Owners Manual
Page 1
For more information regarding Invacare products, parts, and services, please visit www.invacare.com User: BEFORE using this product, read this product. User Manual Perfecto2™Series Oxygen Concentrators without SensO2®, Oxygen Concentrators with SensO2 HomeFill®Compatible Perfecto2 with SensO2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this manual and save for future reference.
For more information regarding Invacare products, parts, and services, please visit www.invacare.com User: BEFORE using this product, read this product. User Manual Perfecto2™Series Oxygen Concentrators without SensO2®, Oxygen Concentrators with SensO2 HomeFill®Compatible Perfecto2 with SensO2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this manual and save for future reference.
Owners Manual
Page 7
... operation of the Perfecto2 concentrator these instructions MUST be followed. ƽ WARNING This section contains important information for the safe operation and use any spark near medical oxygen equipment. DO NOT come in contact with the concentrator while wet. If the concentrator has a damaged cord or plug, if it is required by Invacare. Refer servicing to...
... operation of the Perfecto2 concentrator these instructions MUST be followed. ƽ WARNING This section contains important information for the safe operation and use any spark near medical oxygen equipment. DO NOT come in contact with the concentrator while wet. If the concentrator has a damaged cord or plug, if it is required by Invacare. Refer servicing to...
Owners Manual
Page 8
... left unattended when plugged in use. Perfecto2™Series 8 Part No 1143482 SECTION 1-GENERAL GUIDELINES Operating Information For optimum performance, Invacare recommends that Crush‐Proof oxygen tubing be required for patients using this product and not exceed 50 ft in a typical medical installation. Invacare recommends that each concentrator be on the power cord. These...
... left unattended when plugged in use. Perfecto2™Series 8 Part No 1143482 SECTION 1-GENERAL GUIDELINES Operating Information For optimum performance, Invacare recommends that Crush‐Proof oxygen tubing be required for patients using this product and not exceed 50 ft in a typical medical installation. Invacare recommends that each concentrator be on the power cord. These...
Owners Manual
Page 10
... not in use, the plug provided with the HomeFill home oxygen compressor. Perfecto2™Series 10 Part No 1143482 IOH200 Cabinet Filter *NOTE: This outlet fitting is to the HomeFill ownerʹs manual, part number 1100873, for filling oxygen cylinders with the concentrator should be inserted into the outlet Power Cord *Outlet Fitting fitting...
... not in use, the plug provided with the HomeFill home oxygen compressor. Perfecto2™Series 10 Part No 1143482 IOH200 Cabinet Filter *NOTE: This outlet fitting is to the HomeFill ownerʹs manual, part number 1100873, for filling oxygen cylinders with the concentrator should be inserted into the outlet Power Cord *Outlet Fitting fitting...
Owners Manual
Page 11
... 11 Perfecto2™Series If damage is to be reshipped by common carrier, additional cartons are available from Invacare. Inspect all the components from the carton. 3. Check for any obvious damage to be used immediately, retain containers and packing materials for nicks, dents, scratches or other damages. FIGURE 3.1 Unpacking NOTE: Unless the oxygen concentrator...
... 11 Perfecto2™Series If damage is to be reshipped by common carrier, additional cartons are available from Invacare. Inspect all the components from the carton. 3. Check for any obvious damage to be used immediately, retain containers and packing materials for nicks, dents, scratches or other damages. FIGURE 3.1 Unpacking NOTE: Unless the oxygen concentrator...
Owners Manual
Page 12
...) below 90% efficiency. *Oxygen Output Concentration Levels: Maximum Outlet Pressure: Flow Range: Low Flow Alarm: Average Power Consumption: All Perfecto2 Models while filling a HomeFill System: Up to 6,000 ft (1828 meters) above sea level without degradation of concentration levels. Perfecto2 and Perfecto2W Models: 280 W... the use of concentration levels. From 8,000 ft to 13,129 (4000 meters) below 90% efficiency. 87% to 95.6% for Perfecto2W Models All Perfecto2 Models: Up to 8,000 ft (2438 meters) above sea level without degredation of the Invacare Pediatric Flowmeter IRCPF16 ...
...) below 90% efficiency. *Oxygen Output Concentration Levels: Maximum Outlet Pressure: Flow Range: Low Flow Alarm: Average Power Consumption: All Perfecto2 Models while filling a HomeFill System: Up to 6,000 ft (1828 meters) above sea level without degradation of concentration levels. Perfecto2 and Perfecto2W Models: 280 W... the use of concentration levels. From 8,000 ft to 13,129 (4000 meters) below 90% efficiency. 87% to 95.6% for Perfecto2W Models All Perfecto2 Models: Up to 8,000 ft (2438 meters) above sea level without degredation of the Invacare Pediatric Flowmeter IRCPF16 ...
Owners Manual
Page 15
.../or fumes. Part No 1143482 15 Perfecto2™Series Your concentrator can be easily rolled from room to you about your concentrator and will perform best when operated under the following conditions. SECTION 5-OPERATING INSTRUCTIONS SECTION 5-OPERATING INSTRUCTIONS Introduction Your oxygen concentrator is an electronically operated device that oxygen concentrators are therapeutically equivalent to other than those...
.../or fumes. Part No 1143482 15 Perfecto2™Series Your concentrator can be easily rolled from room to you about your concentrator and will perform best when operated under the following conditions. SECTION 5-OPERATING INSTRUCTIONS SECTION 5-OPERATING INSTRUCTIONS Introduction Your oxygen concentrator is an electronically operated device that oxygen concentrators are therapeutically equivalent to other than those...
Owners Manual
Page 17
... FIGURE 5.3. 6. See Detail "A" in FIGURE 5.3. 9. Place the humidifier bottle/adapter assembly in the humidifier compartment on the concentrator. Attach the cannula/patient supply tubing to the humidifier bottle by turning the wing nut on the humidifier bottle counterclockwise until it... oxygen tubing from the humidifier bottle/adapter assembly to oxygen outlet connector on the concentrator. Attach the humidifier bottle adapter to the humidifier bottle outlet. See Detail "B" in FIGURE 5.3. 7. After assembly, ensure that oxygen is securely fastened. Part No 1143482 17 Perfecto2™...
... FIGURE 5.3. 6. See Detail "A" in FIGURE 5.3. 9. Place the humidifier bottle/adapter assembly in the humidifier compartment on the concentrator. Attach the cannula/patient supply tubing to the humidifier bottle by turning the wing nut on the humidifier bottle counterclockwise until it... oxygen tubing from the humidifier bottle/adapter assembly to oxygen outlet connector on the concentrator. Attach the humidifier bottle adapter to the humidifier bottle outlet. See Detail "B" in FIGURE 5.3. 7. After assembly, ensure that oxygen is securely fastened. Part No 1143482 17 Perfecto2™...
Owners Manual
Page 20
Units with SensO2 Perfecto2™Series 20 Part No 1143482 SECTION 5-OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators - If purity falls below factory preset standards, indicator lights on the control panel will illuminate. Units with SensO2 The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators -
Units with SensO2 Perfecto2™Series 20 Part No 1143482 SECTION 5-OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators - If purity falls below factory preset standards, indicator lights on the control panel will illuminate. Units with SensO2 The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators -
Owners Manual
Page 21
... Failure Call a qualified technician. Switch to FIGURE 5.7. Units without SensO2 NOTE: For this procedure, refer to a back-up supply of oxygen ready. Front Panel Indicators - Call a qualified technician. YELLOW Solid B. Have a back-up time (approximately 30 min.) while waiting for...73% (±3%) GREEN Indicator Light YELLOW Indicator light A. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up supply of oxygen immediately. Units without SensO2 Part No 1143482 21 Perfecto2™Series
... Failure Call a qualified technician. Switch to FIGURE 5.7. Units without SensO2 NOTE: For this procedure, refer to a back-up supply of oxygen ready. Front Panel Indicators - Call a qualified technician. YELLOW Solid B. Have a back-up time (approximately 30 min.) while waiting for...73% (±3%) GREEN Indicator Light YELLOW Indicator light A. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up supply of oxygen immediately. Units without SensO2 Part No 1143482 21 Perfecto2™Series
Owners Manual
Page 22
O2 purity will build to a back-up time (approximately 30 min). Perfecto2™Series 22 Part No 1143482 Switch to a maximum during the initial start warm‐up supply of the Concentrator NOTE: Concentrator may be used during this period. LABEL SYMBOL I/O STATUS SYSTEM OKAY SYSTEM FAILURE INDICATOR LIGHTS (LED) GREEN Indicator Light RED Indicator light Continuous Audible Alarm Sieve-GARD™Compressor Shutdown Call a qualified technician. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of oxygen immediately.
O2 purity will build to a back-up time (approximately 30 min). Perfecto2™Series 22 Part No 1143482 Switch to a maximum during the initial start warm‐up supply of the Concentrator NOTE: Concentrator may be used during this period. LABEL SYMBOL I/O STATUS SYSTEM OKAY SYSTEM FAILURE INDICATOR LIGHTS (LED) GREEN Indicator Light RED Indicator light Continuous Audible Alarm Sieve-GARD™Compressor Shutdown Call a qualified technician. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of oxygen immediately.
Owners Manual
Page 25
... Record Part No 1143482 25 SECTION 6-MAINTENANCE SERIAL NO. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check...Clean Cabinet Filter(s) Check Prescribed L/min. PREVENTIVE MAINTENANCE RECORD ON EACH INSPECTION Record Date of Service Manual. Perfecto2™Series NOTE: 4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years. MODEL...
... Record Part No 1143482 25 SECTION 6-MAINTENANCE SERIAL NO. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check...Clean Cabinet Filter(s) Check Prescribed L/min. PREVENTIVE MAINTENANCE RECORD ON EACH INSPECTION Record Date of Service Manual. Perfecto2™Series NOTE: 4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years. MODEL...
Owners Manual
Page 26
PROBABLE CAUSE: Main Power Loss: 1. Power cord not plugged in. 2. Alarm: Continuous Concentrator not operating, Power Switch On. Unit overheating due to prescribed flowrate. Move oxygen concentrator at outlet. 3. Alarm: Rapid Beep....Beep... Inspect for 60 seconds and then turn power...System Failure: 1. Insufficient power at 0.5 L/min or less. 1. If prescribed flowrate is below 0.5 L/min, use a different outlet. 3. Perfecto2™Series 26 Part No 1143482 SOLUTION: 1. Kinked or blocked tubing, cannula or humidifier. 2. Insert plug into outlet. 2. Flowmeter set at...
PROBABLE CAUSE: Main Power Loss: 1. Power cord not plugged in. 2. Alarm: Continuous Concentrator not operating, Power Switch On. Unit overheating due to prescribed flowrate. Move oxygen concentrator at outlet. 3. Alarm: Rapid Beep....Beep... Inspect for 60 seconds and then turn power...System Failure: 1. Insufficient power at 0.5 L/min or less. 1. If prescribed flowrate is below 0.5 L/min, use a different outlet. 3. Perfecto2™Series 26 Part No 1143482 SOLUTION: 1. Kinked or blocked tubing, cannula or humidifier. 2. Insert plug into outlet. 2. Flowmeter set at...
Owners Manual
Page 27
...required. Correct, clean or replace item. Move to IRC5PO2. Part No 1143482 27 Perfecto2™Series SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: YELLOW or RED light Illuminated. Flowmeter set to blocked air intake. 5. SOLUTION: 1. Move concentrator at 0.5 L/min or less.* 4. GREEN light with 1. Call Service Provider..... * Only applies to another electrical outlet or circuit. 6. Insufficient power at outlet. 6. Flashing. 1. Clean or Replace filters. 2. Low oxygen purity.* 2. Unit overheating due to 1.0 L/min or more. 4a. DO NOT use extension cords. Call Service Provider.
...required. Correct, clean or replace item. Move to IRC5PO2. Part No 1143482 27 Perfecto2™Series SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: YELLOW or RED light Illuminated. Flowmeter set to blocked air intake. 5. SOLUTION: 1. Move concentrator at 0.5 L/min or less.* 4. GREEN light with 1. Call Service Provider..... * Only applies to another electrical outlet or circuit. 6. Insufficient power at outlet. 6. Flashing. 1. Clean or Replace filters. 2. Low oxygen purity.* 2. Unit overheating due to 1.0 L/min or more. 4a. DO NOT use extension cords. Call Service Provider.
Owners Manual
Page 31
... WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to the labor performed by the Invacare Service Department or Invacare Authorized Service Centers. shipments will issue a serialized return authorization. Part No 1143482 31...
... WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to the labor performed by the Invacare Service Department or Invacare Authorized Service Centers. shipments will issue a serialized return authorization. Part No 1143482 31...