Owners Manual
Page 1
User Manual Perfecto2™Series Oxygen Concentrators without SensO2®, Oxygen Concentrators with SensO2 HomeFill®Compatible Perfecto2 with SensO2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.com User: BEFORE using this product, read this product.
User Manual Perfecto2™Series Oxygen Concentrators without SensO2®, Oxygen Concentrators with SensO2 HomeFill®Compatible Perfecto2 with SensO2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.com User: BEFORE using this product, read this product.
Owners Manual
Page 7
... wet. DO NOT place or store product where it is required by the physician's prescription, the concentrator MUST be located in contact with oxygen under pressure. Unplug IMMEDIATELY. If continuous usage is not working properly, if it has been dropped ...any lubricants unless recommended by any spark near medical oxygen equipment. DO NOT disassemble. This includes sparks from static electricity created by Invacare. No user serviceable parts. If the concentrator has a damaged cord or plug, if it can drop into water. DO NOT use of this product. Refer servicing to ...
... wet. DO NOT place or store product where it is required by the physician's prescription, the concentrator MUST be located in contact with oxygen under pressure. Unplug IMMEDIATELY. If continuous usage is not working properly, if it has been dropped ...any lubricants unless recommended by any spark near medical oxygen equipment. DO NOT disassemble. This includes sparks from static electricity created by Invacare. No user serviceable parts. If the concentrator has a damaged cord or plug, if it can drop into water. DO NOT use of this product. Refer servicing to ...
Owners Manual
Page 8
SECTION 1-GENERAL GUIDELINES Operating Information For optimum performance, Invacare recommends that Crush‐Proof oxygen tubing be left unattended when plugged in a typical medical installation. NEVER drop or insert any object or liquid into any opening. DO NOT connect the concentrator in parallel or series with this product and not exceed... Interference This equipment has been tested and found to provide a reasonable protection against electromagnetic interference in . Shorter periods of 30 minutes at a time. Invacare recommends that each concentrator be on the power cord.
SECTION 1-GENERAL GUIDELINES Operating Information For optimum performance, Invacare recommends that Crush‐Proof oxygen tubing be left unattended when plugged in a typical medical installation. NEVER drop or insert any object or liquid into any opening. DO NOT connect the concentrator in parallel or series with this product and not exceed... Interference This equipment has been tested and found to provide a reasonable protection against electromagnetic interference in . Shorter periods of 30 minutes at a time. Invacare recommends that each concentrator be on the power cord.
Owners Manual
Page 10
... used only for connection and operating instructions. Perfecto2™Series 10 Part No 1143482 For more information about the HomeFill, contact your Invacare dealer. The outlet fitting does not affect concentrator performance. IOH200 Cabinet Filter *NOTE: This outlet fitting is to the HomeFill ownerʹs manual, part number 1100873, for filling oxygen...
... used only for connection and operating instructions. Perfecto2™Series 10 Part No 1143482 For more information about the HomeFill, contact your Invacare dealer. The outlet fitting does not affect concentrator performance. IOH200 Cabinet Filter *NOTE: This outlet fitting is to the HomeFill ownerʹs manual, part number 1100873, for filling oxygen...
Owners Manual
Page 11
.... Storage 1. Part No 1143482 11 Perfecto2™Series SECTION 3-HANDLING SECTION 3-HANDLING The concentrator should ALWAYS be reshipped by common carrier, additional cartons are available from Invacare. Unpacking NOTE: For this procedure, refer to be kept in a dry area. 2.... Check for storage until use of the concentrator is to FIGURE 3.1. 1. FIGURE 3.1 Unpacking NOTE: Unless the oxygen concentrator is evident, notify the carrier,...
.... Storage 1. Part No 1143482 11 Perfecto2™Series SECTION 3-HANDLING SECTION 3-HANDLING The concentrator should ALWAYS be reshipped by common carrier, additional cartons are available from Invacare. Unpacking NOTE: For this procedure, refer to be kept in a dry area. 2.... Check for storage until use of the concentrator is to FIGURE 3.1. 1. FIGURE 3.1 Unpacking NOTE: Unless the oxygen concentrator is evident, notify the carrier,...
Owners Manual
Page 12
...177; 0.5 psi (34.5 kPa ± 3.45 kPa) 0.5 to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory. 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected). From 6,000 ft (1828 meters) to 13,129 ft .... 87% to 95.6% for Perfecto2W Models All Perfecto2 Models: Up to 8,000 ft (2438 meters) above sea level without degredation of concentration levels. SECTION 4-TYPICAL PRODUCT PARAMETERS SECTION 4-TYPICAL PRODUCT PARAMETERS Electrical Requirements: Rated Current Input: Sound Level: Altitude: 120 VAC ±...
...177; 0.5 psi (34.5 kPa ± 3.45 kPa) 0.5 to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory. 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected). From 6,000 ft (1828 meters) to 13,129 ft .... 87% to 95.6% for Perfecto2W Models All Perfecto2 Models: Up to 8,000 ft (2438 meters) above sea level without degredation of concentration levels. SECTION 4-TYPICAL PRODUCT PARAMETERS SECTION 4-TYPICAL PRODUCT PARAMETERS Electrical Requirements: Rated Current Input: Sound Level: Altitude: 120 VAC ±...
Owners Manual
Page 14
Perfecto2™Series 14 Part No 1143482 SECTION 4-TYPICAL PRODUCT PARAMETERS Placement: Tubing: Relative Humidity: Time of Operation: Recommended Storage and Shipping Temperature: No closer than three inches from any wall, furniture, draperies, or similar surfaces. 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO NOT pinch). 20 to 60% Up to 24 hours per day. -20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity. *NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes).
Perfecto2™Series 14 Part No 1143482 SECTION 4-TYPICAL PRODUCT PARAMETERS Placement: Tubing: Relative Humidity: Time of Operation: Recommended Storage and Shipping Temperature: No closer than three inches from any wall, furniture, draperies, or similar surfaces. 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO NOT pinch). 20 to 60% Up to 24 hours per day. -20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity. *NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes).
Owners Manual
Page 15
...No 1143482 15 Perfecto2™Series It is intended for increased equipment maintenance. Keep unit at least 3 inches away from room to you about your concentrator and will serve as a reference as a bed or couch, where the air opening may result in environments other types of oxygen directly to ...room on page 12. You may select a room where using your oxygen concentrator would be located in a closet. The air intake of the product or place it on a soft surface, such as you how to avoid smoke...
...No 1143482 15 Perfecto2™Series It is intended for increased equipment maintenance. Keep unit at least 3 inches away from room to you about your concentrator and will serve as a reference as a bed or couch, where the air opening may result in environments other types of oxygen directly to ...room on page 12. You may select a room where using your oxygen concentrator would be located in a closet. The air intake of the product or place it on a soft surface, such as you how to avoid smoke...
Owners Manual
Page 17
... humidifier bottle counterclockwise until it is flowing through the cannula. Attach oxygen tubing from the humidifier bottle/adapter assembly to oxygen outlet connector on the concentrator. After assembly, ensure that oxygen is securely fastened. Place the humidifier bottle/adapter assembly in the humidifier compartment on the...
... humidifier bottle counterclockwise until it is flowing through the cannula. Attach oxygen tubing from the humidifier bottle/adapter assembly to oxygen outlet connector on the concentrator. After assembly, ensure that oxygen is securely fastened. Place the humidifier bottle/adapter assembly in the humidifier compartment on the...
Owners Manual
Page 20
SECTION 5-OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators - If purity falls below factory preset standards, indicator lights on the control panel will illuminate. Units with SensO2 The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. Units with SensO2 Perfecto2™Series 20 Part No 1143482 OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators -
SECTION 5-OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators - If purity falls below factory preset standards, indicator lights on the control panel will illuminate. Units with SensO2 The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. Units with SensO2 Perfecto2™Series 20 Part No 1143482 OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators -
Owners Manual
Page 21
... Failure Call a qualified technician. Units without SensO2 Part No 1143482 21 Perfecto2™Series Call a qualified technician. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up supply of oxygen ready. Units without SensO2 NOTE: For this procedure, refer to 85% (±...
... Failure Call a qualified technician. Units without SensO2 Part No 1143482 21 Perfecto2™Series Call a qualified technician. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up supply of oxygen ready. Units without SensO2 NOTE: For this procedure, refer to 85% (±...
Owners Manual
Page 22
Perfecto2™Series 22 Part No 1143482 O2 purity will build to a back-up time (approximately 30 min). Switch to a maximum during the initial start warm‐up supply of the Concentrator NOTE: Concentrator may be used during this period. LABEL SYMBOL I/O STATUS SYSTEM OKAY SYSTEM FAILURE INDICATOR LIGHTS (LED) GREEN Indicator Light RED Indicator light Continuous Audible Alarm Sieve-GARD™Compressor Shutdown Call a qualified technician. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of oxygen immediately.
Perfecto2™Series 22 Part No 1143482 O2 purity will build to a back-up time (approximately 30 min). Switch to a maximum during the initial start warm‐up supply of the Concentrator NOTE: Concentrator may be used during this period. LABEL SYMBOL I/O STATUS SYSTEM OKAY SYSTEM FAILURE INDICATOR LIGHTS (LED) GREEN Indicator Light RED Indicator light Continuous Audible Alarm Sieve-GARD™Compressor Shutdown Call a qualified technician. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of oxygen immediately.
Owners Manual
Page 23
...the maintenance record guidelines. NOTE: There is one cabinet filter located on environmental conditions. SECTION 6-MAINTENANCE SECTION 6-MAINTENANCE ƽ WARNING The Invacare concentrators are not limited to: high dust, air pollutants, etc. Only qualified personnel should be performed more frequently. NOTE: At a minimum... maintenance. Part No 1143482 23 Perfecto2™Series Power should perform preventive maintenance on the Invacare concentrator. Routine Maintenance NOTE: The following routine maintenance should be disconnected before beginning preventive maintenance on the...
...the maintenance record guidelines. NOTE: There is one cabinet filter located on environmental conditions. SECTION 6-MAINTENANCE SECTION 6-MAINTENANCE ƽ WARNING The Invacare concentrators are not limited to: high dust, air pollutants, etc. Only qualified personnel should be performed more frequently. NOTE: At a minimum... maintenance. Part No 1143482 23 Perfecto2™Series Power should perform preventive maintenance on the Invacare concentrator. Routine Maintenance NOTE: The following routine maintenance should be disconnected before beginning preventive maintenance on the...
Owners Manual
Page 25
... PREVENTIVE MAINTENANCE RECORD ON EACH INSPECTION Record Date of Service Manual. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter...
... PREVENTIVE MAINTENANCE RECORD ON EACH INSPECTION Record Date of Service Manual. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter...
Owners Manual
Page 26
... Perfecto2™Series 26 Part No 1143482 SECTION 7-TROUBLESHOOTING GUIDE SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: Alarm: Short beeps, long pause Concentrator not operating, power switch On. Tripped circuit breaker. Call Service Provider. Move to another electrical outlet or circuit. 3.... cabinet filters. 1b. Power cord not plugged in. 2. Once corrected, turn power Off for kinks or blockages. Move oxygen concentrator at outlet. 3. Kinked or blocked tubing, cannula or humidifier. 2. Inspect house circuit breakers or fuses. PROBABLE CAUSE: Main Power...
... Perfecto2™Series 26 Part No 1143482 SECTION 7-TROUBLESHOOTING GUIDE SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: Alarm: Short beeps, long pause Concentrator not operating, power switch On. Tripped circuit breaker. Call Service Provider. Move to another electrical outlet or circuit. 3.... cabinet filters. 1b. Power cord not plugged in. 2. Once corrected, turn power Off for kinks or blockages. Move oxygen concentrator at outlet. 3. Kinked or blocked tubing, cannula or humidifier. 2. Inspect house circuit breakers or fuses. PROBABLE CAUSE: Main Power...
Owners Manual
Page 27
....* 2. Internal repairs YELLOW Light required. DO NOT use extension cords. SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: YELLOW or RED light Illuminated. Unit overheating due to IRC5PO2. Move concentrator at least three inches from walls, draperies, and furniture. 5.
....* 2. Internal repairs YELLOW Light required. DO NOT use extension cords. SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: YELLOW or RED light Illuminated. Unit overheating due to IRC5PO2. Move concentrator at least three inches from walls, draperies, and furniture. 5.
Owners Manual
Page 31
...not extended to any other person. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to be free from defects in materials and workmanship for a period of three years from date of purchase from Invacare Corporation or a dealer. This warranty..., are excluded from state to be defective, such product shall be refused; For warranty service, please contact Invacare Corporation's Service Department at Invacare Corporation's option. Coverage under this warranty. LIMITED WARRANTY LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED ...
...not extended to any other person. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to be free from defects in materials and workmanship for a period of three years from date of purchase from Invacare Corporation or a dealer. This warranty..., are excluded from state to be defective, such product shall be refused; For warranty service, please contact Invacare Corporation's Service Department at Invacare Corporation's option. Coverage under this warranty. LIMITED WARRANTY LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED ...