Fluke 114,115,117 Manual
Page 2
... limitation of liability may not apply to that Service Center with a description of the problem. Box 1186 5602 BD Eindhoven The Netherlands LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three ...Fluke authorized service center to obtain return authorization information, then send the product to you. 11/99 Fluke Corporation P.O. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. P.O. This warranty does not cover fuses, disposable batteries, or damage from the date of operation...
... limitation of liability may not apply to that Service Center with a description of the problem. Box 1186 5602 BD Eindhoven The Netherlands LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three ...Fluke authorized service center to obtain return authorization information, then send the product to you. 11/99 Fluke Corporation P.O. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. P.O. This warranty does not cover fuses, disposable batteries, or damage from the date of operation...
Fluke 114,115,117 Manual
Page 3
...) position. 1 CAT III meters are designed to protect against transients in fixed-equipment installations at www.fluke.com. Contacting Fluke To contact Fluke, call: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe:...battery-powered, true-rms multimeters (hereafter "the Meter") with a lead in an incorrect terminal. True-rms Multimeters Introduction The Fluke Model 114, Model 115, and Model 117 are in the correct terminals, LEAd is briefly displayed and an audible beep sounds when you move the rotary switch to or from transient impulses. This manual...
...) position. 1 CAT III meters are designed to protect against transients in fixed-equipment installations at www.fluke.com. Contacting Fluke To contact Fluke, call: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe:...battery-powered, true-rms multimeters (hereafter "the Meter") with a lead in an incorrect terminal. True-rms Multimeters Introduction The Fluke Model 114, Model 115, and Model 117 are in the correct terminals, LEAd is briefly displayed and an audible beep sounds when you move the rotary switch to or from transient impulses. This manual...
Fluke 114,115,117 Manual
Page 4
...battery door or Meter case. 2 If in this manual or the protection provided by the Meter might be impaired. • Do not use the Meter or test leads if they appear damaged, or if the Meter is not operating properly. • Always use proper terminals, switch position, and range for measurements. • Verify the Meter's operation... • Use caution with voltages above 30 V ac rms, 42 V ac peak, or 60 V dc. A "WCaution" statement identifies conditions and actions that could damage the Meter or the equipment under test. 114, 115, and 117 Users Manual Safety Information A "XWWarning...
...battery door or Meter case. 2 If in this manual or the protection provided by the Meter might be impaired. • Do not use the Meter or test leads if they appear damaged, or if the Meter is not operating properly. • Always use proper terminals, switch position, and range for measurements. • Verify the Meter's operation... • Use caution with voltages above 30 V ac rms, 42 V ac peak, or 60 V dc. A "WCaution" statement identifies conditions and actions that could damage the Meter or the equipment under test. 114, 115, and 117 Users Manual Safety Information A "XWWarning...
Fluke 114,115,117 Manual
Page 5
... by this function's low input impedance (≈3 kΩ)(114 and 117 only). Symbols B AC (Alternating Current) F DC (Direct Current) X Hazardous voltage N Battery (Low battery when shown on the display.) I Fuse T Double Insulated W Important Information; Refer to measure voltages in circuits that could be damaged by local or national authorities when working alone. • Use only the replacement fuse specified or the protection...
... by this function's low input impedance (≈3 kΩ)(114 and 117 only). Symbols B AC (Alternating Current) F DC (Direct Current) X Hazardous voltage N Battery (Low battery when shown on the display.) I Fuse T Double Insulated W Important Information; Refer to measure voltages in circuits that could be damaged by local or national authorities when working alone. • Use only the replacement fuse specified or the protection...
Fluke 114,115,117 Manual
Page 6
A B C D E 4 Symbol w s R O Y 6 5 4 3 2 1 16 15 VoltAlert 7 14 17 8 9 10 11 12 13 18 Meaning The Meter is a negative value. X Unsafe voltage. The Meter function is set to Diode Test Input is in the VoltAlert™ non-contact voltage detect mode. edy02f.eps Model 117 114, 115, & 117 115 & 117 114, 115, & 117 114, 115, & 117 114, 115, and 117 Users Manual Display No. The Meter function is set to Continuity. Measured input voltage ≥30 V, or voltage overload condition (OL).
A B C D E 4 Symbol w s R O Y 6 5 4 3 2 1 16 15 VoltAlert 7 14 17 8 9 10 11 12 13 18 Meaning The Meter is a negative value. X Unsafe voltage. The Meter function is set to Diode Test Input is in the VoltAlert™ non-contact voltage detect mode. edy02f.eps Model 117 114, 115, & 117 115 & 117 114, 115, & 117 114, 115, & 117 114, 115, and 117 Users Manual Display No. The Meter function is set to Continuity. Measured input voltage ≥30 V, or voltage overload condition (OL).
Fluke 114,115,117 Manual
Page 7
... 114, 115 & 117 L N Battery low warning. 114, 115, & 117 M 610000 mV Indicates the Meter's range selection. 114, 115, & 117 N (Bar graph) Analog display. 114, 115, & 117 O Auto Volts The Meter is rotated to or from any A position. 115 & 117 5 User sets the Meter's range. 114, 115, & 117 P + Bar graph polarity 114, 115, & 117 Q 0L W The input is too large for best resolution. 114, 115, & 117 Manual Manual ranging. Display freezes present reading. 114...
... 114, 115 & 117 L N Battery low warning. 114, 115, & 117 M 610000 mV Indicates the Meter's range selection. 114, 115, & 117 N (Bar graph) Analog display. 114, 115, & 117 O Auto Volts The Meter is rotated to or from any A position. 115 & 117 5 User sets the Meter's range. 114, 115, & 117 P + Bar graph polarity 114, 115, & 117 Q 0L W The input is too large for best resolution. 114, 115, & 117 Manual Manual ranging. Display freezes present reading. 114...
Fluke 114,115,117 Manual
Page 8
... Error Messages Battery must be repaired before the Meter will operate. C Input terminal for all measurements. 114, 115, and 117 Users Manual Terminals 1 V A COM 3 10 A FUSED 2 No. Internal error. Calibration required. B Common (return) terminal for measuring voltage, continuity, resistance, capacitance, frequency and testing diodes. Internal error. The Meter must be repaired before the Meter will operate. 6 Meter calibration is required before it will operate. The Meter must be replaced before it will operate. Description A Input...
... Error Messages Battery must be repaired before the Meter will operate. C Input terminal for all measurements. 114, 115, and 117 Users Manual Terminals 1 V A COM 3 10 A FUSED 2 No. Internal error. Calibration required. B Common (return) terminal for measuring voltage, continuity, resistance, capacitance, frequency and testing diodes. Internal error. The Meter must be repaired before the Meter will operate. 6 Meter calibration is required before it will operate. The Meter must be replaced before it will operate. Description A Input...
Fluke 114,115,117 Manual
Page 9
... turns off at >250 Ω. 114, 115 & 117 114, 115 & 117 R S j Hz (button) I w Diode Test. Auto-V LoZ, AC mV and AC amps are true-rms. AC voltage is ac-coupled. DC-coupled. Frequency from .001 V to 20 A, 30 seconds on the sensed input with a low impedance input. True-rms Multimeters Rotary Switch Positions Rotary Switch Positions Switch Position Measurement Function Model x e Hz (button...
... turns off at >250 Ω. 114, 115 & 117 114, 115 & 117 R S j Hz (button) I w Diode Test. Auto-V LoZ, AC mV and AC amps are true-rms. AC voltage is ac-coupled. DC-coupled. Frequency from .001 V to 20 A, 30 seconds on the sensed input with a low impedance input. True-rms Multimeters Rotary Switch Positions Rotary Switch Positions Switch Position Measurement Function Model x e Hz (button...
Fluke 114,115,117 Manual
Page 10
... all readings. Pressing any button or turning the rotary switch awakens the Meter. K is displayed. • To resume MIN MAX AVG recording, press fagain. • To exit and erase stored readings, press p for 20 minutes. 114, 115, and 117 Users Manual Battery Saver ("Sleep Mode") The Meter automatically enters "Sleep mode" and blanks the display if there is no function change, range change when you override Autorange...
... all readings. Pressing any button or turning the rotary switch awakens the Meter. K is displayed. • To resume MIN MAX AVG recording, press fagain. • To exit and erase stored readings, press p for 20 minutes. 114, 115, and 117 Users Manual Battery Saver ("Sleep Mode") The Meter automatically enters "Sleep mode" and blanks the display if there is no function change, range change when you override Autorange...
Fluke 114,115,117 Manual
Page 11
... Options are canceled when you turn the Meter off . In the Manual Range mode, press qto increment the range. Note You cannot manually change the range in MIN MAX AVG or Display Hold, the Meter beeps twice, indicating an invalid operation and the range does not change. 3. See page 14. bEEP is displayed when enabled. Disables automatic power-down the button indicated in the following pages show...
... Options are canceled when you turn the Meter off . In the Manual Range mode, press qto increment the range. Note You cannot manually change the range in MIN MAX AVG or Display Hold, the Meter beeps twice, indicating an invalid operation and the range does not change. 3. See page 14. bEEP is displayed when enabled. Disables automatic power-down the button indicated in the following pages show...
Fluke 114,115,117 Manual
Page 13
... or ac voltage measurement based on the input applied between the V or + and COM jacks. Measuring AC and DC Millivolts Millivolts AC Millivolts DC DC edy03f.eps Using Auto Volts Selection (114 & 117 only) With the function switch in the l position, the Meter measures ac plus dc millivolts. Press g to switch the Meter to ghost voltages. Measuring AC and...
... or ac voltage measurement based on the input applied between the V or + and COM jacks. Measuring AC and DC Millivolts Millivolts AC Millivolts DC DC edy03f.eps Using Auto Volts Selection (114 & 117 only) With the function switch in the l position, the Meter measures ac plus dc millivolts. Press g to switch the Meter to ghost voltages. Measuring AC and...
Fluke 114,115,117 Manual
Page 16
... (117 only) VoltAlert Hot Neutral VoltAlert RANGE Lo edy09f.eps The Meter measures the frequency of a signal by counting the number of the measured signal is 0 V, 0 A for a stable reading. Frequency works with ac functions only. In frequency, the bar graph and range annunciator indicate the ac voltage or current present. 114, 115, and 117 Users Manual Measuring Frequency (115 & 117 only) XWWarning To avoid...
... (117 only) VoltAlert Hot Neutral VoltAlert RANGE Lo edy09f.eps The Meter measures the frequency of a signal by counting the number of the measured signal is 0 V, 0 A for a stable reading. Frequency works with ac functions only. In frequency, the bar graph and range annunciator indicate the ac voltage or current present. 114, 115, and 117 Users Manual Measuring Frequency (115 & 117 only) XWWarning To avoid...
Fluke 114,115,117 Manual
Page 17
...power connectors or sockets where the actual ac voltage is recessed within the connector itself. Operation may be effected by differences in the "Hi" setting. This setting is not saved when the Meter is no indication, voltage could still be used on other styles of the Meter close to switch the Meter into sleep mode..., hold qwhile turning on cables with shielded wire. The "Hi" setting allows for ac voltage detection on flush mounted wall sockets, power strips, flush mounted industrial outlets and various power cords. The VoltAlert detector works in bare wire applications with voltages...
...power connectors or sockets where the actual ac voltage is recessed within the connector itself. Operation may be effected by differences in the "Hi" setting. This setting is not saved when the Meter is no indication, voltage could still be used on other styles of the Meter close to switch the Meter into sleep mode..., hold qwhile turning on cables with shielded wire. The "Hi" setting allows for ac voltage detection on flush mounted wall sockets, power strips, flush mounted industrial outlets and various power cords. The VoltAlert detector works in bare wire applications with voltages...
Fluke 114,115,117 Manual
Page 18
...useful for example (see below . Testing the Fuse (115 & 117 only) Test fuse as shown below ), the major divisions on the scale represent 0, 15, 30, 45, and 60 V. The number of segments indicates the measured value and is relative to the full-scale value of −30 V turns on an analog meter. In frequency...left. In the 60 V range, for making peak and null adjustments. An input of the selected range. It has an overload indicator (>) to the right and a polarity indicator (+)to the middle of the scale. Because the bar graph is much faster than the digital display, the bar graph is ...
...useful for example (see below . Testing the Fuse (115 & 117 only) Test fuse as shown below ), the major divisions on the scale represent 0, 15, 30, 45, and 60 V. The number of segments indicates the measured value and is relative to the full-scale value of −30 V turns on an analog meter. In frequency...left. In the 60 V range, for making peak and null adjustments. An input of the selected range. It has an overload indicator (>) to the right and a polarity indicator (+)to the middle of the scale. Because the bar graph is much faster than the digital display, the bar graph is ...
Fluke 114,115,117 Manual
Page 19
... install the battery directly into the case, bottom edge first, until it from the case. The battery fits inside the battery door, which is fully seated. True-rms Multimeters Maintenance Fuse edy11f.eps 17 Install and tighten battery door screw. To remove the battery door for battery replacement: 1. Remove the test leads from the Meter before opening the case or battery door. • Use...
... install the battery directly into the case, bottom edge first, until it from the case. The battery fits inside the battery door, which is fully seated. True-rms Multimeters Maintenance Fuse edy11f.eps 17 Install and tighten battery door screw. To remove the battery door for battery replacement: 1. Remove the test leads from the Meter before opening the case or battery door. • Use...
Fluke 114,115,117 Manual
Page 20
... the case or lens/window. Cleaning Wipe the case with an 11 A, 1000 V, FAST fuse having a minimum interrupt rating of 17,000 A. 114, 115, and 117 Users Manual To open the case for fuse replacement: 1. Remove the test leads from its holster. Remove the Meter from the Meter 2. Finally, insert the Meter into its holster. 3. Use only Fluke PN 803293. 6. Remove two...
... the case or lens/window. Cleaning Wipe the case with an 11 A, 1000 V, FAST fuse having a minimum interrupt rating of 17,000 A. 114, 115, and 117 Users Manual To open the case for fuse replacement: 1. Remove the test leads from its holster. Remove the Meter from the Meter 2. Finally, insert the Meter into its holster. 3. Use only Fluke PN 803293. 6. Remove two...
Fluke 114,115,117 Manual
Page 21
... Edition for A input (115 & 117 only 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 °C to + 50 °C Storage: -40 °C to 90 %. Extended specifications are available at 0 % to + 60 °C Temperature Coefficient 0.1 x (specified accuracy)/°C (28 °C) Operating Altitude 2,000 meters Battery 9 Volt Alkaline...
... Edition for A input (115 & 117 only 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 °C to + 50 °C Storage: -40 °C to 90 %. Extended specifications are available at 0 % to + 60 °C Temperature Coefficient 0.1 x (specified accuracy)/°C (28 °C) Operating Altitude 2,000 meters Battery 9 Volt Alkaline...
Fluke 114,115,117 Manual
Page 24
... % + 3 115, 117 99.99 Hz Hz (V or A input)[2] 999.9 Hz 9.999 kHz 50.00 kHz 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 % + 2 115, 117 Notes: [1] All ac ranges except Auto-V LoZ are specified from 1 % to 5 kHz. [3] >10 A...114, 115, and 117 Users Manual Table 1 Accuracy Specifications (cont.) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model AC Amps Truerms[1] (45 Hz to 500 Hz) 6.000 A 10.00 A[3] 20 A for 30 seconds max., 10 minutes rest min. 0.001 A 0.01 A 1.5 % + 3 115, 117 DC Amps 6.000 A 10.00 A[3] 20 A for this and other true-rms meters to display...
... % + 3 115, 117 99.99 Hz Hz (V or A input)[2] 999.9 Hz 9.999 kHz 50.00 kHz 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 % + 2 115, 117 Notes: [1] All ac ranges except Auto-V LoZ are specified from 1 % to 5 kHz. [3] >10 A...114, 115, and 117 Users Manual Table 1 Accuracy Specifications (cont.) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model AC Amps Truerms[1] (45 Hz to 500 Hz) 6.000 A 10.00 A[3] 20 A for 30 seconds max., 10 minutes rest min. 0.001 A 0.01 A 1.5 % + 3 115, 117 DC Amps 6.000 A 10.00 A[3] 20 A for this and other true-rms meters to display...
Fluke 114 Digital Multimeter Datasheet
Page 1
... Technical Data Compact true-rms meter for electrical troubleshooting The Fluke 114 is specified for "go/no-go" testing. The accuracy specifications take the form of: ± ( [ % of 18 °C to 28 °C, with relative humidity at operating temperatures of Reading ] + [ Counts ] ) Maximum voltage between any terminal and earth ground Surge protection Display Bar graph Operating temperature Storage temperature Battery Battery...
... Technical Data Compact true-rms meter for electrical troubleshooting The Fluke 114 is specified for "go/no-go" testing. The accuracy specifications take the form of: ± ( [ % of 18 °C to 28 °C, with relative humidity at operating temperatures of Reading ] + [ Counts ] ) Maximum voltage between any terminal and earth ground Surge protection Display Bar graph Operating temperature Storage temperature Battery Battery...
Fluke 114 Digital Multimeter Datasheet
Page 2
... http://www.fluke.com ©2006 Fluke Corporation. Because inputs below 1 % of range. Ordering Information Fluke-114 Electrical Multimeter Included TL75 Test leads, holster, User's manual and 9 V battery (installed). Fluke Corporation 114 Electrical Multimeter Fluke. PO Box 1186, 5602 BD Eindhoven, The Netherlands For more information call: In... In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other true-rms meters to display non-zero readings when the test leads are shorted together. Accuracy specifications Measurement DC millivolts DC volts Auto volts Range 600.0 mV 6.000 V...
... http://www.fluke.com ©2006 Fluke Corporation. Because inputs below 1 % of range. Ordering Information Fluke-114 Electrical Multimeter Included TL75 Test leads, holster, User's manual and 9 V battery (installed). Fluke Corporation 114 Electrical Multimeter Fluke. PO Box 1186, 5602 BD Eindhoven, The Netherlands For more information call: In... In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other true-rms meters to display non-zero readings when the test leads are shorted together. Accuracy specifications Measurement DC millivolts DC volts Auto volts Range 600.0 mV 6.000 V...