About Error E4
Why the SANYO MEDICOOL come out error E4?
(Posted by mohdjamil88 11 years ago)

Popular Sanyo MPR-1013R Manual Pages

Instruction Manual - Page 1

INSTRUCTION MANUAL Pharmaceutical Refrigerator MPR-513 MPR-513R MPR-1013 MPR-1013R MPR-513R MPR-1013R
Instruction Manual - Page 2

... contact SANYO if any point in this manual may be reproduced in Japan. All rights reserved. CONTENTS PRECAUTIONS FOR SAFE OPERATION CAUTIONS FOR USAGE ENVIRONMENTAL CONDITIONS REFRIGERATOR COMPONENTS INSTALLATION OPTIONAL COMPONENTS START-UP OF UNIT DEFROST CYCLES TEMPERATURE SETTING ALARMS & SAFETY FUNCTIONS ROUTINE MAINTENANCE TROUBLE SHOOTING DISPOSAL OF UNIT SPECIFICATIONS PERFORMANCE...
Instruction Manual - Page 3

... cover should be followed. Be sure to keep this manual in which the electrical components of this unit. < Label on the unit > This mark is prohibited.  ˔ this manual as it contains important safety advice. CAUTION Failure to ... described so that the user complies with this symbol means an instruction must be removed by a qualified engineer or a service personnel only. 2
Instruction Manual - Page 6

...parts such as suffocation may cause current leakage or electric shock. When removing the plug from the introduction of a trapped finger. Do not put the packing plastic bag used for cleaning. Take care not to hot surface. (MPR-1013, MPR-1013R... FOR SAFE OPERATION CAUTION Fix the shelves securely. Incomplete installation may cause electric shock. This may cause injury or damage...
Instruction Manual - Page 7

... the access port. After installation, a rubber cap should ... 12. Plastic and rubber parts can be provided between the...refrigerated by these materials, especially solvents. Check the condenser filter and clean it to clean the unit, wipe it up thoroughly with a cloth soaked in the chamber or compressor failure. (MPR-1013, MPR-1013R... reach the set temperature. 5. Always open . Failure ...
Instruction Manual - Page 10

... and on the bottom of the lower part cover is open. 15. Door switch...90o with pushing the button on . 9. For the replacement, refer to page 10. 3. If this cover ...refrigerating units are mounted inside the cover. Evaporating tray: Defrost water from outside. 5. Air intake vent (in the chamber): Do not block this breaker before the unit starts to be stored in MPR-1013 and MPR-1013R...
Instruction Manual - Page 14

...: Model MPR-S7 sensor MTR-G04 MPR-513 MPR-S7 For model MPR-513R, MPR-1013, and MPR-1013R, the attachment is same. 13 The available recorder and a dedicated fixture is as an optional accessory. If an automatic temperature recorder is required, contact your Sanyo dealer. OPTIONAL COMPONENTS Automatic temperature recorder An automatic temperature recorder is available for this refrigerator as...
Instruction Manual - Page 16

.... 6. Allow the chamber temperature to fall to check the light. DEFROST CYCLES There is installed in the ambient of 35oC and 80% R.H. This will not have any frost which has...the dedicated outlet with 2oC setting in high temperature and high humidity location. Set the desired temperature. Cycle defrost To keep the chamber temperature stable, the refrigeration compressor is cycled on the ...
Instruction Manual - Page 19

...error code has priority on the error display. If the chamber temperature deviates from the set temperature is flashed. E02 and chamber temp. Compressor OFF ----Intermittent tone with the set... When the power to -14oC. Temperature indicator is disconnected or short-circuited. (MPR-1013,1013R only) When power failure or disconnected the power cord. Door alarm When the door...
Instruction Manual - Page 21

...insulation and cause failure. ɾThe compressor and other mechanical parts are completely sealed. Ensure you do not inhale or consume.... These may result in the figure. 2. Cleaning of condenser filter (MPR-1013, 1013R) 1. If the unit is very dirty, use a neutral detergent. ... the power supply to the unit prior to any repair or maintenance of the unit and all accessories. After...
Instruction Manual - Page 22

... below. Remove the screws on the bottom of evaporating tray 1. See Fig. 1. 2. Fig. 1 Clips Fig. 2 For MPR-1013 and MPR-1013R, as shown in Fig. 4, pull out the evaporating tray by pulling out the clips towards front. 3. Replace the tray in the evaporating tray with a neutral detergent and clean water. Then take out the evaporating...
Instruction Manual - Page 23

The fluorescent lamp is placed vertically at the center of the frame (MPR1013 and MPR-1013R) and horizontally at both ends. For MPR-513 and MPR-513R, remove the top shelf, for MPR-1013 and MPR-1013R in the left side of the refrigerator. After replacing a new fluorescent lamp, return it . 6. Attach a new lamp surely. Turn off the fluorescent lamp and then...
Instruction Manual - Page 25

...temp. SPECIFICATIONS  Name Pharmaceutical Refrigerator Model MPR-1013 MPR-1013R MPR-513 MPR-513R ...MPR-1013R), 5 pcs. (MPR-513R) Allowable load; 20 kg/shelf Forced cool air circulation Compressor Hermetic type, Output; 300 W, 1 pc. For chamber cooling; alarm Flashing of temperature display and alarm lamp, Buzzer, (Remote alarm operated) Low temp. "close" Accessories Key set...
Instruction Manual - Page 26

...) 1500 kPa AC 220 V AC 220 V 60 Hz 50 Hz 220 W 205 W AC 230 V 50 Hz 210 W AC 240 V 50 Hz 210 W Model Temperature control range Usable ambient temperature MPR-1013, MPR-1013R +2oC to +14oC -5oC to +35oC Noise level 50 dB (A scale) Maximum pressure 1343 kPa Rated voltage AC 110 V AC 115 V AC...
Instruction Manual - Page 27

... to in order to keep for his and your safety. Safety check sheet 1. Date : Signature : Address, Division : Telephone : Product name: Pharmaceutical Refrigerator Model: MPR-513, 513R MPR-1013, 1013R Serial number: Please decontaminate the unit yourself before servicing. Contamination of the unit a) The unit is safe to work on b) There is some danger (see below . CAUTION Please fill...

Sanyo MPR-1013R Reviews

Do you have an experience with the Sanyo MPR-1013R that you would like to share?
Earn 750 points for your review!
We have not received any reviews for Sanyo yet.

Popular Sanyo MPR-1013R Search Terms

The following terms are frequently used to search for Sanyo MPR-1013R support:

Ask a New Question
Use the box below to post a new question about Sanyo MPR-1013R.
Manuals / Documents
Download any of our Sanyo MPR-1013R manuals for free!

We have the following 1 documents available for the Sanyo MPR-1013R:
  • Instruction Manual

Points & Prizes
  • You can earn points for nearly everything you do on HelpOwl.com
  • You can trade in those points for gift cards at leading retailers such as Amazon.com and Walmart
  • It's that simple!
See How it Works
Create a Free Account

Sanyo Manuals

Find free Sanyo MPR-1013R - Commercial Solutions Lab manuals and user guides available at ManualOwl.com. Try out our unique manual viewer allowing you to interact with manuals from directly within your browser!

Sanyo Reviews

View thousands of Sanyo MPR-1013R - Commercial Solutions Lab user reviews and customer ratings available at ReviewOwl.com.

Contact Information

Complete Sanyo customer service contact information including steps to reach representatives, hours of operation, customer support links and more from ContactHelp.com.

Scoreboard Ratings

See detailed Sanyo customer service rankings, employee comments and much more from our sister site.

Sanyo Recalls

Find comprehensive Sanyo recall information updated hourly on RecallOwl.com.