Owners Manual
Page 1
For more information regarding Invacare products, parts, and services, please visit www.invacare.com User: Before using this product, read this manual and save for future reference. Operator's Manual Platinum™ Series XL, 5, 10 Oxygen Concentrators Standard, with SensO2®and HomeFill® Invacare 5 with SensO2 Model IRC5LXO2 Invacare 10 with SensO2 Model IRC10LXO2 Invacare 5 Model IRC5LX Invacare 10 Model IRC10LX Dealer: This manual MUST be given to the end user.
For more information regarding Invacare products, parts, and services, please visit www.invacare.com User: Before using this product, read this manual and save for future reference. Operator's Manual Platinum™ Series XL, 5, 10 Oxygen Concentrators Standard, with SensO2®and HomeFill® Invacare 5 with SensO2 Model IRC5LXO2 Invacare 10 with SensO2 Model IRC10LXO2 Invacare 5 Model IRC5LX Invacare 10 Model IRC10LX Dealer: This manual MUST be given to the end user.
Owners Manual
Page 11
... be inserted into the outlet fitting. Part No 1118353 11 Platinum™ Series For more information about the HomeFill, contact your Invacare dealer. SECTION 2-FEATURES SECTION 2-FEATURES Oxygen Outlet Flowmeter Oxygen ...Purity Indicator Lights (SensO2 model only)/ Fault and Power Indicator Lights (LX model only) Power Switch Circuit Breaker Elapsed Time Meter REAR VIEW Cabinet Filter (one of two shown) ACCESSORIES (NOT SHOWN): Remote Humidifier/Flowmeter Accessory - IRCPF16 (Platinum 5 and XL...
... be inserted into the outlet fitting. Part No 1118353 11 Platinum™ Series For more information about the HomeFill, contact your Invacare dealer. SECTION 2-FEATURES SECTION 2-FEATURES Oxygen Outlet Flowmeter Oxygen ...Purity Indicator Lights (SensO2 model only)/ Fault and Power Indicator Lights (LX model only) Power Switch Circuit Breaker Elapsed Time Meter REAR VIEW Cabinet Filter (one of two shown) ACCESSORIES (NOT SHOWN): Remote Humidifier/Flowmeter Accessory - IRCPF16 (Platinum 5 and XL...
Owners Manual
Page 13
... 4-TYPICAL PRODUCT PARAMETERS SECTION 4-TYPICAL PRODUCT PARAMETERS Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz Rated Current Input: Sound Level: Platinum 5 and XL: 4.3 A Platinum 10: 4.7 A Platinum 5 and XL: 50 dB Average Platinum 10: 58 dB Average Altitude: Platinum 5 and XL - Up to 4,000 ft (1230 meters) above sea level without degradation of concentration levels.
... 4-TYPICAL PRODUCT PARAMETERS SECTION 4-TYPICAL PRODUCT PARAMETERS Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz Rated Current Input: Sound Level: Platinum 5 and XL: 4.3 A Platinum 10: 4.7 A Platinum 5 and XL: 50 dB Average Platinum 10: 58 dB Average Altitude: Platinum 5 and XL - Up to 4,000 ft (1230 meters) above sea level without degradation of concentration levels.
Owners Manual
Page 14
... (No Accessories Connected) Platinum 5 and XL: 400W Platinum 10: 585W 35 psi ± 3.5 psi (241 kPa ± 24.1 kPa) 0.7 L/min Cabinet (2), Outlet HEPA and Compressor Inlet. Platinum 10 - 2 L/min to 10 L/min Platinum 5 - 0 L/min to 0.5 L/min, Platinum 10 - 0 L/min to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory...
... (No Accessories Connected) Platinum 5 and XL: 400W Platinum 10: 585W 35 psi ± 3.5 psi (241 kPa ± 24.1 kPa) 0.7 L/min Cabinet (2), Outlet HEPA and Compressor Inlet. Platinum 10 - 2 L/min to 10 L/min Platinum 5 - 0 L/min to 0.5 L/min, Platinum 10 - 0 L/min to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory...
Owners Manual
Page 15
...176;C - 35°C) at 15-95% relative humidity Part No 1118353 15 Platinum™ Series Platinum 10: Less than Ambient + 45°F (+25°C) Platinum 5 and XL: Less than Ambient + 4°F (+2°C) Platinum 10: Less than three inches from any wall, furniture, draperies, or similar ...surfaces. 7 ft cannula with UL 1097 ETL certified to 65°C), at 20-60% relative humidity Platinum 5 and XL: Less than Ambient + 35°F (+19°C). SECTION 4-TYPICAL PRODUCT PARAMETERS Operating Ambient Temperature: Exhaust: Oxygen Output Temperature:...
...176;C - 35°C) at 15-95% relative humidity Part No 1118353 15 Platinum™ Series Platinum 10: Less than Ambient + 45°F (+25°C) Platinum 5 and XL: Less than Ambient + 4°F (+2°C) Platinum 10: Less than three inches from any wall, furniture, draperies, or similar ...surfaces. 7 ft cannula with UL 1097 ETL certified to 65°C), at 20-60% relative humidity Platinum 5 and XL: Less than Ambient + 35°F (+19°C). SECTION 4-TYPICAL PRODUCT PARAMETERS Operating Ambient Temperature: Exhaust: Oxygen Output Temperature:...
Owners Manual
Page 17
...cannula back to the level indicated by the manufacturer. 2. Connect Humidifier (If So Prescribed) ƽ WARNING DO NOT overfill humidifier. NOTE: For Platinum 5, XL ‐ recommended humidifiers are high flow: 7900‐10, 7900‐25. 1. DO NOT reverse the oxygen input and output connections. Remove ...feature. If the plug does not fit fully in a polarized outlet only one blade is wider than the other). For Platinum 10, recommended humidifiers are : Invacare 3260, 003‐40, or 006‐40. Humidifier Bottle FIGURE 5.1 Filling the Humidifier Part No 1118353 17...
...cannula back to the level indicated by the manufacturer. 2. Connect Humidifier (If So Prescribed) ƽ WARNING DO NOT overfill humidifier. NOTE: For Platinum 5, XL ‐ recommended humidifiers are high flow: 7900‐10, 7900‐25. 1. DO NOT reverse the oxygen input and output connections. Remove ...feature. If the plug does not fit fully in a polarized outlet only one blade is wider than the other). For Platinum 10, recommended humidifiers are : Invacare 3260, 003‐40, or 006‐40. Humidifier Bottle FIGURE 5.1 Filling the Humidifier Part No 1118353 17...
Owners Manual
Page 26
.... Correct, clean or ing, cannula or humidifier. Reset flowmeter to pre- 0.5 L/min (or less) for 60 seconds and then turn power OFF for scribed flowrate. Platinum™ Series 26 Part No 1118353 Concentrator ers or fuses. If problem recurs, NOT operating, use extension Beeeeeeeep.... Insert plug into outlet. 2. Tripped circuit breaker... CAUSE: SOLUTION: Alarm: SHORT BEEPS, LONG PAUSE Main Power Loss: 1. Insufficient power at 1b. Kinked or blocked tub- use the Pediatric Flowmeter IRCPF16 (Only with Platinum 5 and XL). Alarm: System Failure: 1a.
.... Correct, clean or ing, cannula or humidifier. Reset flowmeter to pre- 0.5 L/min (or less) for 60 seconds and then turn power OFF for scribed flowrate. Platinum™ Series 26 Part No 1118353 Concentrator ers or fuses. If problem recurs, NOT operating, use extension Beeeeeeeep.... Insert plug into outlet. 2. Tripped circuit breaker... CAUSE: SOLUTION: Alarm: SHORT BEEPS, LONG PAUSE Main Power Loss: 1. Insufficient power at 1b. Kinked or blocked tub- use the Pediatric Flowmeter IRCPF16 (Only with Platinum 5 and XL). Alarm: System Failure: 1a.
Owners Manual
Page 30
Invacare Corporation warrants its Platinum XL Oxygen Concentrators when purchased new and unused to the original purchaser who purchases this warranty. Routine maintenance items, such as filters, are excluded from this product when new and unused from Invacare or a dealer, with a copy of the seller's ... THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. Invacare Corporation warrants its Platinum 10 Oxygen Concentrators when purchased new and unused to be free from defects in materials and workmanship for a period ...
Invacare Corporation warrants its Platinum XL Oxygen Concentrators when purchased new and unused to the original purchaser who purchases this warranty. Routine maintenance items, such as filters, are excluded from this product when new and unused from Invacare or a dealer, with a copy of the seller's ... THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. Invacare Corporation warrants its Platinum 10 Oxygen Concentrators when purchased new and unused to be free from defects in materials and workmanship for a period ...