Owners Manual
Page 1
User: BEFORE using this product, read this product. For more information regarding Invacare products, parts, and services, please visit www.invacare.com User Manual Perfecto2™Series Oxygen Concentrators without SensO2®, Oxygen Concentrators with SensO2 HomeFill®Compatible Perfecto2 with SensO2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this manual and save for future reference.
User: BEFORE using this product, read this product. For more information regarding Invacare products, parts, and services, please visit www.invacare.com User Manual Perfecto2™Series Oxygen Concentrators without SensO2®, Oxygen Concentrators with SensO2 HomeFill®Compatible Perfecto2 with SensO2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this manual and save for future reference.
Owners Manual
Page 7
..., the concentrator MUST be located in contact with the concentrator while wet. No user serviceable parts. DO NOT place or store product where it is required by Invacare. DO NOT use of this product. Avoid creation of any spark near medical oxygen equipment. ... shock. Part No 1143482 7 Perfecto2™Series Avoid using while bathing. SECTION 1-GENERAL GUIDELINES SECTION 1-GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Perfecto2 concentrator these instructions MUST be kept away from the oxygen concentrator, tubing and connections, and all...
..., the concentrator MUST be located in contact with the concentrator while wet. No user serviceable parts. DO NOT place or store product where it is required by Invacare. DO NOT use of this product. Avoid creation of any spark near medical oxygen equipment. ... shock. Part No 1143482 7 Perfecto2™Series Avoid using while bathing. SECTION 1-GENERAL GUIDELINES SECTION 1-GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Perfecto2 concentrator these instructions MUST be kept away from the oxygen concentrator, tubing and connections, and all...
Owners Manual
Page 8
... used with this product and not exceed 50 ft in length. Perfecto2™Series 8 Part No 1143482 This does not include normal maintenance items. See maintenance section for user maintenance items. A product should NEVER be ... heated or hot surfaces, including space heaters, stoves and similar electrical appliances. Invacare recommends that each concentrator be left unattended when plugged in parallel or series with EMC limits specified by pulling on and running for patients using this device who are unable to comply with other oxygen concentrators or oxygen therapy devices.
... used with this product and not exceed 50 ft in length. Perfecto2™Series 8 Part No 1143482 This does not include normal maintenance items. See maintenance section for user maintenance items. A product should NEVER be ... heated or hot surfaces, including space heaters, stoves and similar electrical appliances. Invacare recommends that each concentrator be left unattended when plugged in parallel or series with EMC limits specified by pulling on and running for patients using this device who are unable to comply with other oxygen concentrators or oxygen therapy devices.
Owners Manual
Page 10
... your Invacare dealer. IOH200 Cabinet Filter *NOTE: This outlet fitting is to the HomeFill ownerʹs manual, part number 1100873, for filling oxygen cylinders with the concentrator should be used only for connection and operating instructions. Refer to be inserted into the outlet Power Cord *Outlet Fitting fitting. IRCPF16 (Perfecto2 5 Only) HomeFill home oxygen compressor...
... your Invacare dealer. IOH200 Cabinet Filter *NOTE: This outlet fitting is to the HomeFill ownerʹs manual, part number 1100873, for filling oxygen cylinders with the concentrator should be used only for connection and operating instructions. Refer to be inserted into the outlet Power Cord *Outlet Fitting fitting. IRCPF16 (Perfecto2 5 Only) HomeFill home oxygen compressor...
Owners Manual
Page 11
... and packing materials for storage until use of repackaged concentrator. Store the repackaged oxygen concentrator in the upright position to be reshipped by common carrier, additional cartons are available from Invacare. If damage is to prevent cabinet damage while being...oxygen concentrator is evident, notify the carrier, or your local dealer. 2. Inspect all the components from the carton. Remove all loose packing from the carton. 3. Storage 1. SECTION 3-HANDLING SECTION 3-HANDLING The concentrator should ALWAYS be kept in a dry area. 2. Part No 1143482 11 Perfecto2...
... and packing materials for storage until use of repackaged concentrator. Store the repackaged oxygen concentrator in the upright position to be reshipped by common carrier, additional cartons are available from Invacare. If damage is to prevent cabinet damage while being...oxygen concentrator is evident, notify the carrier, or your local dealer. 2. Inspect all the components from the carton. Remove all loose packing from the carton. 3. Storage 1. SECTION 3-HANDLING SECTION 3-HANDLING The concentrator should ALWAYS be kept in a dry area. 2. Part No 1143482 11 Perfecto2...
Owners Manual
Page 12
... of the Invacare Pediatric Flowmeter IRCPF16 Accessory. 0 L/min to 5 L/min For flowrates less than 1 L/min, we recommend the use of concentration levels. From 6,000 ft (1828 meters) to 13,129 ft (4000 meters) below 90% efficiency. *Oxygen Output Concentration Levels: Maximum Outlet Pressure: Flow Range: Low Flow Alarm: Average Power Consumption: All Perfecto2 Models while...
... of the Invacare Pediatric Flowmeter IRCPF16 Accessory. 0 L/min to 5 L/min For flowrates less than 1 L/min, we recommend the use of concentration levels. From 6,000 ft (1828 meters) to 13,129 ft (4000 meters) below 90% efficiency. *Oxygen Output Concentration Levels: Maximum Outlet Pressure: Flow Range: Low Flow Alarm: Average Power Consumption: All Perfecto2 Models while...
Owners Manual
Page 15
... surface, such as you through a nasal cannula. You may select a room where using your oxygen concentrator. Your concentrator can be most convenient. Refer to you use indoors. Part No 1143482 15 Perfecto2™Series He/She should be blocked. Your oxygen concentrator will serve as a reference as a bed or couch, where the air opening may result...
... surface, such as you through a nasal cannula. You may select a room where using your oxygen concentrator. Your concentrator can be most convenient. Refer to you use indoors. Part No 1143482 15 Perfecto2™Series He/She should be blocked. Your oxygen concentrator will serve as a reference as a bed or couch, where the air opening may result...
Owners Manual
Page 17
... 5.3. 7. After assembly, ensure that oxygen is securely fastened. Attach oxygen tubing from the humidifier bottle/adapter assembly to oxygen outlet connector on the humidifier bottle counterclockwise until it is flowing through the cannula. Part No 1143482 17 Perfecto2™Series SECTION 5-OPERATING INSTRUCTIONS Humidifier Bottle..."B" in FIGURE 5.3. 8. See Detail "B" in FIGURE 5.3. 9. Place the humidifier bottle/adapter assembly in the humidifier compartment on the concentrator. Attach the cannula/patient supply tubing to the humidifier bottle by turning the wing nut on the...
... 5.3. 7. After assembly, ensure that oxygen is securely fastened. Attach oxygen tubing from the humidifier bottle/adapter assembly to oxygen outlet connector on the humidifier bottle counterclockwise until it is flowing through the cannula. Part No 1143482 17 Perfecto2™Series SECTION 5-OPERATING INSTRUCTIONS Humidifier Bottle..."B" in FIGURE 5.3. 8. See Detail "B" in FIGURE 5.3. 9. Place the humidifier bottle/adapter assembly in the humidifier compartment on the concentrator. Attach the cannula/patient supply tubing to the humidifier bottle by turning the wing nut on the...
Owners Manual
Page 20
SECTION 5-OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators - If purity falls below factory preset standards, indicator lights on the control panel will illuminate. OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators - Units with SensO2 The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. Units with SensO2 Perfecto2™Series 20 Part No 1143482
SECTION 5-OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators - If purity falls below factory preset standards, indicator lights on the control panel will illuminate. OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators - Units with SensO2 The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. Units with SensO2 Perfecto2™Series 20 Part No 1143482
Owners Manual
Page 21
Have a back-up supply of oxygen immediately. Call a qualified technician. YELLOW Solid B. Auto Shut Down Use Backup RED Call Supplier Normal I/O GREEN FIGURE 5.7 Front Panel Indicators - Units without SensO2 NOTE: For this procedure, refer to FIGURE 5.7. Switch to a back-up supply of the Concentrator NOTE: Concentrator may be used during the initial start warm...
Have a back-up supply of oxygen immediately. Call a qualified technician. YELLOW Solid B. Auto Shut Down Use Backup RED Call Supplier Normal I/O GREEN FIGURE 5.7 Front Panel Indicators - Units without SensO2 NOTE: For this procedure, refer to FIGURE 5.7. Switch to a back-up supply of the Concentrator NOTE: Concentrator may be used during the initial start warm...
Owners Manual
Page 22
LABEL SYMBOL I/O STATUS SYSTEM OKAY SYSTEM FAILURE INDICATOR LIGHTS (LED) GREEN Indicator Light RED Indicator light Continuous Audible Alarm Sieve-GARD™Compressor Shutdown Call a qualified technician. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of oxygen immediately. Perfecto2™Series 22 Part No 1143482 O2 purity will build to a back-up supply of the Concentrator NOTE: Concentrator may be used during this period. Switch to a maximum during the initial start warm‐up time (approximately 30 min).
LABEL SYMBOL I/O STATUS SYSTEM OKAY SYSTEM FAILURE INDICATOR LIGHTS (LED) GREEN Indicator Light RED Indicator light Continuous Audible Alarm Sieve-GARD™Compressor Shutdown Call a qualified technician. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of oxygen immediately. Perfecto2™Series 22 Part No 1143482 O2 purity will build to a back-up supply of the Concentrator NOTE: Concentrator may be used during this period. Switch to a maximum during the initial start warm‐up time (approximately 30 min).
Owners Manual
Page 25
Perfecto2™Series I RC FIGURE 6.2 Preventive Maintenance Record Part No 1143482 25 SECTION 6-MAINTENANCE SERIAL NO. NOTE: 4,380 hours are equivalent to Preventive Maintenance section of .... MODEL NO. PREVENTIVE MAINTENANCE RECORD ON EACH INSPECTION Record Date of Service Manual. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter...
Perfecto2™Series I RC FIGURE 6.2 Preventive Maintenance Record Part No 1143482 25 SECTION 6-MAINTENANCE SERIAL NO. NOTE: 4,380 hours are equivalent to Preventive Maintenance section of .... MODEL NO. PREVENTIVE MAINTENANCE RECORD ON EACH INSPECTION Record Date of Service Manual. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter...
Owners Manual
Page 26
SOLUTION: 1. If problem recurs, use the Pediatric Flowmeter IRCPF16 (Only with Perfecto2 5). Move oxygen concentrator at 0.5 L/min or less. 1. Call Service Provider. Alarm: Rapid Beep....Beep... Flowmeter set at least three inches away from walls, draperies or furniture. 2. If prescribed ...
SOLUTION: 1. If problem recurs, use the Pediatric Flowmeter IRCPF16 (Only with Perfecto2 5). Move oxygen concentrator at 0.5 L/min or less. 1. Call Service Provider. Alarm: Rapid Beep....Beep... Flowmeter set at least three inches away from walls, draperies or furniture. 2. If prescribed ...
Owners Manual
Page 27
...least three inches from walls, draperies, and furniture. 5. DO NOT use extension cords. Internal repairs YELLOW Light required. Flowmeter set to IRC5PO2. Kinked or blocked tubing, cannula or humidifier.* 3. Internal repairs required. Inspect for 60 seconds and then turn power back On. ... cabinet filters. 4b. Low oxygen purity.* 2. GREEN light with 1. Part No 1143482 27 Perfecto2™Series Alarm: Continuous On RED light only. * Only applies to 1.0 L/min or more. 4a. Once corrected, turn power Off for kinks or blockages. Move concentrator at outlet. 6. SECTION 7-...
...least three inches from walls, draperies, and furniture. 5. DO NOT use extension cords. Internal repairs YELLOW Light required. Flowmeter set to IRC5PO2. Kinked or blocked tubing, cannula or humidifier.* 3. Internal repairs required. Inspect for 60 seconds and then turn power back On. ... cabinet filters. 4b. Low oxygen purity.* 2. GREEN light with 1. Part No 1143482 27 Perfecto2™Series Alarm: Continuous On RED light only. * Only applies to 1.0 L/min or more. 4a. Once corrected, turn power Off for kinks or blockages. Move concentrator at outlet. 6. SECTION 7-...
Owners Manual
Page 31
... the entire unit or remove, at purchaser's cost, the defective component part(s) identified, pack the component part(s) in advance. Coverage under this warranty. C.O.D. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to the labor for coverage under this product when new and unused from this warranty shall be repaired or...
... the entire unit or remove, at purchaser's cost, the defective component part(s) identified, pack the component part(s) in advance. Coverage under this warranty. C.O.D. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to the labor for coverage under this product when new and unused from this warranty shall be repaired or...