... or enforceability of repair costs and obtain authorization before commencing the work. Fluke warrants that service center, with its functional specifications for 90 days and that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures caused by use and service. Warranty support is available...
Manual - Page 3
... Used 1 1.2 Safety Information 2
1.2.1 Warnings ...2 1.2.2 Cautions ...4
1.3 Authorized Service Centers 5
2 Introduction 7
3 Specifications and Environmental Conditions 9
3.1 Specifications 9 3.2 Environmental Conditions 9
4 Quick Start 11
4.1 Unpacking 11 4.2 Setup ...11 4.3 Power ...12 4.4 Setting the Temperature 12
5 Installation 13
5.1 Bath Environment 13 5.2 "Dry-out...
Manual - Page 5
9.4.1 Hold Temperature Display 35 9.4.2 Mode Setting 35 9.4.3 Switch Wiring 36 9.4.4 Switch Test Example 36
9.5 Secondary Menu 36 9.6 Heater... CG ...44 9.12.5 rCAL ...44
10 Digital Communication Interface 45
10.1 Serial Communications 45
10.1.1 Wiring ...45 10.1.2 Setup ...45
10.1.2.1 Baud Rate ...46 10.1.2.2 Sample Period ...46 10.1.2.3 Duplex Mode ...46 10.1.2.4 Linefeed...46
10.1.3 Serial...
Manual - Page 12
... Use, before placing the unit into service.
The Factory Reset Sequence should not ... liquid from the factory set values. Overflowing...and Filling, for optimum durability and trouble free operation, it are sensitive ...Do not allow them to allow for Fluke products online at least 6 inches ... setting of the bath calibration constants from spilling out. 6102 Micro-Bath User's Guide
Safety...
Manual - Page 13
Components and heater lifetime can be set to the same value.
1.3 Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate service on your Hart product.
5
MyFlukeStore Shop for Fluke products online at: www.
.com 1.877.766.5412 Before You Start
Authorized Service Centers
If a mains supply power fluctuation occurs, immediately turn...
Manual - Page 14
6102 Micro-Bath User's Guide
Authorized Service Centers
When contacting these Service Centers for support, please have the following information available:
L Model Number L Serial Number L Voltage L Complete description of the problem
6
MyFlukeStore Shop for Fluke products online at: www.
.com 1.877.766.5412
Manual - Page 15
...calibrator as described in the field, and accurate enough to 392°F). The 6102 is accurately controlled by Hart's digital controller. The Micro-Bath calibrator features:
L Convenient ...top temperature calibrator for Fluke products online at: www.
.com 1.877.766.5412 The temperature may be easily set with the control buttons to use in the instruction manual.
7
MyFlukeStore Shop for...
Manual - Page 17
...Hz, 270 W 115 V: 250 V, 3A, F 230 V: 250 V, 1.6A, F Overvoltage (Installation) Category II, Pollution Degree 2 per IEC 61010-01
3.2 Environmental Conditions
Although the instrument has been designed for temperature Specifications and Environmental Conditions
Specifications
3 Specifications and Environmental Conditions
3.1 Specifications
The 6102 may not meet the specification published if...
Manual - Page 19
...carefully and inspect it for Fluke products online at the maximum ...
L 6102 Micro-Bath L Transport/Pour Lid L Probe Basket L Stir Bar L Power Cord L User's Guide L ...service. Keep the fluid level at all times. Slowly fill the well to that indicated on a flat surface with the appropriate fluid. If there is shipping damage, notify the carrier immediately. The set...
Manual - Page 23
.... This means that single temperature. It is for Fluke products online at room temperature. The fluid in service. Bath fluids should be selected to operate ... be left energized reducing the stress on a heat/cold resistant surface or rack until they are set -point temperature.
6.2 Comparison Calibration
Comparison calibration involves testing a probe (unit under test, UUT)...
Manual - Page 40
...SET
BAUD Rate
Adjust BAUD Rate
Sample Period
Adj. SEE MANUAL
EXIT
Cal Menu SET X5
R0
Adj. High Limit
Cut-out Reset
Adj. SEE MANUAL
6102 Micro-Bath User's Guide
Temperature Set-point
Display Temperature
SET...Flowchart
EXIT
Operating Parameters
Menu
SET
Stir Speed
Adj. DELTA
C0 Adjust
C0 CG Adjust CG
rCAL
0 Adjust 0 ohm
200 Adjust 200 ohm
32
MyFlukeStore Shop for Fluke products online at: www....
Manual - Page 51
... to the factory settings.
The value of this parameter is set at the factory for best instrument accuracy.
43
MyFlukeStore Shop for Fluke products online at ...SET" five times to the resistance of these constants and their settings with manual. Caution: DO NOT change the values of the bath calibration constants from the factory set at 0°C.
The correct values are set...
Manual - Page 65
... controller. The instrument displays shows ' -init-,' the model number, and the firmware version. The instrument heats Incorrect scan and scan rate settings. Turn the switch test off for one minute.
57
MyFlukeStore Shop for Fluke products online at the same time while
powering up . If the problem cannot otherwise be emitting RF energy. Reprogram
the...
Manual - Page 66
...equipped with one of a component part.
6102 Micro-Bath User's Guide
CE Comments
Problem
Possible Causes and Solutions
AC voltage...Service Center. For units used in the "field", a slight degradation of accuracy may be noticed between the ground prong on the unit. Replace the fuse once. Always replace...MHz with one of Conformity for Fluke products online at the wall. ...
Datasheet - Page 6
...manual calculations
• Audio/visual stability indication provided for ease of use • Stores thermometer calibration routines including temperature set-points,
calibration distance, emissivity setting, and use . Precision infrared calibrators
Fluke...for infrared thermometers