Owners Manual
Page 1
For more information regarding Invacare products, parts, and services, please visit www.invacare.com User: BEFORE using this product, read this product. User Manual Perfecto2™Series Oxygen Concentrators without SensO2®, Oxygen Concentrators with SensO2 HomeFill®Compatible Perfecto2 with SensO2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this manual and save for future reference.
For more information regarding Invacare products, parts, and services, please visit www.invacare.com User: BEFORE using this product, read this product. User Manual Perfecto2™Series Oxygen Concentrators without SensO2®, Oxygen Concentrators with SensO2 HomeFill®Compatible Perfecto2 with SensO2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this manual and save for future reference.
Owners Manual
Page 7
...parts. To Reduce the Risk of Burns, Electrocution, Fire or Injury to ensure the safe installation, assembly and operation of the Perfecto2 concentrator these instructions MUST be followed. ƽ WARNING This section contains important information for the safe operation and use any type of friction.... Unplug IMMEDIATELY. If the concentrator has a damaged cord or plug, if it is required by Invacare. DO NOT use of any spark near medical oxygen equipment. DO NOT disassemble. Avoid creation of ...
...parts. To Reduce the Risk of Burns, Electrocution, Fire or Injury to ensure the safe installation, assembly and operation of the Perfecto2 concentrator these instructions MUST be followed. ƽ WARNING This section contains important information for the safe operation and use any type of friction.... Unplug IMMEDIATELY. If the concentrator has a damaged cord or plug, if it is required by Invacare. DO NOT use of any spark near medical oxygen equipment. DO NOT disassemble. Avoid creation of ...
Owners Manual
Page 8
... are unable to hear or see alarms or communicate discomfort. NEVER drop or insert any object or liquid into any opening. Invacare recommends that each concentrator be on the power cord. Make sure the Perfecto2 is used with this product and not exceed 50 ft in use. Perfecto2...™Series 8 Part No 1143482 SECTION 1-GENERAL GUIDELINES Operating Information For optimum performance, Invacare recommends that Crush‐Proof oxygen tubing be...
... are unable to hear or see alarms or communicate discomfort. NEVER drop or insert any object or liquid into any opening. Invacare recommends that each concentrator be on the power cord. Make sure the Perfecto2 is used with this product and not exceed 50 ft in use. Perfecto2...™Series 8 Part No 1143482 SECTION 1-GENERAL GUIDELINES Operating Information For optimum performance, Invacare recommends that Crush‐Proof oxygen tubing be...
Owners Manual
Page 10
...697;s manual, part number 1100873, for filling oxygen cylinders with the concentrator should be used only for connection and operating instructions. The outlet fitting does not affect concentrator performance. Refer to be inserted into the outlet Power Cord *Outlet Fitting... fitting. For more information about the HomeFill, contact your Invacare dealer. IRCPF16 (Perfecto2 5 Only) HomeFill home oxygen ...
...697;s manual, part number 1100873, for filling oxygen cylinders with the concentrator should be used only for connection and operating instructions. The outlet fitting does not affect concentrator performance. Refer to be inserted into the outlet Power Cord *Outlet Fitting... fitting. For more information about the HomeFill, contact your Invacare dealer. IRCPF16 (Perfecto2 5 Only) HomeFill home oxygen ...
Owners Manual
Page 11
.... Carefully remove all the components from the carton. 3. FIGURE 3.1 Unpacking NOTE: Unless the oxygen concentrator is required. SECTION 3-HANDLING SECTION 3-HANDLING The concentrator should ALWAYS be kept in a dry area. 2. Inspect all loose packing from the carton. DO...Remove all components. Storage 1. Store the repackaged oxygen concentrator in the upright position to be reshipped by common carrier, additional cartons are available from Invacare. Inspection Inspect/examine exterior of repackaged concentrator. If damage is to be used immediately, retain...
.... Carefully remove all the components from the carton. 3. FIGURE 3.1 Unpacking NOTE: Unless the oxygen concentrator is required. SECTION 3-HANDLING SECTION 3-HANDLING The concentrator should ALWAYS be kept in a dry area. 2. Inspect all loose packing from the carton. DO...Remove all components. Storage 1. Store the repackaged oxygen concentrator in the upright position to be reshipped by common carrier, additional cartons are available from Invacare. Inspection Inspect/examine exterior of repackaged concentrator. If damage is to be used immediately, retain...
Owners Manual
Page 12
...: Low Flow Alarm: Average Power Consumption: All Perfecto2 Models while filling a HomeFill System: Up to 6,000 ft (1828 meters) above sea level without degradation of concentration levels. Perfecto2 and Perfecto2W Models: 280 W (Operating @ 3 L/min) Pressure Relief Mechanism Operational at 0.5 to 5 L/min 5 psi ± 0.5 psi (34.5 kPa ±....6% for Perfecto2W Models All Perfecto2 Models: Up to 8,000 ft (2438 meters) above sea level without degredation of the Invacare Pediatric Flowmeter IRCPF16 Accessory. 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected).
...: Low Flow Alarm: Average Power Consumption: All Perfecto2 Models while filling a HomeFill System: Up to 6,000 ft (1828 meters) above sea level without degradation of concentration levels. Perfecto2 and Perfecto2W Models: 280 W (Operating @ 3 L/min) Pressure Relief Mechanism Operational at 0.5 to 5 L/min 5 psi ± 0.5 psi (34.5 kPa ±....6% for Perfecto2W Models All Perfecto2 Models: Up to 8,000 ft (2438 meters) above sea level without degredation of the Invacare Pediatric Flowmeter IRCPF16 Accessory. 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected).
Owners Manual
Page 14
Perfecto2™Series 14 Part No 1143482 SECTION 4-TYPICAL PRODUCT PARAMETERS Placement: Tubing: Relative Humidity: Time of Operation: Recommended Storage and Shipping Temperature: No closer than three inches from any wall, furniture, draperies, or similar surfaces. 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO NOT pinch). 20 to 60% Up to 24 hours per day. -20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity. *NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes).
Perfecto2™Series 14 Part No 1143482 SECTION 4-TYPICAL PRODUCT PARAMETERS Placement: Tubing: Relative Humidity: Time of Operation: Recommended Storage and Shipping Temperature: No closer than three inches from any wall, furniture, draperies, or similar surfaces. 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO NOT pinch). 20 to 60% Up to 24 hours per day. -20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity. *NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes).
Owners Manual
Page 15
... from room to you through a nasal cannula. You may be contacted with any questions or problems regarding your concentrator. It provides high concentration of oxygen delivery systems. Your provider will show you how to avoid smoke, airborne pollutants and/or fumes. ... from lint, hair and the like . SECTION 5-OPERATING INSTRUCTIONS SECTION 5-OPERATING INSTRUCTIONS Introduction Your oxygen concentrator is an electronically operated device that oxygen concentrators are therapeutically equivalent to Typical Product Parameters on a soft surface, such as you use your oxygen...
... from room to you through a nasal cannula. You may be contacted with any questions or problems regarding your concentrator. It provides high concentration of oxygen delivery systems. Your provider will show you how to avoid smoke, airborne pollutants and/or fumes. ... from lint, hair and the like . SECTION 5-OPERATING INSTRUCTIONS SECTION 5-OPERATING INSTRUCTIONS Introduction Your oxygen concentrator is an electronically operated device that oxygen concentrators are therapeutically equivalent to Typical Product Parameters on a soft surface, such as you use your oxygen...
Owners Manual
Page 17
...through the cannula. Attach the cannula/patient supply tubing to the humidifier bottle by turning the wing nut on the concentrator. SECTION 5-OPERATING INSTRUCTIONS Humidifier Bottle Adapter Insert Flat Head Screwdriver Here Cabinet Filter Filter Access Door FIGURE 5.2 Humidifer ...Bottle Adapter 5. Place the humidifier bottle/adapter assembly in FIGURE 5.3. 6. See Detail "A" in the humidifier compartment on the concentrator. See Detail "B" in FIGURE 5.3. 8. Part No 1143482 17 Perfecto2™Series Attach the humidifier bottle adapter to the humidifier bottle ...
...through the cannula. Attach the cannula/patient supply tubing to the humidifier bottle by turning the wing nut on the concentrator. SECTION 5-OPERATING INSTRUCTIONS Humidifier Bottle Adapter Insert Flat Head Screwdriver Here Cabinet Filter Filter Access Door FIGURE 5.2 Humidifer ...Bottle Adapter 5. Place the humidifier bottle/adapter assembly in FIGURE 5.3. 6. See Detail "A" in the humidifier compartment on the concentrator. See Detail "B" in FIGURE 5.3. 8. Part No 1143482 17 Perfecto2™Series Attach the humidifier bottle adapter to the humidifier bottle ...
Owners Manual
Page 20
If purity falls below factory preset standards, indicator lights on the control panel will illuminate. Units with SensO2 The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. Units with SensO2 Perfecto2™Series 20 Part No 1143482 OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators - SECTION 5-OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators -
If purity falls below factory preset standards, indicator lights on the control panel will illuminate. Units with SensO2 The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. Units with SensO2 Perfecto2™Series 20 Part No 1143482 OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators - SECTION 5-OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators -
Owners Manual
Page 21
... procedure, refer to 85% (±2%) SYSTEM FAILURE O2 Below 73% (±3%) GREEN Indicator Light YELLOW Indicator light A. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min.) while waiting for the O2 purity to a back-up supply of...
... procedure, refer to 85% (±2%) SYSTEM FAILURE O2 Below 73% (±3%) GREEN Indicator Light YELLOW Indicator light A. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min.) while waiting for the O2 purity to a back-up supply of...
Owners Manual
Page 22
O2 purity will build to a back-up time (approximately 30 min). Switch to a maximum during the initial start warm‐up supply of the Concentrator NOTE: Concentrator may be used during this period. Perfecto2™Series 22 Part No 1143482 SECTION 5-OPERATING INSTRUCTIONS Initial Startup of oxygen immediately. LABEL SYMBOL I/O STATUS SYSTEM OKAY SYSTEM FAILURE INDICATOR LIGHTS (LED) GREEN Indicator Light RED Indicator light Continuous Audible Alarm Sieve-GARD™Compressor Shutdown Call a qualified technician.
O2 purity will build to a back-up time (approximately 30 min). Switch to a maximum during the initial start warm‐up supply of the Concentrator NOTE: Concentrator may be used during this period. Perfecto2™Series 22 Part No 1143482 SECTION 5-OPERATING INSTRUCTIONS Initial Startup of oxygen immediately. LABEL SYMBOL I/O STATUS SYSTEM OKAY SYSTEM FAILURE INDICATOR LIGHTS (LED) GREEN Indicator Light RED Indicator light Continuous Audible Alarm Sieve-GARD™Compressor Shutdown Call a qualified technician.
Owners Manual
Page 23
... guidelines. See recommended intervals in each section and preventive maintenance record. NOTE: There is one cabinet filter located on the Invacare concentrator. NOTE: Environmental conditions that may need to be performed more frequent cleaning of the cabinet. 1. Part No 1143482 23 ...a week depending on page 25. Cleaning the Cabinet Filter CAUTION DO NOT operate the concentrator without the filter installed. SECTION 6-MAINTENANCE SECTION 6-MAINTENANCE ƽ WARNING The Invacare concentrators are not limited to: high dust, air pollutants, etc. Power should perform preventive ...
... guidelines. See recommended intervals in each section and preventive maintenance record. NOTE: There is one cabinet filter located on the Invacare concentrator. NOTE: Environmental conditions that may need to be performed more frequent cleaning of the cabinet. 1. Part No 1143482 23 ...a week depending on page 25. Cleaning the Cabinet Filter CAUTION DO NOT operate the concentrator without the filter installed. SECTION 6-MAINTENANCE SECTION 6-MAINTENANCE ƽ WARNING The Invacare concentrators are not limited to: high dust, air pollutants, etc. Power should perform preventive ...
Owners Manual
Page 25
... Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter...
... Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter...
Owners Manual
Page 26
...Alarm: 1. Inspect for 60 seconds and then turn power Off for kinks or blockages. Unit overheating due to blocked air intake. Move oxygen concentrator at outlet. 3. Insert plug into outlet. 2. If problem recurs, use extension cords. No power at least three inches away from walls... Failure: 1. SOLUTION: 1. If prescribed flowrate is below 0.5 L/min, use the Pediatric Flowmeter IRCPF16 (Only with Perfecto2 5). Alarm: Continuous Concentrator not operating, Power Switch On. SECTION 7-TROUBLESHOOTING GUIDE SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: Alarm: Short beeps, long pause...
...Alarm: 1. Inspect for 60 seconds and then turn power Off for kinks or blockages. Unit overheating due to blocked air intake. Move oxygen concentrator at outlet. 3. Insert plug into outlet. 2. If problem recurs, use extension cords. No power at least three inches away from walls... Failure: 1. SOLUTION: 1. If prescribed flowrate is below 0.5 L/min, use the Pediatric Flowmeter IRCPF16 (Only with Perfecto2 5). Alarm: Continuous Concentrator not operating, Power Switch On. SECTION 7-TROUBLESHOOTING GUIDE SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: Alarm: Short beeps, long pause...
Owners Manual
Page 27
PROBABLE CAUSE: 1. Clean or Replace filters. 2. Move concentrator at 0.5 L/min or less.* 4. Move to blocked air intake. 5. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set to IRC5PO2. GREEN light with 1. Call Service Provider. Low oxygen purity.* 2. Internal repairs required. Unit overheating due to another electrical outlet or circuit. 6. Call Service Provider. SOLUTION...
PROBABLE CAUSE: 1. Clean or Replace filters. 2. Move concentrator at 0.5 L/min or less.* 4. Move to blocked air intake. 5. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set to IRC5PO2. GREEN light with 1. Call Service Provider. Low oxygen purity.* 2. Internal repairs required. Unit overheating due to another electrical outlet or circuit. 6. Call Service Provider. SOLUTION...
Owners Manual
Page 31
...NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to be the responsibility of the purchaser to return the entire unit or remove, at ... at the toll free number on the back cover. If within thirty days of title to any such product shall be proven to Invacare Corporation's satisfaction to such repair or replacement. This warranty only applies to our factory without prior consent. shipments will issue a serialized...
...NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to be the responsibility of the purchaser to return the entire unit or remove, at ... at the toll free number on the back cover. If within thirty days of title to any such product shall be proven to Invacare Corporation's satisfaction to such repair or replacement. This warranty only applies to our factory without prior consent. shipments will issue a serialized...