Owners Manual
Page 1
Invacare® Homecare Beds 5411IVC, 5410IVC, 5311IVC, 5310IVC, 5307IVC, 5411LowIVC, 5410LowIVC en User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
Invacare® Homecare Beds 5411IVC, 5410IVC, 5311IVC, 5310IVC, 5307IVC, 5411LowIVC, 5410LowIVC en User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
Owners Manual
Page 3
... Full Electric Bed 33 5.3 Operating the Semi-Electric Bed 34 5.3.1 Using the Emergency Crank 35 5.4 Operating the Manual Bed 36 5.5 Optional Bed Rails 36 6 Maintenance 38 6.1 Service Life 38 6.2 Wear and Tear Items 38 6.3 Cleaning the Bed 39 6.4 Maintenance and Cleaning Checklist 41 7 Troubleshooting 43 7.1 Troubleshooting Mechanical 43 7.2 Troubleshooting Electrical 44 8 Warranty 46 8.1 Limited Warranty 46
... Full Electric Bed 33 5.3 Operating the Semi-Electric Bed 34 5.3.1 Using the Emergency Crank 35 5.4 Operating the Manual Bed 36 5.5 Optional Bed Rails 36 6 Maintenance 38 6.1 Service Life 38 6.2 Wear and Tear Items 38 6.3 Cleaning the Bed 39 6.4 Maintenance and Cleaning Checklist 41 7 Troubleshooting 43 7.1 Troubleshooting Mechanical 43 7.2 Troubleshooting Electrical 44 8 Warranty 46 8.1 Limited Warranty 46
Owners Manual
Page 4
... for efficient, trouble-free use. 1.2 Intended Use The bed is intended to be used as a mattress support system. The bed allows for articulation of the signal words. CAUTION! - It is intended to be used in home care and long term care facilities. 4 1114836-H-05 Invacare® Homecare Beds 1 General 1.1 Symbols Signal words are used in this manual and apply...
... for efficient, trouble-free use. 1.2 Intended Use The bed is intended to be used as a mattress support system. The bed allows for articulation of the signal words. CAUTION! - It is intended to be used in home care and long term care facilities. 4 1114836-H-05 Invacare® Homecare Beds 1 General 1.1 Symbols Signal words are used in this manual and apply...
Owners Manual
Page 5
... on the gearbox (not shown). Dry all components. 4. Make sure all electrical parts (motors, junction box and pendant) are not damaged and that all connection ports are plugged. 3. CAUTION CLEANING INSTRUCTIONS See Owner's Manual 1. Unplug product before storing. A = Pendant cord Overview 2.2 Technical Specifications Full Electric 5410IVC, 5411IVC SHIPPING WEIGHT WEIGHT CAPACITY OCCUPANT SAFE WORKING LOAD...
... on the gearbox (not shown). Dry all components. 4. Make sure all electrical parts (motors, junction box and pendant) are not damaged and that all connection ports are plugged. 3. CAUTION CLEANING INSTRUCTIONS See Owner's Manual 1. Unplug product before storing. A = Pendant cord Overview 2.2 Technical Specifications Full Electric 5410IVC, 5411IVC SHIPPING WEIGHT WEIGHT CAPACITY OCCUPANT SAFE WORKING LOAD...
Owners Manual
Page 7
... install the accessories without reading and understanding all of Injury or Damage To avoid injury or damage from misuse: - Risk of the instructions in the instruction sheet that accompanies them. Contact Dealer or Invacare for current list DESCRIPTION MODEL NUMBER WEIGHT MATTRESS Solace Series, Softform Series 23-31 lbs (10.4-14.1 kg) varies depending on model...
... install the accessories without reading and understanding all of Injury or Damage To avoid injury or damage from misuse: - Risk of the instructions in the instruction sheet that accompanies them. Contact Dealer or Invacare for current list DESCRIPTION MODEL NUMBER WEIGHT MATTRESS Solace Series, Softform Series 23-31 lbs (10.4-14.1 kg) varies depending on model...
Owners Manual
Page 8
..., Injury, Or Damage Improper use of the product with damaged parts could lead to the product malfunctioning, causing injury to the user and/or caregiver. - Continued use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as user manuals, service manuals or instruction sheets supplied with Invacare bed products. 8 1114836...
..., Injury, Or Damage Improper use of the product with damaged parts could lead to the product malfunctioning, causing injury to the user and/or caregiver. - Continued use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as user manuals, service manuals or instruction sheets supplied with Invacare bed products. 8 1114836...
Owners Manual
Page 9
...Replacement parts MUST match original Invacare parts. Risk of Injury or Death To reduce the risk of shock: - If liquid is tightened securely. - Clean up the spill and allow the product or the area around the product, unplug before using the controls again. 1114836-H-05 Safety DANGER! Unplug from the electrical outlet before servicing. - DO NOT use... DO NOT use , make sure all attaching hardware is spilled in these instructions. Risk of the product may cause injury or damage. - After any adjustments, repair or service and before use attachments not ...
...Replacement parts MUST match original Invacare parts. Risk of Injury or Death To reduce the risk of shock: - If liquid is tightened securely. - Clean up the spill and allow the product or the area around the product, unplug before using the controls again. 1114836-H-05 Safety DANGER! Unplug from the electrical outlet before servicing. - DO NOT use... DO NOT use , make sure all attaching hardware is spilled in these instructions. Risk of the product may cause injury or damage. - After any adjustments, repair or service and before use attachments not ...
Owners Manual
Page 10
...(such as tile or wood,) will allow the bed to the Bed Rail Entrapment Risk Notification Guide at any extremities protrude over the side or between the raised bed frame components at www.invacare.com for additional safety information. 10 1114836-H-05 DO NOT move under the bed: - DO..., test the wheel locks to the bed ends before use. - DO NOT transfer occupant if bed is equipped with casters properly locked, some conditions. ALWAYS ensure that the individual utilizing the bed is positioned properly within the confines of small parts or materials: - Risk Of Death, Injury, Or...
...(such as tile or wood,) will allow the bed to the Bed Rail Entrapment Risk Notification Guide at any extremities protrude over the side or between the raised bed frame components at www.invacare.com for additional safety information. 10 1114836-H-05 DO NOT move under the bed: - DO..., test the wheel locks to the bed ends before use. - DO NOT transfer occupant if bed is equipped with casters properly locked, some conditions. ALWAYS ensure that the individual utilizing the bed is positioned properly within the confines of small parts or materials: - Risk Of Death, Injury, Or...
Owners Manual
Page 11
...." Replacement Parts/Accessories DANGER! Do not use this bed in the flat position when left unattended. Risk Of Death, Injury, Or Damage To avoid fire hazards, injury or damage: - Risk Of Death, Injury, Or Damage Patient entrapment from the use ,...User manuals are available at www.invacare.com or your dealer. - Use masked or nasal type oxygen administering equipment only in conjunction with flammable gases. Monitor patient frequently. DO NOT use of any adjustments, repair or service and before use of manual/electric bed. - DANGER! Follow the manufacturer's instructions...
...." Replacement Parts/Accessories DANGER! Do not use this bed in the flat position when left unattended. Risk Of Death, Injury, Or Damage To avoid fire hazards, injury or damage: - Risk Of Death, Injury, Or Damage Patient entrapment from the use ,...User manuals are available at www.invacare.com or your dealer. - Use masked or nasal type oxygen administering equipment only in conjunction with flammable gases. Monitor patient frequently. DO NOT use of any adjustments, repair or service and before use of manual/electric bed. - DANGER! Follow the manufacturer's instructions...
Owners Manual
Page 13
...are not interchangeable. Invacare manuals are not intended nor may cause injury or death. Where gaps occur, patient entrapment is possible and the mattress should be replaced. 13 DANGER!... used for alternative means of safe restraint. - Variations in bed side rail design, width and thickness or firmness of entrapment. Review "A Guide to mattress compression. Follow the manufacturer's instructions...pressure is required to using this bed safety guide. - Monitor patient frequently. Read and understand the User Manual prior to reduce the risk of bed. Use the link located ...
...are not interchangeable. Invacare manuals are not intended nor may cause injury or death. Where gaps occur, patient entrapment is possible and the mattress should be replaced. 13 DANGER!... used for alternative means of safe restraint. - Variations in bed side rail design, width and thickness or firmness of entrapment. Review "A Guide to mattress compression. Follow the manufacturer's instructions...pressure is required to using this bed safety guide. - Monitor patient frequently. Read and understand the User Manual prior to reduce the risk of bed. Use the link located ...
Owners Manual
Page 17
... an environmentally aware manufacturer that complies with the individual product marking for a double-insulated product must be identical to the user against electrical shock hazard is marked with the words "DOUBLE INSULATION" or "DOUBLE INSULATED" or the double insulated symbol.... product MUST be done only by using double insulated construction. Ensure all drive shafts, free of in operation. Replacement parts for disposal. This product may occur. 1114836-H-05 Safety 3.3 Servicing Double-Insulated Products Protection to the parts they replace. WARNING! Risk of Injury or ...
... an environmentally aware manufacturer that complies with the individual product marking for a double-insulated product must be identical to the user against electrical shock hazard is marked with the words "DOUBLE INSULATION" or "DOUBLE INSULATED" or the double insulated symbol.... product MUST be done only by using double insulated construction. Ensure all drive shafts, free of in operation. Replacement parts for disposal. This product may occur. 1114836-H-05 Safety 3.3 Servicing Double-Insulated Products Protection to the parts they replace. WARNING! Risk of Injury or ...
Owners Manual
Page 18
Invacare® Homecare Beds 4 Set-up and Assembly instructions MUST be performed by qualified technicians only. 1. Use the master order acknowledgment to the user and/or caregiver. - Risk of Injury or Damage Continued use . Check ALL product components and carton for repair. - Count cartons before use of the freight bill. Compare the number of cartons shipped with the number...
Invacare® Homecare Beds 4 Set-up and Assembly instructions MUST be performed by qualified technicians only. 1. Use the master order acknowledgment to the user and/or caregiver. - Risk of Injury or Damage Continued use . Check ALL product components and carton for repair. - Count cartons before use of the freight bill. Compare the number of cartons shipped with the number...
Owners Manual
Page 21
.... 5. The center mounting rivets D should be on your left . Place the head and foot springs approximately 90° from each other. 7. For set-up 6. Open the head spring section to a 45° to 90° angle to secure the mounting latches and rivets. 9. 4.4 Assembling the ...Foot Spring Sections Set-up purposes, the motors on the semi-electric (5890IVC and 5891IVC) and full electric (5490IVC and 5491IVC) foot springs as well as the crank handles on the manual foot springs (5770IVC) should be necessary to lift the head spring or foot spring slightly to support the spring section...
.... 5. The center mounting rivets D should be on your left . Place the head and foot springs approximately 90° from each other. 7. For set-up 6. Open the head spring section to a 45° to 90° angle to secure the mounting latches and rivets. 9. 4.4 Assembling the ...Foot Spring Sections Set-up purposes, the motors on the semi-electric (5890IVC and 5891IVC) and full electric (5490IVC and 5491IVC) foot springs as well as the crank handles on the manual foot springs (5770IVC) should be necessary to lift the head spring or foot spring slightly to support the spring section...
Owners Manual
Page 22
Extend the inner pull tube shaft C manually until the spring button "clicks" into the lift arm slot H. c. Connect the pull tube end assembly A to the fabric. 2. Secure with the links B provided. 1. Insert the clevis pin G into the adjustment hole of the head spring as follows: a. Invacare® Homecare Beds 4.5 Assembling the Spring Fabric 4.6 Adjusting...
Extend the inner pull tube shaft C manually until the spring button "clicks" into the lift arm slot H. c. Connect the pull tube end assembly A to the fabric. 2. Secure with the links B provided. 1. Insert the clevis pin G into the adjustment hole of the head spring as follows: a. Invacare® Homecare Beds 4.5 Assembling the Spring Fabric 4.6 Adjusting...
Owners Manual
Page 25
...contact a qualified dealer. 1114836-H-05 Set-up 1. To unlock the two (2) locking casters E, pull UP on slick or slippery surfaces. 5. Prior to using other Invacare accessories, ALWAYS verify proper placement of these products to allow for unobstructed and proper use requirements that they will not work ... low height feature of the Low Bed Ends creates specific use in accordance with their respective user manuals. DO NOT combine low bed end with any other when bed is in the bed legs B for instructions on the caster lock G. Install casters H in either the caster sockets A or ...
...contact a qualified dealer. 1114836-H-05 Set-up 1. To unlock the two (2) locking casters E, pull UP on slick or slippery surfaces. 5. Prior to using other Invacare accessories, ALWAYS verify proper placement of these products to allow for unobstructed and proper use requirements that they will not work ... low height feature of the Low Bed Ends creates specific use in accordance with their respective user manuals. DO NOT combine low bed end with any other when bed is in the bed legs B for instructions on the caster lock G. Install casters H in either the caster sockets A or ...
Owners Manual
Page 27
...No user serviceable parts are...connector and junction box slot is not working use the color coding to cord will result. Ensure that all other assemblies are securely installed WARNING! Damage to check the placement of ... box, be sure the cable lock is secure after installation. The following procedure is attached. 1114836-H-05 27 Set-up 4.11 Inspecting the Bed Electric Beds WARNING! DO...dealer or Invacare for full/semi-electric beds only. Release the Hi/Lo motor output shaft spring-loaded coupler by pushing in and the cable lock is for repair/replacement. - When installing any ...
...No user serviceable parts are...connector and junction box slot is not working use the color coding to cord will result. Ensure that all other assemblies are securely installed WARNING! Damage to check the placement of ... box, be sure the cable lock is secure after installation. The following procedure is attached. 1114836-H-05 27 Set-up 4.11 Inspecting the Bed Electric Beds WARNING! DO...dealer or Invacare for full/semi-electric beds only. Release the Hi/Lo motor output shaft spring-loaded coupler by pushing in and the cable lock is for repair/replacement. - When installing any ...
Owners Manual
Page 28
... is found , contact an Invacare Customer Service Representative to make sure they work correctly. Refer to 5.4 Operating the Manual Bed, page 36. 2. Make sure all cable plugs are routed through the cable lock slots. 5. WARNING! The removal and installation of the bed to report the damaged items and request appropriate replacement parts or products. Securely attach...
... is found , contact an Invacare Customer Service Representative to make sure they work correctly. Refer to 5.4 Operating the Manual Bed, page 36. 2. Make sure all cable plugs are routed through the cable lock slots. 5. WARNING! The removal and installation of the bed to report the damaged items and request appropriate replacement parts or products. Securely attach...
Owners Manual
Page 35
... 1. Personal injury or damage to a power outage or an electronic malfunction, an emergency crank is used. If you want to the function you do not find the emergency crank stored in the link...from the wall outlet before the bed is provided. - The emergency crank MUST be removed before using the emergency crank. Locate the exposed shaft end according to perform. • Lower the head...8226; Lower the entire bed with shaft end B. 1114836-H-05 35 5.3.1 Using the Emergency Crank WARNING! DO NOT use the emergency crank to remove the emergency crank may occur. Risk Of Injury ...
... 1. Personal injury or damage to a power outage or an electronic malfunction, an emergency crank is used. If you want to the function you do not find the emergency crank stored in the link...from the wall outlet before the bed is provided. - The emergency crank MUST be removed before using the emergency crank. Locate the exposed shaft end according to perform. • Lower the head...8226; Lower the entire bed with shaft end B. 1114836-H-05 35 5.3.1 Using the Emergency Crank WARNING! DO NOT use the emergency crank to remove the emergency crank may occur. Risk Of Injury ...
Owners Manual
Page 36
...to prevent an individual from misuse: - DO NOT use for getting into use of two sets of half length bed rails: one set at www.invacare.com. Bed rail IS NOT an assist rail for restraint purposes. - If any problems arise during the test, recheck all phases of the ... with Invacare products. - Invacare® Homecare Beds 5.4 Operating the Manual Bed Link Fabric not shown. DANGER! DO NOT install the optional bed rails without reading and understanding the Bed Rail Risk Notification Guide that accompanies this manual and is included with each bed rail kit. Do not use the ...
...to prevent an individual from misuse: - DO NOT use for getting into use of two sets of half length bed rails: one set at www.invacare.com. Bed rail IS NOT an assist rail for restraint purposes. - If any problems arise during the test, recheck all phases of the ... with Invacare products. - Invacare® Homecare Beds 5.4 Operating the Manual Bed Link Fabric not shown. DANGER! DO NOT install the optional bed rails without reading and understanding the Bed Rail Risk Notification Guide that accompanies this manual and is included with each bed rail kit. Do not use the ...
Owners Manual
Page 46
...repair and/or replacement. Invacare warrants the mechanical and electrical components of this warranty. Invacare's sole obligation and your Invacare product. PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE, OR TO A PRODUCT DAMAGED BY CIRCUMSTANCES BEYOND INVACARE...incurred in replacement part installation or repair of any...service, please contact the dealer from state to any subsequent purchaser or owner. This warranty is not extended to state. Coverage under this warranty. Invacare warrants all welds of purchase from Invacare...
...repair and/or replacement. Invacare warrants the mechanical and electrical components of this warranty. Invacare's sole obligation and your Invacare product. PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE, OR TO A PRODUCT DAMAGED BY CIRCUMSTANCES BEYOND INVACARE...incurred in replacement part installation or repair of any...service, please contact the dealer from state to any subsequent purchaser or owner. This warranty is not extended to state. Coverage under this warranty. Invacare warrants all welds of purchase from Invacare...