Fluke 115 Digital Multimeter Datasheet
Page 1
... LED backlight to work in a compact package. Fluke 115 Multimeter Actual size R N10140 Technical Data Compact true-rms meter for field service technicians The Fluke 115 is specified for A input Display Bar graph Operating temperature Storage temperature Battery Battery life 600 V 6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2 11 A, 1000 V FAST Fuse (Fluke PN 803293) Digital: 6,000 counts, updates 4/sec 33 segments, updates 32/sec -10...
... LED backlight to work in a compact package. Fluke 115 Multimeter Actual size R N10140 Technical Data Compact true-rms meter for field service technicians The Fluke 115 is specified for A input Display Bar graph Operating temperature Storage temperature Battery Battery life 600 V 6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2 11 A, 1000 V FAST Fuse (Fluke PN 803293) Digital: 6,000 counts, updates 4/sec 33 segments, updates 32/sec -10...
Fluke 115 Digital Multimeter Datasheet
Page 2
...-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com ©2006 Fluke Corporation. Frequency is dc coupled, 45 Hz to 1.5 at 4000 counts, decreasing linearly to 5 kHz for ac current. Fluke Corporation 115 Multimeter Ordering information Fluke-115 Multimeter Included TL75 Test leads, holster, User's manual and 9 V battery (installed). PO Box 1186, 5602 BD Eindhoven, The Netherlands For more...
...-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com ©2006 Fluke Corporation. Frequency is dc coupled, 45 Hz to 1.5 at 4000 counts, decreasing linearly to 5 kHz for ac current. Fluke Corporation 115 Multimeter Ordering information Fluke-115 Multimeter Included TL75 Test leads, holster, User's manual and 9 V battery (installed). PO Box 1186, 5602 BD Eindhoven, The Netherlands For more...
Fluke 114,115,117 Manual
Page 2
..., or abnormal conditions of operation or handling. This warranty does not cover fuses, disposable batteries, or damage from the date of the problem. Box 9090 Everett, WA 98206-9090 U.S.A. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of...
..., or abnormal conditions of operation or handling. This warranty does not cover fuses, disposable batteries, or damage from the date of the problem. Box 9090 Everett, WA 98206-9090 U.S.A. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of...
Fluke 114,115,117 Manual
Page 3
...to protect against transients in fixed-equipment installations at the distribution level. When making frequency measurements >1 kHz, the Y symbol is briefly displayed and an audible beep sounds when you to make a measurement with a 6000-count display and a bar graph. Test Lead Alert...-446-5500 Visit Fluke's web site at register.fluke.com. True-rms Multimeters Introduction The Fluke Model 114, Model 115, and Model 117 are battery-powered, true-rms multimeters (hereafter "the Meter") with a lead in an incorrect terminal. All figures show the Model 117. This manual applies to or from...
...to protect against transients in fixed-equipment installations at the distribution level. When making frequency measurements >1 kHz, the Y symbol is briefly displayed and an audible beep sounds when you to make a measurement with a 6000-count display and a bar graph. Test Lead Alert...-446-5500 Visit Fluke's web site at register.fluke.com. True-rms Multimeters Introduction The Fluke Model 114, Model 115, and Model 117 are battery-powered, true-rms multimeters (hereafter "the Meter") with a lead in an incorrect terminal. All figures show the Model 117. This manual applies to or from...
Fluke 114,115,117 Manual
Page 4
... finger guards. • Only use test leads that have the Meter serviced. • Do not apply more than the rated voltage, as specified in doubt, have the same voltage, category, and amperage ratings as the meter and that could damage the Meter or the equipment under test. 114, 115, and 117 Users Manual Safety Information A "XWWarning" statement identifies hazardous...
... finger guards. • Only use test leads that have the Meter serviced. • Do not apply more than the rated voltage, as specified in doubt, have the same voltage, category, and amperage ratings as the meter and that could damage the Meter or the equipment under test. 114, 115, and 117 Users Manual Safety Information A "XWWarning" statement identifies hazardous...
Fluke 114,115,117 Manual
Page 5
... Fluke or a D AC and DC qualified recycler for continuity before use the Auto Volts function to manual J Earth ground ~ Do not dispose of this product as required by this function's low input impedance (≈3 kΩ)(114 and 117 only). Symbols B AC (Alternating Current) F DC (Direct Current) X Hazardous voltage N Battery (Low battery when shown on the display.) I Fuse T Double Insulated W Important Information...
... Fluke or a D AC and DC qualified recycler for continuity before use the Auto Volts function to manual J Earth ground ~ Do not dispose of this product as required by this function's low input impedance (≈3 kΩ)(114 and 117 only). Symbols B AC (Alternating Current) F DC (Direct Current) X Hazardous voltage N Battery (Low battery when shown on the display.) I Fuse T Double Insulated W Important Information...
Fluke 114,115,117 Manual
Page 6
Measured input voltage ≥30 V, or voltage overload condition (OL). 114, 115, and 117 Users Manual Display No. edy02f.eps Model 117 114, 115, & 117 115 & 117 114, 115, & 117 114, 115, & 117 The Meter function is set to Diode Test Input is in the VoltAlert™ non-contact voltage detect mode. The Meter function is set to Continuity. X Unsafe voltage. A B C D E 4 Symbol w s R O Y 6 5 4 3 2 1 16 15 VoltAlert 7 14 17 8 9 10 11 12 13 18 Meaning The Meter is a negative value.
Measured input voltage ≥30 V, or voltage overload condition (OL). 114, 115, and 117 Users Manual Display No. edy02f.eps Model 117 114, 115, & 117 115 & 117 114, 115, & 117 114, 115, & 117 The Meter function is set to Diode Test Input is in the VoltAlert™ non-contact voltage detect mode. The Meter function is set to Continuity. X Unsafe voltage. A B C D E 4 Symbol w s R O Y 6 5 4 3 2 1 16 15 VoltAlert 7 14 17 8 9 10 11 12 13 18 Meaning The Meter is a negative value.
Fluke 114,115,117 Manual
Page 7
... units. 114, 115, & 117 K DC AC Direct current or alternating current 114, 115 & 117 L N Battery low warning. 114, 115, & 117 M 610000 mV Indicates the Meter's range selection. 114, 115, & 117 N (Bar graph) Analog display. 114, 115, & 117 O Auto Volts The Meter is rotated to or from any A position. 115 & 117 5 The Meter selects the range for the selected range. 114, 115, & 117 R LEAd W Test lead alert. User sets the Meter's range. 114, 115...
... units. 114, 115, & 117 K DC AC Direct current or alternating current 114, 115 & 117 L N Battery low warning. 114, 115, & 117 M 610000 mV Indicates the Meter's range selection. 114, 115, & 117 N (Bar graph) Analog display. 114, 115, & 117 O Auto Volts The Meter is rotated to or from any A position. 115 & 117 5 The Meter selects the range for the selected range. 114, 115, & 117 R LEAd W Test lead alert. User sets the Meter's range. 114, 115...
Fluke 114,115,117 Manual
Page 8
..., capacitance, frequency and testing diodes. edy01f.eps Model 115 & 117 114, 115, & 117 114, 115, & 117 bAtt CAL Err EEPr Err F11 Err Error Messages Battery must be repaired before the Meter will operate. Description A Input terminal for measuring ac and dc current to 10 A. Internal error. The Meter must be repaired before the Meter will operate. 114, 115, and 117 Users Manual Terminals 1 V A COM 3 10 A FUSED 2 No. Calibration required.
..., capacitance, frequency and testing diodes. edy01f.eps Model 115 & 117 114, 115, & 117 114, 115, & 117 bAtt CAL Err EEPr Err F11 Err Error Messages Battery must be repaired before the Meter will operate. Description A Input terminal for measuring ac and dc current to 10 A. Internal error. The Meter must be repaired before the Meter will operate. 114, 115, and 117 Users Manual Terminals 1 V A COM 3 10 A FUSED 2 No. Calibration required.
Fluke 114,115,117 Manual
Page 10
... resume MIN MAX AVG recording, press fagain. • To exit and erase stored readings, press p for at least one second or turn the rotary switch. Backlight Press Q to toggle the backlight on . 114, 115, and 117 Users Manual Battery Saver ("Sleep Mode") The Meter automatically enters "Sleep mode" and blanks the display if there is no function change, range change when you override Autorange...
... resume MIN MAX AVG recording, press fagain. • To exit and erase stored readings, press p for at least one second or turn the rotary switch. Backlight Press Q to toggle the backlight on . 114, 115, and 117 Users Manual Battery Saver ("Sleep Mode") The Meter automatically enters "Sleep mode" and blanks the display if there is no function change, range change when you override Autorange...
Fluke 114,115,117 Manual
Page 11
... and Auto is displayed when enabled. To exit Manual Range, press q for at least 1 second or turn the rotary switch. True-rms Multimeters Power-Up Options Button Power-Up Options f p q g Q Turns on the following table while turning the Meter on , it defaults to make basic measurements. Note You cannot manually change . 3. Power-Up Options are canceled when you turn the Meter off . Enables low impedance capacitance measurements. When connecting...
... and Auto is displayed when enabled. To exit Manual Range, press q for at least 1 second or turn the rotary switch. True-rms Multimeters Power-Up Options Button Power-Up Options f p q g Q Turns on the following table while turning the Meter on , it defaults to make basic measurements. Note You cannot manually change . 3. Power-Up Options are canceled when you turn the Meter off . Enables low impedance capacitance measurements. When connecting...
Fluke 114,115,117 Manual
Page 13
... DC DC edy03f.eps Using Auto Volts Selection (114 & 117 only) With the function switch in the l position, the Meter measures ac plus dc millivolts. edy18f.eps With the function switch in the x position, the Meter automatically selects a dc or ac voltage measurement based on the input applied between the V or + and COM jacks. Measuring AC and DC...
... DC DC edy03f.eps Using Auto Volts Selection (114 & 117 only) With the function switch in the l position, the Meter measures ac plus dc millivolts. edy18f.eps With the function switch in the x position, the Meter automatically selects a dc or ac voltage measurement based on the input applied between the V or + and COM jacks. Measuring AC and DC...
Fluke 114,115,117 Manual
Page 14
114, 115, and 117 Users Manual Measuring AC or DC Current (115 & 117) XWWarning To avoid personal injury or damage to the Meter: • Never attempt to make an in-circuit current measurement when the opencircuit potential to earth is >600 V. • Check the Meter's fuse before testing. (See "Testing the Fuse") • Use the proper terminals, switch position, and range for...
114, 115, and 117 Users Manual Measuring AC or DC Current (115 & 117) XWWarning To avoid personal injury or damage to the Meter: • Never attempt to make an in-circuit current measurement when the opencircuit potential to earth is >600 V. • Check the Meter's fuse before testing. (See "Testing the Fuse") • Use the proper terminals, switch position, and range for...
Fluke 114,115,117 Manual
Page 16
... Meter measures the frequency of a signal by counting the number of the measured signal is 0 V, 0 A for frequencies >1 kHz. AC Voltage Frequency AC Current Frequency VoltAlert Hz Press g to turn the frequency measurement function on and off. The trigger level is >1 kHz, the bar graph and Z are unspecified. If the frequency of times the signal crosses a trigger level each second. 114, 115, and 117 Users Manual Measuring Frequency (115...
... Meter measures the frequency of a signal by counting the number of the measured signal is 0 V, 0 A for frequencies >1 kHz. AC Voltage Frequency AC Current Frequency VoltAlert Hz Press g to turn the frequency measurement function on and off. The trigger level is >1 kHz, the bar graph and Z are unspecified. If the frequency of times the signal crosses a trigger level each second. 114, 115, and 117 Users Manual Measuring Frequency (115...
Fluke 114,115,117 Manual
Page 17
... measurements on cables with voltages as low as visual indication when voltage is detected. This setting is not saved when the Meter is turned off or goes into LoZ, (low input impedance) Capacitance mode. The VoltAlert detector works in bare wire applications with ghost voltage, hold qwhile turning on flush mounted wall sockets, power strips, flush mounted industrial outlets and various power cords. True...
... measurements on cables with voltages as low as visual indication when voltage is detected. This setting is not saved when the Meter is turned off or goes into LoZ, (low input impedance) Capacitance mode. The VoltAlert detector works in bare wire applications with ghost voltage, hold qwhile turning on flush mounted wall sockets, power strips, flush mounted industrial outlets and various power cords. True...
Fluke 114,115,117 Manual
Page 18
...bar graph is much faster than the digital display, the bar graph is useful for example (see below . The bar graph is disabled when measuring capacitance. The number of the selected range. In the 60 V range, for making peak and null adjustments. It has an overload indicator (>)...Fuse (115 & 117 only) Test fuse as shown below ), the major divisions on the scale represent 0, 15, 30, 45, and 60 V. In frequency, the bar graph and range annunciator indicates the underlying voltage or current up to the middle of −30 V turns on an analog meter. 114, 115, and 117 Users Manual Using...
...bar graph is much faster than the digital display, the bar graph is useful for example (see below . The bar graph is disabled when measuring capacitance. The number of the selected range. In the 60 V range, for making peak and null adjustments. It has an overload indicator (>)...Fuse (115 & 117 only) Test fuse as shown below ), the major divisions on the scale represent 0, 15, 30, 45, and 60 V. In frequency, the bar graph and range annunciator indicates the underlying voltage or current up to the middle of −30 V turns on an analog meter. 114, 115, and 117 Users Manual Using...
Fluke 114,115,117 Manual
Page 19
Replacing the Battery and Fuse XWWarning To avoid shock, injury, or damage to install the battery directly into the case, bottom edge first, until it from the case. Remove the test leads from the Meter before opening the case or battery door. • Use ONLY a fuse with the amperage, interrupt voltage, and speed ratings specified. Do not attempt to the Meter: •...
Replacing the Battery and Fuse XWWarning To avoid shock, injury, or damage to install the battery directly into the case, bottom edge first, until it from the case. Remove the test leads from the Meter before opening the case or battery door. • Use ONLY a fuse with the amperage, interrupt voltage, and speed ratings specified. Do not attempt to the Meter: •...
Fluke 114,115,117 Manual
Page 20
...115, and 117 Users Manual To open the case for fuse replacement: 1. Separate the case bottom from the case bottom. 4. Use only Fluke PN 803293. 6. Finally, insert the Meter into its holster. 3. Dirt or moisture in the terminals can affect readings. 18 Remove two screws from the case top. 5. To re-assemble the Meter... lens/window. Cleaning Wipe the case with an 11 A, 1000 V, FAST fuse having a minimum interrupt rating of 17,000 A. Do not use abrasives, isopropyl alcohol, or solvents to the case top, then install the two screws. Remove the Meter from its holder and replace it ...
...115, and 117 Users Manual To open the case for fuse replacement: 1. Separate the case bottom from the case bottom. 4. Use only Fluke PN 803293. 6. Finally, insert the Meter into its holster. 3. Dirt or moisture in the terminals can affect readings. 18 Remove two screws from the case top. 5. To re-assemble the Meter... lens/window. Cleaning Wipe the case with an 11 A, 1000 V, FAST fuse having a minimum interrupt rating of 17,000 A. Do not use abrasives, isopropyl alcohol, or solvents to the case top, then install the two screws. Remove the Meter from its holder and replace it ...
Fluke 114,115,117 Manual
Page 21
... 61010-1 2nd Edition for A input (115 & 117 only 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 °C to + 50 °C Storage: -40 °C to 90 %. True-rms Multimeters General Specifications General Specifications Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to...
... 61010-1 2nd Edition for A input (115 & 117 only 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 °C to + 50 °C Storage: -40 °C to 90 %. True-rms Multimeters General Specifications General Specifications Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to...
Fluke 114,115,117 Manual
Page 24
...min. 0.001 A 0.01 A 1.0 % + 3 115, 117 99.99 Hz Hz (V or A input)[2] 999.9 Hz 9.999 kHz 50.00 kHz 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 % + 2 115, 117 Notes: [1] All ac ranges except Auto-V LoZ are specified from 1 % to 1.5 at full...Users Manual Table 1 Accuracy Specifications (cont.) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model AC Amps Truerms[1] (45 Hz to 500 Hz) 6.000 A 10.00 A[3] 20 A for 30 seconds max., 10 minutes rest min. 0.001 A 0.01 A 1.5 % + 3 115, 117 DC Amps 6.000 A 10.00 A[3] 20 A for this and other true-rms meters to display...
...min. 0.001 A 0.01 A 1.0 % + 3 115, 117 99.99 Hz Hz (V or A input)[2] 999.9 Hz 9.999 kHz 50.00 kHz 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 % + 2 115, 117 Notes: [1] All ac ranges except Auto-V LoZ are specified from 1 % to 1.5 at full...Users Manual Table 1 Accuracy Specifications (cont.) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model AC Amps Truerms[1] (45 Hz to 500 Hz) 6.000 A 10.00 A[3] 20 A for 30 seconds max., 10 minutes rest min. 0.001 A 0.01 A 1.5 % + 3 115, 117 DC Amps 6.000 A 10.00 A[3] 20 A for this and other true-rms meters to display...