Question posted by pondith on January 18th, 2017

Fluke Co 220.

My FlukeCO starts up counts down from 010 to 000 and beeps twice and Then shuts of. I have replaced the battery and cleaned the sensor. Is the sensor bad.? Jim Wilson.

Current Answers

Answer #1: Posted by waelsaidani1 on January 19th, 2017 4:53 AM
This answer was accepted by the poster of the original question.
waelsaidani1

Member since:
May 12th, 2013

Points: 19,501,777
1,090
points
If battery power is below the level required to power the Meter, or if the sensor fails, the Meter emits a series of five beeps, followed by a long pause, and another series of five beeps. The Meter then turns itself off. If after replacing the battery, the alarm condition persists, the sensor may have failed. Contact an authorized Fluke service center for information on how to replace the sensor.

Answer #2: Posted by techyagent on January 19th, 2017 4:54 AM
techyagent

Member since:
June 17th, 2014

Points: 4,799,780
100
points
The Meter beeps twice, then performs a self-test. During the selftest, the display counts down from "10" to "0", while the Meter emits a series of 2 beeps followed by a pause. If the Meter passes the self-test, the Meter begins monitoring, which is indicated by a beep every two seconds, and shows a reading on the display.

battery power is below the level required to power the Meter, or if the sensor fails, the Meter emits a series of five beeps, followed by a long pause, and another series of five beeps. The Meter then turns itself off. If after replacing the battery, the alarm condition persists, the sensor may have failed.

I think problem is with the internals

https://www.emlab.com/m/store/Fluke%20CO%20Detector%20User%20Manual.pdf

Thanks


Techygirl

Related Fluke CO-220 Manual Pages

Fluke Air Quality Measurement - Making the Number Add Up Understanding Specifications and Performance of Indoor Air Quality Test - Page 1

... all specs are limited, accurate measurement is important nonetheless. • Instrument specifications. The sensor tech- nologies used to provide credible answers if results or procedures are challenged. Issues in...measure include: • Air temperature • Humidity • Airborne particle sizes and numbers. From the Fluke Digital Library @ www.fluke.com/library
Fluke Air Quality Measurement - Making the Number Add Up Understanding Specifications and Performance of Indoor Air Quality Test - Page 2

... quantify both accurate and complete. tive. Electronic components experience small changes (or "drift")... used to measure some air quality sensors makes their instruments-and what steps ...another period. At calibration, the clock starts over time. The ability to operate at... instruments Only by a temperature coefficient.  Fluke Corporation Making the numbers add up to making...
Fluke Air Quality Measurement - Making the Number Add Up Understanding Specifications and Performance of Indoor Air Quality Test - Page 3

...over time, and in non-contact thermometers. Calibration and, when necessary, replacement of out-of two sensor technologies: the thermocouple or the bead thermistor. In some instances, instrument ...indoor air quality applications.  Fluke Corporation Making the numbers add up -to-date proof of calibration for use one of -date or worn-out sensors will deviate. This range is ...
Fluke Air Quality Measurement - Making the Number Add Up Understanding Specifications and Performance of Indoor Air Quality Test - Page 4

...passes  Fluke Corporation Making the numbers add up: Understanding specifications and performance of a temperature tester may slow response time in the particle count. The sensors cannot be cleaned, so if... electronic circuitry will perform differently as temperatures vary, so the circuitry too must be returned to the manufacturer or sent to an independent testing laboratory to replace the...
Fluke Air Quality Measurement - Making the Number Add Up Understanding Specifications and Performance of Indoor Air Quality Test - Page 5

... Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com ©2006 Fluke Corporation. The gel sensor has a limited life span (two years) and its accuracy can be used as the signal strength of the optical filter/ detector. They can be calibrated...
FE CO-220 Instruction Sheet - Page 1

... ways: • By a reading on the LCD in the world: +1-425-446-5500 Or, visit Fluke's Web site at: www.fluke.com. CO-220 English Instruction Sheet Page 1 ® CO-220 Carbon Monoxide Meter Instruction Sheet Introduction The CO-220 Carbon Monoxide Meter (hereafter referred to as "the Meter") detects the presence of the following...
FE CO-220 Instruction Sheet - Page 2

... On Install the battery. (See "Installing and Replacing the Battery.") To turn the Meter on the display. During the selftest, the display counts down from enclosed ...and dizziness. 800 PPM Dizziness, nausea and convulsions. The Meter beeps twice, then performs a self-test. If the Meter passes the ...beeps followed by a beep every two seconds, and shows a reading on , press the green button....
FE CO-220 Instruction Sheet - Page 3

...220 English Instruction Sheet Page 3 Self-Test The Meter tests the sensor and the battery each time it is turned on the LCD. If battery power is below the level required to replace the sensor. If the Meter detects between 0-3 PPM of five beeps...the maximum CO level it does not establish a new baseline, but starts taking CO measurements. The back-light automatically turns off when sufficient ...
FE CO-220 Instruction Sheet - Page 4

... A LCD (Liquid Crystal Display) B Max/Hold button C ON/OFF button D Sensor (Fluke PN 689497) E Digital reading in parts/million (PPM) F Max/Hold mode (display shows maximum reading) G Low battery indicator H Beeper turned off and then turn it back on the display. CO-220 English Instruction Sheet Page 4 Muting the Beeper To mute the beeper...
FE CO-220 Instruction Sheet - Page 5

CO-220 English Instruction Sheet Page 5 Installing and Replacing the Battery 1 2 5 3 4 NV002F.EPS A Screw B Battery Door C Battery, 9 V NEDA or IEC 6LR61, or equivalent D Battery Connector E Sensor Common Sources of CO Common sources of potentially dangerous levels of CO are: • Poorly maintained furnaces, gas heaters, or fireplaces. • Dirty or plugged ...
FE CO-220 Instruction Sheet - Page 6

Defective/blocked flue. 4. Building not properly pressurized. Wrong fuel (not K-1). 3. Wrong wick or wick height. 4. Maladjusted burner. 4. Not enough air to burn fuel properly. 2. CO-220 English Instruction Sheet Page 6 CO and Appliance Malfunctions The following table identifies typical problems that can produce high levels of CO. Central heating Coal or ...
FE CO-220 Instruction Sheet - Page 7

CO-220 English Instruction Sheet Page 7 Specifications Temperature Operating: Storage: Humidity (non-condensing) < 3 weeks > 3 weeks Measurement range Measurement resolution Accuracy Sensor type Typical sensor life Temperature Coefficient Self Test Battery Typical battery life Certification 0 oC to +50 oC -30 oC to +50 oC 0-99 % RH 0-90 % RH 0 to 1000 PPM 1 PPM for 0 - 200 PPM; 5 PPM for...
FE CO-220 Instruction Sheet - Page 8

... replace the sensor, contact an authorized Fluke service center. Fluke Corporation P.O. During normal use, the Meter does not require adjustment. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of the problem. This warranty does not cover fuses, disposable batteries...

Similar Questions

No Led Display.
My co220 is used very infrequently and when I took it out of case the led display is no displaying a...
(Posted by rgrover56499 1 year ago)
199c Fails To Starts With Lcd Backlight Flash 1 Second, A Sound Of A Relay.
my 199c cannot be started. when i press power on button, the display only flash display background l...
(Posted by jasonguo 3 years ago)
Replacement Co Sensor
Where can a I buy replacement CO sensor for the fluke co-220? P/N 689497
(Posted by Muscente 5 years ago)
Clearing A Error On Startup
Dear Sirs, I have a 353 Clampmeter about 4 yesrs old. On Start up the Screen reads "Cal Error" The ...
(Posted by autowire 10 years ago)
Ask a New Question
Use the box below to post a new question about this Fluke product .
Points & Prizes
  • You can earn points for nearly everything you do on HelpOwl.com
  • You can trade in those points for gift cards at leading retailers such as Amazon.com and Walmart
  • It's that simple!
See How it Works
Create a Free Account

Fluke Manuals

Find free Fluke CO-220 manuals and user guides available at ManualOwl.com. Try out our unique manual viewer allowing you to interact with manuals from directly within your browser!

Fluke Reviews

View thousands of Fluke user reviews and customer ratings available at ReviewOwl.com.

Contact Information

Complete Fluke customer service contact information including steps to reach representatives, hours of operation, customer support links and more from ContactHelp.com.

Scoreboard Ratings

See detailed Fluke customer service rankings, employee comments and much more from our sister site.

Fluke Recalls

Find comprehensive Fluke recall information updated hourly on RecallOwl.com.