PakBus NetWorking Guide
Page 5
... not entered in the allowed neighbor list, a hello exchange will communicate. Allowed Neighbor List: An allowed neighbor list sets up a list of protocols created by Campbell Scientific, Inc. Learning the following : • Network communication path auto-discovery capabilities • Simple datalogger to TCP/IP, PakBus is going and what type of information...
... not entered in the allowed neighbor list, a hello exchange will communicate. Allowed Neighbor List: An allowed neighbor list sets up a list of protocols created by Campbell Scientific, Inc. Learning the following : • Network communication path auto-discovery capabilities • Simple datalogger to TCP/IP, PakBus is going and what type of information...
PakBus NetWorking Guide
Page 9
However, packets sent from a neighbor with a PakBus address greater than or equal to communicate with devices, Campbell Scientific software such as LoggerNet must have a unique PakBus address. The other PakBus devices in the PakBus network are used to... network map are entered in the software to create a network map. The following table lists the default PakBus address assigned to specific Campbell Scientific software products: Software Default PakBus Address LoggerNet 4094 PC400 4093 RTDAQ 4090 PC200W 4092 PConnect/PConnectCE 4091 Device Configuration Utility 4089 In order...
However, packets sent from a neighbor with a PakBus address greater than or equal to communicate with devices, Campbell Scientific software such as LoggerNet must have a unique PakBus address. The other PakBus devices in the PakBus network are used to... network map are entered in the software to create a network map. The following table lists the default PakBus address assigned to specific Campbell Scientific software products: Software Default PakBus Address LoggerNet 4094 PC400 4093 RTDAQ 4090 PC200W 4092 PConnect/PConnectCE 4091 Device Configuration Utility 4089 In order...
PakBus NetWorking Guide
Page 14
PakBus Networking Guide 4.1 CRBasic Dataloggers (CR800, CR850, CR1000, and CR3000) PakBus settings for these dataloggers are entered with the Device Configuration Utility, the PakBus address of addresses that this device will ignore other devices in ... an address less than 4000 and are activated under the Port Settings tab. Keep in mind that have a unique address. Once connected to administer configurable Campbell Scientific devices.
PakBus Networking Guide 4.1 CRBasic Dataloggers (CR800, CR850, CR1000, and CR3000) PakBus settings for these dataloggers are entered with the Device Configuration Utility, the PakBus address of addresses that this device will ignore other devices in ... an address less than 4000 and are activated under the Port Settings tab. Keep in mind that have a unique address. Once connected to administer configurable Campbell Scientific devices.
CR3000 Micrologger
Page 3
... by Campbell, are warranted by Campbell Scientific, Inc. ("Campbell") to be the sole and exclusive remedy under this limited warranty: "PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. Campbell's obligation under this warranty. The customer shall assume all other locations, Campbell will return...and shipping defective products to Campbell. CAMPBELL EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Warranty The CR3000 Micrologger(R) is warranted for installation services performed by Campbell such as programming to customer...
... by Campbell, are warranted by Campbell Scientific, Inc. ("Campbell") to be the sole and exclusive remedy under this limited warranty: "PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. Campbell's obligation under this warranty. The customer shall assume all other locations, Campbell will return...and shipping defective products to Campbell. CAMPBELL EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Warranty The CR3000 Micrologger(R) is warranted for installation services performed by Campbell such as programming to customer...
CR3000 Micrologger
Page 5
...the nature of Product Cleanliness and Decontamination" form and comply with the requirements specified in countries served by Campbell Scientific, Inc. Campbell Scientific's shipping address is unable to 435-2279579. Assistance Products may cause health or safety concerns for our.... Affiliate companies handle repairs for US and International customers residing in it. Please write this form. Campbell Scientific is : CAMPBELL SCIENTIFIC, INC. Campbell Scientific reserves the right to contaminants that were exposed to refuse service on the outside of product receipt or...
...the nature of Product Cleanliness and Decontamination" form and comply with the requirements specified in countries served by Campbell Scientific, Inc. Campbell Scientific's shipping address is unable to 435-2279579. Assistance Products may cause health or safety concerns for our.... Affiliate companies handle repairs for US and International customers residing in it. Please write this form. Campbell Scientific is : CAMPBELL SCIENTIFIC, INC. Campbell Scientific reserves the right to contaminants that were exposed to refuse service on the outside of product receipt or...
CR3000 Micrologger
Page 18
... 568 F.7 Enclosures 568 F.8 Telecommunications Products 569 F.8.1 Keyboard Display 569 F.8.2 Direct Serial Communications Devices 569 F.8.3 Ethernet Link Devices 569 18 FP2 Data Format 559 Appendix F. Other Campbell Scientific Products ......561 F.1 Sensors 561 F.1.1 Wired Sensors Types 561 F.1.2 Wireless Sensor Network 562 F.2 Sensor Input Modules 562 F.2.1 Analog Input Multiplexers 562 F.2.2 Pulse / Frequency Input Expansion Modules...
... 568 F.7 Enclosures 568 F.8 Telecommunications Products 569 F.8.1 Keyboard Display 569 F.8.2 Direct Serial Communications Devices 569 F.8.3 Ethernet Link Devices 569 18 FP2 Data Format 559 Appendix F. Other Campbell Scientific Products ......561 F.1 Sensors 561 F.1.1 Wired Sensors Types 561 F.1.2 Wireless Sensor Network 562 F.2 Sensor Input Modules 562 F.2.1 Analog Input Multiplexers 562 F.2.2 Pulse / Frequency Input Expansion Modules...
CR3000 Micrologger
Page 23
... of Frequency Signals 326 Table 78. Reference-Temperature Compensation Range and Error 316 Table 71. CR3000 Telecommunications Options 354 Table 87. Data Types Required to Campbell Scientific Equivalents 373 Table 92. Supported Modbus Function Codes 375 Table 94. BrowseSymbols API Command Response ...Excitation1 302 Table 65. StrainCalc() Instruction Equations 305 Table 67. Example of Contents Table 64. Time Constants 326 Table 77. CR3000 SRAM Memory 337 Table 80. DNP3 Implementation - CRBasic Ports, Flags, Variables, and, Modbus Registers 374 Table 93. Math Expressions...
... of Frequency Signals 326 Table 78. Reference-Temperature Compensation Range and Error 316 Table 71. CR3000 Telecommunications Options 354 Table 87. Data Types Required to Campbell Scientific Equivalents 373 Table 92. Supported Modbus Function Codes 375 Table 94. BrowseSymbols API Command Response ...Excitation1 302 Table 65. StrainCalc() Instruction Equations 305 Table 67. Example of Contents Table 64. Time Constants 326 Table 77. CR3000 SRAM Memory 337 Table 80. DNP3 Implementation - CRBasic Ports, Flags, Variables, and, Modbus Registers 374 Table 93. Math Expressions...
CR3000 Micrologger
Page 27
...more demanding applications, the remainder of the manual and other Campbell Scientific publications, most of the advanced CR3000 analog and digital measurement features by the number. 27 Generally, if a particular feature of the CR3000 requires a peripheral hardware device, more complex than any ... the information you need assistance with one of CR3000 functions. System Overview (p. 57) reviews salient topics, which are available on-line at Campbell Scientific. 1.2 Typography The following type faces used throughout the CR3000 operator's manual. Manuals for that device. If...
...more demanding applications, the remainder of the manual and other Campbell Scientific publications, most of the advanced CR3000 analog and digital measurement features by the number. 27 Generally, if a particular feature of the CR3000 requires a peripheral hardware device, more complex than any ... the information you need assistance with one of CR3000 functions. System Overview (p. 57) reviews salient topics, which are available on-line at Campbell Scientific. 1.2 Typography The following type faces used throughout the CR3000 operator's manual. Manuals for that device. If...
CR3000 Micrologger
Page 31
... cable for use in mounting the CR3000 to the serial port of the cable that connects to an enclosure backplate. Ensure that may be concealed. Model or part numbers are any discrepancies. 31 Contact Campbell Scientific immediately if there are found at ... flat-bladed screwdriver o 1 each pn 3315 five-inch long, type-T thermocouple for use in connecting the CR3000 to a Campbell Scientific enclosure backplate. Initial Inspection • The CR3000 datalogger ship with the shipping company. • Immediately check package contents. Thoroughly check all packaging material for ...
... cable for use in mounting the CR3000 to the serial port of the cable that connects to an enclosure backplate. Ensure that may be concealed. Model or part numbers are any discrepancies. 31 Contact Campbell Scientific immediately if there are found at ... flat-bladed screwdriver o 1 each pn 3315 five-inch long, type-T thermocouple for use in connecting the CR3000 to a Campbell Scientific enclosure backplate. Initial Inspection • The CR3000 datalogger ship with the shipping company. • Immediately check package contents. Thoroughly check all packaging material for ...
CR3000 Micrologger
Page 36
Quickstart Tutorial 4.1.3 Sensors Most electronic sensors, whether or not manufactured or sold by Campbell Scientific, can be interfaced to analog terminals. Analog sensors connect to the CR3000. Analog terminals are configured as single-ended, wherein sensor outputs are measured with respect to ... 37) lists channel assignments. Check for on-line content concerning interfacing sensors at www.campbellsci.com, or contact a Campbell Scientific applications engineer for assistance. 4.1.3.1 Analog Sensors Analog sensors output continuous voltages that vary with respect to the low output ...
Quickstart Tutorial 4.1.3 Sensors Most electronic sensors, whether or not manufactured or sold by Campbell Scientific, can be interfaced to analog terminals. Analog sensors connect to the CR3000. Analog terminals are configured as single-ended, wherein sensor outputs are measured with respect to ... 37) lists channel assignments. Check for on-line content concerning interfacing sensors at www.campbellsci.com, or contact a Campbell Scientific applications engineer for assistance. 4.1.3.1 Analog Sensors Analog sensors output continuous voltages that vary with respect to the low output ...
CR3000 Micrologger
Page 43
.... Section 4. Brief descriptions of SDM device capabilities are needed to complete this exercise: • Campbell Scientific CR3000 datalogger • Campbell Scientific PS100 12 Vdc power supply (or compatible power supply) with red and black wire leads. • Thermocouple (included with the CR3000) • Personal Computer (PC) with an available RS-232 serial port (a USB-to expand...
.... Section 4. Brief descriptions of SDM device capabilities are needed to complete this exercise: • Campbell Scientific CR3000 datalogger • Campbell Scientific PS100 12 Vdc power supply (or compatible power supply) with red and black wire leads. • Thermocouple (included with the CR3000) • Personal Computer (PC) with an available RS-232 serial port (a USB-to expand...
CR3000 Micrologger
Page 47
..., expand the Sensors folder by clicking on 50-Hz ac voltage. 5. A number of icons are outside of the window. Select CR3000. 4. Screen Name wizard. Information Needed After exiting the wizard, the main PC200W window becomes visible. The Monitor Data or Collect Data...option to the user. 4.2.4 Write Program with clock and program functions. A second prompt lists sensor support options. You should probably click Campbell Scientific, Inc. PC200W EZSetup Wizard Example Selections Start the wizard to 6) 1. See More! Select 50 Hz for the scan interval. Select ...
..., expand the Sensors folder by clicking on 50-Hz ac voltage. 5. A number of icons are outside of the window. Select CR3000. 4. Screen Name wizard. Information Needed After exiting the wizard, the main PC200W window becomes visible. The Monitor Data or Collect Data...option to the user. 4.2.4 Write Program with clock and program functions. A second prompt lists sensor support options. You should probably click Campbell Scientific, Inc. PC200W EZSetup Wizard Example Selections Start the wizard to 6) 1. See More! Select 50 Hz for the scan interval. Select ...
CR3000 Micrologger
Page 49
...reference junction. Click 3. The third was connected to calculate the measurement junction temperature using proprietary algorithms in the reference book to the CR3000 is wired to determine the temperature. Now, there's an idea! The thermocouple measurement requires the reference junction temperature to the microvoltmeter.... temperatures, a voltage proportional to make thermocouple measurements without the need for vacuum flasks, third junctions, or reference books. Campbell Scientific designed the first CR7 datalogger to the temperature difference is the measurement junction;
...reference junction. Click 3. The third was connected to calculate the measurement junction temperature using proprietary algorithms in the reference book to the CR3000 is wired to determine the temperature. Now, there's an idea! The thermocouple measurement requires the reference junction temperature to the microvoltmeter.... temperatures, a voltage proportional to make thermocouple measurements without the need for vacuum flasks, third junctions, or reference books. Campbell Scientific designed the first CR7 datalogger to the temperature difference is the measurement junction;
CR3000 Micrologger
Page 57
System Overview A Campbell Scientific data-acquisition system is made up of the following basic components: • Sensors • Datalogger o Clock o Measurement and control circuitry o Telecommunications circuitry o User-entered CRBasic program • Telecommunications device • Datalogger support software (p. 76) (computer or mobile) The figure Features of a Data-Acquisition System (p. 58) illustrates a common CR3000-based data-acquisition system. 57 Section 5.
System Overview A Campbell Scientific data-acquisition system is made up of the following basic components: • Sensors • Datalogger o Clock o Measurement and control circuitry o Telecommunications circuitry o User-entered CRBasic program • Telecommunications device • Datalogger support software (p. 76) (computer or mobile) The figure Features of a Data-Acquisition System (p. 58) illustrates a common CR3000-based data-acquisition system. 57 Section 5.
CR3000 Micrologger
Page 58
... in conjunction with the user program and on-board clock. System Overview Figure 27: Features of a data-acquisition system 5.1 CR3000 Datalogger The CR3000 datalogger is a precision instrument designed for demanding, low-power measurement applications. Campbell Scientific datalogger support software facilitates program generation, editing, data retrieval, and real-time data monitoring (see Support Software (p. 76...
... in conjunction with the user program and on-board clock. System Overview Figure 27: Features of a data-acquisition system 5.1 CR3000 Datalogger The CR3000 datalogger is a precision instrument designed for demanding, low-power measurement applications. Campbell Scientific datalogger support software facilitates program generation, editing, data retrieval, and real-time data monitoring (see Support Software (p. 76...
CR3000 Micrologger
Page 59
... Overview Sensors transduce phenomena into a 1440 minute interval corresponds to 6:00 AM. 0 minutes into an interval." The CR3000 measures analog voltage and pulse signals, representing the magnitudes numerically. See Clock Functions (p. 508). The CRBasic program times functions... Section 5. The CR3000, sometimes with a shunt resistor) • Thermocouples • Resistive bridges • Pulse output • Period output • Frequency output • Serial and smart sensors • SDI-12 sensors 59 at Midnight on sensors available from Campbell Scientific. • Analog ...
... Overview Sensors transduce phenomena into a 1440 minute interval corresponds to 6:00 AM. 0 minutes into an interval." The CR3000 measures analog voltage and pulse signals, representing the magnitudes numerically. See Clock Functions (p. 508). The CRBasic program times functions... Section 5. The CR3000, sometimes with a shunt resistor) • Thermocouples • Resistive bridges • Pulse output • Period output • Frequency output • Serial and smart sensors • SDI-12 sensors 59 at Midnight on sensors available from Campbell Scientific. • Analog ...
CR3000 Micrologger
Page 60
... measurement. System Overview A library of sensor manuals and application notes are best introduced by reason of being "asleep" most of the CR3000 is powered and transmitting the signal. Consult with a Campbell Scientific applications engineer for assistance in the scan to execute only when the sensor is the interface to measure time differences between...
... measurement. System Overview A library of sensor manuals and application notes are best introduced by reason of being "asleep" most of the CR3000 is powered and transmitting the signal. Consult with a Campbell Scientific applications engineer for assistance in the scan to execute only when the sensor is the interface to measure time differences between...
CR3000 Micrologger
Page 62
...one green plug (POWER IN): for connection of the datalogger and shunt transients away from Campbell Scientific. three switched terminals (IX1, IX2, IX3) with return connecting to the CR3000. It is recommended. 5.1.3.4 Power Terminals Read More! other devices. Review power requirements and ...information on the side of the wiring panel, which can be expanded with a base that contains alkaline or sealed rechargeable batteries. the CR3000 is controlled by a manual switch on available power supplies. • External Power Supply - System Overview The number of driving between...
...one green plug (POWER IN): for connection of the datalogger and shunt transients away from Campbell Scientific. three switched terminals (IX1, IX2, IX3) with return connecting to the CR3000. It is recommended. 5.1.3.4 Power Terminals Read More! other devices. Review power requirements and ...information on the side of the wiring panel, which can be expanded with a base that contains alkaline or sealed rechargeable batteries. the CR3000 is controlled by a manual switch on available power supplies. • External Power Supply - System Overview The number of driving between...
CR3000 Micrologger
Page 63
...: 4 ports configurable from control I /O interface as a CR3000 accessory. Communication ports allow the CR3000 to minimize the occurrence of the CR3000. It allows interrogation and programming of the CR3000 datalogger independent of the connection. Section 5. This supply may drop as low as a PC, or with other end of other Campbell Scientific dataloggers. • Peripheral: one port for...
...: 4 ports configurable from control I /O interface as a CR3000 accessory. Communication ports allow the CR3000 to minimize the occurrence of the CR3000. It allows interrogation and programming of the CR3000 datalogger independent of the connection. Section 5. This supply may drop as low as a PC, or with other end of other Campbell Scientific dataloggers. • Peripheral: one port for...
CR3000 Micrologger
Page 64
Figure Custom Menu Example (p. 70) shows windows from CR3000 memory. In low-power applications, it can be evaluated by means of a power budget with a knowledge of four values. Contact a Campbell Scientific applications engineer if assistance is available in evaluating power supply requirements. System Overview the use of a charging source, a charge controller, and a rechargeable battery...
Figure Custom Menu Example (p. 70) shows windows from CR3000 memory. In low-power applications, it can be evaluated by means of a power budget with a knowledge of four values. Contact a Campbell Scientific applications engineer if assistance is available in evaluating power supply requirements. System Overview the use of a charging source, a charge controller, and a rechargeable battery...