Fluke 8808A Digital Multimeter Datasheet
Page 1
... • 5.5 digit resolution • Basic V dc accuracy of testing: the result is as simple as pushing a button The Fluke 8808A 5.5 digit multimeter has a broad range of functions, measuring volts, ohms and amps with pass/fail indicators eliminates production mistakes, especially those where results ...press multiple buttons to set up a measurement function and range, test limits, or enter other parameters to do. Fluke 8808A Digital Multimeter Extended Specifications Making measurements is either within limits or it makes the measurements you simply press the appropriate setup key....
... • 5.5 digit resolution • Basic V dc accuracy of testing: the result is as simple as pushing a button The Fluke 8808A 5.5 digit multimeter has a broad range of functions, measuring volts, ohms and amps with pass/fail indicators eliminates production mistakes, especially those where results ...press multiple buttons to set up a measurement function and range, test limits, or enter other parameters to do. Fluke 8808A Digital Multimeter Extended Specifications Making measurements is either within limits or it makes the measurements you simply press the appropriate setup key....
Fluke 8808A Digital Multimeter Datasheet
Page 2
8808A Specifications Voltage 100V Setting 90 V to 110 V 120V Setting 108 V to 132 V 220V Setting 198 V to 242 V 240V Setting 216 V to 264 V Frequency 47 Hz to full uncertainty specifications Relative Humidity (non-condensing) Operating Uncontrolled (< 10°C) Power Consumption 15 VA peak (10 W average) Dimensions Height 88 mm (3.46 in) Width 217 mm (8.56 in) Depth 297 mm (11.7 in) Weight 2.1 kg (4.6 Ibs) Display Vacuum Fluorescent Display, segment Environment Temperature Operating 0 rC to 50 rC Storage 40 rC to 70 rC Warm Up hour to 440 Hz.
8808A Specifications Voltage 100V Setting 90 V to 110 V 120V Setting 108 V to 132 V 220V Setting 198 V to 242 V 240V Setting 216 V to 264 V Frequency 47 Hz to full uncertainty specifications Relative Humidity (non-condensing) Operating Uncontrolled (< 10°C) Power Consumption 15 VA peak (10 W average) Dimensions Height 88 mm (3.46 in) Width 217 mm (8.56 in) Depth 297 mm (11.7 in) Weight 2.1 kg (4.6 Ibs) Display Vacuum Fluorescent Display, segment Environment Temperature Operating 0 rC to 50 rC Storage 40 rC to 70 rC Warm Up hour to 440 Hz.
Fluke 8808A Digital Multimeter Datasheet
Page 3
... 0.015 + 0.003 0.015 + 0.004 0.015 + 0.003 0.015 + 0.003 Temperature Coefficient/°C Outside 18 - 28 °C 0.0015 + 0.0005 0.001 + 0.0005 0.0020 + 0.0005 0.0015 + 0.0005 0.0015 + 0.0005 2 Fluke Corporation 8808A Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 3
... 0.015 + 0.003 0.015 + 0.004 0.015 + 0.003 0.015 + 0.003 Temperature Coefficient/°C Outside 18 - 28 °C 0.0015 + 0.0005 0.001 + 0.0005 0.0020 + 0.0005 0.0015 + 0.0005 0.0015 + 0.0005 2 Fluke Corporation 8808A Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 3
Fluke 8808A Digital Multimeter Datasheet
Page 4
AC Voltage Specifications AC Voltage specifications are for ac sinewave signals >5 % of range and For inputs from 1 % to 5 % of range.
AC Voltage Specifications AC Voltage specifications are for ac sinewave signals >5 % of range and For inputs from 1 % to 5 % of range.
Fluke 8808A Digital Multimeter Datasheet
Page 5
Resistance Specifications are for 2-wire resistance plus lead resistance. If REL is not used, add 0.2 for 4-wire resistance function, or 2-wire resistance with REL. Input Characteristics Range Full-Scale (5-1/2 Digits) Slow Resolution Medium Burden Voltage Fast 200 uA 2 mA 20 mA 200 mA 2 A 10 A 199.999 µA 1999.99 µA 19.9999 mA 199.999 mA 1.99999 A 10.0000 A 0.001 µA 0.01 µA 0.1 µA 1 µA 10 µA 100 µA 0.01 µA 0.1 µA 1 µA 10 µA 100 µA 1 mA 0.01 µA 0.1 µA 1 µA 10 µA 100 µA 1 mA Shunt Resistance ...
Resistance Specifications are for 2-wire resistance plus lead resistance. If REL is not used, add 0.2 for 4-wire resistance function, or 2-wire resistance with REL. Input Characteristics Range Full-Scale (5-1/2 Digits) Slow Resolution Medium Burden Voltage Fast 200 uA 2 mA 20 mA 200 mA 2 A 10 A 199.999 µA 1999.99 µA 19.9999 mA 199.999 mA 1.99999 A 10.0000 A 0.001 µA 0.01 µA 0.1 µA 1 µA 10 µA 100 µA 0.01 µA 0.1 µA 1 µA 10 µA 100 µA 1 mA 0.01 µA 0.1 µA 1 µA 10 µA 100 µA 1 mA Shunt Resistance ...
Fluke 8808A Digital Multimeter Datasheet
Page 6
For inputs from 1 % to 5 % of range, add an additional error of 0.1 % of range. Accuracy Accuracy [1] Range 90 days 200 µA 2 mA 20 mA 200 mA 2 A 10 A 23 °C ± 5 °C 0.02 + 0.005 0.015 + 0.005 0.03 + 0.02 0.02 + 0.005 0.05 + 0.02 0.18 + 0.01 Notes: [1] Accuracy given as ± (% of reading + % of range) 1 year 23 °C ± 5 °C 0.03 + 0.005 0.02 + 0.005 0.04 + 0.02 0.03 + 0.008 0.08 + 0.02 0.2 + 0.01 Temperature Coefficient/°C Outside 18 - 28 °C 0.003 + 0.001 0.002 + 0.001 0.005 + 0.001 0.005 + 0.001 0.008 + 0.001 0.008 + 0.001 AC Current ...
For inputs from 1 % to 5 % of range, add an additional error of 0.1 % of range. Accuracy Accuracy [1] Range 90 days 200 µA 2 mA 20 mA 200 mA 2 A 10 A 23 °C ± 5 °C 0.02 + 0.005 0.015 + 0.005 0.03 + 0.02 0.02 + 0.005 0.05 + 0.02 0.18 + 0.01 Notes: [1] Accuracy given as ± (% of reading + % of range) 1 year 23 °C ± 5 °C 0.03 + 0.005 0.02 + 0.005 0.04 + 0.02 0.03 + 0.008 0.08 + 0.02 0.2 + 0.01 Temperature Coefficient/°C Outside 18 - 28 °C 0.003 + 0.001 0.002 + 0.001 0.005 + 0.001 0.005 + 0.001 0.008 + 0.001 0.008 + 0.001 AC Current ...
Fluke 8808A Digital Multimeter Datasheet
Page 7
... Rate Fast Maxiumum Reading 199.99 Resolution 0.01 Diode Test Response Time 100 samples/sec with audible tone Rate Fast Maxiumum Reading 1.9999 V Resolution 0.1 mV 6 Fluke Corporation 8808A Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 7
... Rate Fast Maxiumum Reading 199.99 Resolution 0.01 Diode Test Response Time 100 samples/sec with audible tone Rate Fast Maxiumum Reading 1.9999 V Resolution 0.1 mV 6 Fluke Corporation 8808A Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 7
Fluke 8808A Digital Multimeter Datasheet
Page 8
..., Italian, Simplified Chinese, Japanese), CD Rom with user manual (English). 8 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications Fluke. Printed in U.S.A. 9/2007 3093965 D-EN-N Rev B Pub_ID: 11288-eng Rev 01 Ordering information Models Description 8808A 120V 5.5 Digit Multimeter 8808A 220V 5.5 Digit Multimeter 8808A 100V 5.5 Digit Multimeter 8808A 240V 5.5 Digit Multimeter 8808A/SU includes 8808A package plus, FlukeView Forms basic software, USB to change without notice...
..., Italian, Simplified Chinese, Japanese), CD Rom with user manual (English). 8 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications Fluke. Printed in U.S.A. 9/2007 3093965 D-EN-N Rev B Pub_ID: 11288-eng Rev 01 Ordering information Models Description 8808A 120V 5.5 Digit Multimeter 8808A 220V 5.5 Digit Multimeter 8808A 100V 5.5 Digit Multimeter 8808A 240V 5.5 Digit Multimeter 8808A/SU includes 8808A package plus, FlukeView Forms basic software, USB to change without notice...
User Manual
Page 1
All product names are trademarks of their respective companies. ® 8808A Digital Multimeter Users Manual July 2007 © 2007 Fluke Corporation, All rights reserved.
All product names are trademarks of their respective companies. ® 8808A Digital Multimeter Users Manual July 2007 © 2007 Fluke Corporation, All rights reserved.
User Manual
Page 13
...that allows two properties of a printed Getting Started Manual and a Users Manual on a CD-ROM. 1 Introduction and Specifications Introduction Introduction The Fluke 8808A Digital Multimeter (hereafter referred to be displayed at the same time (e.g., ac voltage in one display and frequency in automated systems. For portability, the Meter... includes a carrying handle that also serves as the Meter) is a 5-1/2 digit dual-display multimeter designed for bench-top operation. The Getting Started Manual contains basic getting started information, contacting...
...that allows two properties of a printed Getting Started Manual and a Users Manual on a CD-ROM. 1 Introduction and Specifications Introduction Introduction The Fluke 8808A Digital Multimeter (hereafter referred to be displayed at the same time (e.g., ac voltage in one display and frequency in automated systems. For portability, the Meter... includes a carrying handle that also serves as the Meter) is a 5-1/2 digit dual-display multimeter designed for bench-top operation. The Getting Started Manual contains basic getting started information, contacting...
User Manual
Page 38
... INPUT V SENSE 4W HI 1000V 750V MAX LO 200 mA MAX 1000 V 600 V CAT I CAT II 300 V 1V LO 500 Vpk 10 A MAX mA 10 A 8808A 5-1/2 DIGIT MULTIMETER 17 16 15 14 13 12 11 10 9 8 7 5 6 Figure 3-1. Front-Panel Features No. All measurements use the INPUT LO terminal as a common input. Name...
... INPUT V SENSE 4W HI 1000V 750V MAX LO 200 mA MAX 1000 V 600 V CAT I CAT II 300 V 1V LO 500 Vpk 10 A MAX mA 10 A 8808A 5-1/2 DIGIT MULTIMETER 17 16 15 14 13 12 11 10 9 8 7 5 6 Figure 3-1. Front-Panel Features No. All measurements use the INPUT LO terminal as a common input. Name...
User Manual
Page 94
...display, proceed as negative, it is 1 A of the voltage reading. Depending on the circumstances, this lead is the same for both circuits. 8808A Users Manual Primary Display REL HOLD Table A-2. For instance, if the external lead resistance is .007 ohm, the "total" common resistance is ...HI 1000V 750V MAX LO 200 mA MAX 1000 V 600 V CAT I CAT II 300 V 1V LO 500 Vpk 10 A MAX mA 10 A 8808A 5-1/2 DIGIT MULTIMETER Figure A-2. A-4 Press Q then E to the test circuit as shown in Figure A-2 and read the measurements on an input signal in the primary display...
...display, proceed as negative, it is 1 A of the voltage reading. Depending on the circumstances, this lead is the same for both circuits. 8808A Users Manual Primary Display REL HOLD Table A-2. For instance, if the external lead resistance is .007 ohm, the "total" common resistance is ...HI 1000V 750V MAX LO 200 mA MAX 1000 V 600 V CAT I CAT II 300 V 1V LO 500 Vpk 10 A MAX mA 10 A 8808A 5-1/2 DIGIT MULTIMETER Figure A-2. A-4 Press Q then E to the test circuit as shown in Figure A-2 and read the measurements on an input signal in the primary display...
User Manual
Page 98
... is the feedback resistor method shown in Figure A-4. For example, determining the leakage current of low-current measurement is needed. 8808A Users Manual Making Low-Level Current Measurements There are many applications where obtaining the utmost accuracy in low-level current measurements is...the burden voltage to be rather low and thus induces an error. Zero Burden Voltage Low-Level Current Measurement eue29.eps A-8 Traditional multimeters make these measurements using a shunt method shown in the shunt resistor is less than the actual value, thus causing an error. IIN...
... is the feedback resistor method shown in Figure A-4. For example, determining the leakage current of low-current measurement is needed. 8808A Users Manual Making Low-Level Current Measurements There are many applications where obtaining the utmost accuracy in low-level current measurements is...the burden voltage to be rather low and thus induces an error. Zero Burden Voltage Low-Level Current Measurement eue29.eps A-8 Traditional multimeters make these measurements using a shunt method shown in the shunt resistor is less than the actual value, thus causing an error. IIN...