FE 63-66-68 Users Manual
Page 4
63/66/68 Infrared Thermometers Using a Contact Temperature Probe (PRB) (66/68 18 HOLD 19 Storing Data (66/68 19 Recalling Data (66/68 19 LOG Clear Function 19 Maintenance 20 Changing the Battery 20 Cleaning the Lens 20 Cleaning the Housing 20 Troubleshooting 21 Accessories 21 CE Certification 22 Specifications 22 ii
63/66/68 Infrared Thermometers Using a Contact Temperature Probe (PRB) (66/68 18 HOLD 19 Storing Data (66/68 19 Recalling Data (66/68 19 LOG Clear Function 19 Maintenance 20 Changing the Battery 20 Cleaning the Lens 20 Cleaning the Housing 20 Troubleshooting 21 Accessories 21 CE Certification 22 Specifications 22 ii
FE 63-66-68 Users Manual
Page 25
...battery 2. Ambient temperature above 40 °C (104 °F) Possible damage by EMF Select target within specifications Replace battery Check and/or replace battery 1. Fluke PN 2148313 • Nylon Soft Carrying Case - Use in area with lower ambient temperature Contact your ...distributor Optional accessories for the thermometers are: • Contact probe (RTD) (66/68) - 63/66/68 Troubleshooting Troubleshooting Symptom Problem ...
...battery 2. Ambient temperature above 40 °C (104 °F) Possible damage by EMF Select target within specifications Replace battery Check and/or replace battery 1. Fluke PN 2148313 • Nylon Soft Carrying Case - Use in area with lower ambient temperature Contact your ...distributor Optional accessories for the thermometers are: • Contact probe (RTD) (66/68) - 63/66/68 Troubleshooting Troubleshooting Symptom Problem ...
FE 63-66-68 Users Manual
Page 26
Specifications Temperature Range Assumes ambient operating temperature of 23 °C (73 °F) to 25 °C (77 °F) 63: -32 °C to 535 °C (-25 °F to 999 °F) 66: -32 °C to 600 °C (-25 °F to 1100 °F) 68: -32 °C to 760 °C (-25 °F to ...1400 °F) Accuracy Target Temperature: Above 510 °C (63) ±1.5 % of reading Above 510 °C (66/68) ±1 % of reading or ±1 °C (±2 °F), whichever is greater 23 °C to 510 °C -18 °C to 23 °C -26 ...
Specifications Temperature Range Assumes ambient operating temperature of 23 °C (73 °F) to 25 °C (77 °F) 63: -32 °C to 535 °C (-25 °F to 999 °F) 66: -32 °C to 600 °C (-25 °F to 1100 °F) 68: -32 °C to 760 °C (-25 °F to ...1400 °F) Accuracy Target Temperature: Above 510 °C (63) ±1.5 % of reading Above 510 °C (66/68) ±1 % of reading or ±1 °C (±2 °F), whichever is greater 23 °C to 510 °C -18 °C to 23 °C -26 ...
FE 63-66-68 Users Manual
Page 27
63/66/68 Specifications Distance to Spot Ratio 63: 12:1 66: 30:1 68: 50:1 Emissivity (Adjustment for surface measured) 63: Pre-set at 0.95 66/68: Digitally adjustable from 0.10 to 1.0 Response Time 500 mSec Repeatability ±0.5 % of reading or ±1 °C (±2 °F), whichever is greater Spectral Response 8 µm ... in x 2.17 in) Battery Type 9 V Alkaline or NiCd Battery Life 63: 10 hours with laser and backlight on 40 hours with laser and backlight off 66/68: 20 hours with laser and backlight on 40 hours with laser and backlight off 23
63/66/68 Specifications Distance to Spot Ratio 63: 12:1 66: 30:1 68: 50:1 Emissivity (Adjustment for surface measured) 63: Pre-set at 0.95 66/68: Digitally adjustable from 0.10 to 1.0 Response Time 500 mSec Repeatability ±0.5 % of reading or ±1 °C (±2 °F), whichever is greater Spectral Response 8 µm ... in x 2.17 in) Battery Type 9 V Alkaline or NiCd Battery Life 63: 10 hours with laser and backlight on 40 hours with laser and backlight off 66/68: 20 hours with laser and backlight on 40 hours with laser and backlight off 23
Fluke 66 Infrared Thermometer Datasheet
Page 2
Specifications Temperature range Accuracy (assumes ambient operating temperature of reading) 8 to 14 µm Digitally adjustable emissivity (from 0.1 to 1.0 by 0.01) 0 to 50 °C (32 to 120 ... display Data logging Probe jack Display hold (7 seconds) Hi/Low alarm LCD backlit Temperature display Display resolution Hard carrying case Locking trigger Tripod mounting Warranty Fluke 66 Fluke 68 -32 to 600 °C (-25 to 1100 °F) -32 to 760 °C (-25 to 1400 °F) -32 to -26 °C (-25 to -15 °...
Specifications Temperature range Accuracy (assumes ambient operating temperature of reading) 8 to 14 µm Digitally adjustable emissivity (from 0.1 to 1.0 by 0.01) 0 to 50 °C (32 to 120 ... display Data logging Probe jack Display hold (7 seconds) Hi/Low alarm LCD backlit Temperature display Display resolution Hard carrying case Locking trigger Tripod mounting Warranty Fluke 66 Fluke 68 -32 to 600 °C (-25 to 1100 °F) -32 to 760 °C (-25 to 1400 °F) -32 to -26 °C (-25 to -15 °...