 2 under 'Using and Adjusting 10:1 Scope Probes'. 1 Using The Test Tool Using a Printer Using a Printer To print a (graphic) hard copy of the present screen, you need to use one of the...
FE 123 & 124 Users Manual
Page 55
... Ni-Cd (standard) or BP130 Ni-MH (extended operating time) battery pack. Adjusting Scope Probes 47 For Fluke 123 use the Fluke BP130 Ni-MH battery pack . 8. Using and Adjusting 10:1 Scope Probes Note The 10:1 voltage probe that the battery pack is placed in the battery compartment as shown in Figure 2-2. Warning To avoid...
... Ni-Cd (standard) or BP130 Ni-MH (extended operating time) battery pack. Adjusting Scope Probes 47 For Fluke 123 use the Fluke BP130 Ni-MH battery pack . 8. Using and Adjusting 10:1 Scope Probes Note The 10:1 voltage probe that the battery pack is placed in the battery compartment as shown in Figure 2-2. Warning To avoid...
FE 123 & 124 Users Manual
Page 56
Open the PROBES submenu. A square wave appears on the screen. 48 Adjust the trimmer screw in the probe housing to normal mode. Fluke 123/124 Users Manual Open the SCOPE INPUTS menu. Return to give an optimum square wave. Open the PROBE AC ADJUST submenu. Highlight PROBE AC ADJUST. Highlight ADJUST 10:1 PROBE.
Open the PROBES submenu. A square wave appears on the screen. 48 Adjust the trimmer screw in the probe housing to normal mode. Fluke 123/124 Users Manual Open the SCOPE INPUTS menu. Return to give an optimum square wave. Open the PROBE AC ADJUST submenu. Highlight PROBE AC ADJUST. Highlight ADJUST 10:1 PROBE.