Page 3
... OF CONTENTS REGISTER YOUR PRODUCT 4 LABEL LOCATION 7 Models 9805 and 9805P...7 SPECIAL NOTES 8 TYPICAL PRODUCT PARAMETERS 9 SECTION 1-GENERAL GUIDELINES 10 Assembling the Lift ...10 Operating the Lift ...10 Using the Sling ...11 Lifting the Patient...11 Transferring the Patient ...12 Performing Maintenance ...12 SECTION 2-FEATURES 13 Hydraulic Lifts...13 Patient Slings ...13 Model Nos. 9042, 9043...
... OF CONTENTS REGISTER YOUR PRODUCT 4 LABEL LOCATION 7 Models 9805 and 9805P...7 SPECIAL NOTES 8 TYPICAL PRODUCT PARAMETERS 9 SECTION 1-GENERAL GUIDELINES 10 Assembling the Lift ...10 Operating the Lift ...10 Using the Sling ...11 Lifting the Patient...11 Transferring the Patient ...12 Performing Maintenance ...12 SECTION 2-FEATURES 13 Hydraulic Lifts...13 Patient Slings ...13 Model Nos. 9042, 9043...
Owners Manual
Page 4
... at www.invacare.com - Safeguard your purchase. 3. Receive updates with product information, maintenance tips, and industry news. 4. or - TABLE OF CONTENTS TABLE OF CONTENTS SECTION 7-OPERATION 26 Operating the Patient Lift ...26 Opening the Legs of the Adjustable Base 26 Raising/Lowering Model Hydraulic Lifts 27 Raising the Lift 27 Lowering the Lift 28 SECTION 8-LIFTING THE...
... at www.invacare.com - Safeguard your purchase. 3. Receive updates with product information, maintenance tips, and industry news. 4. or - TABLE OF CONTENTS TABLE OF CONTENTS SECTION 7-OPERATION 26 Operating the Patient Lift ...26 Opening the Legs of the Adjustable Base 26 Raising/Lowering Model Hydraulic Lifts 27 Raising the Lift 27 Lowering the Lift 28 SECTION 8-LIFTING THE...
Owners Manual
Page 9
...: BASE LENGTH: BOOM HEIGHT AT SLING ATTACHMENT: MIN.: MAX.: RANGE: BOOM OVERALL HEIGHT: WEIGHT CAPACITY: *CASTER SIZE: SHIPPING WEIGHT: (UNIT IN CARTON) 9805 Hydraulic Adjustable Lift 9884 9800A 42¼-inches 22-inches 6½-inches 46¾-inches 17-inches 61-inches 44-inches 73-inches 450 lbs 5-inches 85 lbs... 9805P Hydraulic Adjustable Lift - Painted 9884P N/A 42¼-inches 22-inches 6½-inches 46¾-inches 17-inches 61-inches 44-inches 73-inches 450 lbs 5-...
...: BASE LENGTH: BOOM HEIGHT AT SLING ATTACHMENT: MIN.: MAX.: RANGE: BOOM OVERALL HEIGHT: WEIGHT CAPACITY: *CASTER SIZE: SHIPPING WEIGHT: (UNIT IN CARTON) 9805 Hydraulic Adjustable Lift 9884 9800A 42¼-inches 22-inches 6½-inches 46¾-inches 17-inches 61-inches 44-inches 73-inches 450 lbs 5-inches 85 lbs... 9805P Hydraulic Adjustable Lift - Painted 9884P N/A 42¼-inches 22-inches 6½-inches 46¾-inches 17-inches 61-inches 44-inches 73-inches 450 lbs 5-...
Owners Manual
Page 13
... safety. Slings are interchangeable with any Invacare patient lift. Each sling offers reinforcing at each hook-on point to the lift by the individual's doctor, nurse or medical assistant for convenience of the individual being lifted. Part No 1024492 13 Portable Patient Lift and Sling SECTION 2-FEATURES SECTION 2-FEATURES Hydraulic Lifts NOTE: For this procedure, refer to...
... safety. Slings are interchangeable with any Invacare patient lift. Each sling offers reinforcing at each hook-on point to the lift by the individual's doctor, nurse or medical assistant for convenience of the individual being lifted. Part No 1024492 13 Portable Patient Lift and Sling SECTION 2-FEATURES SECTION 2-FEATURES Hydraulic Lifts NOTE: For this procedure, refer to...
Owners Manual
Page 27
... Hydraulic Lifts NOTE: For this procedure, refer to elevate the boom and the patient. Part No 1024492 27 Portable Patient Lift and Sling DETAIL "A" Shifter Handle Base Closed Base Open SECTION 7-OPERATION Shifter Handle DETAIL "B" Locked Step 1 Step 2 Step 3 FIGURE 7.1 Operating the Patient Lift ...- The control valve 2. The pump handle Raising the Lift The control valve MUST be in the closed position (control valve positioned towards pump handle) to move the...
... Hydraulic Lifts NOTE: For this procedure, refer to elevate the boom and the patient. Part No 1024492 27 Portable Patient Lift and Sling DETAIL "A" Shifter Handle Base Closed Base Open SECTION 7-OPERATION Shifter Handle DETAIL "B" Locked Step 1 Step 2 Step 3 FIGURE 7.1 Operating the Patient Lift ...- The control valve 2. The pump handle Raising the Lift The control valve MUST be in the closed position (control valve positioned towards pump handle) to move the...
Owners Manual
Page 28
...away from pump handle) to lower the boom and the patient. Raising/Lowering Model Hydraulic Lifts - Pump Handle Pump Handle Closed Control Valve Open FIGURE 7.2 Operating the Patient Lift - Raising the Lift Portable Patient Lift and Sling 28 Part No 1024492 NOTE: A safety gate is part of the... hydraulic system that the control valve is opened . SECTION 7-OPERATION Lowering the Lift The control handle MUST be controlled by ...
...away from pump handle) to lower the boom and the patient. Raising/Lowering Model Hydraulic Lifts - Pump Handle Pump Handle Closed Control Valve Open FIGURE 7.2 Operating the Patient Lift - Raising the Lift Portable Patient Lift and Sling 28 Part No 1024492 NOTE: A safety gate is part of the... hydraulic system that the control valve is opened . SECTION 7-OPERATION Lowering the Lift The control handle MUST be controlled by ...
Owners Manual
Page 29
... For this procedure, refer to clear the bed and their weight supported by the lift. NOTE: The manual (crank) lift handle will stay in position until the control valve is opened. 5. NOTE: On a hydraulic lift, the boom will automatically turn back one of the following: • Chains, O-... proceeding further and observe all WARNINGS indicated. Using the steering handle, push the patient lift underneath the bed. 2. SECTION 8-LIFTING THE PATIENT SECTION 8-LIFTING THE PATIENT ƽ WARNING Before lifting or transferring the patient, the base legs MUST be locked in the open end facing...
... For this procedure, refer to clear the bed and their weight supported by the lift. NOTE: The manual (crank) lift handle will stay in position until the control valve is opened. 5. NOTE: On a hydraulic lift, the boom will automatically turn back one of the following: • Chains, O-... proceeding further and observe all WARNINGS indicated. Using the steering handle, push the patient lift underneath the bed. 2. SECTION 8-LIFTING THE PATIENT SECTION 8-LIFTING THE PATIENT ƽ WARNING Before lifting or transferring the patient, the base legs MUST be locked in the open end facing...
Owners Manual
Page 31
... and push from the bed. 8. MOVING THE PATIENT IN THE LIFT FIGURE 8.4 Lifting the Patient Part No 1024492 31 Portable Patient Lift and Sling Turn crank handle counterclockwise (manual) or open control valve (hydraulic) lowering patient so that he/she faces assistant operating the patient lift (FIGURE 8.4). 9. TURNING THE PATIENT DETAIL "B" - Using the steering handles...
... and push from the bed. 8. MOVING THE PATIENT IN THE LIFT FIGURE 8.4 Lifting the Patient Part No 1024492 31 Portable Patient Lift and Sling Turn crank handle counterclockwise (manual) or open control valve (hydraulic) lowering patient so that he/she faces assistant operating the patient lift (FIGURE 8.4). 9. TURNING THE PATIENT DETAIL "B" - Using the steering handles...
Owners Manual
Page 41
...oil (about a drop) will accumulate around the piston from the wheel and swivel bearings. Lubrication The Invacare lift is completely enclosed and sealed with a facial tissue. ƽ WARNING The hydraulic pump is sealed at the factory and if service is no adjustment or maintenance the casters, other ...removed with Neoprene rings to prevent leakage of the Invacare Lift are exposed to extreme moist conditions. The connection between the mast and the base should be cleaned and coated with minimum care and maintenance. All parts of hydraulic oil. however, a six month check and ...
...oil (about a drop) will accumulate around the piston from the wheel and swivel bearings. Lubrication The Invacare lift is completely enclosed and sealed with a facial tissue. ƽ WARNING The hydraulic pump is sealed at the factory and if service is no adjustment or maintenance the casters, other ...removed with Neoprene rings to prevent leakage of the Invacare Lift are exposed to extreme moist conditions. The connection between the mast and the base should be cleaned and coated with minimum care and maintenance. All parts of hydraulic oil. however, a six month check and ...
Owners Manual
Page 44
... Medallion design, and Yes, you purchased your exclusive remedy under this warranty shall be refused; Invacare warrants the products manufactured to be returned for a period of three years on the lift and one year on the back cover, provide dealer's name, address, date of purchase, indicate nature of... other legal rights which vary from state to be defective, such product shall be repaired or replaced, at the address on the slings, hydraulic pump / electric components from whom you can. The defective unit or parts MUST be free from defects in replacement part installation or repair...
... Medallion design, and Yes, you purchased your exclusive remedy under this warranty shall be refused; Invacare warrants the products manufactured to be returned for a period of three years on the lift and one year on the back cover, provide dealer's name, address, date of purchase, indicate nature of... other legal rights which vary from state to be defective, such product shall be repaired or replaced, at the address on the slings, hydraulic pump / electric components from whom you can. The defective unit or parts MUST be free from defects in replacement part installation or repair...