Owners Manual - Page 3
... Label Location 11 3.2 Technical Data 12
4 Setup 15 4.1 Unpacking 15 4.2 Mattress Replacement System Installation 16 4.3 Installing the Side Rails 17 4.4 Installing the Power Unit 17 4.5 Connecting the Hose ...32
6.1 Service Life 32 6.2 Cleaning the System 32 6.3 Disposal 34 6.4 Storing the System 34 6.5 Changing the Filter (if equipped 35
7 Troubleshooting 36
7.1 Troubleshooting 36
8...
Owners Manual - Page 4
... both. Warning indicates a potentially hazardous
situation which, if not avoided, could result in death or serious injury. Gives useful tips, recommendations and information of efficient, trouble-free use.
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE.
4
1148139-H
microAIR® MA90 and MA95 Series
1 General
1.1 Symbols
Signal words...
Owners Manual - Page 5
... Invacare for
repair if product is damaged or malfunctioning.
1148139-H
Safety
DANGER! Risk of Death, Injury or Damage Conditions such as user manuals, service manuals or instruction sheets supplied ...with these instructions and any additional instructional material such as restlessness, mental deterioration and dementia or seizure disorders (uncontrolled body movement), sleeping problems, and...
Owners Manual - Page 8
...
learn about the risks of entrapment. - User manuals are available at www.invacare.com for repair if leakage is required. Work with therapist, physician and other medical staff to the Bed Rail Entrapment Risk Notification Guide at www.invacare.com or your dealer. -
Remove product from service and contact
dealer or Invacare for additional...
Owners Manual - Page 10
... or missing hardware: -
Repair and Service
DANGER! Always maintain and inspect equipment per the instructions in well ventilated area. - Sharp edges or surfaces should be corrected or replaced. - Risk Of Injury Or Damage To avoid injury or damage due to the mattress: - Place product in this manual. Parts showing excessive wear should be replaced. - Keep openings free...
Owners Manual - Page 13
Performance Specifications
MA90Z
MA95Z
Mattress Weight Capacity
Standard - 36 inch:
350 lbs
Bariatric - 42 Inch:
650 lbs
Bariatric - 48 Inch:
1000 lbs
Pressure Zone
Maximum Flow:
50 ± 15 LPM 1275 ± 100 LPM
Maximum Pressure:
35 ± 5 mmHg
Maximum Flow Timer:
30 minutes
15 minutes
Support Surface Inflation Time:
5-10 minutes
25...
Owners Manual - Page 15
...a qualified technician or Invacare for repair. DO NOT use of damage or if the product is complete. - Use proper lifting techniques (lift with damaged parts could lead to the product malfunctioning,... causing injury to the user and/or caregiver. - DO NOT attempt to operate product controls
prior to lift or
handle product safely.
1148139-H
Setup...
Owners Manual - Page 19
... the side of obstruction (i.e. Plug the other end of the following :
a. Keep all moving parts, including the main
frame, mattress deck, and all cord(s) are routed and secured
properly. -... control unit.
3. Call your supplier for a replacement hospital grade cord.
• If the plug is flush with the
control unit connector.
1148139-H
Setup
4.6 Connecting the Power Cord
WARNING! Lift and ...
Owners Manual - Page 24
... recommended that Max Inflate setting be manually disengaged by approximately 20%. After 30 minutes, the Max...LED displays the patient comfort pressure levels and provides a guide to the caregiver to inflate fully (inflation time depends on... (MA95Z Power Unit)
Low Air Loss Mode
1. MA90Z models display levels from 0 to 9 and MA95Z models display the mmHg from Soft (8 ±4 mmHg) to...
Owners Manual - Page 25
...i. Ensure the mattress is in the mattress cover
pocket in the mattress to exit Fowler Transmitter Setup without pressing a key.
• MA95Z Power Unit-Perform the following :
- Cancel-Press the...recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or patient cleaning.
5.1.11 Automatic Wireless Fowler
Manual Operation of the mattress...
Owners Manual - Page 27
... on the control unit. See troubleshooting section of this manual if alarm fails to 5.1 Using the Front Panel, page 21. 1. Periodically test alarm by unplugging the power cord from the wall outlet while unit is plugged in, the LCD screen displays the Run Time in the previously set mode. To display the Run...
Owners Manual - Page 28
...settings.
11. Use a regular pillow to 4-finger width clearance between the air cushions directly underneath the sacral region of the mattress (the end with contracted patients to the desired comfort level. 8. Verify that the patient's feet are towards the foot end of the patient's body. Ensure that there is 3 to help support... with the hose).
4. Set the comfort pressure to provide...
Owners Manual - Page 32
... infectious exposure: - Remove the bedding. 2. Ensure that the control unit is contaminated with safety instructions, maintenance intervals and correct use stated in this manual. This product has an expected lifetime of use and dilution.
1. Do not clean the bed ... or damage from the wall outlet. 5. WARNING! microAIR® MA90 and MA95 Series
6 Maintenance
6.1 Service Life
WARNING!
Owners Manual - Page 37
Problem Fowler Transmitter not working Alarm-Power Failure
Alarm-Hose Disconnected
Cause Low voltage in battery Malfunctioning sensor Sensor not synchronized to power unit
Power cord disconnected No power in the power source
Power outage Blown fuse
Mattress hose disconnected
Troubleshooting
Solution Replace battery Replace fowler Refer to Folwer transmitter setup in this manual Connect power ...
Owners Manual - Page 38
...defects in replacement part installation or repair of any such product. This warranty does not include any such product shall be proven to be repaired or replaced, at the...PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO ADHERE TO THE ENCLOSED INSTRUCTIONS.
Invacare's sole obligation and your Invacare product. C.O.D.
shipments will issue a return authorization.
For warranty service...
Invacare MA90Z Reviews
Do you have an experience with the Invacare MA90Z that you would like to share?
Earn 750 points for your review!
We have not received any reviews for Invacare yet.
Earn 750 points for your review!