Owners Manual
Page 3
... OF CONTENTS REGISTER YOUR PRODUCT 4 LABEL LOCATION 7 Models 9805 and 9805P...7 SPECIAL NOTES 8 TYPICAL PRODUCT PARAMETERS 9 SECTION 1-GENERAL GUIDELINES 10 Assembling the Lift ...10 Operating the Lift ...10 Using the Sling ...11 Lifting the Patient...11 Transferring the Patient ...12 Performing Maintenance ...12 SECTION 2-FEATURES 13 Hydraulic Lifts...13 Patient Slings ...13 Model Nos. 9042, 9043, 9046 and...
... OF CONTENTS REGISTER YOUR PRODUCT 4 LABEL LOCATION 7 Models 9805 and 9805P...7 SPECIAL NOTES 8 TYPICAL PRODUCT PARAMETERS 9 SECTION 1-GENERAL GUIDELINES 10 Assembling the Lift ...10 Operating the Lift ...10 Using the Sling ...11 Lifting the Patient...11 Transferring the Patient ...12 Performing Maintenance ...12 SECTION 2-FEATURES 13 Hydraulic Lifts...13 Patient Slings ...13 Model Nos. 9042, 9043, 9046 and...
Owners Manual
Page 4
...invacare.com - Portable Patient Lift and Sling 4 Part No 1024492 TABLE OF CONTENTS TABLE OF CONTENTS SECTION 7-OPERATION 26 Operating the Patient Lift ...26 Opening the Legs of the Adjustable Base 26 Raising/Lowering Model Hydraulic Lifts 27 Raising the Lift 27 Lowering the Lift 28 SECTION 8-LIFTING... Transferring from a Car to improve product designs based on your product. 5. Invacare can contact you submit will enable Invacare to a Wheelchair 39 SECTION 11-CARE AND MAINTENANCE 41 Hydraulic Pump...41 Lubrication ...41 Wear and Damage ...42 Cleaning ...42 Maintenance Safety ...
...invacare.com - Portable Patient Lift and Sling 4 Part No 1024492 TABLE OF CONTENTS TABLE OF CONTENTS SECTION 7-OPERATION 26 Operating the Patient Lift ...26 Opening the Legs of the Adjustable Base 26 Raising/Lowering Model Hydraulic Lifts 27 Raising the Lift 27 Lowering the Lift 28 SECTION 8-LIFTING... Transferring from a Car to improve product designs based on your product. 5. Invacare can contact you submit will enable Invacare to a Wheelchair 39 SECTION 11-CARE AND MAINTENANCE 41 Hydraulic Pump...41 Lubrication ...41 Wear and Damage ...42 Cleaning ...42 Maintenance Safety ...
Owners Manual
Page 9
... PARAMETERS BASE (ONLY): MAST (ONLY): BASE WIDTH OPEN: CLOSED: BASE HEIGHT: BASE LENGTH: BOOM HEIGHT AT SLING ATTACHMENT: MIN.: MAX.: RANGE: BOOM OVERALL HEIGHT: WEIGHT CAPACITY: *CASTER SIZE: SHIPPING WEIGHT: (UNIT IN CARTON) 9805 Hydraulic Adjustable Lift 9884 9800A 42¼-inches 22-inches 6½-inches 46¾-inches 17-inches 61-inches 44...
... PARAMETERS BASE (ONLY): MAST (ONLY): BASE WIDTH OPEN: CLOSED: BASE HEIGHT: BASE LENGTH: BOOM HEIGHT AT SLING ATTACHMENT: MIN.: MAX.: RANGE: BOOM OVERALL HEIGHT: WEIGHT CAPACITY: *CASTER SIZE: SHIPPING WEIGHT: (UNIT IN CARTON) 9805 Hydraulic Adjustable Lift 9884 9800A 42¼-inches 22-inches 6½-inches 46¾-inches 17-inches 61-inches 44...
Owners Manual
Page 13
...lifting and transferring procedures. H A. The Mast D. The Boom E G B F E. Hydraulic Pump G. Part No 1024492 13 Portable Patient Lift and Sling Lock together. Operation optimizes lift capabilities. Control Valve A FIGURE 2.1 Hydraulic Lifts Patient Slings ƽ WARNING Use a sling that is 450 lbs. The lift ...lift path. It maximizes full travel range. The Castered Base B. Swivel Bar F. Slings attach to -side for the comfort and safety of assistant. Slings are interchangeable with any Invacare patient lift. They are adjustable. Each sling ...
...lifting and transferring procedures. H A. The Mast D. The Boom E G B F E. Hydraulic Pump G. Part No 1024492 13 Portable Patient Lift and Sling Lock together. Operation optimizes lift capabilities. Control Valve A FIGURE 2.1 Hydraulic Lifts Patient Slings ƽ WARNING Use a sling that is 450 lbs. The lift ...lift path. It maximizes full travel range. The Castered Base B. Swivel Bar F. Slings attach to -side for the comfort and safety of assistant. Slings are interchangeable with any Invacare patient lift. They are adjustable. Each sling ...
Owners Manual
Page 27
Opening the Legs of the Adjustable Base Raising/Lowering Model Hydraulic Lifts NOTE: For this procedure, refer to elevate the boom and the patient. There are two controls on page 28. The control valve 2. DETAIL "A" Shifter Handle ...Base Closed Base Open SECTION 7-OPERATION Shifter Handle DETAIL "B" Locked Step 1 Step 2 Step 3 FIGURE 7.1 Operating the Patient Lift - The pump handle Raising the Lift The control valve MUST be in the closed position (control valve positioned towards pump handle) to move the pump handle up and down...
Opening the Legs of the Adjustable Base Raising/Lowering Model Hydraulic Lifts NOTE: For this procedure, refer to elevate the boom and the patient. There are two controls on page 28. The control valve 2. DETAIL "A" Shifter Handle ...Base Closed Base Open SECTION 7-OPERATION Shifter Handle DETAIL "B" Locked Step 1 Step 2 Step 3 FIGURE 7.1 Operating the Patient Lift - The pump handle Raising the Lift The control valve MUST be in the closed position (control valve positioned towards pump handle) to move the pump handle up and down...
Owners Manual
Page 28
... Pump Handle Closed Control Valve Open FIGURE 7.2 Operating the Patient Lift - Raising/Lowering Model Hydraulic Lifts - SECTION 7-OPERATION Lowering the Lift The control handle MUST be controlled by varying the amount that ...maintains a SLOW constant descent of the boom regardless of how far the control valve is opened . NOTE: A safety gate is part of the hydraulic system that the control valve is opened . Raising the Lift Portable Patient Lift and Sling...
... Pump Handle Closed Control Valve Open FIGURE 7.2 Operating the Patient Lift - Raising/Lowering Model Hydraulic Lifts - SECTION 7-OPERATION Lowering the Lift The control handle MUST be controlled by varying the amount that ...maintains a SLOW constant descent of the boom regardless of how far the control valve is opened . NOTE: A safety gate is part of the hydraulic system that the control valve is opened . Raising the Lift Portable Patient Lift and Sling...
Owners Manual
Page 29
...the patient lift underneath the bed. 2. NOTE: On a hydraulic lift, the boom will automatically turn back one of the following: • Chains, O-Rings and D-Rings: After the chains, O-Rings or D-Rings are attached and the S-hooks are secured in both sides of the sling sections (open... position for easy attachment of the swivel bar. Refer to FIGURE 8.1, FIGURE 8.2, FIGURE 8.3, and FIGURE 8.4. 1. Strap Hooks Strap Hooks Chain Hooks FIGURE 8.1 Lifting the Patient 4. The patient should be raised to a ...
...the patient lift underneath the bed. 2. NOTE: On a hydraulic lift, the boom will automatically turn back one of the following: • Chains, O-Rings and D-Rings: After the chains, O-Rings or D-Rings are attached and the S-hooks are secured in both sides of the sling sections (open... position for easy attachment of the swivel bar. Refer to FIGURE 8.1, FIGURE 8.2, FIGURE 8.3, and FIGURE 8.4. 1. Strap Hooks Strap Hooks Chain Hooks FIGURE 8.1 Lifting the Patient 4. The patient should be raised to a ...
Owners Manual
Page 31
... counterclockwise (manual) or open control valve (hydraulic) lowering patient so that he/she faces assistant operating the patient lift (FIGURE 8.4). 9. Pull the patient lift away from the bed and push from the bed. 8. MOVING THE PATIENT IN THE LIFT FIGURE 8.4 Lifting the Patient Part No 1024492 31 Portable Patient Lift and Sling DETAIL "A" - Using the steering handles...
... counterclockwise (manual) or open control valve (hydraulic) lowering patient so that he/she faces assistant operating the patient lift (FIGURE 8.4). 9. Pull the patient lift away from the bed and push from the bed. 8. MOVING THE PATIENT IN THE LIFT FIGURE 8.4 Lifting the Patient Part No 1024492 31 Portable Patient Lift and Sling DETAIL "A" - Using the steering handles...
Owners Manual
Page 41
... six month check and lubrication should be returned to open the hydraulic pump or obtain local service. The casters MUST swivel and roll smoothly. Part No 1024492 41 Portable Patient Lift and Sling All parts of the Invacare Lift are worn, replace these parts IMMEDIATELY. There is completely enclosed ...piston from the wheel and swivel bearings. Keep lift and slings clean and in damage and costly repair. The Invacare Patient Lift is required, the pump unit MUST be noted and reported to keep your patient lift in continuous service. Hydraulic Pump All parts of oil (about a drop...
... six month check and lubrication should be returned to open the hydraulic pump or obtain local service. The casters MUST swivel and roll smoothly. Part No 1024492 41 Portable Patient Lift and Sling All parts of the Invacare Lift are worn, replace these parts IMMEDIATELY. There is completely enclosed ...piston from the wheel and swivel bearings. Keep lift and slings clean and in damage and costly repair. The Invacare Patient Lift is required, the pump unit MUST be noted and reported to keep your patient lift in continuous service. Hydraulic Pump All parts of oil (about a drop...
Owners Manual
Page 44
...LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. Do not return products to be returned for a period of three years on the lift and one year on the back cover, provide dealer's name, address, date of purchase, indicate nature of purchase. This warranty gives ...PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER. This warranty is extended only to be defective, such product shall be repaired or replaced, at the address on the slings, hydraulic pump / electric components from defects in...
...LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. Do not return products to be returned for a period of three years on the lift and one year on the back cover, provide dealer's name, address, date of purchase, indicate nature of purchase. This warranty gives ...PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER. This warranty is extended only to be defective, such product shall be repaired or replaced, at the address on the slings, hydraulic pump / electric components from defects in...