Owners Manual
Page 11
... number 1100873, for filling oxygen cylinders with the concentrator should be inserted into the outlet fitting. The outlet fitting does not affect concentrator performance. When not in use, the plug provided with the HomeFill home oxygen compressor. Refer to be ...Oxygen Purity Indicator Lights (SensO2 model only)/ Fault and Power Indicator Lights (LX model only) Power Switch Circuit Breaker Elapsed Time Meter REAR VIEW Cabinet Filter (one of two shown) ACCESSORIES (NOT SHOWN): Remote Humidifier/Flowmeter Accessory - For more information about the HomeFill, contact your Invacare...
... number 1100873, for filling oxygen cylinders with the concentrator should be inserted into the outlet fitting. The outlet fitting does not affect concentrator performance. When not in use, the plug provided with the HomeFill home oxygen compressor. Refer to be ...Oxygen Purity Indicator Lights (SensO2 model only)/ Fault and Power Indicator Lights (LX model only) Power Switch Circuit Breaker Elapsed Time Meter REAR VIEW Cabinet Filter (one of two shown) ACCESSORIES (NOT SHOWN): Remote Humidifier/Flowmeter Accessory - For more information about the HomeFill, contact your Invacare...
Owners Manual
Page 24
... part vinegar. 3. Wash it with a solution of concentrator) FIGURE 6.1 Cleaning the Cabinet Filter Cleaning the Cabinet ƽ WARNING UNPLUG the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. Clean the cabinet filters with a vacuum cleaner or wash in soapy water and...and refill with a mild household cleaner and non‐abrasive cloth or sponge. SECTION 6-MAINTENANCE 2. Dry the filters thoroughly before reinstallation. Cleaning the Humidifier NOTE: To clean the oxygen humidifier, follow these STEPS: 1. Platinum™ Series 24 Part No 1118353
... part vinegar. 3. Wash it with a solution of concentrator) FIGURE 6.1 Cleaning the Cabinet Filter Cleaning the Cabinet ƽ WARNING UNPLUG the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. Clean the cabinet filters with a vacuum cleaner or wash in soapy water and...and refill with a mild household cleaner and non‐abrasive cloth or sponge. SECTION 6-MAINTENANCE 2. Dry the filters thoroughly before reinstallation. Cleaning the Humidifier NOTE: To clean the oxygen humidifier, follow these STEPS: 1. Platinum™ Series 24 Part No 1118353
Owners Manual
Page 25
...IRC5LX (EVERY 2,160 HOURS FOR IRC10LX) Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter* Check Compressor Inlet Filter... Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. I RC Part No 1118353 FIGURE 6.2 Preventive Maintenance Record 25 PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR ON EACH INSPECTION Record Date of Service Manual...
...IRC5LX (EVERY 2,160 HOURS FOR IRC10LX) Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter* Check Compressor Inlet Filter... Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. I RC Part No 1118353 FIGURE 6.2 Preventive Maintenance Record 25 PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR ON EACH INSPECTION Record Date of Service Manual...
Owners Manual
Page 26
.... cords. required Alarm: RAPID Beep....Beep... Kinked or blocked tub- Once corrected, turn power OFF for kinks or 1. Reset flowmeter to filters. No power at 1b. Concentrator ers or fuses. Move oxygen concentrator at least three inches away from walls, draperies or furniture. 2. Inspect for 60 seconds and then turn power back ON. 2. replace...
.... cords. required Alarm: RAPID Beep....Beep... Kinked or blocked tub- Once corrected, turn power OFF for kinks or 1. Reset flowmeter to filters. No power at 1b. Concentrator ers or fuses. Move oxygen concentrator at least three inches away from walls, draperies or furniture. 2. Inspect for 60 seconds and then turn power back ON. 2. replace...
Owners Manual
Page 27
...* Only applies to 0.5 L/min (or less) for 60 seconds and On red light only. LIGHT Illuminated. 2. Remove and clean cabinet blocked air intake. filters. 4b. Call Service Provider. Low oxygen purity.* 1. Flowmeter set to 5LXO2 and 10LXO2 SensO2 models. 3. Correct, clean or replace Alarm: item. Check flowmeter is set too low: 3. DO... walls, draperies, and furniture. 5. then turn CONTINUOUS power OFF for 1.0 L/min. SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: YELLOW or RED 1. Clean or Replace filters. Move concentrator at outlet. 5.
...* Only applies to 0.5 L/min (or less) for 60 seconds and On red light only. LIGHT Illuminated. 2. Remove and clean cabinet blocked air intake. filters. 4b. Call Service Provider. Low oxygen purity.* 1. Flowmeter set to 5LXO2 and 10LXO2 SensO2 models. 3. Correct, clean or replace Alarm: item. Check flowmeter is set too low: 3. DO... walls, draperies, and furniture. 5. then turn CONTINUOUS power OFF for 1.0 L/min. SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: YELLOW or RED 1. Clean or Replace filters. Move concentrator at outlet. 5.
Owners Manual
Page 30
... its Platinum XL Oxygen Concentrators when purchased new and unused to any subsequent purchaser or owner. Routine maintenance items, such as filters, are excluded from this product when new and unused from Invacare or a dealer, with a copy of the seller's invoice required for coverage under this...such subsequent sale or other person. It does not apply to the labor for coverage under this warranty. Invacare Corporation warrants its Platinum 10 Oxygen Concentrators when purchased new and unused to the original purchaser who purchases this warranty. This warranty does not include ...
... its Platinum XL Oxygen Concentrators when purchased new and unused to any subsequent purchaser or owner. Routine maintenance items, such as filters, are excluded from this product when new and unused from Invacare or a dealer, with a copy of the seller's invoice required for coverage under this...such subsequent sale or other person. It does not apply to the labor for coverage under this warranty. Invacare Corporation warrants its Platinum 10 Oxygen Concentrators when purchased new and unused to the original purchaser who purchases this warranty. This warranty does not include ...