Fluke 190-Series II 4-Channel ScopeMeters Product Datasheet
Page 1
...USB memory device; ITY MANAGEMENT SY ® IFIED TO MEET ISO Models 190-204, 199C, and 225 C shown here QUAL 9001 STEM CERT A new generation of ScopeMeter The 190 Series II include these new capabilities: • 4 independent floating isolated inputs, up ... lock while unattended ScopeMeter 190C Series and 190 Series II Rugged performance, speed and ease of industrial electronics. Choose 200 MHz or 100 MHz bandwidth models. ScopeMeter® 190 Series 190 Series II, 190C Series, and 190C Series with Bus Health Technical Data ScopeMeter Series II 190-104 and 190-204: The first high...
...USB memory device; ITY MANAGEMENT SY ® IFIED TO MEET ISO Models 190-204, 199C, and 225 C shown here QUAL 9001 STEM CERT A new generation of ScopeMeter The 190 Series II include these new capabilities: • 4 independent floating isolated inputs, up ... lock while unattended ScopeMeter 190C Series and 190 Series II Rugged performance, speed and ease of industrial electronics. Choose 200 MHz or 100 MHz bandwidth models. ScopeMeter® 190 Series 190 Series II, 190C Series, and 190C Series with Bus Health Technical Data ScopeMeter Series II 190-104 and 190-204: The first high...
Fluke 190-Series II 4-Channel ScopeMeters Product Datasheet
Page 2
... per channel Time base range 5 ns/div to 5 s/div (in a 1-2-4-sequence. average on each input channel Input voltage CAT II 1000 V, CAT III 600 V rated - traces fade out in ScopeRecord™ roll mode Timing accuracy ± (0.01 % of reading... of channels; A versus B (X-Y-mode); X-Y-mode; all with automatic "Pass/Fail testing", Replay 2 Fluke Corporation ScopeMeter 190 Series II Oscilloscope Modes 190C Series 190 Series II 199C, 225C 196C, 215C 192C 190-204 190-104 Vertical deflection Number of channels 2 2 2 4 4 Bandwidth 200 MHz 100 MHz 60 MHz 200...
... per channel Time base range 5 ns/div to 5 s/div (in a 1-2-4-sequence. average on each input channel Input voltage CAT II 1000 V, CAT III 600 V rated - traces fade out in ScopeRecord™ roll mode Timing accuracy ± (0.01 % of reading... of channels; A versus B (X-Y-mode); X-Y-mode; all with automatic "Pass/Fail testing", Replay 2 Fluke Corporation ScopeMeter 190 Series II Oscilloscope Modes 190C Series 190 Series II 199C, 225C 196C, 215C 192C 190-204 190-104 Vertical deflection Number of channels 2 2 2 4 4 Bandwidth 200 MHz 100 MHz 60 MHz 200...
Fluke 190-Series II 4-Channel ScopeMeters Product Datasheet
Page 3
...PF), Watts, VA, VA reactive, phase (between cursors Single vertical line Min-Max and Average voltage at any record length ScopeMeter 190 Series II Fluke Corporation 3 Each screen has date and time-stamp. Non-interlaced video with markers; N to get optimum frequency resolution in FFT... capturing 100 specified events. Video triggering (on external flash memory drive through USB host port. 190C Series 190 Series II 199C, 225C 196C, 215C 192C 190-204 190-104 Trigger and delay Source Any of complex and dynamic signals like motor drive and control signals. ...
...PF), Watts, VA, VA reactive, phase (between cursors Single vertical line Min-Max and Average voltage at any record length ScopeMeter 190 Series II Fluke Corporation 3 Each screen has date and time-stamp. Non-interlaced video with markers; N to get optimum frequency resolution in FFT... capturing 100 specified events. Video triggering (on external flash memory drive through USB host port. 190C Series 190 Series II 199C, 225C 196C, 215C 192C 190-204 190-104 Trigger and delay Source Any of complex and dynamic signals like motor drive and control signals. ...
Fluke 190-Series II 4-Channel ScopeMeters Product Datasheet
Page 4
... Hz to 20 kHz: Voltmeter ranges Ohms Ranges Accuracy Other meter functions Continuity Diode test Amps Temperature 190C Series 190 Series II 199C, 196C, 192C, 215C, 225C, 190-204, 190-104 Via 4 mm banana inputs, fully isolated from scope inputs and scope ground Up to preset values ... % of specified accuracy for each parameter. Up to 82 °F). Scale factors 1 °C/mV or 1 °F/mV 4 Fluke Corporation ScopeMeter 190 Series II Automatically compares the measurement results to four automatic meter measurements can be made at the same time, using an optional current clamp or shunt...
... Hz to 20 kHz: Voltmeter ranges Ohms Ranges Accuracy Other meter functions Continuity Diode test Amps Temperature 190C Series 190 Series II 199C, 196C, 192C, 215C, 225C, 190-204, 190-104 Via 4 mm banana inputs, fully isolated from scope inputs and scope ground Up to preset values ... % of specified accuracy for each parameter. Up to 82 °F). Scale factors 1 °C/mV or 1 °F/mV 4 Fluke Corporation ScopeMeter 190 Series II Automatically compares the measurement results to four automatic meter measurements can be made at the same time, using an optional current clamp or shunt...
Fluke 190-Series II 4-Channel ScopeMeters Product Datasheet
Page 5
...USB host port. Each recorded sample point contains a minimum, a maximum and an average value, plus a date- ScopeMeter 190 Series II Fluke Corporation 5 Direct storage on 190C Series) Horizontal scale Time from start, time of day Zoom Ranges from start , time of day Zoom Up to ...user selectable Memory depth 27,000 or more Horizontal scale Time from start or absolute time. Recorder Modes 190C Series 190 Series II 199C, 196C, 192C, 215C, 225C, 190-204, 190-104 ScopeRecord™ Roll Mode Dual or multiple input waveform storage mode, using deep memory Source and display...
...USB host port. Each recorded sample point contains a minimum, a maximum and an average value, plus a date- ScopeMeter 190 Series II Fluke Corporation 5 Direct storage on 190C Series) Horizontal scale Time from start, time of day Zoom Ranges from start , time of day Zoom Up to ...user selectable Memory depth 27,000 or more Horizontal scale Time from start or absolute time. Recorder Modes 190C Series 190 Series II 199C, 196C, 192C, 215C, 225C, 190-204, 190-104 ScopeRecord™ Roll Mode Dual or multiple input waveform storage mode, using deep memory Source and display...
Fluke 190-Series II 4-Channel ScopeMeters Product Datasheet
Page 6
... User adjustable, temperature compensated 80 cd/m2 typ. battery) Power Line power Battery power Battery charge indicator 190C Series 199C, 196C, 192C, 215C, 225C, 190 Series II 190-204, 190-104 CAT II 1000 V, CAT III 600 V CAT III 1000 V, CAT IV 600 V Maximum voltage between any contact and... x 86.4 mm (4.54 in x 3.4 in); 144 mm (5.67 in instrument) which can be copied on instrument screen 6 Fluke Corporation ScopeMeter 190 Series II When storing data, this is exchanged. floating voltage Maximum probe voltage Maximum BNC input voltage Maximum voltage on BNC input directly CAT...
... User adjustable, temperature compensated 80 cd/m2 typ. battery) Power Line power Battery power Battery charge indicator 190C Series 199C, 196C, 192C, 215C, 225C, 190 Series II 190-204, 190-104 CAT II 1000 V, CAT III 600 V CAT III 1000 V, CAT IV 600 V Maximum voltage between any contact and... x 86.4 mm (4.54 in x 3.4 in); 144 mm (5.67 in instrument) which can be copied on instrument screen 6 Fluke Corporation ScopeMeter 190 Series II When storing data, this is exchanged. floating voltage Maximum probe voltage Maximum BNC input voltage Maximum voltage on BNC input directly CAT...
Fluke 190-Series II 4-Channel ScopeMeters Product Datasheet
Page 7
... optional OC4USB or PM9080 (optical to PC for CAT IV 600 V, CAT III 1000 V; On-screen battery power indicator. 190 Series II 190-204, 190-104 Up to 7 hours using an OC4USB or PM9080 interface cable). EN61010-1-2001, Pollution Degree 2; UL61010B; up to 2,000..., fully insulated from any measurement input channel FlukeView® ScopeMeter® Software FlukeView ScopeMeter software helps you get more out of waveform data, measurement results, instrument settings and screen copies. ScopeMeter 190 Series II Fluke Corporation 7 On-screen battery power indicator. Three-years (parts...
... optional OC4USB or PM9080 (optical to PC for CAT IV 600 V, CAT III 1000 V; On-screen battery power indicator. 190 Series II 190-204, 190-104 Up to 7 hours using an OC4USB or PM9080 interface cable). EN61010-1-2001, Pollution Degree 2; UL61010B; up to 2,000..., fully insulated from any measurement input channel FlukeView® ScopeMeter® Software FlukeView ScopeMeter software helps you get more out of waveform data, measurement results, instrument settings and screen copies. ScopeMeter 190 Series II Fluke Corporation 7 On-screen battery power indicator. Three-years (parts...
Fluke 190-Series II 4-Channel ScopeMeters Product Datasheet
Page 8
... Hard Shell Carrying Case for 190C Series C290 Hard Shell Carrying Case for Fluke 190 Series II VPS210 Voltage probe set 10:1. Ordering Information 190-204 Color ScopeMeter (200 MHz, 4 channel) 190-204/S Color ScopeMeter (200 MHz, 4 channel), with SCC290-kit 190-104 Color ScopeMeter (100 MHz, 4 channel) 190-104/S Color ScopeMeter (100 MHz, 4 channel), with SCC290-kit 225C Color ScopeMeter (200 MHz/2.5 GS/s) with Bus...
... Hard Shell Carrying Case for 190C Series C290 Hard Shell Carrying Case for Fluke 190 Series II VPS210 Voltage probe set 10:1. Ordering Information 190-204 Color ScopeMeter (200 MHz, 4 channel) 190-204/S Color ScopeMeter (200 MHz, 4 channel), with SCC290-kit 190-104 Color ScopeMeter (100 MHz, 4 channel) 190-104/S Color ScopeMeter (100 MHz, 4 channel), with SCC290-kit 225C Color ScopeMeter (200 MHz/2.5 GS/s) with Bus...
Fluke ScopeMeter Family Datasheet - 120, 190, 190 Series II
Page 3
... pollutants. The only IP-51 rated scopes on the market Rugged and shock-proof, ScopeMeter portable oscilloscopes are double-insulated floating oscilloscopes safety rated for measurements in fixed installations, such...Fluke 190C Series is made any outdoor to utility power conductors • Electricity meters, primary overcurrent protection equipment • Outside and service entrance, service drop from pole to building, run between meter and panel • Overhead line to detached building, underground line to low-voltage (< 1000 V) test equipment. 3 The new Fluke 190 Series II...
... pollutants. The only IP-51 rated scopes on the market Rugged and shock-proof, ScopeMeter portable oscilloscopes are double-insulated floating oscilloscopes safety rated for measurements in fixed installations, such...Fluke 190C Series is made any outdoor to utility power conductors • Electricity meters, primary overcurrent protection equipment • Outside and service entrance, service drop from pole to building, run between meter and panel • Overhead line to detached building, underground line to low-voltage (< 1000 V) test equipment. 3 The new Fluke 190 Series II...
Fluke ScopeMeter Family Datasheet - 120, 190, 190 Series II
Page 8
...ScopeMeter 120 Series: Three-in one simplicity for electrical or electromechanical troubleshooting. • It's an oscilloscope, a multimeter and a paperless recorder in portable oscilloscopes-from 20 MHz to visually check the overall signal quality • Choose 200 MHz or 100 MHz bandwidth models 8 Fluke... with up to 400 ps resolution • CAT II 1000 V/CAT III 600 V safety rated • Plus all the standard features of ScopeMeter including TrendPlot™, Connect-and-View™ triggering and ScopeRecord™ ScopeMeter 190 Series II: Be prepared for anything in a CAT IV world...
...ScopeMeter 120 Series: Three-in one simplicity for electrical or electromechanical troubleshooting. • It's an oscilloscope, a multimeter and a paperless recorder in portable oscilloscopes-from 20 MHz to visually check the overall signal quality • Choose 200 MHz or 100 MHz bandwidth models 8 Fluke... with up to 400 ps resolution • CAT II 1000 V/CAT III 600 V safety rated • Plus all the standard features of ScopeMeter including TrendPlot™, Connect-and-View™ triggering and ScopeRecord™ ScopeMeter 190 Series II: Be prepared for anything in a CAT IV world...
Fluke ScopeMeter Family Datasheet - 120, 190, 190 Series II
Page 9
optional Fluke 190 Series II 190-104 100 MHz 4 • • • • - 190-204 200 MHz 4 • • • • - - 1000 V 600 V 7 hrs, Li-ion - 1000 V 600 V 7 hrs, Li-ion standard standard - optional ... Retail security devices, surveillance and monitoring equipment, RFID • • • • • • • • • Fluke ScopeMeter Quick Selection Guide Fluke 190C Series Features 192C Bandwidth 60 MHz Number of inputs 2 + 1 Dual Input TrendPlot™ • Four Input TrendPlot™ - RS-232 interface to...
optional Fluke 190 Series II 190-104 100 MHz 4 • • • • - 190-204 200 MHz 4 • • • • - - 1000 V 600 V 7 hrs, Li-ion - 1000 V 600 V 7 hrs, Li-ion standard standard - optional ... Retail security devices, surveillance and monitoring equipment, RFID • • • • • • • • • Fluke ScopeMeter Quick Selection Guide Fluke 190C Series Features 192C Bandwidth 60 MHz Number of inputs 2 + 1 Dual Input TrendPlot™ • Four Input TrendPlot™ - RS-232 interface to...
Fluke ScopeMeter Family Datasheet - 120, 190, 190 Series II
Page 10
... Fluke. Mainframes ScopeMeter 190 Series II Fluke 190-104, ScopeMeter Series II 100 MHz 4 Channel Fluke 190-104/S, ScopeMeter Series II 100 MHz 4 Channel with SCC Kit Fluke 190-204, ScopeMeter Series II 200 MHz 4 Channel Fluke 190-204/S, ScopeMeter Series II 200 MHz 4 Channel with SCC Kit ScopeMeter 190C Series Fluke 192C, ScopeMeter 60 MHz Fluke 192C/S, ScopeMeter 60 MHz with SCC Kit Fluke 196C, ScopeMeter 100 MHz Fluke 196C/S, ScopeMeter 100 MHz with SCC Kit Fluke 199C, ScopeMeter 200 MHz Fluke 199C/S, ScopeMeter 200 MHz with SCC Kit Fluke 225C, ScopeMeter...
... Fluke. Mainframes ScopeMeter 190 Series II Fluke 190-104, ScopeMeter Series II 100 MHz 4 Channel Fluke 190-104/S, ScopeMeter Series II 100 MHz 4 Channel with SCC Kit Fluke 190-204, ScopeMeter Series II 200 MHz 4 Channel Fluke 190-204/S, ScopeMeter Series II 200 MHz 4 Channel with SCC Kit ScopeMeter 190C Series Fluke 192C, ScopeMeter 60 MHz Fluke 192C/S, ScopeMeter 60 MHz with SCC Kit Fluke 196C, ScopeMeter 100 MHz Fluke 196C/S, ScopeMeter 100 MHz with SCC Kit Fluke 199C, ScopeMeter 200 MHz Fluke 199C/S, ScopeMeter 200 MHz with SCC Kit Fluke 225C, ScopeMeter...